Effective dose estimation in whole-body multislice CT in paediatric trauma patients

  • Original Article
  • Published:
Pediatric Radiology Aims and scope Submit manuscript

Abstract

Background

The number of multislice CT (MSCT) scans performed in polytraumatized children has increased rapidly. There is growing concern regarding the radiation dose in MSCT and its long-term consequences, especially in children.

Objective

To determine the effective dose to polytraumatized children who undergo whole-body MSCT.

Materials and methods

A total of 51 traumatized children aged 0–16 years underwent a polytrauma protocol CT scan between November 2004 and August 2006 at our institution. The effective dose was calculated retrospectively by a computer program (CT-Expo 1.5, Hannover, Germany).

Results

The mean effective dose was 20.8 mSv (range 8.6–48.9 mSv, SD ± 7.9 mSv). There was no statistically significant difference in the effective dose between male and female patients.

Conclusion

Whole-body MSCT is a superior diagnostic tool in polytraumatized children with 20.8 mSv per patient being a justified mean effective dose. In a potentially life-threatening situation whole-body MSCT provides the clinicians with relevant information to initiate life-saving therapy. Radiologists should use special paediatric protocols that include dose-saving mechanisms to keep the effective dose as low as possible. Further studies are needed to examine and advance dose-saving strategies in MSCT, especially in children.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Brenner D, Elliston CD, Hall EJ et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296

    PubMed  CAS  Google Scholar 

  2. Moss M, McLean D (2006) Pediatric and adult computed tomography practice and patient dose in Australia. Australas Radiol 50:33–40

    Article  PubMed  CAS  Google Scholar 

  3. Fenton SJ, Hansen KW, Meyers RL et al (2004) CT scan and the pediatric trauma patient – are we overdoing it. J Pediatr Surg 39:1877–1881

    Article  PubMed  Google Scholar 

  4. Brenner DJ, Elliston CD (2004) Estimated radiation risks potentially associated with full-body CT screening. Radiology 232:735–738

    Article  PubMed  Google Scholar 

  5. Bundesamt für Strahlenschutz (2006) Jahrsbericht des Bundesamts für Strahlenschutz 2005. Bundesamt für Strahlenschutz, Salzgitter

  6. UNSCEAR (2000) Report of the United Nations Scientific Committee on the Effects of Atomic Radiation, Sources and Effects of Ionizing Radiation. United Nations, New York

    Google Scholar 

  7. Wiest PW, Locken JA, Heintz PH et al (2002) CT scanning: a major source of radiation exposure. Semin Ultrasound CT MR 23:402–410

    Article  PubMed  Google Scholar 

  8. Slovis TL (2002) ALARA conference proceedings: the ALARA concept in pediatric CT – intelligent dose reduction. Pediatr Radiol 32:217–313

    Article  PubMed  Google Scholar 

  9. Menzel HG, Schibilla H, Teunen D (eds) (2000) European guidelines on quality criteria for computed tomography. Publication no. EUR 16262 EN. European Commission, Luxembourg

  10. US Food and Drug Administration (2007) FDA health notification: reducing radiation risk from computed tomography for pediatric and small adult patients. http://www.fda.gov/cdrh/safety/110201-ct.html. Accessed 12 Dec 2008

  11. European Commission. CT safety & efficacy: a broad perspective. A 6th Framework Research Project of the European Commission. http://www.msct.eu. Accessed 12 Dec 2008

  12. Stamm G, Nagel HD (2005) Benutzerhandbuch CT-Expo V 1.5, Nov 2005, 4. Auflage

  13. Stamm G, Nagel HD (2002) CT-expo – a novel program for dose evaluation in CT. Rofo 174:1570–1576

    PubMed  CAS  Google Scholar 

  14. The National Academies (1998/2006) Health risks from exposure to low levels of ionizing radiation: BEIR VII. National Academies Press, Washington DC. http://www.nap.edu. Accessed 12 Dec 2008

  15. The National Academies (1990) Health effects of exposure to low levels of ionizing radiation: BEIR V. National Academies Press, Washington DC

  16. International Commission on Radiological Protection (1990) Recommendations of the International Commission on Radiological Protection. ICRP publication 60. Pergamon, Oxford

