Influence of material choice on the force delivery of bimaxillary tooth positioners on canine malpositions

Der Einfluss der Materialwahl auf die Kraftabgabe von bimaxillären Zahnpositionern bei Fehlstellungen der Eckzähne

  • Original article
  • Published:
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie Aims and scope Submit manuscript

Abstract

Objective

Since their introduction in 1945, tooth positioners have been used to treat a range of malpositions. Although the original appliance was made of natural rubber, today’s tooth positioners are fabricated from various elastic, transparent materials. It was the aim of this study to evaluate the forces and moments produced by current positioners on various upper canine malpositions.

Material and methods

Seven positioners of different materials were tested on 11 upper canine malpositions: 0.25, 0.5, 1 mm supraposition; 0.25, 0.5, 1 mm infraposition; 5°, 10°, 20° mesiorotation; 5° mesioinclination, 5° buccal root torque. We measured forces and moments in vitro after bite closure by 0.5 mm, and opening by 1 mm using a three-dimensional robotic device. All measurements were taken in a temperature-controlled environment at 36 °C.

Results

The forces and moments measured at the canine varied greatly among the different positioners, with the thermoformed EVA positioner showing much greater forces and moments in almost all malpositions. At initial closure, we observed intrusive forces of 6–32 N for suprapositions, 0–11 N intrusive forces for infrapositions, 0–20 Nmm for mesiorotations, 6–12 Nmm for mesioinclinations, and − 23 Nmm to 5 Nmm for buccal root torque. All positioners were most effective on suprapositioned teeth. Very low or negligible correctional forces and moments in conjunction with all infrapositions and 5° with rotation were noted. Labial root torque led to unpredictable moments.

Conclusion

Positioner material plays a major role in delivering forces to the teeth. However, tooth positioners did not induce corrective forces in all the malpositions tested. Clinically relevant correctional forces or moments in conjunction with all suprapositions, rotations of 10° and 20° as well as mesial tipping of the canine were observed.

Zusammenfassung

Ziel

Seit ihrer Einführung 1945 sind Zahnpositioner zur Behandlung einer Reihe von Fehlstellungen eingesetzt worden. Die ursprüngliche Apparatur bestand aus Naturgummi; heute dagegen werden Zahnpositioner aus verschiedenen elastischen, transparenten Materialien hergestellt. Ziel der vorliegenden Arbeit war die Bewertung der Kräfte und Drehmomente, mit denen die derzeit verwendeten Positioner auf verschiedene Fehlstellungen eines oberen Eckzahns einwirken.

Material und Methodik

Sieben Positioner aus verschiedenen Materialien wurden an 11 Fehlstellungen eines oberen Eckzahns getestet: 0,25, 0,5, 1 mm Supraposition; 0,25, 0,5, 1 mm Infraposition; 5°, 10°, 20° Mesiorotation; 5° Mesioinklination, 5° bukkaler Wurzeltorque. Die Kräfte und Drehmomente nach einer Schließung des Bisses um 0,5 mm und einer Öffnung um 1 mm wurden in einer In-vitro-Versuchsanordnung unter Verwendung einer dreidimensionalen robotisierten Vorrichtung getestet. Alle Messungen wurden in einer Umgebung mit einer kontrollierten Temperatur von 36°C vorgenommen.

Ergebnisse

Die am Eckzahn gemessenen Kräfte und Drehmomente wichen für die verschiedenen Positioner stark voneinander ab, wobei der thermogeformte EVA-Positioner für fast alle Fehlstellungen eine erheblich größere Kraft- und Momentabgabe zeigte. Beim initialen Kieferschluss wurden für Suprapositionen Intrusionskräfte von 6–32 N, für Infrapositionen Intrusionskräfte von 0−11 N, für Mesioinklinationen 0−20 Nmm und für bukkalen Wurzeltorque − 23−5 Nmm gemessen. Alle Positioner zeigten sich bei Zähnen mit Suprapositionen am wirksamsten. Sehr geringe oder vernachlässigbare Kräfte/Drehmomente wurden bei allen Infrapositionen und 5° Rotation festgestellt. Beim labialen Wurzeltorque ergaben sich nicht vorhersagbare Drehmomente.

Schlussfolgerung

Die für Positioner verwendeten Materialien spielen eine wichtige Rolle bei der Kraftabgabe an die Zähne. Jedoch wurden durch die Positioner nicht bei allen getesteten Fehlstellungen Korrekturkräfte induziert. Klinisch relevante Korrekturkräfte oder Drehmomente fanden sich für alle Suprapositionen, Rotationen von 10° und 20° sowie für Mesialkippungen des Eckzahns.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

References

  1. Bourauel C, Drescher D, Wolter H (1997) Können die Kraftsysteme von Positionern durch den zusätzlichen Einsatz von Attachments beeinflusst werden? Eine experimentelle Studie. Kieferorthop 11:183–190

