The fundamental contribution of the electromyogram to a high bispectral index: a postoperative observational study

  • Kazuko HayashiEmail author
  • Teiji Sawa
Original Research
  • 25 Downloads

Abstract

The electromyogram (EMG) activity has been reported to falsely increase BIS. Conversely, EMG seems necessary to constitute the high BIS indicative of an awake condition, and may play a fundamental role in calculating BIS, rather than distorting the appropriate BIS. However, exactly how EMG is associated with a high BIS remains unclear. We intended to clarify the respective contributions of EMG and various electroencephalogram (EEG) parameters to high BIS. In 79 courses of anaesthesia, BIS monitor-derived EMG parameters (EMGLOW), and other processed EEG parameters [SEF95 (spectral edge frequency 95%), SynchFastSlow (bispectral parameter), BetaRatio (frequency parameter), total power subtypes in five frequency range], were obtained simultaneously with BIS, every 3 s. These EEG parameters were used for receiver operating characteristic (ROC) analysis of detecting three BIS levels (BIS > 80, BIS > 70, and BIS > 60) to assess their diagnosabilities. A total of 218,418 data points derived from 79 cases were used for analysis. Area under the ROC curve (AUC) was calculated and optimal cut-off (threshold) was determined by Youden index. As the results, for detecting BIS > 80, the AUC of EMGLOW was 0.975 [0.974–0.977] (mean [95% confidence interval]), significantly higher than any other processed EEG parameters such as BetaRatio (0.832 [0.828–0.835]), SEF95 (0.821 [0.817–0.826]) and SynchFastSlow (0.769 [0.764–0.774]) (p < 0.05 each). The threshold of EMGLOW for detecting BIS > 80 was 35.7 dB, with high sensitivity (92.5%) and high specificity (96.5%). Our results suggest EMG contributes considerably to the diagnosis of high BIS, and is particularly essential for determining BIS > 80.

Keywords

Bispectral index Electromyogram Awareness Anaesthesia depth Monitoring 

Notes

Funding

This work was partly supported by a Grant-in-Aid for Scientific Research (C) from the Japan Society for the Promotion of Science (JSPS KAKENHI No. 15K10518), Tokyo, Japan.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

