, Volume 11, Issue 4, pp 597–605 | Cite as

Thermal characterization of resistance and gain of microchannel plate (MCP) detectors for the JENI experiment

  • O. B. ChasselaEmail author
  • A. Grigoriev
  • A. Fedorov
  • N. André
  • E. Le Comte
  • J. Rouzaud
Original Paper

Abstract

The microchannel plate (MCP) has been used for decades as a photon, electron and atoms detector in most of the space instruments dedicated for X-rays, energetic neutral atoms and charged particle imaging. The deep-space missions, as near-future ESA Jupiter Icy moon Explorer (JUICE) mission, expect very low temperature conditions on the destination orbit. Since instruments are usually calibrated on the ground under the “room” temperature, it is very important to know the variation of the detectors’ properties with temperature. The resistance and the gain of MCP detectors, dedicated for the Jovian Energetic Neutrals and Ions (JENI) sensor as part of the Particle Environment Package (PEP) instrument onboard JUICE, were measured as a function of temperature at INTRASPEC TECHNOLOGIES, Toulouse, France over the temperature range of − 50 to + 50 °C and at the CALIPSO-3 facility of the Institut de Recherche en Astrophysique et Planétologie (IRAP), Toulouse, France over the temperature range of − 25 to + 25 °C using samples from PHOTONIS France and PHOTONIS USA. Depending on how the resistance of an MCP detector behaves with temperature, it is possible to properly size a high voltage source or, conversely, to choose a glass technology according to the resistance range required for the MCP detector or their minimum operational bias strip current in relation to the extracted current due to the particle counts. Moreover, since the environment of Jupiter is very severe, the sensors of the PEP package instrument onboard JUICE will operate in the presence of high-energy particles which represent a noise source for them despite the shielding of the instrument against radiation. In particular, the background radiation in the Jovian environment represents a crucial issue for the JENI MCP detectors whose gain can be degraded prematurely if too much of charge is extracted from them. Our measurements show that the gain variation of JENI MCP detectors depends on their characteristics, which is an important result that can be used to optimize their gain performance and lifetime. We also show that the resistance of the JENI MCP detectors increases when the temperature decreases and is influenced by the MCP-resistive heating.

Keywords

MCP Resistive heating Resistance Gain Lifetime 

Notes

Acknowledgements

The contribution of IRAP to the JUICE PEP consortium was supported by the Centre National d’Etudes Spatiales (CNES) through Grants numbers 4500046142 and 4500055493. The authors would like to thank the anonymous referees for their valuable comments which helped to improve this manuscript.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

