Retrieval of Migrated Volume Coils Using Different Clot Retrievers in a Porcine Model

  • Original Article
  • Published:
Clinical Neuroradiology Aims and scope Submit manuscript

Abstract

Purpose

During endovascular treatment of intracranial aneurysms using coils, migration is a serious complication that increases neurological morbidity. The aim of this experimental study was to assess the effectiveness and complications of retrieving volume coils with different, currently available clot retrievers in a porcine model.

Methods

Volume coils of three-dimensional (3D) shape and different sizes were placed in the axillary artery of pigs. By means of 4 different clot retrievers (Trevo ProVue; Solitaire FR; 3D-Separator; ERIC) a total of 40 retrieval maneuvers (10 per retriever) were performed by deploying the retrievers within the migrated coils and trying to trap parts of the coils by advancing the microcatheter. Retrieval rates, retrieval duration, retrieval attempts, and complications were assessed.

Results

Overall coil retrieval was successful in 31 of 40 cases (77.5%). Retrieval rates using the Trevo ProVue (9/10), Solitaire FR (8/10), and the 3D-Separator (9/10) were higher than when using the ERIC (5/10). Duration of retrieval and retrieval attempts were significantly higher using the ERIC (p < 0.05). Complications like inadvertent deployment were only observed in one case using the Solitaire FR. Additional entrapment of the coil-retriever complex at the intermediate catheter was seen in 9 cases (22.5%). There was no case of vasospasm, perforation, or dissection.

Conclusions

Retrieval of migrated volume coils using new-generation clot retrievers is a feasible and effective method. Retrieval rates and duration with the Trevo ProVue, Solitaire FR, and 3D-Separator are superior when compared to the ERIC.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

Abbreviations

PC 400:

Penumbra 400 Coil System

ERIC:

Embolus Retriever with Interlinked Cages

IC:

Intermediate Catheter

DSA:

Digital Subtraction Angiography

WEB:

Woven Endobridge Device

References

  1. Molyneux AJ, Kerr RS, Yu LM, Clarke M, Sneade M, Yarnold JA, Sandercock P; International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group. International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet. 2005;366:809–17.

    Article  Google Scholar 

  2. Scott RB, Eccles F, Molyneux AJ, Kerr RS, Rothwell PM, Carpenter K. Improved cognitive outcomes with endovascular coiling of ruptured intracranial aneurysms: neuropsychological outcomes from the International Subarachnoid Aneurysm Trial (ISAT). Stroke. 2010;41:1743–7.

    Article  Google Scholar 

  3. Li H, Pan R, Wang H, Rong X, Yin Z, Milgrom DP, Shi X, Tang Y, Peng Y. Clipping versus coiling for ruptured intracranial aneurysms: a systematic review and meta-analysis. Stroke. 2013;44:29–37.

    Article  Google Scholar 

  4. Milburn J, Pansara AL, Vidal G, Martinez RC. Initial experience using the Penumbra coil 400: comparison of aneurysm packing, cost effectiveness, and coil efficiency. J Neurointerv Surg. 2014;6:121–4.

    Article  Google Scholar 

  5. Mascitelli JR, Patel AB, Polykarpou MF, Patel AA, Moyle H. Analysis of early angiographic outcome using unique large diameter coils in comparison with standard coils in the embolization of cerebral aneurysms: a retrospective review. J Neurointerv Surg. 2015;7:126–30.

    Article  Google Scholar 

  6. Kulcsár Z, Wanke I, Rüfenacht D, Wetzel SG, Göricke S, Kolia K, Quarfordt S, Calvert J, Hawk H, Baxter B. Safety and effectiveness of large volume coils in the treatment of small aneurysms. J Neurointerv Surg. 2016;8:1260–3.

    Article  Google Scholar 

  7. Berge J, Gariel F, Marnat G, Dousset V. PC400 volumetric coils minimize radiation, reduce procedure time and optimize packing density during endovascular treatment in medium sized cerebral aneurysms. J Neuroradiol. 2016;43:37–42.

    Article  Google Scholar 

  8. Guglielmi G, Viñuela F, Dion J, Duckwiler G. Electrothrombosis of saccular aneurysms via endovascular approach. Part 2: preliminary clinical experience. J Neurosurg. 1991;75:8–14.

    Article  CAS  Google Scholar 

  9. White PM, Lewis SC, Nahser H, Sellar RJ, Goddard T, Gholkar A; HELPS Trial Collaboration. Hydrocoil endovascular aneurysm occlusion and packing study (HELPS trial): procedural safety and operator-assessed efficacy results. AJNR Am J Neuroradiol. 2008;29:217–23.

