Realization of 19F MRI oximetry method using perfluorodecalin

  • Research Article
  • Published:
Magnetic Resonance Materials in Physics, Biology and Medicine Aims and scope Submit manuscript

Abstract

Objective

To identify the technical aspects of the potential use of clinically approved perfluorodecalin (PFD, C10F18) for 19F magnetic resonance imaging (MRI) oximetry method at high magnetic field 7.05 T.

Materials and methods

19F T1 measurements were made on a set of PFD samples with different oxygen contents (0%, 21%, and 100%) at room (21 °C) and body temperature (37 °C). In vivo MRI studies were carried out on one healthy rat and two rats with C6 brain glioma.

Results

The selective excitation of the magnetically equivalent 19F nuclei of CF2 groups of trans-isomer of PFD, which give a doublet at a frequency of about − 140 ppm (in relation the chemical shift of trifluoroacetic acid, which is − 76.55 ppm) should be done for correct implementation of 19F MRI oximetry method. The amount of PFD equal to 30 μl is the optimal for obtaining reliable data on the measured T1 values. In this case, the standard deviation of T1 does not exceed 5%. In vivo MRI studies showed that the values of the partial pressure of oxygen (pO2) decrease from normal values of about 38 mmHg (healthy brain) to almost 0 mmHg at the last stage of tumor growth.

Conclusion

The study showed the feasibility of the successful application of PFD for 19F MRI oximetry method.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Dardzinski BJ, Sotak CH (1994) Rapid tissue oxygen tension mapping using 19F inversion-recovery echo-planar imaging of perfluoro-15-crown-5-ether. Magn Reson Med 32(1):88–97

    Article  CAS  PubMed  Google Scholar 

  2. Mignion L, Magat J, Schakman O, Marbaix E, Gallez B, Jordan BF (2013) Hexafluorobenzene in comparison with perfluoro-15-crown-5-ether for repeated monitoring of oxygenation using 19F MRI in a mouse model. Magn Reson Med 69(1):248–254

    Article  CAS  PubMed  Google Scholar 

  3. Maevsky EI, Gervits LL (2008) Perfluorocarbon-based blood substitute—PERFTORAN Russian experience. Chim Oggi 26(3):34–37

    Google Scholar 

  4. Parhami P, Fung BM (1983) Fluorine-19 relaxation study of perfluoro chemicals as oxygen carriers. J Phys Chem 87:1928–1931

    Article  CAS  Google Scholar 

  5. Jacoby C, Temmea S, Mayenfelsb F et al (2014) Probing different perfluorocarbons for in vivo inflammation imaging by 19F MRI: image reconstruction, biological half-lives and sensitivity. NMR Biomed 27(3):261–271

    Article  CAS  PubMed  Google Scholar 

  6. Kodibagkar VD, Wang X, Mason RP (2008) Physical principles of quantitative nuclear magnetic resonance oximetry. Front Biosci 13:1371–1384

    Article  CAS  PubMed  Google Scholar 

  7. Grobben B, DeDeyn PP, Slegers H (2002) Rat C6 gliomas experimental model system for the study of glioblastoma growth and invasion. Cell Tissue Res 310(3):257–270

    Article  CAS  PubMed  Google Scholar 

  8. Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO (2007) Measurement of signal-to-noise ratios in MR images: Influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging 26:375–385

    Article  PubMed  Google Scholar 

  9. Einstein SA, Weegman BP, Firpo MT, Papas KK, Garwood M (2016) Development and validation of noninvasive magnetic resonance relaxometry for the in vivo assessment of tissue-engineered graft oxygenation. Tissue Eng Part C Methods 22(11):1009–1017

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Carreau A, Hafny-Rahbi BE, Matejuk A, Grillon C, Kieda C (2011) Why is the partial oxygen pressure of human tissues a crucial parameter? Small molecules and hypoxia. J Cell Mol Med 15:1239–1253

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Liu S, Shah SJ, Wilmes LJ et al (2011) Quantitative tissue oxygen measurement in multiple organs using 19F MRI in a rat model. Magn Reson Med 66:1722–1730

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

This work was carried out at the Moscow State University on the equipment of the Collective Using Center “Biospectrotomography” and supported by Russian Foundation for Basic Research Grant no. 17-02-00465

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Mikhail V. Gulyaev.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

All procedures were conducted in accordance with the European Community Council directives 86/609/EEC and the study was approved by the local institutional animal ethics committee.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gulyaev, M.V., Kuznetsova, A.V., Silachev, D.N. et al. Realization of 19F MRI oximetry method using perfluorodecalin. Magn Reson Mater Phy 32, 307–315 (2019). https://doi.org/10.1007/s10334-019-00739-1

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10334-019-00739-1

Keywords

Navigation