References

  1. 1.
    Wiza, J.L.: Microchannel plate detectors. Nucl. Instrum. Methods 162, 587–601 (1979)CrossRefGoogle Scholar
  2. 2.
    Gys, T.: Micro-channel plates and vacuum detectors. Nucl. Instrum. Methods Phys. Res. A 787, 254–260 (2015)CrossRefGoogle Scholar
  3. 3.
    Kraus, H.: Cryogenic detectors and their application to mass spectrometry. Int. J. Mass Spectrom. 215(1–3), 45–58 (2002)CrossRefGoogle Scholar
  4. 4.
    Mackler, D., et al.: Microchannel plate lifetime experiment for the DIS and DES instruments on the magnetospheric multiscale mission. Planet. Space Sci. 161, 91–98 (2018)CrossRefGoogle Scholar
  5. 5.
    Sandel, B.R., Broadfoot, A.L., Shemansky, D.E.: Microchannel plate life tests. Appl. Opt. 16(5), 1435–1437 (1977)CrossRefGoogle Scholar
  6. 6.
    Blasé, R.C., et al.: Microchannel plate detection efficiency to monoenergetic photons between 0.66 and 20 MeV. IEEE Trans Nucl Sci 65(4), 980–988 (2018)CrossRefGoogle Scholar
  7. 7.
    Tulej, M., et al.: Detection efficiency of microchannel plates for e- and π- in the momentum range to 17.5 to 345 MeV/c. Rev. Sci. Instrum. 86, 083310 (2015)CrossRefGoogle Scholar
  8. 8.
    Eberhardt, Edward H.: Gain model for microchannel plates. Appl. Opt. 18(9), 1418–1423 (1979)CrossRefGoogle Scholar
  9. 9.
    Chen, L., et al.: The gain and time characteristics of microchannel plates in various channel geometries. IEEE Trans. Nucl. Sci. 64(4), 1080–1086 (2017)CrossRefGoogle Scholar
  10. 10.
    Fraser, G.W.: The electron detection efficiency of microchannel plates. Nucl. Instrum. Methods Phys. Res. A 206, 445–449 (1983)CrossRefGoogle Scholar
  11. 11.
    Fraser, G.W.: The ion detection efficiency of microchannel plates (MCPs). Int. J. Mass Spectrom. 215, 13–30 (2002)CrossRefGoogle Scholar
  12. 12.
    Tobita, K., et al.: Absolute Detection Efficiency of a Microchannel-Plate Detector for Ions and Neutrals. Jpn. J. Appl. Phys. 26(3), 509–510 (1987)CrossRefGoogle Scholar
  13. 13.
    Rochau, G.A., et al.: Energy dependent sensitivity of microchannel plate detectors. Rev. Sci. Instrum. 77, 10E323 (2006)CrossRefGoogle Scholar
  14. 14.
    Pearson, J.F., et al.: Comparative tests of Russian microchannel plates. Nucl. Instrum. Methods Phys. Res. A 325, 578–583 (1993)CrossRefGoogle Scholar
  15. 15.
    Pearson, J.F., Fraser, G.W., Whiteley, M.J.: Variation of microchannel plate resistance with temperature and applied voltage. Nucl. Instrum. Methods Phys. Res. A 258, 134–147 (2000)Google Scholar
  16. 16.
    Roth, P., Fraser, G.W.: Microchannel plate resistance at cryogenic temperatures. Nucl. Instrum. Methods Phys. Res. A 439, 270–274 (1987)Google Scholar
  17. 17.
    Tremsin, A.S., Vallerga, J.V., Siegmund, O.H.W.: Thermal dependence of electrical characteristics of micromachined silica microchannel plates. Rev. Sci. Instrum. 75(4), 1068 (2004)CrossRefGoogle Scholar
  18. 18.
    Csorba, I.P.: Current gain parameters of microchannel plates. Appl. Opt. 19(22), 3863–3866 (1980)CrossRefGoogle Scholar
  19. 19.
    Slater, D.C., Timothy, J.G.: Microchannel plate modal gain variations with temperature. Rev. Sci. Instrum. 64, 430 (1993)CrossRefGoogle Scholar
  20. 20.
    Siegmund, O.H.W., Coburn, K., Malina, R.F.: Investigation of large format microchannel plate Z configurations. IEEE Trans. Nucl. Sci. NS-32(1), 443–447 (1985)CrossRefGoogle Scholar
  21. 21.
    PHOTONIS Scientific Detectors staff.: Custom MCPs for analytical instruments. https://www.photonis.com. Accessed Mar 2019
  22. 22.
    Hamamatsu Photonics K. K. staff.: MCP (Microchannel plate) and MCP assembly. https://www.hamamatsu.com/resources/pdf/etd/MCP_TMCP0002E.pdf. Accessed May 2019
  23. 23.
    Morenzoni, E., Oba, K., Pedroni, E., Taqqu, D.: Performance of microchannel plates in high magnetic fields. Nucl. Instrum. Methods Phys. Res. A 263, 397–400 (1988)CrossRefGoogle Scholar
  24. 24.
    Oba, K., Rehak, P.: Studies of high-gain micro-channel plate photomultipliers. IEEE Trans. Nucl. Sci. 28(1), 683–688 (1981)CrossRefGoogle Scholar

Copyright information

© CEAS 2019

Authors and Affiliations

  1. 1.Institut de Recherche en Astrophysique et Planétologie, CNRS, UPS, CNES, Université de ToulouseToulouse Cedex 4France

Personalised recommendations