Anterior segment ultrasound biomicroscopy image analysis using ImageJ software: Intra-observer repeatability and inter-observer agreement

  • Azam QureshiEmail author
  • Haoxing Chen
  • Osamah Saeedi
  • Mona A. Kaleem
  • Gianna Stoleru
  • Jordan Margo
  • Sachin Kalarn
  • Janet L. Alexander
Original Paper
  • 154 Downloads

Abstract

Purpose

In this novel study, we demonstrate a standardized imaging and measurement protocol of anterior segment (AS) structures with reliability analysis using ultrasound biomicroscopy (UBM) and ImageJ software.

Methods

Ten pediatric and young adult patients undergoing examination under anesthesia for AS pathology were imaged using UBM. Four trained observers analyzed 20 images using ImageJ. Forty-five structural parameters were measured. Those that relied on the trabecular-iris angle (TIA) as a reference landmark were labeled TIA-dependent (TD) and all others were labeled non-TIA dependent (NTD). Intra-observer repeatability (IOR) and inter-observer agreement (IOA) of measurements were determined using coefficient of variation (CV) and intra-class correlation (ICC) followed by assessment of Bland–Altman plots (BAP) for each pair of observers, respectively.

Results

For NTD parameters, non-ciliary body (CB) related measurements showed CV range 0.60–16.22% and ICC range 0.84–0.89, whereas CB-related parameters showed CV range 2.86–23.40% and ICC range 0.29–0.92. For TD parameters, parameters < 2 degrees removed from reference showed CV range 0.02–5.40% and ICC range 0.89–1.00, whereas parameters > 1 degree removed showed CV range 0.63–27.44% and ICC range 0.22–1.00. No systematic proportional bias was detected by BAPs.

Conclusions

Preplaced landmarks yielded good IOR and IOA in quantitative assessment of AS structures that were NTD and non-CB-related or less removed from the reference. CB-related NTD measurements varied greatly in IOR and IOA, indicating protocol modifications or CB qualitative assessments needed to improve accuracy. Variability in TD measurements increased the further removed from the reference, which supports implementation of a reliable reference landmark to minimize variation.

Keywords

UBM Ultrasound biomicroscopy Anterior segment Pediatric Congenital cataracts Congenital glaucoma 

Notes

Acknowledgements

This work was supported by: (1) The Knights Templar Eye Foundation Career Starter Grants for 2015–2016 awarded to Dr. Janet Alexander and (2) The K23 Career Development Award from the National Eye Institute (K23EY025014) awarded to Dr. Osamah Saeedi. We would like to acknowledge Cierra McKoy for contributing to this work.

Compliance with ethical standards

Conflict of interest

No authors report conflict of interest.

Ethical approval

All procedures performed were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. No animals were included in this study.

Informed consent

This work involved human participants in which appropriate informed consent was obtained.

Supplementary material

10792_2018_882_MOESM1_ESM.pdf (547 kb)
Supplementary material 1 (PDF 546 kb)
10792_2018_882_MOESM2_ESM.pdf (608 kb)
Supplementary material 2 (PDF 608 kb)
10792_2018_882_MOESM3_ESM.pdf (358 kb)
Supplementary material 3 (PDF 357 kb)

