Abstract
The chirp radar-type microwave computed tomograph (CT) measures the temperature change in a human body noninvasively. The paper examines its feasibility. A chirp pulse signal between 1 and 2 GHz is radiated from the transmitting antenna to the phantom. The transmitted waves are detected by the receiving antenna, which is place on the opposite side of the object, and the beat signal between the incident wave and the transmitted wave is produced by the mixer. By spectral analysis of the beat signal, only those signals transmitted on the straight line between the transmitting antenna and the receiving antenna are discriminated from multipath signals. The microwave tomogram can therefore be reconstructed easily using the conventional algorithms for an X-ray CT image. The microwave CT can use the chirp signal to remove the influence of multipath signals caused by diffraction and reflection. The imaging of dielectric materials with complicated structures is thus possible. The experimental results using phantoms show that the spatial resolution of this microwave CT is about 10 mm and that a two-dimensional distribution of temperature change can be measured.
Similar content being viewed by others
References
BOLOMEY, J. Ch., JOFRE, L. and PERONNET, G. (1983) On the possible use of microwave-active imaging for remote thermal sensing. IEEE Trans., MTT-31, 777–781.
JACOBI, J. H. and LARSEN, L. E. (1986) Linear FM pulse compression radar techniques applied to biological imaging. In Medical applications of microwave imaging. LARSEN, L. E. and JACOBI J. H. (Eds.) IEEE Press, 138–147.
MIYAKAWA, M. (1988) A study on the temperature dependency of the permittivity of biological tissues. Proc. 5th Intern. Symp. Hyperthermic Oncology, 728–729.
MtYAKAWA, M. (1990) An attempt of tomographic imaging of temperature change inside a body A specially designed microwave CT. Proc. 3rd Asia-Pacific Microwave Conf., 841–844.
SHEPP, L. A. and LOGAN, B. F. (1974) The Fourier reconstruction of a head section. IEEE Trans., NS-21, 21–43.
YAMAURA, I. (1985) Non-invasive thermometry by means of microwave active imaging. Tech. Rep. IEICE, MW85-81, 81–100.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Miyakawa, M. Tomographic measurement of temperature change in phantoms of the human body by chirp radar-type microwave computed tomography. Med. Biol. Eng. Comput. 31 (Suppl 1), S31–S36 (1993). https://doi.org/10.1007/BF02446647
Received:
Accepted:
Published:
Issue Date: