Abstract
Objective
The aim of this study was to investigate differences between the virtually planned and clinically achieved implant positions in completely template-guided implantations as a function of the type of edentulous space, the residual natural dentition, and the surgical implementation.
Materials and methods
Fifty-six patient cases with a total of 122 implants were evaluated retrospectively. The implantations were completely template-based. The data of the planned implant positions were overlaid with the actual clinical implant positions, followed by measurements of the 3D deviations in terms of coronal (xc) and apical distance, height (xh), and angulation (ang) and statistical analysis.
Results
The mean xc was 1.2 mm (SD 0.7 mm); the mean xa was 1.8 mm (SD 0.9 mm), the mean xh was 0.8 mm (SD 0.7 mm); and the mean ang was 4.8° (SD 3.1). The type of edentulous space and the jaw (maxilla/mandible) had no significant effect on the results in terms of implant positions. The presence of an adjacent natural tooth at the time of implantation had a significant influence on xh (p = 0.04) and ang (p = 0.05). No significant differences were found regarding the surgical approach for any of the parameters examined.
Conclusion
The results of our study are in the same range as those of other studies. Template-guided implantation offers a high degree of accuracy even in the presence of different configurations of the residual dentition or different surgical approaches. A clinical benefit is therefore present, especially from a prosthetic point of view.
Clinical relevance
The clinically achievable accuracy can be described as sufficient for further prosthetic treatment, given the intrinsic and methodological tolerances, making prosthetic rehabilitation safe and predictable.
Similar content being viewed by others
References
Park C, Raigrodski AJ, Rosen J, Spiekerman C, London RM (2009) Accuracy of implant placement using precision surgical guides with varying occlusogingival heights: an in vitro study. J Prosthet Dent 101(6):372–381. https://doi.org/10.1016/S0022-3913(09)60080-9
Behneke A, Burwinkel M, Behneke N (2012) Factors influencing transfer accuracy of cone beam CT-derived template-based implant placement. Clin Oral Implants Res 23(4):416–423. https://doi.org/10.1111/j.1600-0501.2011.02337.x
Nickenig HJ, Wichmann M, Hamel J, Schlegel KA, Eitner S (2010) Evaluation of the difference in accuracy between implant placement by virtual planning data and surgical guide templates versus the conventional free-hand method - a combined in vivo - in vitro technique using cone-beam CT (part II). J Craniomaxillofac Surg 38(7):488–493. https://doi.org/10.1016/j.jcms.2009.10.023
Rungcharassaeng K, Caruso JM, Kan JYK, Schutyser F, Boumans T (2015) Accuracy of computer-guided surgery: a comparison of operator experience. J Prosthet Dent 114(3):407–413. https://doi.org/10.1016/j.prosdent.2015.04.004
Vercruyssen M, Cox C, Coucke W, Naert I, Jacobs R, Quirynen M (2014) A randomized clinical trial comparing guided implant surgery (bone- or mucosa-supported) with mental navigation or the use of a pilot-drill template. J Clin Periodontol 41(7):717–723. https://doi.org/10.1111/jcpe.12231
Schneider D, Marquardt P, Zwahlen M, Jung RE (2009) A systematic review on the accuracy and the clinical outcome of computer-guided template-based implant dentistry. Clin Oral Implants Res 20(Suppl 4):73–86. https://doi.org/10.1111/j.1600-0501.2009.01788.x
Tahmaseb A, Wismeijer D, Coucke W, Derksen W (2014) Computer technology applications in surgical implant dentistry: a systematic review. Int J Oral Maxillofac Implants 29(Suppl):25–42. https://doi.org/10.11607/jomi.2014suppl.g1.2
Reyes A, Turkyilmaz I, Prihoda TJ (2015) Accuracy of surgical guides made from conventional and a combination of digital scanning and rapid prototyping techniques. J Prosthet Dent 113(4):295–303. https://doi.org/10.1016/j.prosdent.2014.09.018
Kernen F et al (2015) Accuracy of three-dimensional printed templates for guided implant placement based on matching a surface scan with CBCT. Clin Implant Dent Relat Res
Kuhl S et al (2015) Technical accuracy of printed surgical templates for guided implant surgery with the coDiagnostiX software. Clin Implant Dent Relat Res 17(Suppl 1):e177–e182. https://doi.org/10.1111/cid.12152
Geng W, Liu C, Su Y, Li J, Zhou Y (2015) Accuracy of different types of computer-aided design/computer-aided manufacturing surgical guides for dental implant placement. Int J Clin Exp Med 8(6):8442–8449