References

  1. 1.
    Whitham EM, Pope KJ, Fitzgibbon SP, et al. Scalp electrical recording during paralysis: quantitative evidence that EEG frequencies above 20 Hz are contaminated by EMG. Clin Neurophysiol. 2007;118:1877–8.CrossRefGoogle Scholar
  2. 2.
    Goncharova II, McFarland DJ, Vaughan TM, Wolpaw JR. EMG contamination of EEG: spectral and topographical characteristics. Clin Neurophysiol. 2003;114:1580–93.CrossRefGoogle Scholar
  3. 3.
    Kamata K, Aho AJ, Hagihira S, Yli-Hankala A, Jantti V. Frequency band of EMG in anaesthesia monitoring. Br J Anaesth. 2011;107:822–3.CrossRefGoogle Scholar
  4. 4.
    Dahaba AA. Different conditions that could result in the bispectral index indicating an incorrect hypnotic state. Anesth Analg. 2005;101:765–73.CrossRefGoogle Scholar
  5. 5.
    Rampil IJ. A primer for EEG signal processing in anesthesia. Anesthesiology. 1998;89:980–1002.CrossRefGoogle Scholar
  6. 6.
    Morimoto Y, Hagihira S, Koizumi Y, Ishida K, Matsumoto M, Sakabe T. The relationship between bispectral index and electroencephalographic parameters during isoflurane anesthesia. Anesth Analg. 2004;98:1336–40.CrossRefGoogle Scholar
  7. 7.
    Sleigh JW, Steyn-Ross DA, Steyn-Ross ML, Williams ML, Smith P. Comparison of changes in electroencephalographic measures during induction of general anesthesia: influence of the gamma frequency band and electromyogram signal. Br J Anaesth. 2001;86:50–8.CrossRefGoogle Scholar
  8. 8.
    Dahaba AA, Bornemann H, Hopfgartner E, et al. Effect of sugammadex or neostigmine neuromuscular block reversal on bispectral index monitoring of propofol/remifentanil anaesthesia. Br J Anaesth. 2012;108:602–6.CrossRefGoogle Scholar
  9. 9.
    Aho AJ, Kamata K, Yli-Hankala A, Lyytikäinen LP, Kulkas A, Jäntti V. Elevated BIS and entropy values after sugammadex or neostigmine: an electroencephalographic or electromyographic phenomenon? Acta Anaesthesiol Scand. 2012;56:465–73.CrossRefGoogle Scholar
  10. 10.
    Russell IF. The ability of bispectral index to detect intraoperative wakefulness during total intravenous anaesthesia compared with the isolated forearm technique. Anaesthesia. 2013;68:502–11.CrossRefGoogle Scholar
  11. 11.
    Andrzejowski JC, Carroll TA. Inappropriate elevation of bispectral index and disruption of neurosurgery after irrigation induced facial nerve irritation. Br J Anaesth. 2007;99:750–1.CrossRefGoogle Scholar
  12. 12.
    Liu N, Chazot T, Huybrechts I, Law-Koune J-D, Barvais L, Fischler M. The influence of a muscle relaxant bolus on bispectral and Datex-Ohmeda entropy values during propofol remifentanil induced loss of consciousness. Anesth Analg. 2005;101:1713–8.CrossRefGoogle Scholar
  13. 13.
    Vivien B, Di Maria S, Ouattara A, Langeron O, Coriat P, Riou B. Overestimation of Bispectral Index in sedated intensive care unit patients revealed by administration of muscle relaxant. Anesthesiology. 2003;99:9–17.CrossRefGoogle Scholar
  14. 14.
    Panousis P, Heller AR, Burghardt M, Bleyl JU, Koch T. The effects of electromyographic activity on the accuracy of the Narcotrend® monitor compared with the Bispectral Index during combined anaesthesia. Anaesthesia. 2007;62:868–74.CrossRefGoogle Scholar
  15. 15.
    Bruhn J, Bouillon TW, Shafer SL. Electromyographic activity falsely elevates the bispectral index. Anesthesiology. 2000;92:1485–7.CrossRefGoogle Scholar
  16. 16.
    Vasella FC, Frascarolo P, Spahn DR, Magnusson L. Antagonism of neuromuscular blockade but not muscle relaxation affects depth of anaesthesia. Br J Anaesth. 2005;94:742–7.CrossRefGoogle Scholar
  17. 17.
    Schuller PJ, Newell S, Strickland PA, Barry JJ. Response of bispectral index to neuromuscular block in awake volunteers. Br J Anaesth. 2015;115(1):i95–103.CrossRefGoogle Scholar
  18. 18.
    Messner M, Beese U, Romstock J, Dinkel M, Tschaikowsky K. The bispectral index declines during neuromuscular block in fully awake persons. Anesth Analg. 2003;97:488–91.CrossRefGoogle Scholar