    Article  CAS  Google Scholar 

  10. Henkes H, Fischer S, Weber W, Miloslavski E, Felber S, Brew S, Kuehne D. Endovascular coil occlusion of 1811 Intracranial aneurysms: early angiographic and clinical results. Neurosurgery. 2004;54:268–85.

    Article  Google Scholar 

  11. Leslie-Mazwi TM, Heddier M, Nordmeyer H, Stauder M, Velasco A, Mosimann PJ, Chapot R. Stent retriever use for retrieval of displaced microcoils: a consecutive case series. AJNR Am J Neuroradiol. 2013;34:1996–9.

    Article  CAS  Google Scholar 

  12. Ding D, Liu KC. Management strategies for intraprocedural coil migration during endovascular treatment of intracranial aneurysms. J Neurointerv Surg. 2014;6:428–31.

    Article  Google Scholar 

  13. Lee C‑Y. Use of wire as a snare for endovascular retrieval of displaced or stretched coils: rescue from a technical complication. Neuroradiology. 2011;53:31–5.

    Article  Google Scholar 

  14. Fiorella D, Albuquerque FC, Deshmukh VR, McDougall CG. Monorail snare technique for the recovery of stretched platinum coils: technical case report. Neurosurgery. 2005;57:E210.

    Article  Google Scholar 

  15. Dinc H, Kuzeyli K, Kosucu P, Sari A, Cekirge S. Retrieval of prolapsed coils during endovascular treatment of cerebral aneurysms. Neuroradiology. 2006;48:269–72.

    Article  Google Scholar 

  16. Henkes H, Lowens S, Preiss H, Reinartz J, Miloslavski E, Kühne D. A new device for endovascular coil retrieval from intracranial vessels: alligator retrieval device. AJNR Am J Neuroradiol. 2006;27:327–9.

    CAS  PubMed  Google Scholar 

  17. O’Hare A, Brennan P, Thornton J. Retrieval of a migrated coil using an X6 MERCI device. Interv Neuroradiol. 2009;15:99–102.

    Article  Google Scholar 

  18. O’Hare AM, Rogopoulos AM, Stracke PC, Chapot RG. Retrieval of displaced coil using a Solitaire(®) stent. Clin Neuroradiol. 2010;20:251–4.

    Article  Google Scholar 

  19. Hopf-Jensen S, Hensler HM, Preiß M, Müller-Hülsbeck S. Solitaire® stent for endovascular coil retrieval. J Clin Neurosci. 2013;20:884–6.

    Article  CAS  Google Scholar 

  20. Singh DP, Kwon SC, Huang L, Lee WJ. Retrieval of distally migrated coils with detachable Intracranial stent during coil embolization of cerebral aneurysm. J Cerebrovasc Endovasc Neurosurg. 2016;18:48–54.

    Article  Google Scholar 

  21. Nas OF, Kacar E, Kaya A, Erdogan C, Hakyemez B. Retrieval of a dislocated coil and stent-assisted coiling by Solitaire(®) stent during endovascular treatment of an intracranial aneurysm. Diagn Interv Imaging. 2016;97:381–4.

    Article  CAS  Google Scholar 

  22. Liu KC, Ding D, Starke RM, Geraghty SR, Jensen ME. Intraprocedural retrieval of migrated coils during endovascular aneurysm treatment with the Trevo Stentriever device. J Clin Neurosci. 2014;21:503–6.

    Article  Google Scholar 

  23. Nikoubashman O, Pjontek R, Brockmann MA, Tolba R, Wiesmann M. Retrieval of migrated coils with stent retrievers: an animal study. AJNR Am J Neuroradiol. 2015;36:1162–6.

    Article  CAS  Google Scholar 

  24. Simgen A, Ley D, Roth C, Cattaneo GF, Mühl-Benninghaus R, Müller A, Körner H, Kim YJ, Scheller B, Reith W, Yilmaz U. Evaluation of occurring complications after flow diverter treatment of elastase-induced aneurysm in rabbits using micro-CT and MRI at 9.4 T. Neuroradiology. 2016;58:987–96.

    Article  Google Scholar 

  25. Kabbasch C, Mpotsaris A, Chang DH, Hiß S, Dorn F, Behme D, Onur O, Liebig T. Mechanical thrombectomy with the Trevo ProVue device in ischemic stroke patients: does improved visibility translate into a clinical benefit? J Neurointerv Surg. 2016;8:778–82.

