Abstract
Perforation of the atretic pulmonary valve with balloon dilation in infants with pulmonary atresia with intact ventricular septum (PA-IVS) is standard initial therapy for right-ventricular (RV) decompression. This procedure often results in adequate pulmonary blood flow, thus eliminating the need for neonatal surgery. Nonetheless, the incidence of RV outflow-tract complications and mortality for this intervention is significant. We report our experience with retrograde snare-guided radiofrequency (RF) perforation in an attempt to improve accuracy and decrease procedural complications. Medical records were reviewed for the period between March 2007 and May 2010 for all patients with PA-IVS who presented to the catheterization laboratory for attempted RF perforation in infancy. Specific details reviewed included demographics, preprocedural echocardiographic (echo) data, procedural technique and complications, pre- and post-RV pressures and pulmonary valve gradients, need for surgical intervention in the neonatal period, and short- to medium-term follow-up. Eleven neonates with PA-IVS underwent RF perforation using a retrograde snare-guided technique during the study period. The pulmonary valve was successfully perforated and the wire snared in all 11 patients. Six of 11 atretic valves were crossed on the first attempt with low energy (5 W × 2 s). No episodes of tamponade or RV/PA perforation occurred as confirmed by echocardiogram performed immediately after the procedure. There was no ductal spasm with retrograde catheter manipulations. Sequential dilation of the perforated valve was not necessary. RV pressures decreased from 169 % systemic before dilation to 93 % after dilation (p < 0.001) with a residual pulmonary valve gradient of 16 ± 6 mm Hg. Eight of 11 patients (73 %) were discharged without surgery at an average 16 days after the intervention. Three patients required Blalock–Taussig shunts to augment pulmonary blood flow. Femoral artery thrombus occurred in 4 of 11 patients treated with anticoagulation, and 2 patients had atrial arrhythmias during the procedure. No other complications occurred. No preprocedural echo parameters predicted requirement for surgical shunt before hospital discharge. Retrograde snare-guided pulmonary valve perforation in infants with PA-IVS is safe and may decrease the incidence of significant procedural complications.
Similar content being viewed by others
References
Agnoletti G, Piechaud J, Bonhoeffer P et al (2003) Perforation of the atretic pulmonary valve: long-term follow-up. J Am Col Cardiol 41:1399–1403
Alwi M, Geeth K, Bilkis A et al (2000) Pulmonary atresia with intact ventricular septum: percutaneous radiofrequency assisted valvotomy and balloon dilation versus surgical valvotomy and Blalock Taussig shunt. J Am Coll Cardiol 35:468–476
Drighil A, Aljufan M, Slimi A et al (2010) Echocardiographic determinants of successful balloon dilation in pulmonary atresia with intact ventricular septum. Eur J Echocardiogr 11:172–175
Gibbs JL, Blackburn ME, Uzun O et al (1997) Laser valvotomy with balloon valvoplasty for pulmonary atresia with intact ventricular septum: five years’ experience. Heart 77:225–228
Humpl T, Soderberg B, McCrindle B et al (2003) Percutaneous balloon valvotomy in pulmonary atresia with intact ventricular septum—Impact on patient care. Circulation 108:826–832
Mi Y, Chau A, Chiu C et al (2005) Evolution of the management approach for pulmonary atresia with intact ventricular septum. Heart 91:657–663
Moller J (2010) Operative and interventional procedures in 1039 neonates with pulmonary valvular atresia and intact ventricular septum. A multi-institutional study. Prog Pediatr Cardiol 29:15–18
Qureshi S (2006) Catheterization in neonates with pulmonary atresia with intact ventricular septum. Cathet Cardiovasc Diagn 67:924–931
Qureshi S, Rosenthal E, Tynan M et al (1991) Transcatheter laser-assisted balloon pulmonary valve dilation in pulmonic valve atresia. Am J Cardiol 67:428–431
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Lamers, L., Garn, B., Ellsworth, E. et al. Decreased Incidence of Right-Ventricular Outflow Tract Complications Using a Retrograde Snare Technique for Radiofrequency Pulmonary Valve Perforation. Pediatr Cardiol 33, 1275–1280 (2012). https://doi.org/10.1007/s00246-012-0295-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00246-012-0295-7