Dialytic interval and the timing of electrocardiographic screening for subcutaneous cardioverter-defibrillator placement in chronic hemodialysis patients

  • Published:
Journal of Interventional Cardiac Electrophysiology Aims and scope Submit manuscript

    We’re sorry, something doesn't seem to be working properly.

    Please try refreshing the page. If that doesn't work, please contact support so we can address the problem.

Abstract

Background

Hemodialysis (HD) patients have a high risk of sudden death but limited vascular access and high complication rates from transvenous implantable cardioverter-defibrillators (ICDs). Subcutaneous ICDs (S-ICD) may be an alternative, but dynamic ECG changes may result in inappropriate shocks. This study aims to define the screen failure rate for S-ICD in patients pre- and post-HD.

Methods

ECG waveforms were obtained using electrodes mimicking the S-ICD sensing vectors in an unselected test group of chronic HD patients and a control group of ICD-eligible non-dialysis patients. Participants passed screening if their QRS and T-waves fit within the screening template in supine and standing positions in any lead. Test group participants were screened before and after HD and control group patients were screened at two separate time points. HD patients were stratified into the four following groups: (A) passed screening before and after HD, (B) failed screening before but passed after HD, (C) passed screening before but failed after HD, and (D) failed screening before and after HD. Patients in group A passed the screening for ICD implantation, and patients in groups B, C, and D failed the screening for ICD implantation. Control patients were similarly classified by pass/fail status at the two assessment points.

Results

Of the 76 patients enrolled, 51 were HD patients and 25 were controls. Of the 51 HD patients, 43 (84%) were in group A, four participants (8%) were in group B, one (2%) was in group C, and three participants (6%) were in group D. There were no differences in any of the clinical or demographic variables between the pass and fail test HD groups. None of the 25 controls failed the screening at either time point (p = 0.047 vs HD patients).

Conclusions

Overall, HD patients were more likely to fail S-ICD screening compared to non-HD patients (16 v 0%, p = 0.047) and are more likely to do so prior to HD. Patients on HD should be screened at multiple time points around the dialytic interval to reduce the risk of inappropriate shocks.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Collins AJ, Foley RN, Herzog C, Chavers B, Gilbertson D, Herzog C, et al. US Renal Data System 2012 Annual Data Report. Am J Kidney Dis. 2013;61:A7, e1–476.

    Article  Google Scholar 

  2. Eknoyan G, Beck GJ, Cheung AK, Daugirdas JT, Greene T, Kusek JW, et al. Effect of dialysis dose and membrane flux in maintenance hemodialysis. N Engl J Med. 2002;347:2010–9.

    Article  PubMed  Google Scholar 

  3. Foley RN, Parfrey PS, Sarnak MJ. Clinical epidemiology of cardiovascular disease in chronic renal disease. Am J Kidney Dis. 1998;32:S112–9.

    Article  PubMed  CAS  Google Scholar 

  4. Bleyer AJ, Hartman J, Brannon PC, Reeves-Daniel A, Satko SG, Russell G. Characteristics of sudden death in hemodialysis patients. Kidney Int. 2006;69:2268–73.

    Article  PubMed  CAS  Google Scholar 

  5. Charytan DM, Patrick AR, Liu J, Setoguchi S, Herzog CA, Brookhart MA, et al. Trends in the use and outcomes of implantable cardioverter-defibrillators in patients undergoing dialysis in the United States. Am J Kidney Dis. 2007;58:409–17.

    Article  Google Scholar 

  6. Dasgupta A, Montalvo J, Medendorp S, Lloyd-Jones DM, Ghossein C, Goldberger J, et al. Increased complication rates of cardiac rhythm management devices in ESRD patients. Am J Kidney Dis. 2007;49:656–63.

    Article  PubMed  Google Scholar 

  7. Aggarwal A, Wang Y, Rumsfeld JS, Curtis JP, Heidenreich PA, National Cardiovascular Data R. Clinical characteristics and in-hospital outcome of patients with end-stage renal disease on dialysis referred for implantable cardioverter-defibrillator implantation. Heart Rhythm. 2009;6:1565–71.

    Article  PubMed  Google Scholar 

  8. Hreybe H, Razak E, Saba S. Effect of end-stage renal failure and hemodialysis on mortality rates in implantable cardioverter-defibrillator recipients. Pacing Clin Electrophysiol. 2007;30:1091–5.

    Article  PubMed  Google Scholar 

  9. Charytan DM, Patrick AR, Liu J, Setoguchi S, Herzog CA, Brookhart MA, et al. Trends in the use and outcomes of implantable cardioverter-defibrillators in patients undergoing dialysis in the United States. Am J Kidney Dis. 2011;58:409–17.

    Article  PubMed  Google Scholar 

  10. Bardy GH, Smith WM, Hood MA, Crozier IG, Melton IC, Jordaens L, et al. An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med. 2010;363:36–44.

    Article  PubMed  CAS  Google Scholar 

  11. Weiss R, Knight BP, Gold MR, Leon AR, Herre JM, Hood M, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation. 2013;128:944–53.

    Article  PubMed  Google Scholar 

  12. El-Chami MF, Levy M, Kelli HM, Casey M, Hoskins MH, Goyal A, Langberg JJ, Patel A, Delurgio D, Lloyd MS, Leon AR, Merchant FM. Outcome of subcutaneous implantable cardioverter defibrillator implantation in patients with end-stage renal disease on dialysis. J Cardiovasc Electrophysiol 2015.

  13. Herzog CA, Mangrum JM, Passman R. Sudden cardiac death and dialysis patients. Semin Dial. 2008;21:300–7.

    Article  PubMed  Google Scholar 

  14. Poulikakos DM, Malik MPM. Challenges of ECG monitoring and ECG interpretations in dialysis units. J Electrocardiol. 2016;49:855–9.

    Article  PubMed  Google Scholar 

  15. Groh CA, Sharma S, Pelchovitz DJ, Bhave PD, Rhyner J, Verma N, et al. Use of an electrocardiographic screening tool to determine candidacy for a subcutaneous implantable cardioverter-defibrillator. Heart Rhythm. 2014;11:1361–6.

    Article  PubMed  Google Scholar 

  16. Pun PH, Hellkamp AS, Sanders GD, Middleton JP, Hammill SC, Al-Khalidi HR, et al. Primary prevention implantable cardioverter defibrillators in end-stage kidney disease patients on dialysis: a matched cohort study. Nephrol Dial Transplant. 2015;30:829–35.

    Article  PubMed  Google Scholar 

  17. Olde Nordkamp LR, Warnaars JL, Kooiman KM, de Groot JR, Rosenmoller BR, Wilde AA, et al. Which patients are not suitable for a subcutaneous ICD: incidence and predictors of failed QRS-T-wave morphology screening. J Cardiovasc Electrophysiol. 2014;25:494–9.

    Article  PubMed  Google Scholar 

  18. Randles DA, Hawkins NM, Shaw M, Patwala AY, Pettit SJ, Wright DJ. How many patients fulfil the surface electrocardiogram criteria for subcutaneous implantable cardioverter-defibrillator implantation? Europace. 2014;16:1015–21.

    Article  PubMed  Google Scholar 

  19. Wu VC, Lin LY, Wu KD. QT interval dispersion in dialysis patients. Nephrology (Carlton). 2005;10:109–12.

    Article  Google Scholar 

  20. Ansari A, Kaupke CJ, Vaziri ND, Miller R, Barbari A. Cardiac pathology in patients with end-stage renal disease maintained on hemodialysis. Int J Artif Organs. 1993;16:31–6.

    Article  PubMed  CAS  Google Scholar 

  21. Dillon JJ, DeSimone CV, Sapir Y, Somers VK, Dugan JL, Bruce CJ, et al. Noninvasive potassium determination using a mathematically processed ECG: proof of concept for a novel “blood-less, blood test”. J Electrocardiol. 2015;48:12–8.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Frenova Renal Research.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Rod Passman.

Ethics declarations

Conflict of interest

Rod Passman: Honoraria, Medtronic, Biotronik; royalties, UpToDate. All other authors declare that they have no conflict of interest.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

You, C., Sharma, S., Bavishi, A. et al. Dialytic interval and the timing of electrocardiographic screening for subcutaneous cardioverter-defibrillator placement in chronic hemodialysis patients. J Interv Card Electrophysiol 52, 179–184 (2018). https://doi.org/10.1007/s10840-018-0343-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10840-018-0343-1

Keywords

Navigation