Introduction

Drug metabolism studies are challenging. This is especially true for the screening for drug related compounds in complex samples of in vivo studies with high matrix burden including the routinely used samples of plasma, urine, bile, and faeces, but also of whole blood and of extracts of organs.

Depending on applied drug doses and the pharmacokinetics of the analytes, signals of interest with exceptionally low intensity need to be screened out of matrix interferences of endogenous and/or co-administered additives with high signal intensity. Additionally the absence of radiotracers, which is the case in ‘first-in-man’ studies, results in challenging situations when metabolites needed to be reliably identified and elucidated in early stages of pharmaceutical development. With the background of regulatory guidelines [1, 2] and the continuous progress of analytical technologies the pivotal role of 14C-metabolism studies was recently revisited with the proposal that regulatory requirements can be even better met without conducting the traditional animal radiolabelled studies [3]. However the discussion evolves, it underlines the increasing importance of reliable high sensitivity screening methodologies in ‘real life samples’, especially in cold in vivo samples, for the future drug development process.

Many analytical approaches using different mass spectrometry techniques have been developed for the screening and identification of drug related compounds in drug metabolism studies. Unit resolution platforms like triple quadrupole [4] or ion trap instruments [58] used targeted approaches which required multiple runs of samples or predicted MRM- or untargeted enhanced survey scans to trigger product ion scanning within one analytical run. The data were successfully used for both metabolite identification and quantification of the parent drug and semi quantification of metabolites of in vivo and in vitro studies.

The high resolution hybrid systems, such as the LTQ-Orbitrap have effectively demonstrated their capabilities for parallel data collection, mass defect filtration and exact mass ion extraction for both in vitro and in vivo drug metabolism studies with and without data/information dependent acquisition methodologies [922]. With sophisticated generic settings including peak intensity dependent triggering and inclusion/exclusion list dependent triggering of MS/MS or MSn experiments, in vivo drug metabolism studies were performed with high reliability with respect to metabolite coverage.

The MSE acquisition technique represents another approach and provides a completely untargeted and unbiased methodology which can be used with fast acquisition hybrid quadrupole time-of-flight (QTOF) instruments in combination with the separation power of ultra high performance liquid chromatography (UPLC). The advantage of this technique is the simplified data collection which is (1) independent of exclusion lists for high abundant matrix interferences that may impair the efficiency of data dependent MS/MS-triggering and which is (2) independent of generic or predictive inclusion lists which are biased and therefore potentially not comprehensive. MSE data enable the analysis for any compound of interest, irrespective of its signal intensities or those of co-eluting matrix ions.

This technology was previously shown to be universally applicable not only for drug metabolism studies [2326], but also for impurity analysis [27], environmental/waste water analysis [28, 29], and proteomics [30]. We here report the use of the UPLC-QTOF/MSE approach as general ‘first-line’ screening for in vitro and in vivo studies. Published drug metabolism data of the recently approved drugs Dabigatran [31] and Linagliptin [32] were based on this technology. For both drugs the alignment of parent-ions with fragment ion species were achieved by matching the extracted ion chromatograms (XICs) with exactly the same peak shape. The chromatographic resolution therefore is an important factor in the process. The MSE approach perfectly works for samples with clean matrix from in vitro studies and with in vivo samples with advantageous analyte/matrix ratio. However, for the identification of low level metabolites dedicated MS/MS-experiments are needed due to strong interferences of co eluting matrix ions.

The utilization of ion mobility separation (IMS), as an additional principle of differentiation, will play a key role for the further development of the MSE acquisition methodology. Based on the collisional cross section as key parameter, ions are separated by differences of individual size to charge ratios in combination with the interaction of a buffer gas. The ion mobility separation of structural isomers, polymeric conformers and enantiomers in applications of proteomics, glycomics and metabolomics was successfully demonstrated with various instrumental setups [3335].

In this publication we present the combination of UPLC and travelling wave ion mobility separation (TWIMS) together with the MSE acquisition methodology as new approach for the reliable, unbiased and fast analysis of drugs and its metabolites which is tailored for the analysis in complex biological sample material.

Only parts of the parent compound structure and only reduced mass spectrometry data are allowed to be presented. However, this does not detract from the demonstration of the viability and utility of the general approach outlined here.