References

  1. 1.
    Hussein TR, Shalaby SM, Elbakary MA, Elseht RM, Gad RE (2014) Ultrasound biomicroscopy as a diagnostic tool in infants with primary congenital glaucoma. Clin Ophthalmol (Auckland, NZ) 8:1725Google Scholar
  2. 2.
    Mérula RV, Cronemberger S, Diniz Filho A, Calixto N (2008) New comparative ultrasound biomicroscopic findings between fellow eyes of acute angle closure and glaucomatous eyes with narrow angle. Arquivos Brasileiros de Oftalmologia 71(6):793–798CrossRefGoogle Scholar
  3. 3.
    Prata TS, Dorairaj S, De Moraes CG et al (2013) Is preoperative ciliary body and iris anatomical configuration a predictor of malignant glaucoma development? Clin Exp Ophthalmol 41(6):541–545CrossRefGoogle Scholar
  4. 4.
    Pavlin CJ, Harasiewicz K, Foster FS (1992) Ultrasound biomicroscopy of anterior segment structures in normal and glaucomatous eyes. Am J Ophthalmol 113(4):381–389CrossRefGoogle Scholar
  5. 5.
    Kobayashi H, Hirose M, Kobayashi K (2000) Ultrasound biomicroscopic analysis of pseudophakic pupillary block glaucoma induced by Soemmering’s ring. Br J Ophthalmol 84(10):1142–1146CrossRefGoogle Scholar
  6. 6.
    Marigo FA, Finger PT, McCormick SA et al (2000) Iris and ciliary body melanomas: ultrasound biomicroscopy with histopathologic correlation. Arch Ophthalmol 118(11):1515–1521CrossRefGoogle Scholar
  7. 7.
    Pavlin CJ, Harasiewicz K, Sherar MD, Foster FS (1991) Clinical use of ultrasound biomicroscopy. Ophthalmology 98(3):287–295CrossRefGoogle Scholar
  8. 8.
    Pavlin CJ, Foster FS (1992) Ultrasound biomicroscopy in glaucoma. Acta Ophthalmol 70(S204):7–9CrossRefGoogle Scholar
  9. 9.
    Dada T, Sihota R, Gadia R, Aggarwal A, Mandal S, Gupta V (2007) Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment. J Cataract Refract Surg 33(5):837–840CrossRefGoogle Scholar
  10. 10.
    Wang Z, Huang J, Lin J, Liang X, Cai X, Ge J (2014) Quantitative measurements of the ciliary body in eyes with malignant glaucoma after trabeculectomy using ultrasound biomicroscopy. Ophthalmology 121(4):862–869CrossRefGoogle Scholar
  11. 11.
    Oh J, Shin HH, Kim JH, Kim HM, Song JS (2007) Direct measurement of the ciliary sulcus diameter by 35-megahertz ultrasound biomicroscopy. Ophthalmology 114(9):1685–1688CrossRefGoogle Scholar
  12. 12.
    Narayanaswamy A, Sakata LM, He MG et al (2010) Diagnostic performance of anterior chamber angle measurements for detecting eyes with narrow angles: an anterior segment OCT study. Arch Ophthalmol 128(10):1321–1327CrossRefGoogle Scholar
  13. 13.
    Tello C, Liebmann J, Potash SD, Cohen H, Ritch R (1994) Measurement of ultrasound biomicroscopy images: intraobserver and interobserver reliability. Invest Ophthalmol Vis Sci 35(9):3549–3552Google Scholar
  14. 14.
    Nonaka A, Kondo T, Kikuchi M et al (2006) Angle widening and alteration of ciliary process configuration after cataract surgery for primary angle closure. Ophthalmology 113(3):437–441CrossRefGoogle Scholar
  15. 15.
    Feizi S, Jafarinasab MR, Karimian F, Hasanpour H, Masudi A (2014) Central and peripheral corneal thickness measurement in normal and keratoconic eyes using three corneal pachymeters. J Ophthalmic Vis Res 9(3):296–304.  https://doi.org/10.4103/2008-322x.143356 Google Scholar
  16. 16.
    Kurimoto Y, Park M, Sakaue H, Kondo T (1997) Changes in the anterior chamber configuration after small-incision cataract surgery with posterior chamber intraocular lens implantation. Am J Ophthalmol 124(6):775–780CrossRefGoogle Scholar
