Imaging study of pseudo-CT images of superposed ultrasound deformation fields acquired in radiotherapy based on step-by-step local registration

  • Original Article
  • Published:
Medical & Biological Engineering & Computing Aims and scope Submit manuscript

Abstract

The purpose of this study is to create a new pseudo-computed tomography (CT) imaging approach under superposed ultrasound (US) deformation fields based on step-by-step local registration. Scanned CT and US 3D image datasets of three patients with postoperative cervical carcinoma were selected, including CT (CTsim) and US images (USsim) acquired during simulated positioning process and cone beam CT (CBCT) and US images for positioning verification (USpv) acquired after treatment for 10 times. Regions of interest such as urinary bladders were segmented out and accepted local registration to obtain different deformation fields. These deformation fields were successively performed according to their order and then applied to localized CT images to obtain pseudo-CT (CTps). After filtering, we obtained the final correct pseudo-CT (CTpsf). The pseudo-CT based on the mask of the whole imaging region of US images (WCTps) were acquired as control. Then, we compared CTpsf, CTps, WCTps, and CBCT in terms of their similarity in anatomical structure and differences in pseudo-CT and CTsim in terms of dosimetry. Structural similarity degree between CTpsf and CBCT was larger compared with that between CTps and WCTps. Target regions and dosages of endangered organs between CTpsf and CTsim were different under the same calculation conditions based on the Monte Carlo algorithm. Compared with the VMAT plan of CTsim, the pass rate of CTpsf in γ analysis under the standards of 2% dosage difference and 2-mm distance difference was 91.8%. The imaging quality of CTpsf was better compared with WCTps and CTps. It exhibited high similarity with CBCT in anatomical structure and had favorable application prospect in adaptive radiotherapy.

The local deformation registration is performed between the ultrasound images based on different regions of interest, and then stepwise applied to localized CT images to obtain pseudo-CT. After filtering, the corrected pseudo CT image is obtained.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Takemura A, Shoji S, Ueda S, Kurata Y, Kumano T, Takamatsu S, Suzuki M (2009) Effect of daily setup errors on individual dose distribution in conventional radiotherapy: an initial study[J]. Radiol Phys Technol. 2(2):151–158

    Article  PubMed  Google Scholar 

  2. Yan D, Ziaja E, Jaffray D, Wong J, Brabbins D, Vicini F, Martinez A (1998) The use of adaptive radiation therapy to reduce setup error: a prospective clinical study[J]. Int J Radiat Oncol Biol Phys 41(3):715–720

    Article  PubMed  CAS  Google Scholar 

  3. Yan H, Zhen X, Cerviño L, Jiang SB, Jia X (2013) Progressive cone beam CT dose control in image-guided radiation therapy[J]. Med Phys 40(6):060701

    Article  PubMed  PubMed Central  Google Scholar 

  4. Meroni S, Mongioj V, Giandini T, Bonfantini F, Cavallo A, Carrara M, Stucchi C, Cavatorta C, Pignoli E (2016) EP-1822: limits and potentialities of the use of CBCT for dose calculation in adaptive radiotherapy[J]. Radiother Oncol 119:S854–S855

    Article  Google Scholar 

  5. Pennec X, Cachier P, Ayache N (2001) Tracking brain deformations in time-sequences of 3D US images[J]. 24(4–5):801–813

  6. Andreasen D, Edmund JM, Zografos V et al (2016) Computed tomography synthesis from magnetic resonance images in the pelvis using multiple random forests and auto-context features[C]//SPIE medical imaging. International Society for Optics and Photonics 9784:978417

  7. Cao X, Yang J, Gao Y, Guo Y, Wu G, Shen D (2017) Dual-core steered non-rigid registration for multi-modal images via bi-directional image synthesis[J]. Med Image Anal 41:18–31

    Article  PubMed  PubMed Central  Google Scholar 

  8. Li M, Ballhausen H, Hegemann NS, Ganswindt U, Manapov F, Tritschler S, Roosen A, Gratzke C, Reiner M, Belka C (2015) A comparative assessment of prostate positioning guided by three-dimensional ultrasound and cone beam CT[J]. Radiat Oncol 10(1):82–83

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  9. Van d MS, Camps SM, van Elmpt WJ et al (2016) Simulation of pseudo-CT images based on deformable image registration of ultrasound images: a proof of concept for transabdominal ultrasound imaging of the prostate during radiotherapy.[J]. Med Phys 43(4):1913–1920

