An analytical approach for the simulation of realistic low-dose fluoroscopic images

  • Sai Gokul HariharanEmail author
  • Norbert Strobel
  • Christian Kaethner
  • Markus Kowarschik
  • Rebecca Fahrig
  • Nassir Navab
Original Article
  • 27 Downloads

Abstract

Purpose

The quality of X-ray images plays an important role in computer-assisted interventions. Although learning-based denoising techniques have been shown to be successful in improving the image quality, they often rely on pairs of associated low- and high-dose X-ray images that are usually not possible to acquire at different dose levels in a clinical scenario. Moreover, since data variation is an important requirement for learning-based methods, the use of phantom data alone may not be sufficient. A possibility to address this issue is a realistic simulation of low-dose images from their related high-dose counterparts.

Method

We introduce a novel noise simulation method based on an X-ray image formation model. The method makes use of the system parameters associated with low- and high-dose X-ray image acquisitions, such as system gain and electronic noise, to preserve the image noise characteristics of low-dose images.

Results

We have compared several corresponding regions of the associated real and simulated low-dose images—obtained from two different imaging systems—visually as well as statistically, using a two-sample Kolmogorov–Smirnov test at 5% significance. In addition to being visually similar, the hypothesis that the corresponding regions—from 80 pairs of real and simulated low-dose regions—belonging to the same distribution has been accepted in 81.43% of the cases.

Conclusion

The results suggest that the simulated low-dose images obtained using the proposed method are almost indistinguishable from real low-dose images. Since extensive calibration procedures required in previous methods can be avoided using the proposed approach, it allows an easy adaptation to different X-ray imaging systems. This in turn leads to an increased diversity of the training data for potential learning-based methods.

Keywords

Noise simulation X-ray imaging Simulating low-dose X-ray images 

Notes

Acknowledgements

This work was supported by Siemens Healthineers AG. The concepts and results presented in this paper are based on research and are not commercially available.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

