Empirical formula for the HPGe-detector efficiency dependence on energy and distance

  • I. PylypchynetsEmail author
  • A. Lengyel
  • O. Parlag
  • V. Maslyuk
  • I. Potoki
Article
  • 18 Downloads

Abstract

The full-energy-peak efficiency calibration for a HPGe detector has been constructed in the γ-ray energy range from Eγ = 50–1500 keV, using 8 standard point sources of gamma radiation, namely Na-22, Co-57, Co-60, Cd-109, Ba-133, Cs-137, Eu-152, Am-241. The measuring system equipped with sample holder to investigate the source-to-detector distance (x) dependence for 5–24.5 cm range. The universal empirical formula was proposed to calculate the detector efficiency depending on energy Eγ and distance (x). The formula was tested for several measurements and broadened energy rage up to 3 MeV.

Keywords

Efficiency calibration Radionuclides Gamma spectrometry HPGe detector Geometric dependence of the efficiency 

Notes

Acknowledgements

The authors would like to express sincere thanks to the Kyiv Regional Research and Production Center for Standardization, Metrology and Certification “KIVOBLSTANDARTMETROLOGIYA” for providing certified gamma-ray samples for the experiment. The authors acknowledge Kushtan I.M. for the technical support of the experiment.

References

  1. 1.
    Hosoda Masahiro, Tokonami Shinji, Tazoe Hirofumi et al (2013) Activity concentrations of environmental samples collected in Fukushima Prefecture immediately after the Fukushima nuclear accident. Sci Rep.  https://doi.org/10.1038/srep02283 Google Scholar
  2. 2.
    Jackel B, Westmeier W, Patzelt P (1987) On the photopeak efficiency of germanium gamma-ray detectors. Nucl Instrum Methods Phys Res A261:543–548CrossRefGoogle Scholar
  3. 3.
    Debertin K, Helmer RG (1988) Gamma and X-ray spectrometry with semiconductor detectors. Elsevier, AmsterdamGoogle Scholar
  4. 4.
    Knoll GF (2010) Radiation detection and measurement, 4th edn. Wiley, New YorkGoogle Scholar
  5. 5.
    Dias MS, Cardoso V, Vanin VR, Koskinas MF (2004) Combination of nonlinear function and mixing method for fitting HPGe efficiency curve in the 59–2754 keV energy range. Appl Radiat Isot 60:683–687CrossRefGoogle Scholar
  6. 6.
    Mowatt RS (1969) A semi-empirical efficiency curve for a Ge(Li) detector in energy range 50–1400 keV. Nucl Instrum Methods 70:237–244CrossRefGoogle Scholar
  7. 7.
    Nair C, Junghans AR, Erhard M et al (2005) Absolute efficiency calibration of hpge detectors for photoactivation experiments. Annual Reports—Forschungcentrum Dresden- Rossebdorf, FZD. https://www.hzdr.de/FWK/jb05/PDF/s41.pdf
  8. 8.
    Lengyel A, Parlag OA, Maslyuk VT (2009) Semiempirical description of Ge(Li) and HPGe detectors efficiency for photofission experiments. Uzhgorod Univ Sci Her Ser Phys 24:95–99Google Scholar
  9. 9.
    Potoki IS (2012) Parametric description of the absolute efficiency of semiconductor detectors for measuring of bulk sample activity. Uzhgorod Univ Sci Her Ser Phys 31:196–201Google Scholar
  10. 10.
    Longoria IC, Benitez JS (1996) Full energy peak efficiency of a Ge detector as a function of energy and distance. Appl Radiat Isot 47(3):339–343CrossRefGoogle Scholar
  11. 11.
    Barrera M, Casas-Ruiz M, Alonso JJ, Vidal J (2017) Precise determination of HPGe detector efficiency for gamma spectrometry measurements of environmental samples with variable geometry and density. Nukleonika 62(1):47–59CrossRefGoogle Scholar
  12. 12.
    Dryak P, Kovar P (2006) Experimental and MC determination of HPGe detector efficiency in the 40–2754 keV energy range for measuring point source geometry with the source-to-detector distance of 25 cm. Appl Radiat Isot 64:1346–1349CrossRefGoogle Scholar
  13. 13.
    NuDat 2.6. National Nuclear Data Center, Brookhaven National Laboratory. www.nndc.bnl.gov/nudat2/. Accessed 2 Oct 2017
  14. 14.
    James F, Ross M (1967) Function minimization and error analysis. MINUIT D506. CEREN Computer Centre Program library, pp 1–47Google Scholar
  15. 15.
    Lepy MC (2010) Detection efficiency. Presented to IAEA-ALMERA Technical Visit at Laboratoire National Henri Becquerel. http://www.nucleide.org/ICRM_GSWG/Training/Efficiency.pdf

Copyright information

© Akadémiai Kiadó, Budapest, Hungary 2019

Authors and Affiliations

  1. 1.Institute of Electron PhysicsUkrainian National Academy of SciencesUzhgorodUkraine

Personalised recommendations