Materials and methods

Chemicals and reagents

The drug compound of interest (M0) was synthesized by the medicinal chemistry group of Boehringer Ingelheim Pharma GmbH & Co. KG, Ingelheim, Germany. Chemical purity was >97 %. For lock mass correction a solution of leucin enkephalin (Sigma Aldrich, Stadt, Germany) at a concentration of 2 μg/ml in acetonitrile/0.1 % aqueous formic acid (1:1) was used. Acetonitrile, methanol, water and formic acid were of analytical grade or higher purity and were obtained from Sigma Aldrich (Steinheim, Germany).

Biological sample material

The study protocol of in vivo experiments was in full conformance with the respective animal experimentation application approved by the local animal welfare officer of Boehringer Ingelheim Pharma GmbH & Co. KG, Biberach, Germany, as well as by the responsible supervisory authority, Tübingen, Germany. Boehringer Ingelheim Pharma GmbH & Co. KG is accredited by the Association for Assessment and Accreditation of Laboratory Animal Care International (AAALAC).

Male and female rats (n = 3, stain: Crl:WI(HAN)) received a single oral dose of 50 mg/kg of the drug compound dissolved in water after an overnight fast (≥10 h). Plasma (sampling times: 0, 0.25, 0.5, 1, 6, 24 h), urine and faeces (fractions 0–48 h) were sampled. Bile fractions (0–6 h) were obtained from bile duct cannulated rats (n = 2). Sample aliquots of urine and faeces samples of individuals were pooled proportionally according to the relative weight of individual sample fractions. Plasma was pooled proportionally to sampling times to yield a sample in which concentrations of analytes were proportional to AUC’s (AUC = area under the concentration time curve) within the given time interval [36, 37]. The pooled plasma (0–24 h) was processed by solid phase extraction using OASIS HLB cartridges (200 mg, 6 cc, Waters, Milford, MA, USA) and 0.1 % aqueous formic acid/acetonitrile (1:9) as eluent. Pooled faeces samples were exhaustively extracted with acetonitrile/water (1:1). Pooled urine and bile was diluted 1:10 with aqueous formic acid (0.1 %) before analysis. All samples and sample extracts were stored at −20 °C until analysis.

UPLC/TWIMS/MSE QTOF mass spectrometry

Metabolite profiles of the sample material were analyzed by ultra high performance liquid chromatography coupled to traveling wave ion mobility separation QTOF mass spectrometry using the MSE acquisition methodology. Volumes of 5 μl of sample material were injected on column. The UPLC-system (Acquity UPLC, Sample Manager 2777, Binary solvent manager, Column Heater/Cooler, all Waters, Milford, MA, USA) was coupled to a Synapt G2 mass spectrometer (Waters MS Technologies, Manchester, UK). Samples were analyzed by reverse phase UPLC using 150 × 3.0 mm columns (Ascentis express Phenyl-Hexyl, 2.7 μm, Supelco, USA) with a gradient of aqueous ammonium formate (0.05 M, pH 8) as eluent A versus acetonitrile/methanol (1:1) as eluent B with a flow rate of 0.6 ml/min (gradient: 5 % B at 0 min, 5 % B at 1 min, 30 % B at 20 min, 99 % B at 23 min, 99 % B at 32 min). The column temperature was 40 °C.

