Abstract
A simple dose-saving procedure for the diagnosis of congenital hip dysplasia employing a diaphragm in near-focus position is presented. Doses to the skin and the gonads are measured with a phantom, comparing the new and other techniques. With optimal techniques the skin dose is smaller than 2 mR and the gonadal doses lie around 0.1 mR for both male and female patients.
Similar content being viewed by others
References
Correll, R., Plischke, U., Dresel, H., Joos, M.: Gonadenschutz bei Beckenaufnahmen. Röntgenblätter29, 137 (1976)
Fendel, H., Goebel, W.: Radiation exposure of infants in the radiological examination of the hip. Ann. Radiol. (Paris)11, 282 (1968)
Frik, W.: Reihenuntersuchungen der Hüftgelenke bei Säuglingen und Strahlenschutz. Fortschr. Röntgenstr.116, 453 (1972)
Heinrich, H., Schuster, W.: Reduction of dose by filtration in pediatric fluoroscopy and fluorography. Ann. Radiol. (Paris)19, 57 (1976)
Hofer, H.: Strahlenschutz bei Hüftgelenksröntgenaufnahmen an Kleinkindern in der täglichen Praxis. Orthop. Traumatol.13, 11 (1966)
Keller, G., Schmidt, T., Siegert, A.: Zum Aspekt der Hüftgelenksdysplasie in Klinik und Röntgenologie. Z. Orthop.115, 9 (1977)
Krepler, P., Havranek, Ch., Nava, N.: Ein spezieller Gonadenschutz für Hüftvergleichsaufnahmen bei Säuglingen mit dosimetrischen Studien. Röntgenpraxis29, 271 (1976)
Reiss, K., Mika, N.: Tabellen zur Röntgendiagnostik I. Erlangen: Siemens AG 1973