Abstract
Background
Hemodialysis (HD) patients have a high risk of sudden death but limited vascular access and high complication rates from transvenous implantable cardioverter-defibrillators (ICDs). Subcutaneous ICDs (S-ICD) may be an alternative, but dynamic ECG changes may result in inappropriate shocks. This study aims to define the screen failure rate for S-ICD in patients pre- and post-HD.
Methods
ECG waveforms were obtained using electrodes mimicking the S-ICD sensing vectors in an unselected test group of chronic HD patients and a control group of ICD-eligible non-dialysis patients. Participants passed screening if their QRS and T-waves fit within the screening template in supine and standing positions in any lead. Test group participants were screened before and after HD and control group patients were screened at two separate time points. HD patients were stratified into the four following groups: (A) passed screening before and after HD, (B) failed screening before but passed after HD, (C) passed screening before but failed after HD, and (D) failed screening before and after HD. Patients in group A passed the screening for ICD implantation, and patients in groups B, C, and D failed the screening for ICD implantation. Control patients were similarly classified by pass/fail status at the two assessment points.
Results
Of the 76 patients enrolled, 51 were HD patients and 25 were controls. Of the 51 HD patients, 43 (84%) were in group A, four participants (8%) were in group B, one (2%) was in group C, and three participants (6%) were in group D. There were no differences in any of the clinical or demographic variables between the pass and fail test HD groups. None of the 25 controls failed the screening at either time point (p = 0.047 vs HD patients).
Conclusions
Overall, HD patients were more likely to fail S-ICD screening compared to non-HD patients (16 v 0%, p = 0.047) and are more likely to do so prior to HD. Patients on HD should be screened at multiple time points around the dialytic interval to reduce the risk of inappropriate shocks.
Similar content being viewed by others
References
Collins AJ, Foley RN, Herzog C, Chavers B, Gilbertson D, Herzog C, et al. US Renal Data System 2012 Annual Data Report. Am J Kidney Dis. 2013;61:A7, e1–476.
Eknoyan G, Beck GJ, Cheung AK, Daugirdas JT, Greene T, Kusek JW, et al. Effect of dialysis dose and membrane flux in maintenance hemodialysis. N Engl J Med. 2002;347:2010–9.
Foley RN, Parfrey PS, Sarnak MJ. Clinical epidemiology of cardiovascular disease in chronic renal disease. Am J Kidney Dis. 1998;32:S112–9.
Bleyer AJ, Hartman J, Brannon PC, Reeves-Daniel A, Satko SG, Russell G. Characteristics of sudden death in hemodialysis patients. Kidney Int. 2006;69:2268–73.
Charytan DM, Patrick AR, Liu J, Setoguchi S, Herzog CA, Brookhart MA, et al. Trends in the use and outcomes of implantable cardioverter-defibrillators in patients undergoing dialysis in the United States. Am J Kidney Dis. 2007;58:409–17.
Dasgupta A, Montalvo J, Medendorp S, Lloyd-Jones DM, Ghossein C, Goldberger J, et al. Increased complication rates of cardiac rhythm management devices in ESRD patients. Am J Kidney Dis. 2007;49:656–63.
Aggarwal A, Wang Y, Rumsfeld JS, Curtis JP, Heidenreich PA, National Cardiovascular Data R. Clinical characteristics and in-hospital outcome of patients with end-stage renal disease on dialysis referred for implantable cardioverter-defibrillator implantation. Heart Rhythm. 2009;6:1565–71.
Hreybe H, Razak E, Saba S. Effect of end-stage renal failure and hemodialysis on mortality rates in implantable cardioverter-defibrillator recipients. Pacing Clin Electrophysiol. 2007;30:1091–5.
Charytan DM, Patrick AR, Liu J, Setoguchi S, Herzog CA, Brookhart MA, et al. Trends in the use and outcomes of implantable cardioverter-defibrillators in patients undergoing dialysis in the United States. Am J Kidney Dis. 2011;58:409–17.
Bardy GH, Smith WM, Hood MA, Crozier IG, Melton IC, Jordaens L, et al. An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med. 2010;363:36–44.
Weiss R, Knight BP, Gold MR, Leon AR, Herre JM, Hood M, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation. 2013;128:944–53.
El-Chami MF, Levy M, Kelli HM, Casey M, Hoskins MH, Goyal A, Langberg JJ, Patel A, Delurgio D, Lloyd MS, Leon AR, Merchant FM. Outcome of subcutaneous implantable cardioverter defibrillator implantation in patients with end-stage renal disease on dialysis. J Cardiovasc Electrophysiol 2015.
Herzog CA, Mangrum JM, Passman R. Sudden cardiac death and dialysis patients. Semin Dial. 2008;21:300–7.
Poulikakos DM, Malik MPM. Challenges of ECG monitoring and ECG interpretations in dialysis units. J Electrocardiol. 2016;49:855–9.
Groh CA, Sharma S, Pelchovitz DJ, Bhave PD, Rhyner J, Verma N, et al. Use of an electrocardiographic screening tool to determine candidacy for a subcutaneous implantable cardioverter-defibrillator. Heart Rhythm. 2014;11:1361–6.
Pun PH, Hellkamp AS, Sanders GD, Middleton JP, Hammill SC, Al-Khalidi HR, et al. Primary prevention implantable cardioverter defibrillators in end-stage kidney disease patients on dialysis: a matched cohort study. Nephrol Dial Transplant. 2015;30:829–35.
Olde Nordkamp LR, Warnaars JL, Kooiman KM, de Groot JR, Rosenmoller BR, Wilde AA, et al. Which patients are not suitable for a subcutaneous ICD: incidence and predictors of failed QRS-T-wave morphology screening. J Cardiovasc Electrophysiol. 2014;25:494–9.
Randles DA, Hawkins NM, Shaw M, Patwala AY, Pettit SJ, Wright DJ. How many patients fulfil the surface electrocardiogram criteria for subcutaneous implantable cardioverter-defibrillator implantation? Europace. 2014;16:1015–21.
Wu VC, Lin LY, Wu KD. QT interval dispersion in dialysis patients. Nephrology (Carlton). 2005;10:109–12.
Ansari A, Kaupke CJ, Vaziri ND, Miller R, Barbari A. Cardiac pathology in patients with end-stage renal disease maintained on hemodialysis. Int J Artif Organs. 1993;16:31–6.
Dillon JJ, DeSimone CV, Sapir Y, Somers VK, Dugan JL, Bruce CJ, et al. Noninvasive potassium determination using a mathematically processed ECG: proof of concept for a novel “blood-less, blood test”. J Electrocardiol. 2015;48:12–8.
Acknowledgements
The authors would like to thank Frenova Renal Research.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
Rod Passman: Honoraria, Medtronic, Biotronik; royalties, UpToDate. All other authors declare that they have no conflict of interest.
Rights and permissions
About this article
Cite this article
You, C., Sharma, S., Bavishi, A. et al. Dialytic interval and the timing of electrocardiographic screening for subcutaneous cardioverter-defibrillator placement in chronic hemodialysis patients. J Interv Card Electrophysiol 52, 179–184 (2018). https://doi.org/10.1007/s10840-018-0343-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-018-0343-1