Abstract
Background
Automated devices collecting quantitative measurements of pupil size and reactivity are increasingly used for critically ill patients with neurological disease. However, there are limited data on the effect of ambient light conditions on pupil metrics in these patients. To address this issue, we tested the range of pupil reactivity in healthy volunteers and critically ill patients in both bright and dark conditions.
Methods
We measured quantitative pupil size and reactivity in seven healthy volunteers and seven critically ill patients with the Neuroptics-200 pupillometer in both bright and dark ambient lighting conditions. Bright conditions were created by overhead LED lighting in a room with ample natural light. Dark conditions consisted of a windowless room with no overhead light source. The primary outcome was the Neurological Pupil Index (NPi), a composite metric ranging from 0 to 5 in which > 3 is considered normal. Secondary outcomes included resting and constricted pupil size, change in pupil size, constriction velocity, dilation velocity, and latency. Results were analyzed with multi-level linear regression to account for both inter- and intra-subject variability.
Results
Fourteen subjects underwent ten pupil readings each in bright and dark conditions, yielding 280 total measurements. In healthy subjects, median NPi in bright and dark conditions was 4.2 and 4.3, respectively. In critically ill subjects, median NPi was 2.85 and 3.3, respectively. Multi-level linear regression demonstrated significant differences in pupil size, pupil size change, constriction velocity, and dilation velocity in various light levels in healthy patients, but not NPi. In the critically ill, NPi and pupil size change were significantly affected.
Conclusion
Ambient light levels impact pupil parameters in both healthy and critically ill subjects. Changes in NPi under different light conditions are small and more consistent in healthy subjects, but significantly differ in the critically ill. Practitioners should standardize lighting conditions to maximize measurement reliability.
Similar content being viewed by others
References
Fisher CM. The neurological examination of the comatose patient. Acta Neurol Scand. 1969;45:1–56.
Mcnealy DE, Plum F. Brainstem dysfunction with supratentorial mass lesions. Arch Neurol. 1962;7:10–32.
Chesnut RM, Gautille T, Blunt BA, Klauber MR, Marshall LE. The localizing value of asymmetry in pupillary size in severe head injury: relation to lesion type and location. Neurosurgery. 1994;34:840–5.
Marmarou A, Lu J, Butcher I, McHugh GS, Murray GD, Steyerberg EW, et al. Prognostic value of the glasgow coma scale and pupil reactivity in traumatic brain injury assessed pre-hospital and on enrollment: an impact analysis. J Neurotrauma. 2007;24:270–80.
Andrews BT, Pitts LH. Functional recovery after traumatic transtentorial herniation. Neurosurgery. 1991;29:227–31.
Sakas DE, Bullock MR, Teasdale GM. One-year outcome following craniotomy for traumatic hematoma in patients with fixed dilated pupils. J Neurosurg. 1995;82:961–5.
Tien HC, Cunha JR, Wu SN, Chughtai T, Tremblay LN, Brenneman FD, et al. Do trauma patients with a glasgow coma scale score of 3 and bilateral fixed and dilated pupils have any chance of survival? J Trauma. 2006;60:274–8.
Litvan I, Saposnik G, Maurino J, Gonzalez L, Saizar R, Sica RE, et al. Pupillary diameter assessment: need for a graded scale. Neurology. 2000;54:530–1.
Couret D, Boumaza D, Grisotto C, Triglia T, Pellegrini L, Ocquidant P, et al. Reliability of standard pupillometry practice in neurocritical care: an observational, double-blinded study. Crit Care. 2016;20:99.
Meeker M, Du R, Bacchetti P, Privitera CM, Larson MD, Holland MC, et al. Pupil examination: validity and clinical utility of an automated pupillometer. J Neurosci Nurs. 2005;37:34–40.
R Core Team. R: A language and environment for statistical computing. R Foundation for Statistical Computing, Vienna, Austria, 2017. http://www.R-project.org/.
White D, Rabago-Smith M. Genotype-phenotype associations and human eye color. J Hum Genet. 2011;56:5–7.
Kawasaki A. Disorders of pupillary function, accomodation and lacrimation. In: Miller NNN, editor. Walsh and hoyt’s clinical neuro-ophthalmology. Philadelphia: Lippincott Williams & Wilkins; 2005. p. 740–805.
Shirozu K, Setoguchi H, Tokuda K, Karashima Y, Ikeda M, Kubo M, et al. The effects of anesthetic agents on pupillary function during general anesthesia using the automated infrared quantitative pupillometer. J Clin Monit Comput. 2017;31:291–6.
Loewenfeld I. Pupillary changes related to age. In: Thompson HS, editor. Topics in neuro-ophthalmology. Baltimore: Williams & Williams; 1979. p. 124–50.
Digre KB. Principles and techniques of examination of the pupils, accomodation, and lacrimation. In: Miller NRN, Nancy J, editors. Walsh and hoyt’s clinical neuro-ophthalmology. Philadelphia: LIppincott Williams & Wilkins; 2005.
Olson DM, Stutzman S, Saju C, Wilson M, Zhao W, Aiyagari V. Interrater reliability of pupillary assessments. Neurocrit Care. 2016;24:251–7.
Lara F, Bernal-Molina P, Fernandez-Sanchez V, Lopez-Gil N. Changes in the objective amplitude of accommodation with pupil size. Optom Vis Sci. 2014;91:1215–20.
Ba-Ali S, Lund-Andersen H, Ahmadi H, Brondsted AE. Effect of intermittent versus continuous light exposure on pupillary light response, as evaluated by pupillometry. Front Neurol. 2017;8:746.
Lowenstein OLI. Influence of retinal adaptation upon the pupillary reflex to light in normal man. II. Effect of adaptation to dim illumination upon pupillary reflexes elicted by bright light. Am J Ophthalmol. 1961;51:644–54.
Funding
American Brain Foundation Crowdsourcing Grant
Author information
Authors and Affiliations
Contributions
CO, MD, MPHS were responsible for study design, analysis, interpretation of results, and the writing of the manuscript. MH, BS assisted in statistical analysis of results and entering demographic and clinical data for patients. SS, MD, PHD oversaw study design and provided expertise regarding analysis and interpretation of results.
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical Approval
The study has been approved by the institutional ethics committee. For this type of retrospective study, formal consent is not required.
Electronic Supplementary Material
Below is the link to the electronic supplementary material.
Rights and permissions
About this article
Cite this article
Ong, C., Hutch, M. & Smirnakis, S. The Effect of Ambient Light Conditions on Quantitative Pupillometry. Neurocrit Care 30, 316–321 (2019). https://doi.org/10.1007/s12028-018-0607-8
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12028-018-0607-8