First Interlaboratory Comparison on Calibration of Temperature-Controlled Enclosures in Turkey

  • TEMPMEKO 2016
  • Published:
International Journal of Thermophysics Aims and scope Submit manuscript

Abstract

The number of accredited laboratories in the field of calibration of temperature-controlled enclosures has been increasing in Turkey. One of the main criteria demonstrating the competence of a calibration laboratory is successful participation in interlaboratory comparisons. Therefore, TUBITAK UME Temperature Laboratory organized the first interlaboratory comparison on “Calibration of Temperature-Controlled Enclosures” in Turkey as a pilot laboratory between January and November, 2013. Forty accredited laboratories which provide routine calibration services to the industry in this field participated in the comparison. The standards used during the comparison was a climatic chamber for the measurements at \(-40\, {^{\circ }}\hbox {C},\,-20\, {^{\circ }}\hbox {C}, 40\, {^{\circ }}\hbox {C}\) and \(100\, {^{\circ }}\hbox {C}\) and an oven for the measurements at \(200\, {^{\circ }}\hbox {C}\). The protocol of the comparison was prepared considering guide EURAMET cg-20 and BS EN/IEC standards 600068-3-5 and 600068-3-11. During the comparison measurements, each participant had the liberty to choose the most convenient calibration points in terms of their accreditation scope among the values mentioned above and carried out on-site measurements at UME. The details and the results of this comparison are given in the paper. Determination of the statistical consistency of the results with the uncertainties given by the participants can be assessed by the method of \(E_{n}\) value assessment for each laboratory. \(E_{n}\) values for all measurement results based on the results of pilot and participating laboratories were calculated.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. ISO/IEC 17025, General Requirements for the Competence of Testing and Calibration Laboratories (2005)

  2. Euramet cg-20, Calibration of Temperature and/or Humidity Enclosures, Version 4.0 (02/2015)

  3. BS EN/IEC 60068-3-5, Environmental Testing—Part 3–5: Confirmation of the Performance of Temperature Chambers. BS EN (2002), IEC (2001)

  4. BS EN/IEC 60068-3-11, Environmental Testing—Part 3–11: Calculation of Uncertainty of Conditions in Climatic Test Chambers (2007)

  5. DKD-R5-7, Calibration of Climatic Chambers (2004) [English Translation (2007)]

  6. EA-4/02 M, Evaluation of the Uncertainty of Measurement in Calibration (2013)

  7. ISO/IEC 17043, Conformity Assessment—General Requirements for Proficiency Testing (2010)

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to A. Uytun.

Additional information

Selected Papers of the 13th International Symposium on Temperature, Humidity, Moisture and Thermal Measurements in Industry and Science.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Uytun, A., Kalemci, M. First Interlaboratory Comparison on Calibration of Temperature-Controlled Enclosures in Turkey. Int J Thermophys 38, 168 (2017). https://doi.org/10.1007/s10765-017-2303-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1007/s10765-017-2303-x

Keywords

Navigation