Abstract

The purpose of this study was to evaluate the push-out load-bearing capacity of threaded fiber-reinforced composite (FRC) devices for use as bone-anchored devices. The purpose was also to evaluate the possibility to use bioactive glass (BAG) granules on the experimental FRC devices in terms the mechanical behavior.

Three experimental FRC devices (n = 15) were fabricated for the study: (a) threaded device with smooth surface; (b) threaded device with BAG granules (S53P4, Vivoxid Ltd, Turku, Finland) and supplementary retention grooves, and (c) unthreaded device with BAG granules. Threaded titanium devices were used as controls. The FRC devices were prepared from a light-polymerized dimethacrylate resin reinforced with preimpregnated unidirectional and bidirectional E-glass fibers (EverStick, StickTech Ltd, Turku, Finland). Experimental and control devices were embedded into dental plaster to simulate bone before the mechanical push-out test was carried out. ANOVA and Weibull analysis were used for the statistical evaluation. Threaded FRC devices had significantly higher push-out strength than the threaded titanium device (p < .001). The push-out forces exceeding 2,500 N were measured for threaded FRC devices with supplementary grooves and BAG coating. No thread failures were observed in any FRC devices. The unthreaded FRC devices with BAG lost 70% of glass particles during the test, while no BAG particles were lost from threaded FRC devices. It can be concluded that threaded FRC devices can withstand high push-out forces in the dental plaster without a risk of thread failure under physiological load.

Keywords

Bioactive Glass Dental Implant Weibull Modulus Bite Force Weibull Analysis 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