    Article  Google Scholar 

  10. Camps S, Meer SVD, Verhaegen F, et al. (2016) Various approaches for pseudo-CT scan creation based on ultrasound to ultrasound deformable image registration between different treatment time points for radiotherapy treatment plan adaptation in prostate cancer patients[J]. 2(3):035018

  11. Shrimali V, Anand RS, Kumar V (2009) Current trends in segmentation of medical ultrasound B-mode images: a review[J]. IETE Tech Rev 26(1):8–17

    Article  Google Scholar 

  12. Klein S, Staring M, Murphy K, Viergever MA, Pluim J (2010) Elastix: a toolbox for intensity-based medical image registration[J]. IEEE Trans Med Imaging 29(1):196–204

    Article  PubMed  Google Scholar 

  13. Shamonin DP, Bron EE, Lelieveldt BPF et al (2013) Fast parallel image registration on CPU and GPU for diagnostic classification of Alzheimer's disease[J]. Front Neuroendocrinol 7(50):50–60

    Google Scholar 

  14. Robinson D, Liu D, Steciw S, Field C, Daly H, Saibishkumar EP, Fallone G, Parliament M, Amanie J (2012) An evaluation of the clarity 3D ultrasound system for prostate localization[J]. J Appl Clin Med Phys 13(4):100–112

    Article  PubMed Central  Google Scholar 

  15. Zhang Z, Liu F, Tsui H, Lau Y, Song X (2014) A multiscale adaptive mask method for rigid intraoperative ultrasound and preoperative CT image registration[J]. Med Phys 41(10):102903

    Article  PubMed  Google Scholar 

  16. Kadir T, Zisserman A, Brady M (2004) An affine invariant salient region detector[C]// Proc European Conference on Computer Vision. 228–241

  17. Lu B, Wang H, Lin Z (2011) High order Gaussian curvature flow for image smoothing[C]// international conference on multimedia technology. IEEE:5888–5891

  18. Vik T, Kabus S, Berg J V, et al. Validation and comparison of registration methods for free-breathing 4D lung CT[J]. Proc Spie, 2008, 6914(6914):69142P-1-69142P-10

  19. Riyahi S, Choi W, Bhooshan N, Tan S, Zhang H, Lu W (2016) Comparison of registration methods for modeling pathologic response of esophageal cancer to chemoradiation therapy[J]. Med Phys 43(6):3331–3339

    Article  Google Scholar 

  20. Dong C, Wang R, Meng X et al (2014) A comparison of liver protection among 3-D conformal radiotherapy, intensity-modulated radiotherapy and RapidArc for hepatocellular carcinoma[J]. Radiat Oncol 9(1):48

    Article  Google Scholar 

  21. Robertson SP, Weiss E, Hugo GD (2012) Localization accuracy from automatic and semi-automatic rigid registration of locally-advanced lung cancer targets during image-guided radiation therapy[J]. Med Phys 39(1):330–341

    Article  PubMed  Google Scholar 

  22. Geraily G, Mirzapour M, Mahdavi SR et al (2014) Monte Carlo study on beam hardening effect of physical wedges[J]. Iran J Radiat Res 12(3):249–256

    Google Scholar 

  23. Aubry J, Pouliot J, Beaulieu L (2008) Correction of megavoltage cone-beam CT images for dose calculation in the head and neck region[J]. Med Phys 35(3):900–907

    Article  PubMed  Google Scholar 

  24. Zhang J, Zhang W, Lu J (2015) A correction algorithm for Kilovoltage cone-beam computed tomography dose calculations in cervical Cancer patients[J]. Med Phys 42(6):3242–3242

    Article  Google Scholar 

Download references

Funding

This work was supported by the Natural Science Foundation of Jiangsu Province Research of China (grant no. BK20151181).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Xinye Ni.

Ethics declarations

Ethics approval and consent to participate

The protocol of this study was approved by the medical ethics committee of Second People’s Hospital of Changzhou, Nanjing Medical University (2017-002-01).

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Not applicable.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Sun, H., Lin, T., Xie, K. et al. Imaging study of pseudo-CT images of superposed ultrasound deformation fields acquired in radiotherapy based on step-by-step local registration. Med Biol Eng Comput 57, 643–651 (2019). https://doi.org/10.1007/s11517-018-1912-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11517-018-1912-2

Keywords

Navigation