Journal of Medical Systems

, Volume 35, Issue 5, pp 779–786 | Cite as

A Study of Heating Duration and Scanning Path in Focused Ultrasound Surgery

  • Dehui Li
  • Guofeng Shen
  • Hui Luo
  • Jinfeng Bai
  • Yazhu Chen
Original Paper


A conventional method of avoiding normal tissue overheating during focused ultrasound surgery (FUS) is to apply equal heating duration for each sonication in between cooling intervals. However, this method is time-consuming and expensive. A novel method with unequal heating duration in different scanning paths without cooling intervals was investigated in this paper. This method was compared with the conventional method through the ablation of a 10 × 10 cm2 target. The simulation results indicated that the new method was able to reduce treatment time by more than 50%, with higher than 95% coverage index, and more uniform thermal dose distribution. The method was further verified through the ablation of a circular lesion. Ex vivo experiments were also performed to confirm the simulation results. Both simulation and experimental results have proven that the new method can increase heating efficiency, and are a promising new approach in FUS.


Focused ultrasound surgery Unequal heating duration Scanning path Heating efficiency 



This study was supported by the National Natural Science Foundation of China (No. 30800246), the Ph.D. Programs Foundation of Ministry of Education of China (No. 20070248082), and Shanghai Key Technologies R&D Program of China (No. 09441900500).


  1. 1.
    Wu, F., Wang, Z. B., Chen, W. Z., Zou, J. Z., Bai, J., Zhu, H., Li, K. Q., Xie, F. L., Jin, C. B., Su, H. B., and Gao, G. W., Extracorporeal focused ultrasound surgery for treatment of human solid carcinomas: early Chinese clinical experience. Ultrasound Med. Biol 30(2):245–260, 2004.CrossRefGoogle Scholar
  2. 2.
    Illing, R. O., Kennedy, J. E., Wu, F., ter Haar, G. R., Protheroe, A. S., Friend, P. J., Gleeson, F. V., Cranston, D. W., Phillips, R. R., and Middleton, M. R., The safety and feasibility of extracorporeal high-intensity focused ultrasound (HIFU) for the treatment of liver and kidney tumours in a Western population. Br. J. Cancer 93(8):890–895, 2005.CrossRefGoogle Scholar
  3. 3.
    Kennedy, J. E., High-intensity focused ultrasound in the treatment of solid tumours. Nat. Rev. Cancer 5(4):321–327, 2005.CrossRefGoogle Scholar
  4. 4.
    Damianou, C., and Hynynen, K., Focal spacing and near-field heating during pulsed high temperature ultrasound therapy. Ultrasound Med. Biol. 19(9):777–787, 1993.CrossRefGoogle Scholar
  5. 5.
    McDannold, N., Jolesz, F., and Hynynen, K., Determination of the Optimal Delay between Sonications during Focused Ultrasound Surgery in Rabbits by Using MR Imaging to Monitor Thermal Buildup in Vivo. Radiology 211(2):419–426, 1999.Google Scholar
  6. 6.
    Fan, X., and Hynynen, K., Ultrasound surgery using multiple sonications - Treatment time considerations. Ultrasound Med. Biol. 22(4):471–482, 1996.CrossRefGoogle Scholar
  7. 7.
    Daum, D. R., and Hynynen, K., A 256-element ultrasonic phased array system for the treatment of large volumes of deep seated tissue. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 46(5):1254–1268, 1999.CrossRefGoogle Scholar
  8. 8.
    Wan, H., VanBaren, P., Ebbini, E. S., and Cain, C. A., Ultrasound surgery: comparison of strategies using phased array systems. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 43(6):1085–1098, 1996.CrossRefGoogle Scholar
  9. 9.
    Fan, X., and Hynynen, K., Control of the necrosed tissue volume during noninvasive ultrasound surgery using a 16-element phased array. Med. Phys. 22,(3):297–306, 1995.CrossRefGoogle Scholar
  10. 10.
    Cheng, T. Y., Ju, K. C., Ho, C. S., Chen, Y. Y., and Chang, H., Split-focused ultrasound transducer with multidirectional heating for breast tumor thermal surgery. Med. Phys. 35(4):1387–1397, 2008.CrossRefGoogle Scholar
  11. 11.
    Daum, D. R., and Hynynen, K., Thermal dose optimization via temporal switching in ultrasound surgery. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 45(1):208–215, 1998.CrossRefGoogle Scholar
  12. 12.
    Liu, H. L., Shih, T. C., Chen, W. S., and Ju, K. C., A novel strategy to increase heating efficiency in a split-focus ultrasound phased array. Med. Phys. 34(7):2957–2967, 2007.CrossRefGoogle Scholar
  13. 13.
    Leibel, S. A., and Philips, T. L., Textbook of radiation oncology. Saunders W. B, Philadelphia, pp. 1–1674, 2004.Google Scholar
  14. 14.
    Liu, H. L., Lin, W. L., and Chen, Y. Y., A fast and conformal heating scheme for producing large thermal lesions using a 2D ultrasound phased array. Int. J. Hyperthermia. 23(1):69–82, 2007.CrossRefMathSciNetGoogle Scholar
  15. 15.
    Caulfield, R. E., Yin, X., Juste, J., and Hynynen, K., A novel phase assignment protocol and driving system for a high-density focused ultrasound array. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 54(4):793–801, 2007.CrossRefGoogle Scholar
  16. 16.
    O’Neil, H. T., Theory of focusing radiators. J. Acoust. Soc. Am. 21(5):516–526, 1949.CrossRefGoogle Scholar
  17. 17.
    Ebbini, E. S., and Cain, C. A., Multiple-focus ultrasound phased-array pattern synthesis: Optimal driving-signal distributions for hyperthermia. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 36(5):540–548, 1989.CrossRefGoogle Scholar
  18. 18.
    Pennes, H. H., Analysis of tissue and arterial blood temperatures in the resting human forearm. J. Appl. Physiol. 1(2):93–122, 1948.Google Scholar
  19. 19.
    Zhang, C., and Chen, Y., Investigation of a spherical-section ultrasound phased array for hepatic ablation. J. Zhejiang Univ. Sci. 8(8):1237–1245, 2007.CrossRefMATHGoogle Scholar
  20. 20.
    Sapareto, S. A., and Dewey, W. C., Thermal dose determination in cancer therapy. Int. J. Radiat. Oncol. Biol. Phys. 10(6):787–800, 1984.CrossRefGoogle Scholar
  21. 21.
    Zhou Y., Kargl S., Kim K., and Hwang J. H., “Comparison of pathway in high intensity focused ultrasound lesion production”, POMA, vol.2, 020002, New Orleans, Louisiana, 2008.Google Scholar

Copyright information

© Springer Science+Business Media, LLC 2010

Authors and Affiliations

  • Dehui Li
    • 1
    • 2
  • Guofeng Shen
    • 1
    • 2
  • Hui Luo
    • 1
    • 2
  • Jinfeng Bai
    • 1
    • 2
  • Yazhu Chen
    • 1
    • 2
  1. 1.Biomedical Instrument InstituteShanghai Jiao Tong UniversityShanghaiChina
  2. 2.Med-X Research InstituteShanghai Jiao Tong UniversityShanghaiChina

Personalised recommendations