References

  1. 1.
    R. ADELL, U. LEKHOLM, B. POCKLER and P. I. BRÅNEMARK, Int. J. Oral. Surg. 6 (1981) 387CrossRefGoogle Scholar
  2. 2.
    J. A. EKELUND, L. W. LINDQUIST, G. E. CARLSSON and T. JEMT, Int. J. Prosthodont. 16 (2003) 602Google Scholar
  3. 3.
    T. JEMT, U. LEKHOLM and R. ADELL, Int. J. Oral Maxillofac. Implants 4 (1980) 211Google Scholar
  4. 4.
    J. E. LEMONS, Implant Dent. 7 (1998) 351CrossRefGoogle Scholar
  5. 5.
    T. ALBREKTSSON, P. I. BRÅNEMARK, H. A. HANSSON and J. A. LINDSTOM, Acta Orthop. Scand. 52 (1981) 155CrossRefGoogle Scholar
  6. 6.
    B. FRIBERG, T. JEMT and U. LEKHOLM, Int. J. Oral Maxillofac. Implants. 6 (1991)142Google Scholar
  7. 7.
    S. YAMADA, S. SEKIYA, K. YAMANOUCHI, H. KITAMURA, M. OHSHIMA and T. SATO, Bull. Tokyo Dent. Coll. 30 (1989)187Google Scholar
  8. 8.
    O. BAHAT, Int. J. Oral Maxillofac. Implants. 7 (1992) 459Google Scholar
  9. 9.
    M. NEVINS and B. LANGER, Int. J. Oral Maxillofac. Implants. 8 (1993) 428Google Scholar
  10. 10.
    R. J. LAZZARA, S. S. PORTER, T. TESTORI, J. GALANTE and L. ZETTERQVIST, J. Esthet. Dent. 10 (1998) 280CrossRefGoogle Scholar
  11. 11.
    U. GRUNDER, N. BOITEL, M. IMOBERDORF, K. MEYENBERG, T. MEIER and C. ANDREONI, Compend Contin. Educ. Dent. 20 (1999) 628Google Scholar
  12. 12.
    B. CHEHROUDI, T. R. GOULD and D. M. BRUNETTE, J. Biomed. Mater. Res. 26 (1992) 493CrossRefGoogle Scholar
  13. 13.
    B. CHEHROUDI, D. MCDONNELL and D. M. BRUNETTE, J. Biomed. Mater. Res. 34 (1997) 279CrossRefGoogle Scholar
  14. 14.
    C. HALLGREN, H. REIMERS, J. GOLD and A. WENNERBERG, J. Biomed. Mater. Res. 57 (2001) 485CrossRefGoogle Scholar
  15. 15.
    J. LI, H. LIAO, B. FARTASH, L. HERMANSSON and T. JOHNSSON, Biomaterials. 18 (1997) 691CrossRefGoogle Scholar
  16. 16.
    Y. T. SUL, C. B. JOHANSSON, Y. JEONG, K. ROSER, A. WENNERBERG and T. ALBREKTSSON, J. Mater. Sci. Mater. Med. 12 (2001) 1025CrossRefGoogle Scholar
  17. 17.
    L. VIDYASAGAR and P. APSE, Baltic Dent. Maxillofac. J. 6 (2004) 51Google Scholar
  18. 18.
    J. T. KRAUSER, C. BONER and N. BONER, Cah Prothese. 71 (1990) 56Google Scholar
  19. 19.
    Y. L. CHANG, D. LEW, J. B. PARK and J. C. KELLER, J. Oral Maxillofac. Surg. 57 (1999) 1096CrossRefGoogle Scholar
  20. 20.
    C. K. CHANG, J. S. WU, D. L. MAO and C. X. DING, J. Biomed. Mater. Res. 56 (2001) 17CrossRefGoogle Scholar
  21. 21.
    J. SCHROOTEN and J. A. HELSEN, Biomaterials. 21 (2000) 1461CrossRefGoogle Scholar
  22. 22.
    N. MORITZ, S. ROSSI, E. VEDEL, T. TIRRI, H. YLANEN, H. ARO and T. NÄRHI, J. Mater. Sci. Mater. Med. 15 (2004) 795CrossRefGoogle Scholar
  23. 23.
    C. E. MISCH, Z. QU and M. W. BIDEZ, J. Oral Maxillofac. Surg. 57 (1999) 700CrossRefGoogle Scholar
  24. 24.
    J. E. LEMONS, Implant Dent. 7 (1998) 351CrossRefGoogle Scholar
  25. 25.
    J. B. BRUNSKI, Adv. Dent. Res. 13 (1999) 99Google Scholar
  26. 26.
    A. M. Ballo, L. Lassila, T. Närhi and P. K. Vallittu, J. Contemp Dent Pract. 2007 (In press)Google Scholar
  27. 27.
    P. K. Vallittu and J. Assila, Oral Rehabil. 19 (1992) 225Google Scholar
  28. 28.
    P. K. VALLITTU, J. Oral Rehabil. 25 (1998) 100CrossRefGoogle Scholar
  29. 29.
    K. CHUNG, T. LIN, and F. WANG, J. Oral Rehabil. 25 (1998) 214CrossRefGoogle Scholar
  30. 30.
    M. A. FREILICH, J. P. DUNCAN, E. K. ALARCON, and K. A. ECKROTE, J. Prosthet. Dent. 88 (2002) 449CrossRefGoogle Scholar
  31. 31.
    M. BEHR, M. ROSENTRITT, R. LANG, and G. HANDEL, Clin. Oral. Implants Res. 12 (2001) 174CrossRefGoogle Scholar
  32. 32.
    J. P. DUNCAN, M. A. FREILICH and C. J. LATVIS, J. Prosthet. Dent. 84 (2000) 200CrossRefGoogle Scholar
  33. 33.
    S. L. EVANS, C. M. HUNT and S. AHUJA, J. Mater. Sci-Mater. Med. 13 (2002) 1143CrossRefGoogle Scholar
  34. 34.
    J. PALUSSIERE. et al., Eur spine J. 14 (2005) 982CrossRefGoogle Scholar
  35. 35.
    G. S. ANDREASSEN, P. R. HOINESS, I. SKRAAMM, O. GRANLUND and L. ENGEBRETSEN, Arch. Orthop. Trauma Surg. 124 (2004) 161CrossRefGoogle Scholar
  36. 36.
    N. MORITZ, S. ROSSI, E. VEDEL, T. TIRRIT, H. YLANEN, H. ARO and T. NÄRHI, J. Mater. Sci. Mater. Med. 15 (2004) 795CrossRefGoogle Scholar
  37. 37.
    L. B. LUM and J. F. OSIER, J. Oral Implantol. 18 (1992) 349Google Scholar
  38. 38.
    T. ALBREKTSSON, A. WENNERBERG, Int. J. Prosthodont. 17 (2004) 536Google Scholar
  39. 39.
    A. PIATTELLI, A. SCARANO, L. FAVERO, G. IEZZI, G. PETRONE, G. A. FAVERO, J. Periodontol. 74 (2003) 385CrossRefGoogle Scholar
  40. 40.
    R. MATTILA, M. PUSKA, L. LASSIILA and P. K. VALLITTU, J. Mater. Sci.: Mater. Med. 41 (2006) 4321Google Scholar
  41. 41.
    H. Yuehuei, “Mechanical properties of Bone. Mechanical Testing of Bone and the bone-Implant Interface” (CRC Press; 2000) p.43Google Scholar
  42. 42.
    D. BUSER, T. NYDEGGER, T. OXLAND, D. L. COCHRAN, R. K. SCHENK, H. P. HIRT, D. SNETIVY and L. P. NOLTE, J. Biomed. Mater. Res. 45 (1999) 75CrossRefGoogle Scholar
  43. 43.
    T. M. Van EIJDEN, Arch. Oral Biol. 36 (1991) 535CrossRefGoogle Scholar

Copyright information

© Springer Science+Business Media, LLC 2007

Authors and Affiliations

  • Ahmed Mansour Ballo
    • 1
  • Lippo V. Lassila
    • 1
  • Pekka K. Vallittu
    • 1
  • Timo O. Närhi
    • 1
  1. 1.Department of Prosthetic Dentistry and Biomaterials Science, Institute of DentistryUniversity of TurkuTurkuFinland