Table 7 FS-120 powder collected from the OCM verification brick

From: The Mars Science Laboratory Organic Check Material

Sample n 3-FP nmol/ 0.05cc* 1-FN nmol/ 0.05cc*
Average 1σ standard deviation Range Average 1σ standard deviation Range
blank 1 BDL    BDL   
12.5 mm depth 2 0.20 0.11 0.16 11.2 13.8 19.5
17.5 mm depth 2 0.10 0.01 0.01 14.3 16.5 23.3
24.5 mm depth 2 0.14 0.01 0.01 18.7 17.7 25.1
34.5 mm depth 2 0.11 0.01 0.02 19.2 17.9 25.4
41 mm depth 2 0.11 0.00 0.01 21.8 22.6 32.0
48 mm depth 2 0.32 0.32 0.46 21.9 24.4 34.5
50 mm depth 2 0.11 0.03 0.04 18.9 19.9 28.1
spot 1§ 2 0.16 0.11 0.16 10.1 7.2 10.1
spot 2§ 2 0.12 0.07 0.09 14.5 11.4 16.1
spot 3§ 2 0.09 0.06 0.08 17.9 22.0 31.2
spot 4§ 2 0.24 0.01 0.01 52.2 53.8 76.1
  1. §See Fig. 11 for locations on brick
  2. *Dopant concentrations are based on a calibration using FS120 doped with 3 FP
  3. BDL—below detection limits