Introduction

The neutron activation method is one of the standard methods for the plasma diagnostics at tokamaks. The activation detectors (targets of chosen isotopes) are placed in exposition locations, where the neutron yield from fusion plasma (D–D or D–T) is measured. After the activation the samples are delivered with a pneumatic transport system to a gamma-ray spectrometer. A deconvolution of the spectra allows calculating the primary neutron yield in given points.

A valuable supplement to the conventional neutron activation method is a delayed neutron activation method (e.g. at JET tokamak [1]) which uses fissionable isotopes as activated targets. Fissionable isotopes as 235U, 238U or 232Th, irradiated with neutrons, emit prompt neutrons (~ 99%) and later the so-called delayed neutrons (~ 1%) from the fission products which decay with the beta-emission. The decay chain can be generally written as

$${}_{{\text{Z}}}^{{\text{A}}} {\text{X}}_{{\text{N}}} \mathop{\longrightarrow}\limits^{\upbeta}{}_{{{\text{Z}} + 1}}^{{\text{A}}} {\text{Y}}_{{{\text{N}} - 1}} \to {}_{{{\text{Z}} + 1}}^{{\text{A}}} {\text{Y}}_{{{\text{N}} - 2}}^{{\prime }} + {\text{n}}$$

where the X nucleus is the delayed neutron precursor and Y is the delayed neutron emitter. The delay is determined by the half-life time of the precursor. Usually, six groups of the delayed neutrons (originated by different fission products) are distinguished [2] and the total decay in time is described by

$$S(t) = r_{\text{f}} \;\sum\limits_{i = 1}^{6} {\nu_{{{\text{d}}i}} \,\lambda_{{{\text{d}}i}} \,\exp ( - \lambda_{{{\text{d}}i}} t)}$$

where rf is the number of fission reactions in the sample, νdi is the yield of the ith delayed neutron group, λdi is the decay constant of this group. The corresponding half-life time can range in some cases up to 1 or 2 min. For example, the 235U fission products being the precursors for the delayed neutrons, have the average half-life times from ~ 0.2 to ~ 56 s. The average energy of the delayed neutrons is in the range of (300–600 keV) [2, 3].

Conceptual Design of the Device and Modelling of the Radiation Transport

A system of the detection of delayed neutrons must have a high sensitivity for delayed neutrons escaping from the fissionable sample and a low sensitivity for accompanied γ radiation which can enter the counters. In our design a number of thermal neutron detectors (3He type), are placed in a neutron moderator, and surrounded a central cavity for the sample (Fig. 1). The cavity is surrounded by a bismuth layer which prevents counters from γ radiation. The next polyethylene layer which contains 3He detectors is used to slow down delayed neutrons to thermal energies. The external layers consisting of polyethylene, cadmium, and a B4C, prevent outside background neutrons entering the detectors.

Fig. 1
figure 1

MCNP model of the DET-12 device (vertical and horizontal cross sections) [5]. a Vertical section. b Horizontal section

Detectors are connected to separate electronic lines (H.V. supply, preamplifier, amplifier, counting electronics) and finally to a control PC unit with the data acquisition and analysing systems.

All elements of the experimental set-up have been optimized by detailed Monte Carlo calculations, i.e. type of the fissionable material, geometry and size of the target sample, number and location of 3He detectors, shape and sizes of all covers. The set-up has been modelled using MCNP numerical code for radiation transport [4] and details of calculations are presented in the reports [3, 5, 6].

DET-12 device has been optimized for measurement of delayed neutrons produced during fission reactions of 235U induced by thermal and fast neutrons and of 238U and 232Th induced by fast neutrons. These two last nuclides are characterized by energy threshold for neutron-induced fission reaction (about 1 MeV for 238U and 0.5 MeV for 232Th) which is a valuable feature for the detection of fast neutrons.

Size and shape of the target sample were optimized taking into account the typical shape of the pneumatic system applied at JET [1] and a magnitude of neutron current through the sample surface. The cylindrical sample of height 1.8 cm and diameter 1.8 cm has been chosen.

Twelve 3He neutron detectors: 1 diameter, 30 cm length (25 cm active), 5 atm. pressure have been chosen as the optimum number. The total efficiency of detection, calculated as a sum from all 12 detectors has been evaluated from the MCNP simulations on 24%. The simulations were performed either for all 16 detectors or for 12 detectors, i.e. when the detectors in corners are removed (Fig. 1b). The contribution of these detectors is smaller than of the others. The efficiency decreases less than it results only from the ratio of 12/16. As expected MCNP calculations confirmed, the contribution of the corner detectors is smaller than of the others. A decrease of the efficiency, ratio ε(N12)/ε(N16) ≈ 0.87, is smaller than that which would result from the ratio of numbers of the detectors, 12/16 = 0.75.

Final dimensions for the optimized DET-12 device are:

  • Square horizontal size: 58 cm × 58 cm, height 74 cm

  • Central hole for the pneumatic transport: 6 cm × 6 cm

  • Consecutive layers from the hole towards outside (see Fig. 1): (6) bismuth: 2 cm, (4) polyethylene (moderator of delayed neutrons): 12 cm, (3) B4C (absorber): 3.8 cm, (2) Cadmium (absorber): 0.2 cm, (1) Polyethylene (external protection): 8 cm.

Electronic System for Neutron Detection and Data Acquisition

The expected intensity of counts in the 3He detectors will be rather low. Therefore, it has been decided to use a system in which three 3He detectors are coupled to one preamplifier. Then the preamplifiers in a usual way are connected to spectroscopy amplifiers. Four lines are connected with a digitizer card in a PC. A complete scheme of the system is shown in Fig. 2 [5]. All the electronic blocks were connected (including a preparation of dedicated cables for the groups of detectors) after preparatory operation tests and H.V. and amplification adjustments were made.

Fig. 2
figure 2

Neutron detection and data acquisition system [5]

Detection Efficiency of DET-12 Device [5]

Detection efficiency of the DET-12 device was tested using the 252Cf source of activity = 8.777 kBq placed in the middle of the measuring channel of DET-12, i.e. in the position presumed for the activated sample to be measured. Spontaneous fission of 252Cf brings the neutron emission with the coefficient of 11.627% [7]. This results in the neutron emission rate of 1020 cps. The obtained average count rate of the DET-12 device is 191.4 ± 0.2 cps. Thus, the neutron detection efficiency can be assessed at 18.8%. Comparing the experimental results with those obtained from the Monte Carlo simulations the difference of the estimations is 20%. Probable reason may be inaccurate knowledge of the active volume of the detector and its gas composition, which were used for numerical calculations. The Monte Carlo simulations do not include neither dead time nor efficiency of the detectors (problem of the detector active volume). Real composition of the gas in the detectors may be different from the one assumed in the MC simulations—manufacturer does not release even principal information and additionally some admixtures, like CO2 or Ar as a quench gas, (below 1%) are generally known to be present.

Summary

Measurement of delayed neutrons which are emitted from fissionable samples irradiated in a neutron field originated from thermonuclear plasma is one of types of the neutron activation method applicable for a neutron diagnostic of fusion plasma.

The experimental device (DET-12) has been designed, optimized and constructed for the measurements of delayed neutrons from the neutron induced fission reactions. A small amount of delayed neutrons generated during the fission reactions dedicates this method to the measurement of very high neutron yields generated in large fusion devices. Analytical methods, MCNP calculations and experiments indicate that DET-12 device can works with good statistics with minimal neutron fluxes about 108–109 n/cm2s.