    Article  Google Scholar 

  26. Mpotsaris A, Bussmeyer M, Weber W. Mechanical thrombectomy with the penumbra 3D separator and lesional aspiration: technical feasibility and clinical outcome. Clin Neuroradiol. 2014;24:245–50.

    Article  CAS  Google Scholar 

  27. Behme D, Kowoll A, Mpotsaris A, Hader C, Hechelhammer L, Weber J, Weber W. Multicenter clinical experience in over 125 patients with the Penumbra Separator 3D for mechanical thrombectomy in acute ischemic stroke. J Neurointerv Surg. 2016;8:8–12.

    Article  Google Scholar 

  28. Kahles T, Garcia-Esperon C, Zeller S, Hlavica M, Añon J, Diepers M, Nedeltchev K, Remonda L. Mechanical thrombectomy using the new ERIC retrieval device is feasible, efficient, and safe in acute ischemic stroke: a swiss stroke center experience. AJNR Am J Neuroradiol. 2016;37:114–9.

    Article  CAS  Google Scholar 

  29. Raoult H, Redjem H, Bourcier R, Gaultier-Lintia A, Daumas-Duport B, Ferré JC, Eugène F, Fahed R, Bartolini B, Piotin M, Desal H, Gauvrit JY, Blanc R. Mechanical thrombectomy with the ERIC retrieval device: initial experience. J Neurointerv Surg. 2016 May 17. [Epub ahead of print]

    Article  PubMed  Google Scholar 

  30. Roth C, Junk D, Papanagiotou P, Keuler A, Körner H, Schumacher M, Reith W. A comparison of 2 stroke devices: the new Aperio clot-removal device and the solitaire AB/FR. AJNR Am J Neuroradiol. 2012;33:1317–20.

    Article  CAS  Google Scholar 

  31. Luo CB, Chang FC, Teng MM, Guo WY, Chang CY. Stent management of coil herniation in embolization of internal carotid aneurysms. AJNR Am J Neuroradiol. 2008;29:1951–5.

    Article  Google Scholar 

  32. Kwon HJ, Chueh JY, Puri AS, Koh HS. Early detachment of the Solitaire stent during thrombectomy retrieval: an in vitro investigation. J Neurointerv Surg. 2015;7:114–7.

    Article  Google Scholar 

  33. Yub Lee S, Won Youn S, Kyun Kim H, Rok Do Y. Inadvertent detachment of a retrievable Intracranial Stent: review of manufacturer and user facility device experience. Neuroradiol J. 2015;28:172–6.

    Article  Google Scholar 

  34. Masoud H, Nguyen TN, Martin CO, Holloway WE, Ambekar S, Yavagal DR, Haussen DC, Nogueira R, Lozano DJ, Puri A, Quateen A, Iancu D, Abraham MG, Chen M, Mehta S, Malisch T, Marden F, Novakovic R, Roy D, Weill A, Norbash AM. Inadvertent Stent retriever detachment: a multicenter case series and review of device experience FDA reports. Interv Neurol. 2016;4:75–82.

    Article  Google Scholar 

Download references

Acknowledgements

We would like to thank the team of the Institute for Clinical & Experimental Surgery, Saarland University, for supporting this study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Andreas Simgen.

Ethics declarations

Conflict of interest

A. Simgen, T. Tomori, R. Mühl-Bennighaus, H. Bomberg, U. Yilmaz, H. Körner, M.W. Laschke, M.D. Menger and W. Reith declare that they have no competing interests.

Ethical standards

The experiments were approved by the local animal protection committee of the Saarland University and conducted in accordance with the European legislation on the protection of animals (Directive 2010/63/EU) and the NIH guidelines on the care and use of laboratory animals (NIH publication #85-23 Rev. 1985).

Additional information

A. Simgen: Interventions, angiographic evaluation and manuscript writing. T. Tomori: Animal care and manuscript writing. R. Mühl-Benninghaus: Animal care and CT scanning. H. Bomberg: Statistical analysis. U. Yilmaz: Angiographic evaluation. H. Körner: Angiographic evaluation. M.W. Laschke: Animal care and manuscript writing. M.D. Menger: Project development and manuscript writing. W. Reith: Project development, interventions, angiographic evaluation and manuscript writing.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Simgen, A., Tomori, T., Mühl-Benninghaus, R. et al. Retrieval of Migrated Volume Coils Using Different Clot Retrievers in a Porcine Model. Clin Neuroradiol 28, 593–600 (2018). https://doi.org/10.1007/s00062-017-0600-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00062-017-0600-1

Keywords

Navigation