  19. 19.
    Chakrabarti D, Surve RM, Bs D, Masapu D. Intraoperative aberrant bispectral index values due to facial nerve monitoring. J Clin Anesth. 2017;37:61–2.CrossRefGoogle Scholar
  20. 20.
    Hayashi K. Inappropriately low bispectral index of the elderly during emergence from sevoflurane anesthesia. J Clin Anesth. 2016;34:279–81.CrossRefGoogle Scholar
  21. 21.
    Sigl JC, Chamoun NG. An introduction to bispectral analysis for the electroencephalogram. J Clin Monit Comput. 1994;10:392–404.CrossRefGoogle Scholar
  22. 22.
    Hagihira S, Takashina M, Mori T, Mashimo T, Yoshiya I. Practical issues in bispectral analysis of electroencephalographic signals. Anesth Analg. 2001;93:966–70.CrossRefGoogle Scholar
  23. 23.
    Araki R, Hayashi K, Sawa T. Dopamine D2-receptor antagonist droperidol deepens sevoflurane anesthesia. Anesthesiology 2018;128:754–63.CrossRefGoogle Scholar
  24. 24.
    Hayashi K, Mukai N, Sawa T. Simultaneous bicoherence analysis of occipital and frontal electroencephalograms in awake and anesthetized subjects. Clin Neurophysiol. 2014;125:194–201.CrossRefGoogle Scholar
  25. 25.
    Hayashi K, Sawa T, Matsuura M. Anesthesia depth-dependent features of electroencephalographic bicoherence spectrum during sevoflurane anesthesia. Anesthesiology. 2008;108:841–50.CrossRefGoogle Scholar
  26. 26.
    Hayashi K, Tsuda N, Sawa T, Hagihira S. Ketamine increases the frequency of electroencephalographic bicoherence peak on the alpha spindle area induced with propofol. Br J Anaesth. 2007;99:389–95.CrossRefGoogle Scholar
  27. 27.
    Obuchowski NA, McClish DK. Sample size determination for diagnostic accuracy studies involving binormal ROC curve indices. Stat Med. 1997;16:1529–42.CrossRefGoogle Scholar
  28. 28.
    Johansen JW, Sebel PS. Development and clinical application of electroencephalographic bispectrum monitoring. Anesthesiology. 2000;93:1336–44.CrossRefGoogle Scholar
  29. 29.
    Miller A, Sleigh JW, Barnard J, Steyn-Ross DA. Does bispectral analysis of the electroencephalogram add anything but complexity? Br J Anaesth. 2004;92:8–13.CrossRefGoogle Scholar
  30. 30.
    Schneider G, Schöniger S, Kochs E. Does bispectral analysis add anything but complexity? BIS sub-components may be superior to BIS for detection of awareness. Br J Anaesth. 2004;93:596–7.CrossRefGoogle Scholar
  31. 31.
    Dressler O, Schneider G, Stockmanns G, Kochs EF. Awareness and the EEG power spectrum: analysis of frequencies. Br J Anaesth. 2004;93:806–9.CrossRefGoogle Scholar
  32. 32.
    Zand F, Hadavi SM, Chohedri A, Sabetian P. Survey on the adequacy of depth of anaesthesia with bispectral index and isolated forearm technique in elective Caesarean section under general anaesthesia with sevoflurane. Br J Anaesth. 2014;112:871–8.CrossRefGoogle Scholar
  33. 33.
    Jordan D, Steiner M, Kochs EF, Schneider G. A program for computing the prediction probability and the related receiver operating characteristic graph. Anesth Analg. 2010;111:1416–21.CrossRefGoogle Scholar
  34. 34.
    Hino M, Mihara T, Miyazaki S, et al. Development and validation of a risk scale for emergence agitation after general anesthesia in Children: a prospective observational study. Anesth Analg. 2017;125:550–55.CrossRefGoogle Scholar
  35. 35.
    Mathews DM, Clark L, Johansen J, Matute E, Seshagiri CV. Increases in electroencephalogram and electromyogram variability are associated with an increased incidence of intraoperative somatic response. Anesth Analg. 2012;114:759–70.CrossRefGoogle Scholar
  36. 36.
    Sawa T, EEG analyzer f-PIS monitor. http://anesth-kpum.org/blog_ts/?p=559, (Ver.46_SR) and p = 373, (Ver.45_SR). Accessed 6 Dec 2017.

Copyright information

© Springer Nature B.V. 2019

Authors and Affiliations

  1. 1.Department of AnesthesiologyKyoto Chubu Medical CenterNantan, KyotoJapan
  2. 2.Department of Anesthesiology and Critical CareKyoto Prefectural University of MedicineKyotoJapan

Personalised recommendations