Abstract
Purpose
During endovascular treatment of intracranial aneurysms using coils, migration is a serious complication that increases neurological morbidity. The aim of this experimental study was to assess the effectiveness and complications of retrieving volume coils with different, currently available clot retrievers in a porcine model.
Methods
Volume coils of three-dimensional (3D) shape and different sizes were placed in the axillary artery of pigs. By means of 4 different clot retrievers (Trevo ProVue; Solitaire FR; 3D-Separator; ERIC) a total of 40 retrieval maneuvers (10 per retriever) were performed by deploying the retrievers within the migrated coils and trying to trap parts of the coils by advancing the microcatheter. Retrieval rates, retrieval duration, retrieval attempts, and complications were assessed.
Results
Overall coil retrieval was successful in 31 of 40 cases (77.5%). Retrieval rates using the Trevo ProVue (9/10), Solitaire FR (8/10), and the 3D-Separator (9/10) were higher than when using the ERIC (5/10). Duration of retrieval and retrieval attempts were significantly higher using the ERIC (p < 0.05). Complications like inadvertent deployment were only observed in one case using the Solitaire FR. Additional entrapment of the coil-retriever complex at the intermediate catheter was seen in 9 cases (22.5%). There was no case of vasospasm, perforation, or dissection.
Conclusions
Retrieval of migrated volume coils using new-generation clot retrievers is a feasible and effective method. Retrieval rates and duration with the Trevo ProVue, Solitaire FR, and 3D-Separator are superior when compared to the ERIC.
Similar content being viewed by others
Abbreviations
- PC 400:
-
Penumbra 400 Coil System
- ERIC:
-
Embolus Retriever with Interlinked Cages
- IC:
-
Intermediate Catheter
- DSA:
-
Digital Subtraction Angiography
- WEB:
-
Woven Endobridge Device
References
Molyneux AJ, Kerr RS, Yu LM, Clarke M, Sneade M, Yarnold JA, Sandercock P; International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group. International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet. 2005;366:809–17.
Scott RB, Eccles F, Molyneux AJ, Kerr RS, Rothwell PM, Carpenter K. Improved cognitive outcomes with endovascular coiling of ruptured intracranial aneurysms: neuropsychological outcomes from the International Subarachnoid Aneurysm Trial (ISAT). Stroke. 2010;41:1743–7.
Li H, Pan R, Wang H, Rong X, Yin Z, Milgrom DP, Shi X, Tang Y, Peng Y. Clipping versus coiling for ruptured intracranial aneurysms: a systematic review and meta-analysis. Stroke. 2013;44:29–37.
Milburn J, Pansara AL, Vidal G, Martinez RC. Initial experience using the Penumbra coil 400: comparison of aneurysm packing, cost effectiveness, and coil efficiency. J Neurointerv Surg. 2014;6:121–4.
Mascitelli JR, Patel AB, Polykarpou MF, Patel AA, Moyle H. Analysis of early angiographic outcome using unique large diameter coils in comparison with standard coils in the embolization of cerebral aneurysms: a retrospective review. J Neurointerv Surg. 2015;7:126–30.
Kulcsár Z, Wanke I, Rüfenacht D, Wetzel SG, Göricke S, Kolia K, Quarfordt S, Calvert J, Hawk H, Baxter B. Safety and effectiveness of large volume coils in the treatment of small aneurysms. J Neurointerv Surg. 2016;8:1260–3.
Berge J, Gariel F, Marnat G, Dousset V. PC400 volumetric coils minimize radiation, reduce procedure time and optimize packing density during endovascular treatment in medium sized cerebral aneurysms. J Neuroradiol. 2016;43:37–42.
Guglielmi G, Viñuela F, Dion J, Duckwiler G. Electrothrombosis of saccular aneurysms via endovascular approach. Part 2: preliminary clinical experience. J Neurosurg. 1991;75:8–14.
