Half-value layer measurement: simple process method using radiochromic film

  • T. Gotanda
  • T. Katsuda
  • R. Gotanda
  • A. Tabuchi
  • K. Yamamoto
  • T. Kuwano
  • H. Yatake
  • K. Kashiyama
  • Y. Takeda
Scientific Papers

Abstract

Although the half-value layer (HVL) is one of the important parameters for QA and QC, constant monitoring has not been performed because the measurements using an ionization chamber (IC) are time-consuming and complicated. To solve these problems, the use of radiochromic film (GAFCHROMIC XR TYPE R: GAF-R) with step-shaped aluminum (Al) filters, referred to herein as the simple process method, has been developed. The measurement X-ray tube voltages were 120 kV, 100 kV, and 80 kV. The Al filter area, the full exposure area, and the unexposed area were set on the GAF-R so as to obtain correct data. The HVL was evaluated using the density attenuation ratio. The HVLs obtained using the GAFR and an IC dosimeter were compared. HVLs with X-ray tube voltages of 120 kV, 100 kV, and 80 kV using the GAF-R were 4.10 mm, 3.55 mm and 2.97 mm, respectively. The difference ratios of the HVLs using the GAF-R and the IC were 1.2%, 7.6%, and 10.0%, respectively. The HVL at 120 kV can be routinely and quickly measured using the simple process method. Therefore, an IC dosimeter is not needed for HVL measurements for QA and QC. However, the HVL measurements of low energy (100 kV and 80 kV) need attention.

Key words

half-value layer radiochromic film effective energy quality assurance quality control 

