Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals

  • Original Article
  • Published:
Pediatric Radiology Aims and scope Submit manuscript

Abstract

Background

Little is known about exposure differences among hospitals. Large differences might identify outliers using excessive exposure.

Objective

We used the newly described exposure index and deviation index to compare the difference in existing radiographic exposures for neonatal portable chest radiographs among four academic children’s hospitals.

Materials and methods

For each hospital we determined the mean exposure index. We also set target exposure indices and then measured the deviation from this target.

Results

There was not a large difference in exposure index among sites. No site had an exposure index mean that was more than twice or less than half that of any other site. For all four sites combined, 92% of exposures had a deviation index within the range from −3 to +3. Thus exposures at each hospital were consistently within a reasonable narrow spectrum.

Conclusion

Mean exposure index differences are caused by operational differences with mean values that varied by less than 50% among four hospitals. Ninety-two percent of all exposures were between half and double the target exposure. Although only one vendor’s equipment was used, these data establish a practical reference range of exposures for neonatal portable radiographs that can be recommended to other hospitals for neonatal chest radiographs.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Goske MJ, Applegate KE, Boylan J et al (2008) The Image Gently campaign: working together to change practice. AJR 190:273–274

    Article  PubMed  Google Scholar 

  2. Don S (2004) Radiosensitivity of children: potential for overexposure in CR and DR and magnitude of doses in ordinary radiographic examinations. Pediatr Radiol 34(Suppl 3):S167–172, discussion S234-241

    Article  PubMed  Google Scholar 

  3. Uffmann M, Schaefer-Prokop C (2009) Digital radiography: the balance between image quality and required radiation dose. Eur J Radiol 72:202–208

    Article  PubMed  Google Scholar 

  4. Vaño E, Fernández JM, Ten Ignacio J et al (2007) Transition from screen-film to digital radiography: evolution of patient radiation doses at projection radiography. Radiology 243:461–466

    Article  PubMed  Google Scholar 

  5. International Commission on Radiological Protection (2004) Managing patient dose in digital radiology: a report of the International Commission on Radiological Protection. Ann ICRP 34:1–73

    Google Scholar 

  6. Seibert JA, Shelton DK, Moore EH (1996) Computed radiography X-ray exposure trends. Acad Radiol 3:313–318

    Article  PubMed  CAS  Google Scholar 

  7. Schaetzing R (2004) Management of pediatric radiation dose using Agfa computed radiography. Pediatr Radiol 34(Suppl 3):S207–S214

    Article  PubMed  Google Scholar 

  8. Shepard SJ, Wang J, Flynn M et al (2009) An exposure indicator for digital radiography: AAPM Task Group 116 (executive summary). Med Phys 36:2898–2914. Erratum (2010). Med Phys 37:405

    Article  Google Scholar 

  9. (2008) Medical electrical equipment—exposure index of digital x-ray imaging systems. Part 1: definitions and requirements for general radiography, International Electrotechnical Commission (IEC), international standard IEC 62494–1:2008–08 Geneva, Switzerland

  10. Cooper M, Cohen M, Piersall K et al (2011) Using the exposure index to monitor radiation exposure for portable chest radiographs in neonates. Pediatr Radiol 41:592–601

    Article  PubMed  Google Scholar 

  11. Cohen MD (2010) Quality assurance: potential use for the newly described exposure index in clinical practice. J Am Coll Radiol 7:748–749

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Mervyn D. Cohen.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Cohen, M.D., Markowitz, R., Hill, J. et al. Quality assurance: a comparison study of radiographic exposure for neonatal chest radiographs at 4 academic hospitals. Pediatr Radiol 42, 668–673 (2012). https://doi.org/10.1007/s00247-011-2290-1

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00247-011-2290-1

Keywords

Navigation