Improved MRI of the neonatal heart: feasibility study using a knee coil

  • Technical Innovation
  • Published:
Pediatric Radiology Aims and scope Submit manuscript

Abstract

As a first step towards the development of a neonatal cardiac MRI (CMRI) coil, we investigated the utilization, handling and image quality of a knee coil for CMRI in infants at 3 Tesla. Scout imaging and cine imaging with optimized pulse sequence parameters were performed in two neonates and the resulting images were evaluated for diagnostic purposes. All images were of high diagnostic quality. Complete examinations took less than five minutes. The excellent signal-to-noise ratio provided by this coil allowed all cine studies to be acquired with only three averages. In conclusion, a knee coil can be successfully applied in neonates for CMRI, but the coil design could be improved further to facilitate routine clinical use.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Pennell DJ, Sechtem UP, Higgins CB et al (2004) Clinical indications for cardiovascular magnetic resonance (CMR): Consensus Panel report. Eur Heart J 25:1940–1965

    Article  PubMed  Google Scholar 

  2. Groves AM (2011) Cardiac magnetic resonance in the study of neonatal haemodynamics. Semin Fetal Neonatal Med 16:36–41

    Article  PubMed  Google Scholar 

  3. Duerden RM, Pointon KS, Habib S (2006) Review of clinical cardiac MRI. Imaging 18:178–186

    Google Scholar 

  4. Gutberlet M, Spors B, Grothoff M et al (2004) Comparison of different cardiac MRI sequences at 1.5/3.0 T with respect to signal-to-noise and contrast-to-noise ratios: initial experience. Röfo 176:801–808

    PubMed  CAS  Google Scholar 

  5. Pamboucas C, Nihoyannopoulos P (2006) Cardiovascular magnetic resonance at 3 Tesla: advantages, limitations and clinical potential. Hellenic J Cardiol 47:170–173

    PubMed  Google Scholar 

  6. Schick F (2005) Whole-body MRI at high field: technical limits and clinical potential. Eur Radiol 15:946–959

    Article  PubMed  Google Scholar 

  7. Wen H, Denison TJ, Singeman RW et al (1997) The intrinsic signal-to-noise ratio in human cardiac imaging at 1.5, 3 and 4 T. J Magn Reson 125:65–71

    Article  PubMed  CAS  Google Scholar 

  8. Wieben O, Francois C, Reeder SB (2008) Cardiac MRI of ischemic heart disease at 3 T: potential and challenges. Eur J Radiol 65:15–28

    Article  PubMed  Google Scholar 

  9. Groves AM, Chiesa G, Durighel G et al (2011) Functional cardiac MRI in preterm and term newborns. Arch Dis Child Fetal Neonatal Ed 96:86–91

    Article  Google Scholar 

  10. Kawel N, Valsangiacomo-Buechel E, Hoop R et al (2010) Preoperative evaluation of pulmonary artery mophology and pulmonary circulation in neonates with pulmonary atresia—usefulness of MR angiography in clinical routine. J Cardiovasc Magn Reson 12:52–59

    Article  PubMed  Google Scholar 

  11. Merchant N, Groves A, Larkman DJ et al (2009) A patient care system for early 3.0 Tesla magnetic resonance imaging of very low birth weight infants. Early Hum Dev 85:779–783

    Article  PubMed  CAS  Google Scholar 

Download references

Acknowledgements

We thank Mrs. Traudel Hansen (RN) for expert care and monitoring of the neonates during the CMRI examinations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Carsten Rickers.

Electronic supplementary material

Below is the link to the electronic supplementary material.

ESM 1

(AVI 21100 kb)

Rights and permissions

Reprints and permissions

About this article

Cite this article

Helle, M., Jerosch-Herold, M., Voges, I. et al. Improved MRI of the neonatal heart: feasibility study using a knee coil. Pediatr Radiol 41, 1429–1432 (2011). https://doi.org/10.1007/s00247-011-2215-z

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00247-011-2215-z

Keywords

Navigation