    Google Scholar 

  2. Bowman SJ, Carano A (2002) Short-term, intensive use of the tooth positioner in case finishing. J Clin Orthod 36:216–219

    PubMed  Google Scholar 

  3. Cottingham LL (1969) Gnathologic clear plastic positioner. Am J Orthod 55:23–31

    Article  PubMed  Google Scholar 

  4. Craig RG, Godwin WC (1967) Physical properties of materials for custom-made mouth protectors. J Mich State Dent Assoc 49:34–40

    PubMed  Google Scholar 

  5. Elsasser WA (1950) Some observations of the history and uses of the Kesling positioner. Am J Orthod 36:368–374

    Article  PubMed  Google Scholar 

  6. Kamada K, Tamura T, Hama Y et al (1984) Case reports on tooth positioners using LTV vinyl silicone rubber. J Nihon Univ Sch Dent 26:11–29

    Article  PubMed  Google Scholar 

  7. Kesling HD (1946) Coordinating the predetermined pattern and tooth positioner with conventional treatment. Am J Orthod Oral Surg 32:285–293

    Article  PubMed  Google Scholar 

  8. Kesling HD (1945) The philosophy of the tooth positioning appliance. Am J Orthod Oral Surg 31:297–304

    Article  Google Scholar 

  9. Lee BW (1965) Relationship between tooth-movement rate and estimated pressure applied. J Dent Res 44:1053–1057

    Article  PubMed  Google Scholar 

  10. Moore DJ, Glaser ZR, Tabacco MJ, Linebaugh MG (1977) Evaluation of polymeric materials for maxillofacial prosthetics. J Prosthet Dent 38:319–326

    Article  PubMed  Google Scholar 

  11. Nishiyama M, Kamada K, Horiuchi S (1977) A new construction of tooth positioner with LTV vinyl silicone rubber. J Nihon Univ Sch Dent 19:93–102

    Article  PubMed  Google Scholar 

  12. Park Y, Hartsfield JK, Katona TR, Roberts WE (2008) Tooth positioner effects on occlusal contacts and treatment outcomes. Angle Orthod 78(6):1050–1056

    Article  PubMed  Google Scholar 

  13. Permann I, Bantleon HP, Muchitsch AP, Droschl H (1989) Der Positioner – eine Möglichkeit zur Feineinstellung der Okklusion, objektiviert durch Kondymetermessungen der Diskrepanz zwischen IKP und RKP. Fortschr Kieferorthop 50:530–539

    Article  PubMed  Google Scholar 

  14. Posselt U (1956) Hinge opening axis of the mandible. Acta Odontol Scand 57(14):49–63

    Article  Google Scholar 

  15. Prestel B, Drescher D, Bourauel C (1995) Kräfte und Drehmomente beim Einsatz von Positionern. Kieferorthopädie 9:117–120

    Google Scholar 

  16. Reitan K (1957) Some factors determining the evaluation of forces in orthodontics. Am J Orthod 43:32–45

    Article  Google Scholar 

  17. Rost D, Schwarze CW, Hilgers RD (1995) Die Kraftabgabe von Positionern bei unter-schiedlicher Schneidezahnprotrusion. Eine In-vitro-Untersuchung. Fortschr Kieferorthop 56:104–109

    Article  PubMed  Google Scholar 

  18. Sander FG, Fröhls H (1989) Qualitative and quantitative Bestimmung des settling-Effekts durch die Positioner-Therapie. Inf Orthod Kieferorthop 21:311–318

    PubMed  Google Scholar 

  19. Smith R, Storey E (1952) The importance of force in orthodontics. Aust J Dent 56:291–304

    Google Scholar 

  20. Vorhies JM (1960) Short, intensive use of tooth positioners and an appraisal of the results. Angle Orthod 30:248–54

    Google Scholar 

  21. Warunek SP, Sorensen SE, Cunat JJ, Green LJ (1989) Physical and mechanical properties of elastomers in orthodontic positioners. Am J Orthod Dentofacial Orthop 95:388–400

    Article  PubMed  Google Scholar 

  22. Wells NE (1970) Application of the positioner appliance in the orthodontic treatment. Am J Orthod 58:351–366

    Article  PubMed  Google Scholar 

  23. Wichelhaus A, Hüffmeier S, Sander FG (2003) Dynamic functional force measurements on an anterior bite plane during the night. J Orofac Orthop 64:417–425

    Article  PubMed  Google Scholar 

Download references

Interessenkonflikt

Der korrespondierende Autor gibt an, dass kein Interessenkonflikt besteht.

Conflict of interest

The corresponding author states that there are no conflicts of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to L. Brauchli.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Lochmatter, D., Steineck, M. & Brauchli, L. Influence of material choice on the force delivery of bimaxillary tooth positioners on canine malpositions. J Orofac Orthop 73, 104–115 (2012). https://doi.org/10.1007/s00056-011-0067-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00056-011-0067-7

Keywords

Schlüsselwörter

Navigation