References

  1. 1.
    Amiot C, Girard C, Chanussot J, Pescatore J, Desvignes M (2016) Spatio-temporal multiscale denoising of fluoroscopic sequence. IEEE Trans Med Imaging 35(6):1565–1574CrossRefGoogle Scholar
  2. 2.
    Cerciello T, Bifulco P, Cesarelli M, Fratini A (2012) A comparison of denoising methods for X-ray fluoroscopic images. Biomed Signal Process Control 7(6):550–559CrossRefGoogle Scholar
  3. 3.
    Chan CL, Sullivan BJ, Sahakian AV, Katsaggelos AK, Swiryn S, Hueter DC, Frohlich T (1990) Simulation of quantum mottle in digital angiographic images. Biomed Image Process 1245:104–111 SPIECrossRefGoogle Scholar
  4. 4.
    Chen Y, Pock T (2017) Trainable nonlinear reaction diffusion: a flexible framework for fast and effective image restoration. IEEE Trans Pattern Anal Mach Intell 39(6):1256–1272CrossRefGoogle Scholar
  5. 5.
    Dabov K, Foi A, Katkovnik V, Egiazarian K (2009) BM3D image denoising with shape-adaptive principal component analysis. In: Signal processing with adaptive sparse structured representationsGoogle Scholar
  6. 6.
    Gislason-Lee AJ, Kumcu A, Kengyelics SM, Brettle DS, Treadgold LA, Sivananthan M, Davies AG (2015) How much image noise can be added in cardiac x-ray imaging without loss in perceived image quality? J Electron Imaging 24(5):051006CrossRefGoogle Scholar
  7. 7.
    Hariharan SG, Kaethner C, Strobel N, Kowarschik M, Fahrig R, Navab N (2018) Estimation of noise stabilization parameters for low-dose X-ray images corrupted by Poisson–Gaussian noise. (manuscript in preparation) Google Scholar
  8. 8.
    Hariharan SG, Strobel N, Kaethner C, Kowarschik M, Demirci S, Albarqouni S, Fahrig R, Navab N (2018) A photon recycling approach to the denoising of ultra-low dose X-ray sequences. Int J CARS 13:847–854.  https://doi.org/10.1007/s11548-018-1746-2 CrossRefGoogle Scholar
  9. 9.
    Hariharan SG, Strobel N, Kowarschik M, Fahrig R, Navab N (2018) Simulation of realistic low dose fluoroscopic images from their high dose counterparts. In: Bildverarbeitung für die Medizin 2018, Springer, pp 80–85Google Scholar
  10. 10.
    Holdsworth D, Gerson R, Fenster A (1990) A time-delay integration charge-coupled device camera for slot-scanned digital radiography. Med Phys 17(5):876–886CrossRefGoogle Scholar
  11. 11.
    Kostadin D, Alessandro F, Karen E (2007) Video denoising by sparse 3D transform-domain collaborative filtering. In: European signal processing conference, vol 149Google Scholar
  12. 12.
    Kuan DT, Sawchuk AA, Strand TC, Chavel P (1985) Adaptive noise smoothing filter for images with signal-dependent noise. IEEE Trans Pattern Anal Mach Intell 2:165–177CrossRefGoogle Scholar
  13. 13.
    Luisier F, Blu T, Unser M (2007) A new sure approach to image denoising: interscale orthonormal wavelet thresholding. IEEE Trans Image Process 16(3):593–606CrossRefGoogle Scholar
  14. 14.
    Makitalo M, Foi A (2013) Optimal inversion of the generalized anscombe transformation for poisson-gaussian noise. IEEE Trans Image Process 22(1):91–103CrossRefGoogle Scholar
  15. 15.
    Massey FJ Jr (1951) The Kolmogorov–Smirnov test for goodness of fit. J Am Stat Assoc 46(253):68–78CrossRefGoogle Scholar
  16. 16.
    Matviychuk Y, Mailhé B, Chen X, Wang Q, Kiraly A, Strobel N, Nadar M (2016) Learning a multiscale patch-based representation for image denoising in X-ray fluoroscopy. In: Proceedings of international conference on image processing, pp 2330–2334Google Scholar
  17. 17.
    Schmidt U, Roth S (2014) Shrinkage fields for effective image restoration. In: Proceedings of the IEEE conference on computer vision and pattern recognition, pp 2774–2781Google Scholar
  18. 18.
    Starck JL, Murtagh FD, Bijaoui A (1998) Image processing and data analysis: the multiscale approach. Cambridge University Press, CambridgeCrossRefGoogle Scholar
  19. 19.
    Veldkamp WJ, Kroft LJ, van Delft JPA, Geleijns J (2009) A technique for simulating the effect of dose reduction on image quality in digital chest radiography. J Digit Imaging 22(2):114–125CrossRefGoogle Scholar
  20. 20.
    Wolterink JM, Leiner T, Viergever MA, Isgum I (2017) Generative adversarial networks for noise reduction in low-dose CT. IEEE Trans Med Imaging 36:2536–2545CrossRefGoogle Scholar
  21. 21.
    Yang K, Huang SY, Packard NJ, Boone JM (2010) Noise variance analysis using a flat panel X-ray detector: a method for additive noise assessment with application to breast ct applications. Med Phys 37(7):3527–3537CrossRefGoogle Scholar
  22. 22.
    Zhang K, Zuo W, Chen Y, Meng D, Zhang L (2017) Beyond a Gaussian denoiser: residual learning of deep cnn for image denoising. IEEE Trans Image Process 26:3142–3155CrossRefGoogle Scholar

Copyright information

© CARS 2019

Authors and Affiliations

  1. 1.Computer Aided Medical ProceduresTechnische Universität MünchenMunichGermany
  2. 2.Siemens Healthineers AG, Advanced TherapiesForchheimGermany
  3. 3.Fakultät für ElektrotechnikHochschule für angewandte Wissenschaften Würzburg-SchweinfurtSchweinfurtGermany
  4. 4.Whiting School of EngineeringJohns Hopkins UniversityBaltimoreUnited States
  5. 5.Pattern Recognition LabFriedrich-Alexander-Universität Erlangen-NürnbergErlangenGermany

Personalised recommendations