The combined Z-spray/lockspray interface of the Synapt G2 [38, 39] was operated in positive ion mode with nitrogen as cone and desolvation gas at flow rates of 20 L/h and 1,000 L/h, respectively. The source temperature and desolvation temperature were set to 120 °C and 550 °C, respectively. The capillary, sampling cone and extraction cone voltages were set to 3.0 kV, 20 V and 4 V, respectively. The first quadrupole operated in non-resolving RF-only mode. The collision energy of the trap cell was set to 4 eV. Mass spectral data over the range m/z 50–1200 were acquired in two separate channels using two continuously interlaced collision energies of the transfer cell, at 0 eV and 25 eV, with a switching frequency of 2 Hz. The trap, IMS and transfer traveling wave devices were operated with traveling wave amplitudes and velocities of 8.0 V and 508 m/s, 40.0 V and 1000 m/s, and 3.0 V and 450 m/s, respectively. The trap/transfer cells were operated with argon at a pressure of 2.5 e-2 mbar. Nitrogen as drift gas was introduced into the IMS cell to maintain a pressure of 3.1 mbar. The time of flight analyzer was operated using V-mode optics with an average mass resolution of m/Δm = 20,000 (full-width at half-maximum definition). The TOF-pusher and trap-gate frequencies were 18.5 kHz and 92.5 Hz, respectively. Thus, individual IMS cycles of 10.8 ms duration were analyzed with a pusher period of 54 μs resulting in 200 TOF pushes/IMS cycle. Mass data were lock mass calibrated with leucine-enkephalin (m/z 556.2771). The calibrant was introduced with a flow rate of 10 μl/min and was measured for 0.5 s with a frequency of 1/min over the full chromatographic run time. Lock mass data were acquired in a third separate channel. Data acquisition and processing were performed using Masslynx (V4.1, SCN714), Driftscope (V2.1) and MSE viewer (V1.2).

Results and discussion

Pooled samples of plasma, urine, bile and faeces after oral administration of a drug compound to rats were investigated in this study. Native urine and bile were diluted; plasma and faeces were extracted before analysis.

The investigated compound (M0), a tetrahydropyranyl (THP) derivative with substituted heteroaromatic system, was subject to extensive phase I metabolism (Fig. 1). Metabolites were formed by initial oxidation of the THP-moiety in 2-, 3- and 4-position (metabolite fraction M8). Two major metabolites, the ring-open carboxylic acid/alcohol (M1) and the diol (M2), were formed by initial oxidation in 2-position of the THP-moiety followed by further oxidation (leading to M1) or reduction (M2), respectively. Adducts with urea were identified in urine (M3, M4). Minor metabolites were formed by oxidation of the heteroaromatic system (M5, M6, M7). Phase II transformations played a subordinate role.

Fig. 1
figure 1

Metabolism pathways of M0

Basic principles of metabolite screening by UPLC/TWIMS/MSE QTOF-MS

In general, samples were analyzed for drug related compounds by two independent, orthogonal separation techniques including UPLC (ultra high performance liquid chromatography) and TWIMS (travelling wave ion mobility separation) followed by detection with high resolution QTOF mass spectrometry using the MSE acquisition approach. The instrument was operated in positive ion mode with the first quadrupole operating non-resolving, in RF-only mode.

The triwave geometry of the instruments ion mobility cell [38, 39] allows fragmentation in both the trap (pre IMS separation) and the transfer regions (post IMS separation). In this application, transfer fragmentation was performed. According to the MSE acquisition methodology the collision energy was alternately switched with a frequency of 2 Hz between two settings ‘low (=0 eV)’ and ‘high (=25 eV)’ which allowed the parallel detection of [M+H]+-ions with only minimal fragmentation and of only fragment ions, respectively (Fig. 2). [M+H]+-ions and fragments of all, endogenous compounds and drug related compounds of interest, were acquired in two separated data sets ‘low’ and ‘high’, respectively. Ions from an individual compound e.g. [M+H]+-ions in the data ‘low’ and their specific fragments in the data ‘high’, are aligned by the retention time (retention time alignment).

Fig. 2
figure 2

Synapt G2, combined UPLC with IMS and MSE with interlaced collision energies

In combination with ion mobility separation, [M+H]+-ions from the RF-only quadrupole are first collected in the trap cell of the instruments triwave device and then periodically released in packages into the IMS cell (Fig. 2). After separation due to differences in collisional cross sections, they are transported into the transfer cell. Here, with interlaced collision energies, the [M+H]+-ions were either transported with a low collision energy (0 eV) without or minimum fragmentation to the TOF detector or subjected to additional fragmentation with a higher collision energy (25 eV). With collision energy applied after ion mobility separation, fragments retain the original drift time of the parent ions. As a result, all ions of an individual compound e.g. [M+H]+, and its fragments are aligned not only by retention time, but additionally by drift time (=retention time and drift time alignment).

In summary, all ion species of the two data sets ‘high’ (product ions) and ‘low’ (parent ions) were characterized by the four dimensions intensity (I), mass (m/z), retention time (rt) and drift time (dt). Ions of individual compounds are aligned by specific combinations of retention- and drift times.