  17. 17.
    Kanellopoulos AJ, Asimellis G (2014) Clear-cornea cataract surgery: pupil size and shape changes, along with anterior chamber volume and depth changes. A Scheimpflug imaging study. Clin Ophthalmol (Auckland, NZ) 8:2141CrossRefGoogle Scholar
  18. 18.
    Goldsmith JA, Li Y, Chalita MR, Westphal V, Patil CA, Rollins AM, Izatt JA, Huang D (2005) Anterior chamber width measurement by high-speed optical coherence tomography. Ophthalmology 112(2):238–244CrossRefGoogle Scholar
  19. 19.
    Piñero DP, Plaza Puche AB, Alió JL (2008) Corneal diameter measurements by corneal topography and angle-to-angle measurements by optical coherence tomography: evaluation of equivalence. J Cataract Refract Surg 34(1):126–131.  https://doi.org/10.1016/j.jcrs.2007.10.010 CrossRefGoogle Scholar
  20. 20.
    Chua J, Muen WJ, Reddy A, Brookes J (2012) The masquerades of a childhood ciliary body medulloepithelioma: a case of chronic uveitis, cataract, and secondary glaucoma. Case reports in ophthalmological medicineGoogle Scholar
  21. 21.
    Németh J, Csákány B, Pregun T (1996) Ultrasound biomicroscopic morphometry of the anterior eye segment before and after one drop of pilocarpine. Int Ophthalmol 20(1–3):39–42Google Scholar
  22. 22.
    Kobayashi H, Ono H, Kiryu J, Kobayashi K, Kondo T (1999) Ultrasound biomicroscopic measurement of development of anterior chamber angle. Br J Ophthalmol 83:559–562CrossRefGoogle Scholar
  23. 23.
    Wang D, He M, Wu L, Yaplee S, Singh K, Lin S (2012) Differences in iris structural measurements among American Caucasians, American Chinese, and mainland Chinese. Clin Exp Ophthalmol 40(2):162–169CrossRefGoogle Scholar
  24. 24.
    Lee RY, Huang G, Porco TC, Chen YC, He M, Lin SC (2013) Differences in iris thickness among African Americans, Caucasian Americans, Hispanic Americans, Chinese Americans, and Filipino-Americans. J Glaucoma 22(9):673–678CrossRefGoogle Scholar
  25. 25.
    Leung CK, Palmiero PM, Weinreb RN, Li H, Sbeity Z, Dorairaj S, Leung D, Liu S, Liebmann LM, Congdon N, Lam DS, Ritch R (2010) Comparisons of anterior segment biometry between Chinese and Caucasians using anterior segment optical coherence tomography. Br J Ophthalmol 94(9):1184–1189CrossRefGoogle Scholar
  26. 26.
    Mérula RV, Cronemberger S, Diniz Filho A, Calixto N (2008) New comparative ultrasound biomicroscopic findings between fellow eyes of acute angle closure and glaucomatous eyes with narrow angle. Arq Bras Oftalmol 71(6):793–798CrossRefGoogle Scholar
  27. 27.
    Hussein TR, Shalaby SM, Elbakary MA, Elseht RM, Gad RE (2014) Ultrasound biomicroscopy as a diagnostic tool in infants with primary congenital glaucoma. Clin Ophthalmol (Auckland, NZ) 8:1725Google Scholar
  28. 28.
    Kanellopoulos AJ, Asimellis G (2014) Clear-cornea cataract surgery: pupil size and shape changes, along with anterior chamber volume and depth changes. A Scheimpflug imaging study. Clin Ophthalmol (Auckland, NZ) 8:2141CrossRefGoogle Scholar

Copyright information

© Springer Science+Business Media B.V., part of Springer Nature 2018

Authors and Affiliations

  1. 1.University of Maryland School of MedicineBaltimoreUSA
  2. 2.University Hospitals Case Medical CenterClevelandUSA
  3. 3.University of Maryland Eye AssociatesBaltimoreUSA
  4. 4.Department of OphthalmologyUniversity of MarylandBaltimoreUSA

Personalised recommendations