On the accuracy of cochlear duct length measurement in computed tomographic images

  • G. Jakob Lexow
  • Marcel Kluge
  • Nils-Claudius Gellrich
  • Thomas Lenarz
  • Omid Majdani
  • Thomas S. Rau
Otology

Abstract

Purpose

Patient specific selection of cochlear implants would benefit from pre-operative knowledge of cochlear length. Several methods for its measurement or estimation have been described in literature. This study focused on the achievable accuracy in clinically available imaging.

Methods

Five simplified cochlea models milled into porcine bone were scanned in water using clinical cone beam computed tomography. Due to their well-known dimensions these phantoms served as gold standard for the length measurements. Each phantom was measured ten times using the custom software Comet. In addition, cochleae in ten image datasets taken indiscriminately from clinical routine were measured ten times each to test the precision under realistic conditions. The results were also compared to estimations based on the diameter of the basal turn (A value) as described in literature.

Results

Measurement accuracy of the phantoms’ lengths was high (average error: − 0.2 mm; standard deviation: 0.3 mm). The pooled standard deviation for the measurements in clinical datasets was 0.6 mm. Errors resulted mainly from problems locating the helicotrema. The estimations differed on average − 1.7 to + 0.4 mm from the manual measurements and had standard deviations between 0.5 and 0.6 mm depending on the algorithm.

Conclusions

The program Comet was successfully used to accurately measure the length of the cochlea models in clinically available imaging. The lower image quality of patient scans reduced the precision of the measurement. Estimations using the A value are a quicker alternative for averagely sized cochleae in cases where the lack of accuracy is tolerable.

Keywords

Cochlear implant In-vitro verification Midmodiolar view 

Notes

Acknowledgements

The presented work was supported by the German Research Foundation (DFG) through the cluster of excellence “Hearing4all” EXC 1077/1 and by the German Federal Ministry of Education and Research (BMBF, FKZ 13GW0019E). Responsibility for the contents of this publication lies with the authors.

Compliance with ethical standards

Conflict of interest

G. Jakob Lexow, Marcel Kluge, Nils-Claudius Gellrich, Thomas Lenarz, Omid Majdani and Thomas S. Rau declare that they have no conflict of interest.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

