Abstract
To develop a portable device for measuring the donor corneal transparency and validate its efficacy for corneal evaluation in the eye-banks and for research. The transparency device (TD) has a light source, a detachable system for corneal insertion and a base for light transmission. The probe detects the transmitted light which is measured by a lux-meter. A contact lens was set as ‘control’ to reduce the light scattering concern, an empty petri-plate as ‘blank’ and the cornea as ‘sample’. Two experts and non-experts (masked) observed the corneas for subjective analysis which was then compared using the TD. The parameters observed were scars, foreign-body, stromal-deformities, folds, thickness and opacity which were then converted to a relative overall percentage by the observer. Twenty corneas were evaluated for correlation, five tissues to obtain standard-deviation and twenty-four pairs for a comparative study. Experts mimicked the eye-banks with long-term experience while non-experts mimicked the emerging eye-banks. Subjective values by the experts closely resembled the measurements by TD. The average correlation between the experts and the non-experts to TD was 0.985 and 0.960 respectively. TD showed higher reproducibility than experts followed by the non-experts. The comparative study showed that increase in thickness reduces the transparency. TD is portable, easy, efficient, maintains sterility and less expensive hence the emerging eye-banks and researchers can use to raise their standards and evaluate the transparency for in vitro tests and comparative studies. The suitable transparency for the cornea deemed for clinical applications was found to be >75 %.
References
Hassell JR, Birk David E (2010) The molecular basis of corneal transparency. Exp Eye Res 91(3):326–335
Kara R (2011) The eye: the physiology of human perception. Britannica educational publishing, Chicago, Illinois USA, pp 1–288
Thuret G, Manissolle C, Campos-Guyotat L, Guyotat D, Gain P (2005) Animal compound-free medium and poloxamer for human corneal organ culture and deswelling. Invest Ophthalmol Vis Sci 46(3):816–822
Timothy JF (1997) Basis of transparency of the normal cornea. Eye 11:465–471
Ventura L, Torres de Jesus G, Camilo Dablas de Oliveira G, Souza SJF (2005) Portable light transmission measuring system for preserved corneas. Biomed Eng Online 4(70)
Acknowledgments
The authors would like to thank Claudia Breda, Davide Camposampiero, Elisa Zanetti, Irina Negura and Mattia Baruzzo from The Veneto Eye Bank Foundation for their subjective evaluations and support with corneal tissues.
Conflict of interest
The authors have NO conflicting or financial interests.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Parekh, M., Ferrari, S., Ruzza, A. et al. A portable device for measuring donor corneal transparency in eye banks. Cell Tissue Bank 15, 7–13 (2014). https://doi.org/10.1007/s10561-013-9372-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10561-013-9372-2