Abstract

Objective

Adenosine, which has been used for a myocardial perfusion scan, shows rapid clearance from blood because of its short half-life of <10 seconds. This simulation study evaluates influences of modes of radionuclide injection on ventricular adenosine concentration when one intravenous injection line is used.

Methods

Assuming that radionuclide injection is a unit impulse, time-activity curves were measured in the left ventricle (LV) and fitted by a gamma function. Typical patterns of concentration fluctuation when adenosine infusion was temporarily modified were calculated by the convolution integral of input function and unit impulse response. Variation of concentration was measured by experiments using continuous99mTc injection and co-infusion of water via a three-way stopcock. Modes of co-infusion with various infusion speeds and volumes were examined.

Results

Intermission of adenosine infusion and rapid displacement by radionuclide co-injection significantly influenced the adenosine concentration in LV. Intermission of adenosine infusion for 2 seconds caused a 15% decrease in the adenosine concentration in the left ventricle. When a square-shaped input was assumed, a three-fold higher concentration of adenosine for 3 seconds created by radionuclide injection resulted in a +42% increase in the LV concentration. Based on a measured input function, radionuclide injection using three-way stopcock through one route caused a two- to three-fold increase in the steady concentration in the vein just after injection. When 0.5 ml of radionuclide was slowly co-injected, with three ways opened, it caused a relatively low fluctuation, creating a +34% to -47% change in concentration of LV. A flush of radionuclide with physiological saline significantly increased the adenosine concentration in LV, when short half-lives were assumed.

Conclusion

An intravenous adenosine and radiopharmaceutical injection in the same line is feasible. However, the fluctuation of concentration depends significantly on the mode of injection. To minimize the fluctuation, a slow injection of a small volume of a myocardial imaging agent via a co-injection route, with three ways opened, is recommended.

Key words

adenosine myocardial perfusion imaging adverse effect simulation 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Klocke FJ, Baird MG, Lorell BH, Bateman TM, Messer JV, Berman DS, et al. ACC/AHA/ASNC guidelines for the clinical use of cardiac radionuclide imaging—executive summary: a report of the American College of Cardiology/ American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASNC Committee to Revise the 1995 Guidelines for the Clinical Use of Cardiac Radionuclide Imaging).J Am Coll Cardiol 2003; 42: 1318–1333.PubMedCrossRefGoogle Scholar
  2. 2.
    Nishimura S, Nishimura T, Yamazaki J, Doi O, Konishi T, Iwasaki T, et al. Comparison of myocardial perfusion imaging by thallium-201 single-photon emission computed tomography with SUNY4001 (adenosine) and exercise —crossover clinical trial at multi-center.KAKU IGAKU (Jpn J Nucl Med) 2004; 41: 143–154.Google Scholar
  3. 3.
    Sakata Y, Nishimura T, Yamazaki J, Nishimura S, Kajiya T, Kodama K, et al. Diagnosis of coronary artery disease by thallium-201 myocardial scintigraphy with intravenous infusion of SUNY4001 (adenosine) in effort angina pectoris —the clinical trial report at multi-center: phase II.KAKU IGAKU (Jpn J Nucl Med) 2004; 41: 123–132.Google Scholar
  4. 4.
    Yamazaki J, Nishimura T, Nishimura S, Kajiya T, Kodama K, Kato K. The diagnostic value for ischemic heart disease of thallium-201 myocardial scintigraphy by intravenous infusion of SUNY4001 (adenosine)—the report of clinical trial at multi-center: phase III.KAKU IGAKU (Jpn J Nucl Med) 2004; 41: 133–142.Google Scholar
  5. 5.
    Klabunde RE. Dipyridamole inhibition of adenosine metabolism in human blood.Eur J Pharmacol 1983; 93: 21- 26.PubMedCrossRefGoogle Scholar
  6. 6.
    Moser GH, Schrader J, Deussen A. Turnover of adenosine in plasma of human and dog blood.AmJPhysiol 1989; 256: C799–806.Google Scholar
  7. 7.
    Hidaka T, Yamaki A, Takagi H, N. I, Ohno T. Effect of SUNY 4001 on coronary blood flow in anesthetized open-chest dogs [in Japanese, also appeared in the pharmaceutical interview form of Daiichi Pharmaceutical Co., June 2005].Pharma Medica 2003; 21: 161–167.Google Scholar
  8. 8.
    Kajiya T, Kurogane H, Kijima M, Aizawa T, Tanaka T, Yamazaki J, et al. The effect of continuous infusion on coronary blood flow velocity of patients with effort angina pectoris: comparison between normal and severely stenotic arteries in the same patients [in Japanese].Kokyu to Junkan 2004; 52: 529–535.Google Scholar
  9. 9.
    Cerqueira MD, Verani MS, Schwaiger M, Heo J, Iskandrian AS. Safety profile of adenosine stress perfusion imaging: results from the Adenoscan Multicenter Trial Registry.J Am Coll Cardiol 1994; 23: 384–389.PubMedCrossRefGoogle Scholar
  10. 10.
    Korkmaz ME, Mahmarian JJ, Guidry GW, Verani MS. Safety of single-site adenosine thallium-201 scintigraphy.Am J Cardiol 1994; 73: 200–204.PubMedCrossRefGoogle Scholar
  11. 11.
    Cave V, Heo J, Cassel D, Iskandrian B, Iskandrian AS. Side effects during adenosine thallium imaging with single-port or double-port infusion protocols.Am Heart J 1992; 124: 610–613.PubMedCrossRefGoogle Scholar

Copyright information

© Springer 2006

Authors and Affiliations

  • Kenichi Nakajima
    • 1
  • Junichi Taki
    • 1
  • Tatsuya Yoneyama
    • 1
  • Makoto Fukuoka
    • 1
  • Daiki Kayano
    • 1
  • Norihisa Tonami
    • 1
  1. 1.Department of Nuclear MedicineKanazawa University HospitalKanazawaJAPAN

Personalised recommendations