  17. Pierce DA, Preston DL (2000) Radiation-related cancer risks at low doses among atomic bomb survivors. Radiat Res 154:178–186

    Article  PubMed  CAS  Google Scholar 

  18. Mettler FA, Wiest PW, Locken JA et al (2000) CT scanning: patterns of use and dose. J Radiol Prot 20:353–359

    Article  PubMed  Google Scholar 

  19. Federal Ministry for the Environment, Nature Conservation and Nuclear Safety (2004) Umweltradioaktivität und Strahlenbelastung im Jahr 2003 (Parliamentary Report), publication 15/3889. German Bundestag, Berlin

  20. Hörmann M, Scharitzer M, Philipp M et al (2003) First experience with multidetector CT in traumatized children. Eur J Radiol 48:125–132

    Article  PubMed  Google Scholar 

  21. Wedegärtner U, Lorenzen M, Nagel HD et al (2004) Diagnostic imaging in polytrauma: comparison of radiation exposure from whole-body MSCT and conventional radiography with organ-specific CT. Rofo 176:1039–1044

    PubMed  Google Scholar 

  22. Huda W, Vance A (2007) Patient radiation doses from adults and pediatric CT. AJR 188:540–546

    Article  PubMed  Google Scholar 

  23. Semelka RC, Armao DM, Elias J Jr et al (2007) Imaging strategies to reduce the risk of radiation in CT studies, including selective substitution with MRI. J Magn Reson Imaging 25:900–909

    Article  PubMed  Google Scholar 

  24. Albrecht T, von Schlippenbach J, Stahel PF et al (2004) The role of whole body spiral CT in the primary work-up of polytrauma patients – comparison of conventional radiography and abdominal sonography. Rofo 176:1142–1150

    PubMed  CAS  Google Scholar 

  25. Greess H, Wolf H, Suess S et al (2004) Automatic exposure control to reduce the dose in subsecond multislice spiral CT: phantom measurements and clinical results. Rofo 176:862–869

    PubMed  CAS  Google Scholar 

  26. Greess H, Lutze J, Noemayr A et al (2004) Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation. Eur Radiol 14:995–999

    Article  PubMed  CAS  Google Scholar 

  27. Papadakis AE, Perisinakis K, Damilakis J (2007) Angular on-line tube current modulation in multidetector CT examinations of children and adults: The influence of different scanning parameters on dose reduction. Med Phys 34:2864–2874

    Article  PubMed  Google Scholar 

  28. Ravenel JG, Scalzetti EM, Huda W et al (2001) Radiation exposure and image quality in chest CT examinations. AJR 177:279–284

    PubMed  CAS  Google Scholar 

  29. Paterson A, Frush DP, Donnelly LF (2001) Helical CT of the body: are settings adjusted for pediatric patients. AJR 176:297–301

    PubMed  CAS  Google Scholar 

  30. Huda W, Chamberlain CC, Rosenbaum AE et al (2001) Radiation doses to infants and adults undergoing head CT examination. Med Phys 28:393–399

    Article  PubMed  CAS  Google Scholar 

  31. Honnef D, Wildberger JE, Stargardt A et al (2004) Multislice spiral CT (MSCT) in pediatric radiology: dose reduction for chest and abdomen examinations. Rofo 176:1021–1030

    PubMed  CAS  Google Scholar 

  32. Huda W (2007) Radiation doses and risks in chest computed tomography examinations. Proc Am Thorac Soc 4:316–320

    Article  PubMed  Google Scholar 

  33. Paterson A, Frush DP (2007) Dose reduction in paediatric MDCT: general principles. Clin Radiol 62:507–517

    Article  PubMed  CAS  Google Scholar 

  34. McNitt-Gray MF (2002) AAPM/RSNA physics tutorial for residents: topics in CT. Radiation dose in CT. Radiographics 22:1541–1553

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Robin D. Munk.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Munk, R.D., Strohm, P.C., Saueressig, U. et al. Effective dose estimation in whole-body multislice CT in paediatric trauma patients. Pediatr Radiol 39, 245–252 (2009). https://doi.org/10.1007/s00247-008-1091-7

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00247-008-1091-7

Keywords

Navigation