Abstract
Purpose
The aim of this study was to analyze the influence of several factors on the temperature in the work chamber during shoulder arthroscopy procedures in order to identify danger zones when using radiofrequency.
Methods
Intraarticular temperature was measured intraoperatively using system with special probe that directly measured the temperature in 22 patients. Data collection was prospective. The main parameters studied were the measurement of the temperature depending on: localization of the procedure (glenohumeral or subacromial), the use of coagulation or ablation, the number of portals, the pressure of the arthropump, the time of use, the blood pressure and the temperature of the operating room.
Results
Ninety-three recordings were made. No complications were identified. Addition of a portal reduces the average elevation of 3.8 °C (p < 0.05). Ambient temperature above 19.15 °C with two portals leads to an average increase of 13.3 °C (p < 0.05). Increasing the pressure of the arthropump of 10 mmHg increases the temperature of 0.8 °C (p < 0.05). No significant difference was found on the change in blood pressure, location and mode of use.
Conclusions
These results show the interest of controlling these factors when performing shoulder arthroscopy procedure. This study identifies situations of high joint risk when using radiofrequency and thus to prevent secondary complications such as burns and massive chondrolyses.
Similar content being viewed by others
References
Statistiques activité MCO (indicateur hospi-diag) “Arthroscopies de l'épaule” par GHM (Groupement Homogène de Malades), Données de l’Agence technique de l’information sur l’hospitalisation. http://www.scansante.fr/
Pell RF, Uhl RL (2004) Complications of thermal ablation in wrist arthroscopy. Arthroscopy 20(Suppl 2):84–86. https://doi.org/10.1016/j.arthro.2004.04.005
Lubowitz JH, Poehling GG (2007) Glenohumeral thermal capsulorrhaphy is not recommended—shoulder chondrolysis requires additional research. Arthroscopy 23:687. https://doi.org/10.1016/j.arthro.2007.05.001
Good CR, Shindle MK, Kelly BT et al (2007) Glenohumeral chondrolysis after shoulder arthroscopy with thermal capsulorrhaphy. Arthroscopy 23:797.e1–797.e5. https://doi.org/10.1016/j.arthro.2007.03.092
Hanypsiak BT, Faulks C, Fine K et al (2004) Rupture of the biceps tendon after arthroscopic thermal capsulorrhaphy. Arthroscopy 20(Suppl 2):77–79. https://doi.org/10.1016/j.arthro.2004.04.015
Obrzut SL, Hecht P, Hayashi K et al (1998) The effect of radiofrequency energy on the length and temperature properties of the glenohumeral joint capsule. Arthroscopy 14:395–400
Toth AP, Warren RF, Petrigliano FA et al (2011) Thermal shrinkage for shoulder instability. HSS J 7:108–114. https://doi.org/10.1007/s11420-010-9187-7
Lecoq C, Branfaux M (2014) Brulures cutanées en rapport avec le matériel d’electro-coagulation en chirurgie arthroscopiqe, Session Orthorisq, Société Francaise de Chirurgie Orthopédique et Traumatologique
Ryu JH, Savoie FH (2010) Postarthroscopic glenohumeral chondrolysis of the shoulder. Sports Med Arthrosc 18:181–187. https://doi.org/10.1097/JSA.0b013e3181eb6ca3
ANSM (Agence National de Sécurité du médicament et des produits de santé) (2015) Systèmes d’électrocoagulation en arthroscopie: Rappel de bonne utilisation pour prévenir le risque de brûlure cutanée. http://www.ansm.sante.fr
Horstman CL, McLaughlin RM (2006) The use of radiofrequency energy during arthroscopic surgery and its effects on intraarticular tissues. Vet Comp Orthop Traumatol 19:65–71
Voss JR, Lu Y, Edwards RB et al (2006) Effects of thermal energy on chondrocyte viability. Am J Vet Res 67:1708–1712. https://doi.org/10.2460/ajvr.67.10.1708