Data evaluation, basic principles

The general concept of screening for metabolites by diagnostic drug related fragments by rt/dt-aligned MSE data is discussed in the following:

The basis for the metabolite screening with the MSE methodology is a sound understanding of the fragmentation pathways of the parent drug’s [M+H]+. Drug related compounds can be identified by extracted ion chromatograms (XICs) of diagnostic fragments in the high collision energy data (‘data high’) which are specific for defined substructures of the parent compound. Associated [M+H]+-ions are identified with their XICs in the low collision energy data (‘data low’) by correlation with exactly matching retention times and chromatographic profiles and drift times and vice versa. Independent of the intensity, any known fragment or — with respect to putative biotransformations — any expected product ion can be used for this approach. This methodology is presented in the following sections.

The product ion spectrum of the parent drug’s [M+H]+, M0, is shown in Fig. 3. The base ion was formed by loss of the THP-residue, [M-THP+H]+, and was used as diagnostic fragment for metabolites with unchanged heterocyclic moiety. In Fig. 4 the search for metabolites with [M-THP+H]+ in urine is presented. The extracted ion chromatogram of [M-THP+H]+ in the data ‘high’ is displayed in the upper trace (Fig. 4, XIC A). The chromatograms below show the XICs of the correlated [M+H]+-ions of the parent compound M0 (XIC B) and the metabolites M1 (XIC C: M0+2O), M2 (XIC D: M0+O+2H), M3, M4 (XIC E: M0+CH2N2O) and M8 (XIC F: M0+O) in the data ‘low’. Metabolite specific ions of both data ‘high’ and ‘low’ exhibit exactly the same peak shape. The higher the chromatographic resolution, the better the alignment of metabolite related peaks and the discrimination of matrix signals.

Fig. 3
figure 3

MS/MS data of the parent compound’s [M+H]+

Fig. 4
figure 4

Identification of drug related compound in rat urine by correlation of XICs of the diagnostic fragment [M-THP+H]+ with metabolite [M+H]+-ions. Upper trace: data ‘high’: XIC of [M-THP+H]+, light gray: TIC. Lower traces: data ‘low’: XIC of parent drug and metabolite’s [M+H]+, light gray: TIC

With sample material of in vivo studies a complete separation of interferences is only achieved in rare cases due to the high matrix burden. Thus, the unambiguous structure assignment based on only retention time aligned ‘high’ and ‘low’ data alone is challenging, sometimes impossible. With the introduction of the additional dimension of separation by ion mobility, completely purified MS- and MS/MS data are achievable. In Fig. 5 a common situation of a strong interference with matrix compounds with nearly identical chromatographic profile to the compound of interest is shown. The parent compound, M0, co eluted with a polyethylenglycol isomer, [M+H]+ m/z 564, at the retention time of 17.9 min. The IMS data can be used to unambiguously resolve these peaks and align their fragments without the need for dedicated MS/MS investigations. Figure 6 shows the XIC data ‘high’ (2 upper traces) and ‘low’ (2 lower traces) of the peak top at 17.9 min (average, ±1.5 s) which is now plotted versus the drift time dimension. Again, the XIC of the diagnostic fragment, [M0-THP+H]+ (XIC A) can be used as indicator for the drug related compounds. The drift time for this fragment was 4.4 ms. The aligned [M+H]+ of the parent compound was unambiguously identified by the corresponding XIC in the data ‘low’ (XIC C). The [M+H]+ of the PEG-isomer, separated by IMS was detected with a drift time of 7.6 ms (XIC D). Its fragment C6H13O3 +, m/z 133, which, in a similar manner, can be used as general diagnostic fragment for PEG-isomers (see below) co eluted with the same drift time at 7.6 ms (XIC B).