Turbush SK, Turkyilmaz I (2012) Accuracy of three different types of stereolithographic surgical guide in implant placement: an in vitro study. J Prosthet Dent 108(3):181–188. https://doi.org/10.1016/S0022-3913(12)60145-0
Raico Gallardo YN et al (2016) Accuracy comparison of guided surgery for dental implants according to the tissue of support: a systematic review and meta-analysis. Clin Oral Implants Res
Behneke A, Burwinkel M, Knierim K, Behneke N (2012) Accuracy assessment of cone beam computed tomography-derived laboratory-based surgical templates on partially edentulous patients. Clin Oral Implants Res 23(2):137–143. https://doi.org/10.1111/j.1600-0501.2011.02176.x
Zhao XZ, Xu WH, Tang ZH, Wu MJ, Zhu J, Chen S (2014) Accuracy of computer-guided implant surgery by a CAD/CAM and laser scanning technique. Chin J Dent Res 17(1):31–36
Naziri E, Schramm A, Wilde F (2016) Accuracy of computer-assisted implant placement with insertion templates. GMS Interdiscip Plast Reconstr Surg DGPW. 5: p. Doc15
Beretta M, Poli PP, Maiorana C (2014) Accuracy of computer-aided template-guided oral implant placement: a prospective clinical study. J Periodontal Implant Sci 44(4):184–193. https://doi.org/10.5051/jpis.2014.44.4.184
Alzoubi F, Massoomi N, Nattestad A (2016) Accuracy assessment of immediate and delayed implant placements using CAD/CAM surgical guides. J Oral Implantol
Scherer U, Stoetzer M, Ruecker M, Gellrich NC, von See C (2015) Template-guided vs. non-guided drilling in site preparation of dental implants. Clin Oral Investig 19(6):1339–1346. https://doi.org/10.1007/s00784-014-1346-7
Schnutenhaus S, Edelmann C, Rudolph H, Luthardt RG (2016) Retrospective study to determine the accuracy of template-guided implant placement using a novel nonradiologic evaluation method. Oral Surg Oral Med Oral Pathol Oral Radiol 121(4):e72–e79. https://doi.org/10.1016/j.oooo.2015.12.012
Cohen J (1988) Statistical power analysis for the behavioral sciences. 2nd ed, vol xxi. L. Erlbaum Associates, Hillsdale 567 p
Vercruyssen M et al (2008) The use of CT scan based planning for oral rehabilitation by means of implants and its transfer to the surgical field: a critical review on accuracy. J Oral Rehabil 35(6):454–474. https://doi.org/10.1111/j.1365-2842.2007.01816.x
Schneider D, Schober F, Grohmann P, Hammerle CHF, Jung RE (2015) In-vitro evaluation of the tolerance of surgical instruments in templates for computer-assisted guided implantology produced by 3-D printing. Clin Oral Implants Res 26(3):320–325. https://doi.org/10.1111/clr.12327
Van Assche N, van Steenberghe D, Quirynen M, Jacobs R (2010) Accuracy assessment of computer-assisted flapless implant placement in partial edentulism. J Clin Periodontol 37(4):398–403. https://doi.org/10.1111/j.1600-051X.2010.01535.x
Ersoy AE, Turkyilmaz I, Ozan O, McGlumphy EA (2008) Reliability of implant placement with stereolithographic surgical guides generated from computed tomography: clinical data from 94 implants. J Periodontol 79(8):1339–1345. https://doi.org/10.1902/jop.2008.080059
Arisan V, Karabuda ZC, Ozdemir T (2010) Accuracy of two stereolithographic guide systems for computer-aided implant placement: a computed tomography-based clinical comparative study. J Periodontol 81(1):43–51. https://doi.org/10.1902/jop.2009.090348
Platzer S, Bertha G, Heschl A, Wegscheider WA, Lorenzoni M (2013) Three-dimensional accuracy of guided implant placement: indirect assessment of clinical outcomes. Clin Implant Dent Relat Res 15(5):724–734. https://doi.org/10.1111/j.1708-8208.2011.00406.x
Cassetta M, Giansanti M, di Mambro A, Calasso S, Barbato E (2013) Accuracy of two stereolithographic surgical templates: a retrospective study. Clin Implant Dent Relat Res 15(3):448–459. https://doi.org/10.1111/j.1708-8208.2011.00369.x
Funding
The authors declare that no conflict of interest exists as per the guidelines of the International Committee of Medical Journal Editors. There was no funding for this study. Within the framework of another prospective study, Camlog and Swissmeda provided materials. The first-named author has held presentations for Camlog and Swissmeda.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
The study was approved by the Ethics Committee of the University of Ulm (decision no. 339/14 dated April 12, 2014).
Informed consent
For this type of study, formal consent is not required.