Ho E, Cofield RH, Balm MR et al (1999) Neurologic complications of surgery for anterior shoulder instability. J Shoulder Elbow Surg 8:266–270
Greis PE, Burks RT, Schickendantz MS, Sandmeier R (2001) Axillary nerve injury after thermal capsular shrinkage of the shoulder. J Shoulder Elbow Surg 10:231–235. https://doi.org/10.1067/mse.2001.114260
Troxell CR, Morgan CD, Rajan S et al (2011) Dermal burns associated with bipolar radiofrequency ablation in the subacromial space. Arthroscopy 27:142–144. https://doi.org/10.1016/j.arthro.2010.10.013
Lu Y, Bogdanske J, Lopez M et al (2005) Effect of simulated shoulder thermal capsulorrhaphy using radiofrequency energy on glenohumeral fluid temperature. Arthroscopy 21:592–596. https://doi.org/10.1016/j.arthro.2005.02.013
Zoric BB, Horn N, Braun S, Millett PJ (2009) Factors influencing intra-articular fluid temperature profiles with radiofrequency ablation. J Bone Joint Surg Am 91:2448–2454. https://doi.org/10.2106/JBJS.H.01552
Huber M, Eder C, Mueller M et al (2013) Temperature profile of radiofrequency probe application in wrist arthroscopy: monopolar versus bipolar. Arthroscopy 29:645–652. https://doi.org/10.1016/j.arthro.2012.11.006
McKeon B, Baltz MS, Curtis A, Scheller A (2007) Fluid temperatures during radiofrequency use in shoulder arthroscopy: a cadaveric study. J Shoulder Elbow Surg 16:107–111. https://doi.org/10.1016/j.jse.2006.01.012
Huynh V, Barbier O, Bajard X et al (2017) Subacromial temperature profile during bipolar radiofrequency use in shoulder arthroscopy. Comparison of Coblation(®) vs. VAPR(®). Orthop Traumatol Surg Res 103:489–491. https://doi.org/10.1016/j.otsr.2017.02.011
Good CR, Shindle MK, Griffith MH et al (2009) Effect of radiofrequency energy on glenohumeral fluid temperature during shoulder arthroscopy. J Bone Joint Surg Am 91:429–434. https://doi.org/10.2106/JBJS.G.01261
Davies H, Wynn-Jones H, De Smet T et al (2009) Fluid temperatures during arthroscopic subacromial decompression using a radiofrequency probe. Acta Orthop Belg 75:153–157
Barker SL, Johnstone AJ, Kumar K (2012) In vivo temperature measurement in the subacromial bursa during arthroscopic subacromial decompression. J Shoulder Elbow Surg 21:804–807. https://doi.org/10.1016/j.jse.2011.09.024
Ghostine B, el Khoury FI, Roue J, Dagher E (2014) Influence de la radiofréquence sur la température intra-articulaire en arthroscopie de l’épaule. Revue de Chirurgie Orthopédique et Traumatologique 100:e6. https://doi.org/10.1016/j.rcot.2014.09.296
Li S, Chien S, Brånemark P-I (1999) Heat shock-induced necrosis and apoptosis in osteoblasts. J Orthop Res 17:891–899. https://doi.org/10.1002/jor.1100170614
Züger BJ, Ott B, Mainil-Varlet P et al (2001) Laser solder welding of articular cartilage: tensile strength and chondrocyte viability. Lasers Surg Med 28:427–434. https://doi.org/10.1002/lsm.1070
Caffey S, McPherson E, Moore B et al (2005) Effects of radiofrequency energy on human articular cartilage. Am J Sports Med 33:1035–1039. https://doi.org/10.1177/0363546504271965
Kaplan LD, Ernsthausen JM, Bradley JP et al (2003) The thermal field of radiofrequency probes at chondroplasty settings. Arthroscopy 19:632–640. https://doi.org/10.1016/S0749-8063(03)00128-2
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This study was the subject of a presentation to the French Arthroscopy Society in December 2017.
Rights and permissions
About this article
Cite this article
Chivot, M., Airaudi, S., Galland, A. et al. Analysis of parameters influencing intraarticular temperature during radiofrequency use in shoulder arthroscopy. Eur J Orthop Surg Traumatol 29, 1205–1210 (2019). https://doi.org/10.1007/s00590-019-02419-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00590-019-02419-1