Fig. 5
figure 5

Urine after administration of M0. Upper trace: data ‘high’: XIC of [M-THP+H]+, light gray: TIC. Lower traces: data ‘low’: XIC of parent drug’s and PEG’s [M+H]+, light gray: TIC, Inserted figure: mass data at rt = 17.9 min, data ‘low’

Fig. 6
figure 6

Urine after administration of M0. Mass data ‘high’ and ‘low’ at rt = 17.9 min ±1.5 s, plotted versus the drift time. Upper 2 traces: data ‘high’: XIC of [M-THP+H]+, and PEG fragment, light gray: TIC Lower 2 traces: data ‘low’: XIC of parent drug’s and PEG’s [M+H]+, light gray: TIC

Figure 7 demonstrates the improvement of the data quality of MS- (data ‘low’, left) and MS/MS-spectra (data ‘high’, right) from rt-aligned data to combined rt/dt-aligned data. The [M+H]+-ion species of the parent drug and the PEG-isomer were separated from one another and were additionally completely stripped of other interferences by applying appropriate retention time and drift time windows of ±1.5 s and ±1.5 ms, respectively. Due to the conservation of the [M+H]+-parent ion drift times, the parent compound- and PEG-fragments were also completely separated and cleaned from one another and clear of other matrix interferents. Thus, the combined rt/dt-aligned MS- and MS/MS-spectra can directly be used for structure elucidation without compromise.

Fig. 7
figure 7

Urine after administration of M0. Mass data at rt = 17.9 min before and after mobility separation (rt/dt-windows: 1.5 s/1.5 ms). Left: MS-spectra of data ‘low’, drift time separation of [M+H]+-parent ions at 4.4 and 7.6 ms. Right: MS/MS-spectra of data ‘high’, remaining [M+H]+- and fragment ions with conserved drift time separation of [M+H]+ parent ions

A second representative met-ID example is presented using the metabolite fraction M8 which consisted of oxidation products of the THP moiety. The XIC of the [M+H]+ showed a broad peak at 14.0 min (see Fig. 4, XIC F). In Fig. 8, the mass spectra of the data ‘low’ of the peak top at 14.0 min (±1.5 s) without and with ion mobility separation is presented: Without IMS the [M+H]+ together with its in source fragment [M-THP+H]+ were present with high abundant interferences of co eluting matrix compounds (upper spectrum). The ions were separated and aligned by IMS into drug related ions (dt = 3.2 ms and 4.5 ms) and matrix signals (dt = 3.9 ms and 6.9 ms). The spectra are shown with rt/dt-windows of ±1.5 s/±1.5 ms. The process of ion alignment is demonstrated again in Fig. 9: Here the ‘high’ and ‘low’ mass data of the peak top at 14.0 min (±1.5 s) are plotted versus the drift time; XICs which significantly contributed to the total ion current are selected. The XIC of the diagnostic fragment, [M-THP+H]+ of the data ‘high’ showed two peaks at the drift times 3.2 ms and 4.5 ms (upper trace, XIC A). In the data ‘low’ the XIC of the M8’s [M+H]+ (at 4.5 ms, XIC B) could be unambiguously aligned with the diagnostic fragment of the data ‘high’. In this example, the second peak at 3.2 ms (XIC C) corresponded to the XIC of the [M-THP+H]+ itself, which was formed by in-source fragmentation and therefore resolved in the IMS cell from the parent ion prior to the transfer cell fragmentation process. The prominent matrix ions were completely separated with drift times of 3.9 ms and 6.9 ms, respectively (XIC D and E).

Fig. 8
figure 8

Urine after administration of M0. Mass data ‘low’ at retention time of metabolite fraction M8 at rt = 14.0 min ± 1.5 s. Upper spectrum: Mass data over all drift times: 0–10 ms. Lower spectra: Mass data of metabolite ions and prominent matrix signals at defined drift time windows (±1.5 ms)

Fig. 9
figure 9

Urine after administration of M0. M8-data ‘high’ and ‘low’ at rt = 14.0 min ± 1.5 s, plotted versus the drift time. Upper trace: data ‘high’: XIC of the diagnostic [M-THP+H]+-ion, light gray: TIC Lower traces: data ‘low’: XICs of the [M+H]+-ion of M8 and its in-source fragment [M-THP+H]+ together with XICs of additional matrix signals with significant contribution to the TIC, light gray: TIC