References

  1. 1.
    Siedband, M., Baiter, S., Brateman, L., Britt, W., Deutsch, G., Dobrin, R., Duerkes, R., Fields, T., Lawrence, D., Lin, P., Morgan, T., Properzio, W., Rossi, R., Rothenberg, L., Scheid, C., Starchman, D., Vucich, J., Waggener, R. and. Winkler, N.Basic Quality Control In Diagnostic Radiology, AAPM report No. 4, 1977.Google Scholar
  2. 2.
    Japanese Industrial Standards Committee,Methods of calibration for exposure meters, air kerma meters, air absorbed dose meters and dose-equivalent meters, JIS Z 4511, 2005.Google Scholar
  3. 3.
    Butson, M.J., Yu, P.K.N., Cheung, T. and Metcalfe, P., Radiochromic film for medical radiation dosimetry, Materials Science and Engineering R, 41:61–120, 2003.CrossRefGoogle Scholar
  4. 4.
    Cheung, T., Butson, M.J. and Yu, P.K.N.,Reflection spectrometry analysis of irradiated GAFCHROMIC XR type R radiochromic films, Appl Radiat Isot, 63(1):127–9, 2005.CrossRefPubMedGoogle Scholar
  5. 5.
    Gotanda, R., Katsuda, T., Gotanda, T., Eguchi, M., Takewa, S., Tabuchi, A. and Yatake, H.,Computed Tomography Phantom for Radiochromic Film Dosimetry, Australas Phys Eng Sci Med, 30(3): 194–199, 2007.PubMedCrossRefGoogle Scholar
  6. 6.
    Gotanda, R., Katsuda, T., Gotanda, T., Tabuchi, A., Yatake, H. and Takeda, Y.,Dose Distribution in Pediatric CT Head Examination: Phantom Study, IFMBE Proceedings 22: 775–779, 2008.CrossRefGoogle Scholar
  7. 7.
    Gotanda, R., Katsuda, T., Gotanda, T., Tabuchi, A., Yatake, H. and Takeda, Y.,Dose Distribution in Pediatric CT Head Examination using a New Phantom with Radiochromic Film, Australas Phys Eng Sci Med, 31(4):339–344, 2008.PubMedGoogle Scholar
  8. 8.
    Gotanda, T., Gotanda, R., Tabuchi, A., Kuwano, T., Yatake, H., Yamamoto, K. and Katsuda, T.,Visualization of Radiation Exposure for Computed Tomography, Health Sciences, 24, 394–403, 2008.Google Scholar
  9. 9.
    International Specialty Products web site. Available at: www.ispcorp.com 2006.Google Scholar
  10. 10.
    C. L. Fletcher and J. A. Mills,An assessment of Gaf Chromic film for measuring 50 kV and 100 kV percentage depth dose curves, Phys Med Biol, 53:209–218, 2008.CrossRefGoogle Scholar
  11. 11.
    Rink, A., Vitkin, I. A. and Jaffray, D. A.Energy dependence (75 kVp to 18 MV) of radiochromic films assessed using a real-time optical dosimeter, Med Phys 34:458–463, 2007.CrossRefPubMedGoogle Scholar
  12. 12.
    Gotanda, T., Katsuda, T., Gotanda, R., Tabuchi, A., Yamamoto, K., Kuwano, T., Yatake, H. and Takeda, Y.,Evaluation of effective energy for QA and QC: measurement of half-value layer using radiochromic film density, Australas. Phys. Eng. Sci. Med. 32(1), 2009.Google Scholar
  13. 13.
    Yimin, Z., Assen A. K., Vivek, M., Ali, S. M. and Jeffrey, F. W.Quantitative evaluation of radiochromic film response for two-dimensional dosimetry, Med. Phys. 24(2):223–231, 1997.CrossRefGoogle Scholar
  14. 14.
    Dini, S.A., Koona, R.A., Ashburn, J.R. and Ali, S.M.,Dosimetric evaluation of GAFCHROMIC XR type T and XR type R films, Am Coll Med Phys 6(1):114–134, 2005.Google Scholar
  15. 15.
    Butson, M.J., Cheung, T. and Yu, P.K.N.,XR type-R radiochromic film x-ray energy response, Phys Med Biol, 50(15):195–199, 2005.CrossRefGoogle Scholar
  16. 16.
    Butson, M.J., Cheung, T. and Yu, P.K.N.,Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range, Appl. Radiat. Isot. 64(1):60–2, 2006.CrossRefPubMedGoogle Scholar
  17. 17.
    Cheung, T., Butson, M.J. and Yu, P.K.N.,Post-irradiation colouration of Gafchromic EBT radiochromic film, Phys. Med. Biol. 50:281–285, 2005.CrossRefGoogle Scholar
  18. 18.
    Seltzer, S. M. and Hubbell, J. H.,Table and graphs of photon mass attenuation coefficients and mass energy-absorption coefficients for photon energies 1 keV to 20 MeV for elements Z=1 to 92 and some dosimetric materials, National Institute of Standards and Technology, 1995.Google Scholar

Copyright information

© Australasian College of Physical Scientists and Engineers in Medicine 2009

Authors and Affiliations

  • T. Gotanda
    • 1
    • 2
  • T. Katsuda
    • 3
  • R. Gotanda
    • 1
  • A. Tabuchi
    • 1
    • 4
  • K. Yamamoto
    • 1
    • 5
  • T. Kuwano
    • 1
    • 6
  • H. Yatake
    • 1
    • 7
  • K. Kashiyama
    • 1
    • 8
  • Y. Takeda
    • 1
  1. 1.Graduate School of Health SciencesOkayama UniversityOkayama-city, OkayamaJapan
  2. 2.Department of RadiologyOishi HospitalHiroshimaJapan
  3. 3.Faculty of Health Care SciencesHimeji Dokkyo UniversityHimejiJapan
  4. 4.Kawasaki Medical School Kawasaki HospitalOkayamaJapan
  5. 5.Department of RadiologyOsaka Cancer Prevention and Detection CenterOsakaJapan
  6. 6.Department of Diagnostic RadiologyOsaka Medical Center for Cancer and Cardiovascular DiseasesOsakaJapan
  7. 7.Department of RadiologyKaizuka City HospitalOsakaJapan
  8. 8.Department of Diagnostic RadiologyOsaka General CenterOsakaJapan

Personalised recommendations