References

  1. 1.
    Roland PS, Wright CG (2006) Surgical aspects of cochlear implantation: mechanisms of insertional trauma. Adv Otorhinolaryngol 64:11–30.  https://doi.org/10.1159/000094642 PubMedGoogle Scholar
  2. 2.
    Verberne J, Risi F, Campbell L, Chambers S, O’Leary S (2016) The effect of scala tympani morphology on basilar membrane contact with a straight electrode array. Otol Neurotol 38:47–53.  https://doi.org/10.1097/MAO.0000000000001259 CrossRefGoogle Scholar
  3. 3.
    Hardy M (1938) The length of the organ of Corti in man. Am J Anat 62:291–311.  https://doi.org/10.1002/aja.1000620204 CrossRefGoogle Scholar
  4. 4.
    Ulehlová L, Voldrich L, Janisch R (1987) Correlative study of sensory cell density and cochlear length in humans. Hear Res 28:149–151.  https://doi.org/10.1016/0378-5955(87)90045-1 CrossRefPubMedGoogle Scholar
  5. 5.
    Adunka O, Unkelbach MH, Mack MG, Radeloff A, Gstoettner W (2005) Predicting basal cochlear length for electric-acoustic stimulation. Arch Otolaryngol Head Neck Surg 131:488–492.  https://doi.org/10.1001/archotol.131.6.488 CrossRefPubMedGoogle Scholar
  6. 6.
    Würfel W, Lanfermann H, Lenarz T, Majdani O (2014) Cochlear length determination using cone beam computed tomography in a clinical setting. Hear Res 316:65–72.  https://doi.org/10.1016/j.heares.2014.07.013 CrossRefPubMedGoogle Scholar
  7. 7.
    Johnston JDA, Scoffings D, Chung M, Baguley D, Donnelly NP, Axon PR, Gray RF, Tysome JR (2016) Computed tomography estimation of cochlear duct length can predict full insertion in cochlear implantation. Otol Neurotol 37:223–228.  https://doi.org/10.1097/MAO.0000000000000955 CrossRefPubMedGoogle Scholar
  8. 8.
    Meng J, Li S, Zhang F, Li Q, Qin Z (2016) Cochlear size and shape variability and implications in cochlear implantation surgery. Otol Neurotol 37:1307–1313.  https://doi.org/10.1097/MAO.0000000000001189 CrossRefPubMedGoogle Scholar
  9. 9.
    Mistrík P, Jolly C (2016) Optimal electrode length to match patient specific cochlear anatomy. Eur Ann Otorhinolaryngol Head Neck Dis 133:S68–S71.  https://doi.org/10.1016/j.anorl.2016.05.001 CrossRefPubMedGoogle Scholar
  10. 10.
    Koch RW, Elfarnawany M, Zhu N, Ladak HM, Agrawal SK (2017) Evaluation of cochlear duct length computations using synchrotron radiation phase-contrast imaging. Otol Neurotol 38:1.  https://doi.org/10.1097/MAO.0000000000001410 CrossRefGoogle Scholar
  11. 11.
    Lexow GJ, Rau TS, Gellrich N-C, Lenarz T, Majdani O (2015) Cochlea-Längenmessung in rotierenden, midmodiolaren Schichtansichten—Darstellung der Methode. In: Hahn HK, Kikinis R, Klein J et al (eds) Proc. 14. Jahrestagung der Deutschen Gesellschaft für Computer- und Roboterassistierte Chirurgie (CURAC). Fraunhofer MEVIS, Bremen, pp 335–336Google Scholar
  12. 12.
    Lexow GJ, Schurzig D, Gellrich N-C, Lenarz T, Majdani O, Rau TS (2016) Visualization, measurement and modelling of the cochlea using rotating midmodiolar slice planes. Int J Comput Assist Radiol Surg.  https://doi.org/10.1007/s11548-016-1374-7 Google Scholar
  13. 13.
    Van Wermeskerken GKA, Prokop M, Van Olphen AF, Albers FWJ (2007) Intracochlear assessment of electrode position after cochlear implant surgery by means of multislice computer tomography. Eur Arch Otorhinolaryngol 264:1405–1407.  https://doi.org/10.1007/s00405-007-0389-7 CrossRefPubMedGoogle Scholar
  14. 14.
    Lecerf P, Bakhos D, Cottier J-P, Lescanne E, Trijolet JP, Robier A (2011) Midmodiolar reconstruction as a valuable tool to determine the exact position of the cochlear implant electrode array. Otol Neurotol 32:1075–1081.  https://doi.org/10.1097/MAO.0b013e318229d4dd CrossRefPubMedGoogle Scholar
  15. 15.
    Malherbe TK, Hanekom T, Hanekom JJ (2016) Constructing a three-dimensional electrical model of a living cochlear implant user’s cochlea. Int J Numer Method Biomed Eng 32:e02751.  https://doi.org/10.1002/cnm.2751 CrossRefGoogle Scholar
  16. 16.
    Escudé B, James C, Deguine O, Cochard N, Eter E, Fraysse B (2006) The size of the cochlea and predictions of insertion depth angles for cochlear implant electrodes. Audiol Neurotol 11:27–33.  https://doi.org/10.1159/000095611 CrossRefGoogle Scholar
  17. 17.
    Alexiades G, Dhanasingh A, Jolly C (2015) Method to estimate the complete and two-turn cochlear duct length. Otol Neurotol 36:904–907.  https://doi.org/10.1097/MAO.0000000000000620 CrossRefPubMedGoogle Scholar
  18. 18.
    Avci E, Nauwelaers T, Lenarz T, Hamacher V, Kral A (2014) Variations in microanatomy of the human cochlea. J Comp Neurol 522:3245–3261.  https://doi.org/10.1002/cne.23594 CrossRefPubMedPubMedCentralGoogle Scholar
  19. 19.
    Verbist BM, Skinner MW, Cohen LT, Leake PA, James C, Boëx C, Holden TA, Finley CC, Roland PS, Roland JT, Haller M, Patrick JF, Jolly CN, Faltys MA, Briaire JJ, Frijns JHM (2010) Consensus panel on a cochlear coordinate system applicable in histologic, physiologic, and radiologic studies of the human cochlea. Otol Neurotol 31:722–730.  https://doi.org/10.1097/MAO.0b013e3181d279e0 CrossRefPubMedPubMedCentralGoogle Scholar
  20. 20.
  21. 21.
    MED-EL (2013) Electrode arrays: designed for atraumatic implantation providing superior hearing performance. http://www.medel.com/data/pdf/21617.pdf. Accessed 19 Feb 2018
  22. 22.
    Cochlear (2014) Cochlear Nucleus CI422 with Slim Straight - Technical Specifications. http://www.cochlear.com/wps/wcm/connect/e9bc4ab6-8b75-4cf4-bcda-fb7b2757efbb/FUN1356+ISS2+MAR14+CI422+Specifications+3.pdf?MOD=AJPERES. Accessed 19 Feb 2018
  23. 23.
    Martinez-Monedero R, Niparko JK, Aygun N (2011) Cochlear coiling pattern and orientation differences in cochlear implant candidates. Otol Neurotol 32:1086–1093.  https://doi.org/10.1097/MAO.0b013e31822a1ee2 CrossRefPubMedGoogle Scholar
  24. 24.
    Erixon E, Rask-Andersen H (2013) How to predict cochlear length before cochlear implantation surgery. Acta Otolaryngol 133:1258–1265.  https://doi.org/10.3109/00016489.2013.831475 CrossRefPubMedGoogle Scholar
  25. 25.
    Yoo SK, Wang G, Rubinstein JT, Vannier MW (2000) Three-dimensional geometric modeling of the cochlea using helico-spiral approximation. IEEE Trans Biomed Eng 47:1392–1402.  https://doi.org/10.1109/10.871413 CrossRefPubMedGoogle Scholar
  26. 26.
    Rivas A, Cakir A, Hunter JB, Labadie RF, Zuniga MG, Wanna GB, Dawant BM, Noble JH (2017) Automatic cochlear duct length estimation for selection of cochlear implant electrode arrays. Otol Neurotol 38:339–346.  https://doi.org/10.1097/MAO.0000000000001329 CrossRefPubMedPubMedCentralGoogle Scholar
  27. 27.
    Franke-Trieger A, Mürbe D (2015) Estimation of insertion depth angle based on cochlea diameter and linear insertion depth: a prediction tool for the CI422. Eur Arch Otorhinolaryngol 272:3193–3199.  https://doi.org/10.1007/s00405-014-3352-4 CrossRefPubMedGoogle Scholar

Copyright information

© Springer-Verlag GmbH Germany, part of Springer Nature 2018

Authors and Affiliations

  1. 1.Department of Otolaryngology and Cluster of Excellence “Hearing4all” EXC 1077/1Hannover Medical SchoolHanoverGermany
  2. 2.Department of Craniomaxillofacial SurgeryHannover Medical SchoolHanoverGermany

Personalised recommendations