Cardiovascular Engineering

, Volume 7, Issue 1, pp 39–42 | Cite as

Gyro-effect Stabilizes Unstable Permanent Maglev Centrifugal Pump

  • Kun-Xi Qian
Original Paper
  • 52 Downloads

Abstract

According to Earnshaw’s Theorem (1839), the passive maglev cannot achieve stable equilibrium and thus an extra coil is needed to make the rotor electrically levitated in a heart pump. The author had developed a permanent maglev centrifugal pump utilizing only passive magnetic bearings, to keep the advantages but to avoid the disadvantages of the electric maglev pumps. The equilibrium stability was achieved by use of so-called “gyro-effect”: a rotating body with certain high speed can maintain its rotation stably. This pump consisted of a rotor (driven magnets and an impeller), and a stator with motor coil and pump housing. Two passive magnetic bearings between rotor and stator were devised to counteract the attractive force between the motor coil iron core and the rotor driven magnets. Bench testing with saline demonstrated a levitated rotor under preconditions of higher than 3,250 rpm rotation and more than1 l/min pumping flow. Rotor levitation was demonstrated by 4 Hall sensors on the stator, with evidence of reduced maximal eccentric distance from 0.15 mm to 0.07 mm. The maximal rotor vibration amplitude was 0.06 mm in a gap of 0.15 mm between rotor and stator. It concluded that Gyro-effect can help passive maglev bearings to achieve stabilization of permanent maglev pump; and that high flow rate indicates good hydraulic property of the pump, which helps also the stability of passive maglev pump.

Keywords

Permanent maglev centrifugal pump Gyro-effect Passive magnetic bearings Equilibrium stability 

References

  1. 1.
    Bramm G, Novak P, Olsen DB. A free-floating body as the motor of a centrifugal blood pump for LVAD or TAH. Euro Soc Artif Organs Proc 1981;8:41–5Google Scholar
  2. 2.
    Akamazu T. Centrifugal blood pump with a magnetically suspended impeller. Artif Organs 1992;16:305–8Google Scholar
  3. 3.
    Mueller J, Nuesser P, Weng Y. The first ventricular cardiac assist device without any wear and tear in a clinical multi-center trial. ASAIO J 2003;49:162CrossRefGoogle Scholar
  4. 4.
    Goldowsky M. Mini hemoreliable axial flow LVAD with magnetic bearings. Part 1,2,3. ASAIO J 2002;48:96–101PubMedCrossRefGoogle Scholar
  5. 5.
    Hiroyuki O, Michiko M, Toru M. Novel maglev pump with a combined magnetic bearing. ASAIO J 2005;51(1):50–5 January/FebruaryCrossRefGoogle Scholar
  6. 6.
    Hoshi H, Asama H, Shinshi T, Ohuchi K, Nakamura M, Mizuno T, i Arai H, Shimokohbe A, Takatani S. Disposable magnetically levitated centrifugal blood pump: design and in vitro performance. Artif Organs 2005;29(7) 520–6Google Scholar
  7. 7.
    Sinha PK. Electromagnetic suspention: dynamics and control. IEEE Control Eng Ser, 1987;30, p 290Google Scholar
  8. 8.
    Earnshaw S. On the nature of molecular forces which regulate the constitution of luminiferous ether. Trans Camb Philol Soc 1839;7:97–112Google Scholar
  9. 9.
    Braunbeck W. Freischwebende koerper im elektrischen und magnetischen Feld. Z Phys 1939;112:764–9CrossRefGoogle Scholar
  10. 10.
    Harrigan RM. Levitron device. U.S. Patent. $382245. May 3, 1983Google Scholar
  11. 11.
    Qian KX, Zeng P, Ru WM, Yuan HY, Feng ZG, LI L. Permanent magnetic-levitation of rotating impeller: a decisive breakthrough in the centrifugal pump. J Med Eng Tech 2002;26(1):36–8CrossRefGoogle Scholar
  12. 12.
    Qian KX, Zeng P, Ru WM, Yuan HY. New concept and new design of permanent maglev rotary artificial heart blood pumps. J Med Eng Phys 2006;28(4):383–388CrossRefGoogle Scholar
  13. 13.
    Qian KX, Zeng P, Ru WM, Yuan HY. A novel permanent maglev impeller TAH: most requirements on blood pumps have been satisfied. J Biomater Appl 2003;18(1):53–62PubMedCrossRefGoogle Scholar
  14. 14.
    Qian KX. Novel magnetic spring and magnetic bearing. IEEE Trans Magnetics 2003;39:559–601CrossRefGoogle Scholar
  15. 15.
    Wan FK, Qian KX. Permanent maglev impeller blood pump and analysis of its rotator suspension. Beijing J Biomed Eng 2005;24(3):199–203Google Scholar
  16. 16.
    Li H, Qian KX. Experimental studies of a stimulating human cardiovascular system. Beijing J Biomed Eng 2005;24(5):405–7Google Scholar
  17. 17.
    Zeng P, Qian KX. Control of brush-less dc motor without position sensor for artificial heart. J Jiangsu Univ 1999;20:41–3Google Scholar
  18. 18.
    Zeng P. Closed loop speed regulation system of sensor-less brush-less DC motor of artificial heart. J Jiangsu Univ 2000;21:247–50Google Scholar
  19. 19.
    Wang DF, Zhou XQ, Qian KX, Lick SD, Deyo DJ, Chambers S, Zwischenberg JB. Compact para-corporeal right heart and lung assist device for total right heart and respiratory support in awake sheep. ASAIO J 2005;51(2):56ACrossRefGoogle Scholar
  20. 20.
    Wang DF, Loran DB, Lick SD, Scott SK, Campbell KM, Donald DJ, Qian KX, Chambers SD, Zwischenberger JB. Development of compact paracorporeal pumped artificial lung for total assistance of pulmonary circulation and respiration. 21st Annual Meeting of Houston Society for Engineering in Medicine and Biolory 12–13 February 2004, Houston, TexasGoogle Scholar
  21. 21.
    Qian KX, Zeng P, Ru WM, Yuan HY. A durable impeller pump with rolling bearing and purge system. J Med Eng Techn 2001;25(6):273–5CrossRefGoogle Scholar
  22. 22.
    Qian KX, Zeng P, Ru WM, Yuan HY, Feang ZG, Li L. Toward a durable impeller pump with mechanical bearings. ASAIO J 2002;48:290–2PubMedCrossRefGoogle Scholar

Copyright information

© Springer Science+Business Media, LLC 2007

Authors and Affiliations

  1. 1.Biomedical Engineering InstituteJiang-su UniversityZhenjiangChina

Personalised recommendations