Measurements of the photonuclear neutron yield of 15 MV medical linear accelerator

  • A. Alfuraih
  • M. P. W. Chin
  • N. M. Spyrou
Facilities

Abstract

High energy medical linear accelerators (>10 MV) are increasingly used in radiotherapy. At such high photon energies neutron production via photonuclear reactions in the heavy elements which compose the linac head is inevitable. Neutrons from linacs can contribute to an additional dose to staff, patients and the general public. Our intention is two-fold; to provide shielding against the neutron contamination and to establish the depthdose curve of thermal neutrons within human tissue, with an aim to utilise linacs in boron neuron capture therapy (BNCT). In our studies neutron measurements were undertaken, with a Varian Clinac 2100C/D linear accelerator operating at 15 MV nominal energy, by irradiating 18 cm thick 30×30 cm2 block of tissue equivalent material. Measurements were taken using indium and aluminum activation foil at the centre of the block. Our results show that by leaving the linac jaws open neutron production is increased compared to the case when these are shut, for one minute exposure at 400MU. In this work we present a comparison between our results and existing literature and attempt to explore some sharp contrasts.

Keywords

Neutron Yield Boron Neutron Capture Therapy Neutron Fluence Photonuclear Reaction Neutron Production 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    NCRP, Neutron Contamination from Medical Electron Accelerators, NCRP Report 79, 1984.Google Scholar
  2. 2.
    J. S. Laughlin, A. Reid, L. Zeitz, J. Ding, Proc. Conf. on Neutrons from Electron Medical Accelerators, H. T. Heaton II and R. Jacobs (Eds), NBS Special Publication 554, 1979, p. 1.Google Scholar
  3. 3.
    R. C. Mccall, W. P. Swanson, Neutron sources and their characteristics, Proc. Conf. on Neutrons from Electron Medical Accelerators, H. T. Heaton II and R. Jacobs (Eds), NBS Special Publication 554, 1979, p. 75.Google Scholar
  4. 4.
    R. C. Mccall, SALC Report SALC-PUB 4480, Stanford, California, 1987.Google Scholar
  5. 5.
    F. Bush, Br. J. Radiol., 14 (1946) 1914.Google Scholar
  6. 6.
    H. M. Garnica-Garza, Phys. Med. Biol., 50 (2005) 531.CrossRefGoogle Scholar
  7. 7.
    H. S. Kim, Y. H. Park, B. C. Koo, J. W. Kwon, J.-S. Lee, H. S. Choi, Radiat. Prot. Dosim., 123 (2007) 323.CrossRefGoogle Scholar
  8. 8.
    X. S. Mao, K. R. Kase, J. C. Liu, W. R. Nelson, J. H. Kleck, S. Johnsen, Health Phys., 72 (1997) 524.CrossRefGoogle Scholar