Skip to main content
Log in

Fixed points as part of the KT-500 and KT-650 temperature calibrators

  • Published:
Measurement Techniques Aims and scope

Abstract

The results of a measurement of the indium, tin, and zinc fixed points in the KT-500 and KT-650 temperature calibrators are presented. It is shown that the error in reproducing the temperature does not exceed 0.003°C for indium and tin and 0.014°C for zinc. It is concluded that it is possible to use these points to check second-class and third-class standard resistance thermometers.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Yu. V. Shevelev and V. Ya. Cherepanov, Izmer. Tekh., No. 2, 39 (2004); Measurement Techniques, 47, No. 2, 172 (2004).

  2. Supplementary Information for the International Temperature Scale of 1990, Paviilon de Breteuil, Serves. BIPM (1990).

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to A. V. Kryukov.

Additional information

__________

Translated from Izmeritel’naya, Tekhnika, No. 6, pp. 57–59, June, 2007.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kryukov, A.V., Kurilenok, K.V., Polunin, S.P. et al. Fixed points as part of the KT-500 and KT-650 temperature calibrators. Meas Tech 50, 668–670 (2007). https://doi.org/10.1007/s11018-007-0128-x

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11018-007-0128-x

Key words

Navigation