, Volume 8, Issue 4, pp 239–243 | Cite as

Optimal dose of injection in activation study with O-15 water and PET

  • Norihiro Sadato
  • Yoshiharu Yonekura
  • Yasuhiro Magata
  • Sadahiko Nishizawa
  • Koichi Ishizu
  • Hidehiko Okazawa
  • Tatsuro Tsuchida
  • Fumiko Tanaka
  • Nagara Tamaki
  • Satoshi Sasayama
  • Hiroshi Shibasaki
  • Junji Konishi
Original Article

Abstract

In activation studies with the bolus method for O-15 water and PET, the radiotracer concentration may reach the limits of the system in terms of dead time correction and accidental coincidence. To obtain the optimal injection dose of O-15 water, we performed a normal volunteer study to evaluate the relationship between the injected dose and the radioactivity concentration in the brain and a phantom study to evaluate the performance of the PET scanner (PCT3600W) under high count rate conditions and the effect of averaging on the signal to noise ratio for the PET images.

A linear relationship was noted between the injected dose (normalized for each body weight: x) and the mean radiotracer concentration in the brain measured by PET (y) (y = 2.52 + 30.1x, n = 64, r = 0.87, p < 0.001). The percent error in the measurement of radioactivity with PET was within ± 5% in the 100 to 2000 nCi/m/ (3.7–74 KBq/mZ) range. Below 100 nCi/mZ (3.7 KBq/mZ), the percent error increased due to the rapid increase in noise in the reconstructed images. Over 1000 nCi/ mZ (37 KBq/mZ), on the other hand, the noise was almost unchanged.

With our PET scanner, the optimal range of the radiotracer concentration in the brain is below 1000 nCi/mZ (37 KBq/mZ), corresponding to an injection dose of 33 mCi (1.22 GBq)/60 kg body weight. With the same total dose, the increment of number of repeated measurements for averaging provided the better signal to noise ratio. In designing a paradigm for an activation PET study, the injection dose and the number of repeated measurements for averaging should be considered.

Key words

PET activation study O-15 water optimum dose signal to noise ratio 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Fox PT, Burton H, Raichle ME. Mapping human somatosensory cortex with positron emission tomography.J Neurosurg 67: 34–43, 1987.PubMedCrossRefGoogle Scholar
  2. 2.
    Fox PT, Mintun MA, Raichle ME, Herscovitch P. A non-invasive approach to quantitative functional brain mapping with H2 15O and positron emission tomography.J Cereb Blood Flow Metab 4: 329–333, 1984.PubMedGoogle Scholar
  3. 3.
    Mazziotta JC, Huang S-C, Phelps ME, Carson RE, MacDonald NS, Mahoney K. A noninvasive positron computed tomography technique using oxygen-15 labeled water for the evaluation of neurobehavioral task batteries.J Cereb Blood Flow Metab 5: 70–78, 1985.PubMedGoogle Scholar
  4. 4.
    Karp JS, Daube-Witherspoon ME, Hoffman EJ, Lewellen TK, Links JM, Wong W-H, et al. Performance standards in positron emission tomography.J Nucl Med 32: 2342–2350, 1991.PubMedGoogle Scholar
  5. 5.
    Mukai T, Senda M, Yonekura Y, Tamaki N, Yamamoto K, Nishizawa S, et al. System design and performance of a newly developed high resolution PET scanner using double wobbling mode.J Nucl Med 29: 877, 1988.Google Scholar
  6. 6.
    Endo M, Hukuda H, Suhara T, Matsumoto T, Iinuma T, Yamasaki T, et al. Design and performance of PCT-3600W (15-slice type): a whole-body positron emission tomograph.J Nucl Med 32: 1061, 1991.Google Scholar
  7. 7.
    Sadato N, Yonekura Y, Senda M, Iwasaki Y, Matoba N, Tamaki N, et al. PET and the autoradiographic method with continuous inhalation of oxygen-15-gas: theoretical analysis and comparison with conventional steady-state methods.J Nucl Med 34: 1672–1680, 1993.PubMedGoogle Scholar
  8. 8.
    Shibasaki H, Sadato N, Lyshkpw H, Yonekura Y, Honda M, Nagamine T, et al. Both primary motor cortex and supple-mentary motor area play an important role in complex finger movement.Brain 116: 1387–1398, 1993.PubMedCrossRefGoogle Scholar
  9. 9.
    Sadato N, Yonekura Y, Senda M, Magata Y, Iwasaki Y, Matoba N, et al. Noninvasive measurement of regional cerebral blood flow change with H2 15O and positron emission tomography using a mechanical injector and a standard arterial input function.IEEE Trans Med Imag 12: 703–710, 1993.CrossRefGoogle Scholar
  10. 10.
    Hoffman EJ, Huang S, Phelps ME, Kuhl DE. Quantitation in positron emission computed tomography: 4. Effect of accidental coincidences.J Comput Assist Tomogr 5: 391–400, 1981.PubMedCrossRefGoogle Scholar
  11. 11.
    Iida H, Kanno I, Miura S. Rapid measurement of cerebral blood flow with positron emission tomography.In Exploring Brain Functional Anatomy With Positron Tomography, Chadwick DJ, Whelan J (ed.), Chichester, Wiley, pp. 23–42, 1991.Google Scholar
  12. 12.
    Evans AC, Thompson CJ, Marrett S, Meyer E, Mazza M. Performance evaluation of the PC-2048: a new 15-slice encoded-crystal PET scanner for neurological studies.IEEE Trans Med Imag 10: 90–98, 1991CrossRefGoogle Scholar
  13. 13.
    KearfottKJ. Absorbed dose estimates forpositron emission tomography (PET): C15O,11CO, and CO15O.J Nucl Med 23: 1031–1037, 1982.PubMedGoogle Scholar
  14. 14.
    Steinmetz H, Furst G, Freund H-J. Variation of perisylvian and calcarine anatomic landmarks within stereotaxic pro-portional coordinates.AJNR 11: 1123–1130, 1990.PubMedGoogle Scholar

Copyright information

© Springer-Verlag 1994

Authors and Affiliations

  • Norihiro Sadato
    • 1
  • Yoshiharu Yonekura
    • 2
  • Yasuhiro Magata
    • 1
  • Sadahiko Nishizawa
    • 1
  • Koichi Ishizu
    • 1
  • Hidehiko Okazawa
    • 1
  • Tatsuro Tsuchida
    • 1
  • Fumiko Tanaka
    • 1
  • Nagara Tamaki
    • 1
  • Satoshi Sasayama
    • 1
  • Hiroshi Shibasaki
    • 2
  • Junji Konishi
    • 1
  1. 1.Department of Nuclear MedicineKyoto University Faculty of MedicineJapan
  2. 2.Department of Brain PathophysiologyKyoto University Faculty of MedicineKyotoJapan

Personalised recommendations