White PM, Lewis SC, Nahser H, Sellar RJ, Goddard T, Gholkar A; HELPS Trial Collaboration. Hydrocoil endovascular aneurysm occlusion and packing study (HELPS trial): procedural safety and operator-assessed efficacy results. AJNR Am J Neuroradiol. 2008;29:217–23.
Henkes H, Fischer S, Weber W, Miloslavski E, Felber S, Brew S, Kuehne D. Endovascular coil occlusion of 1811 Intracranial aneurysms: early angiographic and clinical results. Neurosurgery. 2004;54:268–85.
Leslie-Mazwi TM, Heddier M, Nordmeyer H, Stauder M, Velasco A, Mosimann PJ, Chapot R. Stent retriever use for retrieval of displaced microcoils: a consecutive case series. AJNR Am J Neuroradiol. 2013;34:1996–9.
Ding D, Liu KC. Management strategies for intraprocedural coil migration during endovascular treatment of intracranial aneurysms. J Neurointerv Surg. 2014;6:428–31.
Lee C‑Y. Use of wire as a snare for endovascular retrieval of displaced or stretched coils: rescue from a technical complication. Neuroradiology. 2011;53:31–5.
Fiorella D, Albuquerque FC, Deshmukh VR, McDougall CG. Monorail snare technique for the recovery of stretched platinum coils: technical case report. Neurosurgery. 2005;57:E210.
Dinc H, Kuzeyli K, Kosucu P, Sari A, Cekirge S. Retrieval of prolapsed coils during endovascular treatment of cerebral aneurysms. Neuroradiology. 2006;48:269–72.
Henkes H, Lowens S, Preiss H, Reinartz J, Miloslavski E, Kühne D. A new device for endovascular coil retrieval from intracranial vessels: alligator retrieval device. AJNR Am J Neuroradiol. 2006;27:327–9.
O’Hare A, Brennan P, Thornton J. Retrieval of a migrated coil using an X6 MERCI device. Interv Neuroradiol. 2009;15:99–102.
O’Hare AM, Rogopoulos AM, Stracke PC, Chapot RG. Retrieval of displaced coil using a Solitaire(®) stent. Clin Neuroradiol. 2010;20:251–4.
Hopf-Jensen S, Hensler HM, Preiß M, Müller-Hülsbeck S. Solitaire® stent for endovascular coil retrieval. J Clin Neurosci. 2013;20:884–6.
Singh DP, Kwon SC, Huang L, Lee WJ. Retrieval of distally migrated coils with detachable Intracranial stent during coil embolization of cerebral aneurysm. J Cerebrovasc Endovasc Neurosurg. 2016;18:48–54.
Nas OF, Kacar E, Kaya A, Erdogan C, Hakyemez B. Retrieval of a dislocated coil and stent-assisted coiling by Solitaire(®) stent during endovascular treatment of an intracranial aneurysm. Diagn Interv Imaging. 2016;97:381–4.
Liu KC, Ding D, Starke RM, Geraghty SR, Jensen ME. Intraprocedural retrieval of migrated coils during endovascular aneurysm treatment with the Trevo Stentriever device. J Clin Neurosci. 2014;21:503–6.
Nikoubashman O, Pjontek R, Brockmann MA, Tolba R, Wiesmann M. Retrieval of migrated coils with stent retrievers: an animal study. AJNR Am J Neuroradiol. 2015;36:1162–6.
Simgen A, Ley D, Roth C, Cattaneo GF, Mühl-Benninghaus R, Müller A, Körner H, Kim YJ, Scheller B, Reith W, Yilmaz U. Evaluation of occurring complications after flow diverter treatment of elastase-induced aneurysm in rabbits using micro-CT and MRI at 9.4 T. Neuroradiology. 2016;58:987–96.
Kabbasch C, Mpotsaris A, Chang DH, Hiß S, Dorn F, Behme D, Onur O, Liebig T. Mechanical thrombectomy with the Trevo ProVue device in ischemic stroke patients: does improved visibility translate into a clinical benefit? J Neurointerv Surg. 2016;8:778–82.