Data evaluation, implementation of automated data mining tools

The general approach of the data evaluation process as discussed in the previous section is laborious and time consuming. For an accelerated routine analysis of the 4D ‘high’ and ‘low’ data the MSE viewer was used as an automated software tool. In a first step of data processing, the included apex convolution algorithm reduced the raw data, i.e. the continuum data with respect to the dimension, mass (m/z), retention time (rt) and drift time (dt), to single discrete component data (Fig. 10). Consequently, a particular ion of interest can be displayed as a single data point in a 3 dimensional data space which is defined by the axes m/z, rt, and dt. In this 3D data space, signals are resolved which are differentiated in at least one of the parameter m/z, rt or dt. The fourth dimension which is the intensity (I) is not displayed in Fig. 10. In the second step of data processing, the MSE viewer allows the extraction of purified MS- and MS/MS data by appropriate settings of retention time and drift time windows. The drift time is implemented as a filter and the display can be toggled to show the data with and without dt-separation.

Fig. 10
figure 10

4D Apex convolution principle: Left picture: continuum 4D raw data which is defined by intensity (I), mass (m/z), retention time (rt), and drift time (dt). Mid picture: continuum data is convoluted to discrete components. Right picture: display of convoluted data in the m/z-/rt-/dt-data space: signals are separated when differentiated in at least one of the dimensions m/z, rt or dt

With regard to the worst case of in vivo met-ID studies, the very low abundance metabolite signals are resolved in the 3D-data space (m/z vs. rt vs. dt) and can be identified as drug related signals, even in the presence of very intense co eluting matrix ions.

The above process is demonstrated in the following, using a common challenging, in vivo met-ID example: the analysis of an extract of rat faeces with low concentrations of drug related compounds together with very high matrix burden.

The continuum raw data ‘high’ and ‘low’ were converted to discrete component data by the peak detection software for further metabolite analysis. In Fig. 11 the screening for drug related compounds in the faeces sample with the diagnostic [M-THP+H]+ fragment is presented: The upper trace shows the ‘XIC’ of the [M-THP+H]+ of the peak detected component data ‘high’. The lower traces shows the automatically rt/dt-aligned [M+H]+-ions of the parent drug (M0) and the metabolites M1, M2, M5 and M6 of the peak detected component data ‘low’. Due to the orthogonality of UPLC and TWIMS, interfering ions of unwanted matrix are efficiently discriminated. The spectra are shown with rt- and dt-windows of ±1.5 s and ±0.05 ms, respectively.

Fig. 11
figure 11

Identification of drug related compound in rat faeces after 4D apex convolution of ‘high’ and ‘low’ data: rt/dt-alignment of component signals (rt/dt-windows: ±1.5 s/±0.05 ms). Upper trace: [M-THP+H]+ component peaks versus retention time. Lower traces: parent drug’s and metabolite’s [M+H]+ component peaks after rt/dt-alignment versus retention time

In the Figs. 12 and 13 the comparison of continuum raw data at the peak tops of the [M+H]+-XICs of the metabolites M1 and M6 versus the rt/dt-aligned MS data is presented. The continuum ‘high’ and ‘low’ raw data (Figs. 12 and 13, left) demonstrate the analytical challenge with low abundant analyte signals and high matrix interferences which is a common situation with in vivo sample material, especially in samples of faeces or bile. After apex convolution in combination with rt/dt-alignment the MS- and MS/MS-spectra (Figs. 12 and 13, right) were completely purified from all matrix signals. The extracted data were identical to directed MS/MS-spectra of selected [M+H]+-parent ions and can be used without compromises for structure elucidation and further downstream processes. The advantage over directed analytical approaches is the simplified setup which is completely unbiased: Without compromise, rt/dt-aligned MS- and MS/MS-data from any compound of interest can be extracted retrospectively at any time. The shown examples of the metabolites M1 and M6 are representative for all analytes including M0, M2, M3, M5. The rt/dt-alignment resulted to cleaned MS- and MS/MS spectra (data not shown here).

