Availability of intraarterial pressure waveforms from catheter-manometer systems during surgery

  • Published:
Journal of Clinical Monitoring Aims and scope Submit manuscript

Abstract

We determined how often and for how long usable pressure waveforms were unavailable from a radial intraarterial pressure cannula during anesthesia and surgery in 41 patients. During cardiac surgery with a continuous flush system, usable arterial pressure was unavailable 8.7% of the time. It was unavailable 9.1% of the time during noncardiac surgical procedures with a continuous flush system, and 14.7% of the time in systems without continuous flush. Thus, the use of a continuous flush device improves intraarterial pressure availability. Artifact is the principal contributor to unavailability, followed by flushing and blood sampling. With rare exceptions the use of a Riva-Rocci cuff for occasional return-to-flow maneuvers on the same arm as the intraarterial cannula reduces intraarterial pressure availability only slightly, certainly not enough to detract from its usefulness in providing an estimation of systolic pressure during intraarterial pressure monitoring.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Prys-Roberts C: Invasive monitoring of the circulation. In: Saidman LJ, Smith NT, eds. Monitoring in anesthesia. Boston: Butterworth, 1984:83–93

    Google Scholar 

  2. Smith NT, Wesseling KH, de Wit B: Evaluation of two prototype devices producing noninvasive, pulsatile, calibrated blood pressure measurement from a finger. J Clin Monit 1985; 1:17–29

    Article  PubMed  Google Scholar 

  3. Penaz J. Voigt A, Teichmann W: Beitrag zur fortlaufenden indirekten Blutdruckmessung. Zschr Inn Med 1976; 31:1030–1033

    CAS  Google Scholar 

  4. Hellige, G: Ventrikeldruckmessung mit konventionellen Kathetermanometersystemen. Teil 1. Basic Res Cardiol 1976; 71:319–336

    Article  PubMed  CAS  Google Scholar 

  5. Hellige, G: Ventrikeldruckmessung mit konventionellen Kathetermanometersystemen. Teil II. Basic Res Cardiol 1976; 71:389–406

    Article  PubMed  CAS  Google Scholar 

  6. Bruner JMR: Handbook of blood pressure monitoring. Littleton, MA: PSG, 1978

    Google Scholar 

  7. Bruner JMR, Krenis LR, Kunsman JM, Sherman AP: Comparison of direct and indirect methods of measuring arterial blood pressure, part I. Med Instrum 1981; 15:11–21

    PubMed  CAS  Google Scholar 

  8. Bruner JMR, Krenis LR, Kunsman JM, Sherman AP: Comparison ot direct and indirect methods of measuring arterial blood pressure, part II. Med Instrum 1981; 15:97–101

    PubMed  CAS  Google Scholar 

  9. Bruner JMR, Krenis LR, Kunsman JM, Sherman AP: Comparison ot direct and indirect methods of measuring arterial blood pressure, part III. Med Instrum 1981; 14:182–188

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Wesseling, K.H., Smith, N.T. Availability of intraarterial pressure waveforms from catheter-manometer systems during surgery. J Clin Monitor Comput 1, 11–16 (1985). https://doi.org/10.1007/BF02832684

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02832684

Keywords

Navigation