Radiation protective (RP) eyewear effectively protects crystalline lenses from radiation exposure. A drawback of RP eyewear is the angular dependence of the shielding effect, which results from the design of the eyewear. In this study, 21 models of RP eyewear with different designs and lead equivalences were assessed. Each piece of RP eyewear was hung on a Styrofoam phantom that imitated the head, and a 0.125-cc ionization chamber dosimeter was placed at the position of the crystalline lens. The differences in angular dependence of the shielding effect were evaluated by changing the irradiation angle, and parameters that improved the angular dependence of the shielding effect—sufficient lead equivalence, large coverage design, and minimum gap between the crystalline lens and the RP eyewear—were identified. Thus, the findings highlight the importance of selecting RP eyewear according to the angular distribution and the nature of radiation exposure in the workplace for radiation workers.

Keywords

Angular dependence Crystalline lens Interventional radiology Radiation protective eyewear Shielding rate 

Notes

Acknowledgements

This research was supported by an NSR Grant for Radiation Safety Regulation Research Strategic Promotion 2017, Japan.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

This article does not contain any studies performed with human or animal participants.

References

  1. 1.
    Valentin J. Avoidance of radiation injuries from medical interventional procedures. Ann ICRP. 2000;30(2):7–67.CrossRefGoogle Scholar
  2. 2.
    International Commission on Radiological Protection. ICRP Publication 118 statement on tissue reactions/early and late effects of radiation in normal tissues and organs—threshold doses for tissue reactions in a radiation protection context. ICRP Publ. 2012;3:313–27.Google Scholar
  3. 3.
    Principi S, Delgado C, Ginjaume M, Beltran M, Rovira JJ, Duch MA. Eye lens dose in interventional cardiology. Radiat Prot Dosimetry. 2015;165(1–4):289–93.CrossRefGoogle Scholar
  4. 4.
    Anchali K, Suphot S, Kousuke M. The current status of eye lens dose measurement in interventional cardiology personnel in Thailand. Radiol Phys Technol. 2017;10(2):142–7.CrossRefGoogle Scholar
  5. 5.
    Nachiappan AC, Horn GL, Spann SC, Mayo RC, Wynne DM, Archer BR, Whigham CJ, Hancock JA. Operator radiation dose reduction during fluoroscopic interventional procedures. J Am Coll Radiol. 2015;12(5):527–30.CrossRefGoogle Scholar
  6. 6.
    O’Connor U, Walsh C, Gallagher A, Dowling A, Guiney M, Ryan JM, McEniff N, O'Reilly G. Occupational radiation dose to eyes from interventional radiology procedures in light of the new eye lens dose limit from the International Commission on Radiological Protection. Br J Radiol. 2015;88(1049):20140627.CrossRefGoogle Scholar
  7. 7.
    Bahruddin NA, Hashim S, Karim MKA, Sabarudin A, Ang WC, Salehhon N, Bakar KA. Radiation dose to physicians' eye lens during interventional radiology. J Phys. 2016;694(1):012035 (Conference Series).Google Scholar
  8. 8.
    Carinou E, Ferrari P, Bjelac OC, Gingaume M, Merce MS, O'Connor U. Eye lens monitoring for interventional radiology personnel: dosemeters, calibration and practical aspects of H p (3) monitoring. A 2015 review. J Radiol Prot. 2015;35(3):R17–34.CrossRefGoogle Scholar
  9. 9.
    Principi S, Ginjaume M, Duch MA, Sánchez RM, Fernández JM, Vano E. Influence of dosemeter position for the assessment of eye lens dose during interventional cardiology. Radiat Prot Dosimetry. 2015;164(1–2):79–83.CrossRefGoogle Scholar
  10. 10.
    McVey S, Sandison A, Sutton DG. An assessment of RP eyewear in interventional radiology. J Radiol Prot. 2013;33(3):647–59.CrossRefGoogle Scholar
  11. 11.
    Farah J, Struelens L, Dabin J, Koukorava C, Donadille L, Jacob S, Schnelzer M, Auvinen A, Vanhavere F, Clairand I. A correlation study of eye lens dose and personal dose equivalence for interventional cardiologists. Radiat Prot Dosimetry. 2013;157(4):561–9.CrossRefGoogle Scholar
  12. 12.
    Bart D, van Rooijen BD, de Haan MW, Das M, Arnoldussen CW, de Graaf R, van Zwam WH, Backes WH, Jeukens CR. Efficacy of radiation safety glasses in interventional radiology. Cardiovasc Intervent Radiol. 2014;37:1149–55.CrossRefGoogle Scholar
  13. 13.
    Domienik J, Bissinger A, Grabowicz W, Jankowski Ł, Kręcki R, Makowski M, Masiarek K, Plewka M, Lubiński A, Peruga JZ. The impact of various protective tools on the dose reduction in the eye lens in an Interventional Cardiology-Clinical Study. J Radiol Prot. 2016;36(2):309–18.CrossRefGoogle Scholar
  14. 14.
    Hu P, Kong Y, Chen B, Liu Q, Zhuo W, Liu H. Shielding effect of lead glasses on radiologists' eye lens exposure in interventional procedures. Radiat Prot Dosimetry. 2017;20:136–40.Google Scholar
  15. 15.
    Magee JS, Martin CJ, Sandblom V, Carter MJ, Almén A, Cederblad Å, Jonasson P, Lundh C. Derivation and application of dose reduction factors for protective eyewear worn in interventional radiology and cardiology. J Radiol Prot. 2014;34(4):811–23.CrossRefGoogle Scholar
  16. 16.
    Sturchio GM, Newcomb RD, Molella R, Varkey P, Hagen PT, Schueler BA. Protective eyewear selection for interventional fluoroscopy. Health Phys. 2013;104(2 Suppl 1):S11–S16.CrossRefGoogle Scholar
  17. 17.
    Clairand I, Bordy JM, Daures J, Debroas J, Denozière M, Donadille L, Ginjaume M, Itié C, Koukorava C, Krim S, Lebacq AL, Martin P, Struelens L, Sans-Mercé M, Tosic M, Vanhavere F. Active personal dosemeters in interventional radiology: tests in laboratory conditions and in hospitals. Radiat Prot Dosimetry. 2011;144(1–4):453–8.CrossRefGoogle Scholar
  18. 18.
    Hu P, Kong Y, Chen B, Liu Q, Zhuo W, Liu H. Shielding effect of lead glasses on radiologists' eye lens exposure in interventional procedures. Radiat Prot Dosimetry. 2017;174(1):136–40.PubMedGoogle Scholar

Copyright information

© Japanese Society of Radiological Technology and Japan Society of Medical Physics 2019

Authors and Affiliations

  1. 1.Department of Health Sciences, Graduate School of Medical ScienceKyushu UniversityFukuoka CityJapan
  2. 2.Department of Health Sciences, Faculty of Medical SciencesKyushu UniversityFukuokaJapan
  3. 3.Department of RadiologyNational Hospital Organization Kanagawa HospitalKanagawaJapan
  4. 4.Department of Radiological TechnologyInternational University of Health and Welfare Narita HospitalChibaJapan
  5. 5.Department of Diagnostic RadiologyHiroshima UniversityHiroshimaJapan
  6. 6.Faculty of NursingTokyo Healthcare UniversityTokyoJapan

Personalised recommendations