Fig. 12
figure 12

Metabolite M1 in faeces (rt: 7.06 min), Data ‘low’: a continuum data, b after apex convolution and rt/dt-alignment (rt: 18.147 ± 1.5 s, dt: 4.38 ± 0.05 ms). Data ‘high’: c continuum data, d after apex convolution and rt/dt-alignment (rt: 18.147 ± 1.5 s, dt: 4.38 ± 0.05 ms)

Fig. 13
figure 13

Metabolite M6 in faeces (rt: 12.16 min), Data ‘low’: a continuum data, b after apex convolution and rt/dt-alignment (rt: 18.147 ± 1.5 s, dt: 4.38 ± 0.05 ms). Data ‘high’: c continuum data, d after apex convolution and rt/dt-alignment (rt: 18.147 ± 1.5 s, dt: 4.38 ± 0.05 ms)

To further demonstrate the universality of this approach an excursus to the analysis of the PEG-isomers is presented. As discussed earlier the parent compound co eluted with a PEG-isomer with nearly the same retention time at 17.9 min (Fig. 5). Figure 7 showed the alignment of the [M+H]+, m/z 564 (data ‘low’), of the PEG-isomer with m/z 133.0865 (C6H13O3 +) of the data ‘high’, which is a common fragment of this compound class [40]. This general PEG-related fragment can be used as diagnostic ion for PEG-isomers in the original data ‘high’ and ‘low’ for the identification of all PEG-isomers which were present in the urine sample (Fig. 14): The XIC of m/z 133.0865 is shown in the upper trace; a superimposition of aligned [M+H]+-XICs of the PEG-isomers was correlated in the lower trace. Data processing with the MSE viewer by apex convolution and rt/dt-alignment resulted also for this PEG-isomers to completely purified MS- and MS/MS-data for further structure elucidation/assignment (data not presented).

Fig. 14
figure 14

Urine after administration of M0. Data ‘high’and ‘low’. Upper trace: data ‘high’: XIC of PEG fragment, m/z 133, light gray: TIC. Lower trace: data ‘low’: superimposition XICs of PEG-isomers [M+H]+, light gray: TIC

Summary and conclusions

This new approach for the screening and elucidation of drugs and their metabolites in complex matrix consists of the combination of two orthogonal separation techniques, UPLC (ultra high performance liquid chromatography) and TWIMS (traveling wave ion mobility separation), together with high resolution QTOF mass spectrometry using the MSE acquisition methodology. Two mass data sets were acquired separately in two channels using interlaced collision energies detecting [M+H]+ parent ions and product ions in parallel. All ion species of the two data sets were characterized by the four dimensions intensity (I), mass (m/z), retention time (rt) and drift time (dt). The raw data were processed by the apex convolution software for the reduction of the continuum m/z-, rt- and dt-data to discrete components. As a result, ions that differ in at least one of the dimensions m/z, rt or dt were resolved. Using fragmentation after ion mobility separation, [M+H]+ ion species of the low collision energy channel are aligned with their related fragments of the high collision energy channel by virtue of possessing the same combination of retention time and drift time. Thus, associated MS- and MS/MS-data for compounds of interest are extractable with appropriate rt- and dt-windows. Due to the orthogonality of UPLC and TWIMS, unwanted interfering ions, e.g. high intensity matrix signals which are generally common in mass data of in vivo sample material, were efficiently discriminated. As result, completely purified MS- and MS/MS-data for any compounds can be extracted from the mass data for further evaluation. This principle supersedes the data dependent scan methodologies which are laborious and difficult to set up in complex sample materials with low concentrations of analytes and high matrix burden. The MSE viewer as a tailored software tool was used for automated data mining.

This contribution refers to the application of UPLC/TWIMS/MSE QTOF Mass Spectrometry to drug metabolism studies. Its general and universal applicability is demonstrated by the analysis of in vivo sample material of especially faeces containing highest concentrations of matrix components and low concentrations of analytes. The MS- and MS/MS data of drug related compounds was completely purified from matrix signals and could be used for further processing, e.g. structure elucidation/structure assignment, without compromise.

In general the unbiased ‘high/low’ 4D data sets open the field for retrospective analysis of any compound at any time including drug metabolites, up and down regulated endogenous compounds, small molecules and proteins without the need for sample reanalysis which is always questionable with respect to stability and storage conditions. Thus, the 4D data sets can be used as comprehensive data repositories during the development process of a drug including early in vitro investigations, first in man trials, drug metabolism studies in samples of toxicology species and late stage 14C human ADME trials. This generic approach is particularly useful in cases where the coverage of human metabolites need to be retrospectively demonstrated in previously analyzed preclinical samples of toxicology testing.