Mpotsaris A, Bussmeyer M, Weber W. Mechanical thrombectomy with the penumbra 3D separator and lesional aspiration: technical feasibility and clinical outcome. Clin Neuroradiol. 2014;24:245–50.
Behme D, Kowoll A, Mpotsaris A, Hader C, Hechelhammer L, Weber J, Weber W. Multicenter clinical experience in over 125 patients with the Penumbra Separator 3D for mechanical thrombectomy in acute ischemic stroke. J Neurointerv Surg. 2016;8:8–12.
Kahles T, Garcia-Esperon C, Zeller S, Hlavica M, Añon J, Diepers M, Nedeltchev K, Remonda L. Mechanical thrombectomy using the new ERIC retrieval device is feasible, efficient, and safe in acute ischemic stroke: a swiss stroke center experience. AJNR Am J Neuroradiol. 2016;37:114–9.
Raoult H, Redjem H, Bourcier R, Gaultier-Lintia A, Daumas-Duport B, Ferré JC, Eugène F, Fahed R, Bartolini B, Piotin M, Desal H, Gauvrit JY, Blanc R. Mechanical thrombectomy with the ERIC retrieval device: initial experience. J Neurointerv Surg. 2016 May 17. [Epub ahead of print]
Roth C, Junk D, Papanagiotou P, Keuler A, Körner H, Schumacher M, Reith W. A comparison of 2 stroke devices: the new Aperio clot-removal device and the solitaire AB/FR. AJNR Am J Neuroradiol. 2012;33:1317–20.
Luo CB, Chang FC, Teng MM, Guo WY, Chang CY. Stent management of coil herniation in embolization of internal carotid aneurysms. AJNR Am J Neuroradiol. 2008;29:1951–5.
Kwon HJ, Chueh JY, Puri AS, Koh HS. Early detachment of the Solitaire stent during thrombectomy retrieval: an in vitro investigation. J Neurointerv Surg. 2015;7:114–7.
Yub Lee S, Won Youn S, Kyun Kim H, Rok Do Y. Inadvertent detachment of a retrievable Intracranial Stent: review of manufacturer and user facility device experience. Neuroradiol J. 2015;28:172–6.
Masoud H, Nguyen TN, Martin CO, Holloway WE, Ambekar S, Yavagal DR, Haussen DC, Nogueira R, Lozano DJ, Puri A, Quateen A, Iancu D, Abraham MG, Chen M, Mehta S, Malisch T, Marden F, Novakovic R, Roy D, Weill A, Norbash AM. Inadvertent Stent retriever detachment: a multicenter case series and review of device experience FDA reports. Interv Neurol. 2016;4:75–82.
Acknowledgements
We would like to thank the team of the Institute for Clinical & Experimental Surgery, Saarland University, for supporting this study.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
A. Simgen, T. Tomori, R. Mühl-Bennighaus, H. Bomberg, U. Yilmaz, H. Körner, M.W. Laschke, M.D. Menger and W. Reith declare that they have no competing interests.
Ethical standards
The experiments were approved by the local animal protection committee of the Saarland University and conducted in accordance with the European legislation on the protection of animals (Directive 2010/63/EU) and the NIH guidelines on the care and use of laboratory animals (NIH publication #85-23 Rev. 1985).
Additional information
A. Simgen: Interventions, angiographic evaluation and manuscript writing. T. Tomori: Animal care and manuscript writing. R. Mühl-Benninghaus: Animal care and CT scanning. H. Bomberg: Statistical analysis. U. Yilmaz: Angiographic evaluation. H. Körner: Angiographic evaluation. M.W. Laschke: Animal care and manuscript writing. M.D. Menger: Project development and manuscript writing. W. Reith: Project development, interventions, angiographic evaluation and manuscript writing.
Rights and permissions
About this article
Cite this article
Simgen, A., Tomori, T., Mühl-Benninghaus, R. et al. Retrieval of Migrated Volume Coils Using Different Clot Retrievers in a Porcine Model. Clin Neuroradiol 28, 593–600 (2018). https://doi.org/10.1007/s00062-017-0600-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00062-017-0600-1