Abstract

Aims

Limitations imposed by the coronary sinus venous anatomy triggered the transseptal approach for endocardial LV lead placement. The alignment of the interatrial septum (IAS) and its neighborhood anatomy does not favor transseptal puncture from the pre-pectoral area. Locating and advancing a pre-pectoral LV lead delivery catheter (PDC) through an opening created in the IAS via femoral transseptal puncture (FTP) is time consuming and technically difficult. We describe a method where the PDC is snare coupled to the femoral transseptal apparatus (FTA). When the FTA is advanced into the left atrium (LA) the coupled PDC follows.

Methods

The catheter of a 25-mm loop snare kit is replaced with the PDC (SelectSite®). The snare loop is positioned in the right common iliac vein from the pre-pectoral access. The PDC is coupled to the FTA by advancing the transseptal apparatus through the open snare loop. After conventional FTP, the FTA is withdrawn back into the right atrium (RA) over an extra support wire positioned in the LA. The PDC with open snare loop is pulled over the FTA up to the RA. The PDC is advanced to close the snare loop on the extra support wire immediately distal to the tip of the dilator close to the puncture site. The PDC is deflected to align with the FTA. The snare coupled catheters are gently advanced across the IAS into the LA. The PDC is released from the FTA by advancing the snare and opening the loop; the snare is then removed from the PDC. The PDC is deflected and advanced into the left ventricle (LV). After positioning the 4.1 Fr lumen less LV lead, the PDC is sliced and removed.

Results

The PDC snare coupled to the FTA was advanced into the LA in all five patients, however, access was lost during catheter manipulation in the one right-sided case. Endocardial LV lead was successfully positioned in all five patients.

Conclusion

Snare coupling the pre-pectoral SelectSite® catheter to the FTA is technically simple, reliable and a safe method for transseptal endocardial LV lead placement for left pre-pectoral implantation.

This is a preview of subscription content, log in to check access.

Access options

Buy single article

Instant unlimited access to the full article PDF.

US$ 39.95

Price includes VAT for USA

Subscribe to journal

Immediate online access to all issues from 2019. Subscription will auto renew annually.

US$ 99

This is the net price. Taxes to be calculated in checkout.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

References

  1. 1.

    Leon, A. R., Abraham, W. T., Curtis, A. B., et al. (2005). Safety of transvenous cardiac resynchronization system implantation in patients with chronic heart failure: combined results of over 2,000 patients from a multi-center study program. Journal of the American College of Cardiology, 46, 2348–56.

  2. 2.

    Gras, D., Böcker, D., Lunati, M., et al. (2007). Implantation of cardiac resynchronization therapy systems in the CARE-HF trial: procedural success rate and safety. Europace, 9, 516–22.

  3. 3.

    Koos, R., Sinha, A. M., Markus, K., Breithardt, O. A., Mischke, K., Zarse, M., et al. (2004). Comparison of left ventricular lead placement via the coronary venous approach versus lateral thoracotomy in patients receiving cardiac resynchronization therapy. The American Journal of Cardiology, 94, 59–63.

  4. 4.

    Daoud, E. G., Kalbfleisch, S. J., Hummel, J. D., Weiss, R., Augustini, R. S., Duff, S. B., et al. (2002). Implantation techniques and chronic lead parameters of biventricular pacing dual-chamber defibrillators. Journal of Cardiovascular Electrophysiology, 13, 964–70.

  5. 5.

    Jais, P., Takahashi, A., Garrigue, S., Yamane, T., Hocini, M., Shah, D. C., Barold, S. S., et al. (2000). Mid-term follow-up of endocardial biventricular pacing. Pacing and Clinical Electrophysiology, 23, 1744–1747.

  6. 6.

    Leclercq, F., Hager, F. X., Macia, J. C., Mariottini, C. J., Pasquie, J. L., & Grolleau, R. (1999). Left ventricular lead insertion using a modified transseptal catheterization technique: a totally endocardial approach for permanent biventricular pacing in end-stage heart failure. Pacing and Clinical Electrophysiology, 22, 1570–1575.

  7. 7.

    Ji, S., Cesario, D. A., Swerdlow, C. D., & Shivkumar, K. (2004). Left ventricular endocardial lead placement using a modified transseptal approach. Journal of Cardiovascular Electrophysiology, 15, 234–236.

  8. 8.

    Lau, E. W. (2009). Achieving permanent left ventricular pacing-options and choice. Pacing and Clinical Electrophysiology, 32(11), 1466–77. Epub 2009 Aug 31.

  9. 9.

    Jais, P., Douard, H., Shah, D. C., Barold, S., Barat, J. L., & Clementy, J. (1998). Endocardial biventricular pacing. Pacing and Clinical Electrophysiology, 21, 2128–2131.

  10. 10.

    Van Gelder, B. M., Scheffer, M. G., Meijer, A., & Bracke, F. A. (2007). Transseptal endocardial left ventricular pacing: an alternative technique for coronary sinus lead placement in cardiac resynchronization therapy. Heart Rhythm, 4, 454–460.

  11. 11.

    Lau, E. W. (2009). A streamlined technique of trans-septal endocardial left ventricular lead placement. Journal of Interventional Cardiac Electrophysiology, 26, 73–81.

  12. 12.

    Morgan, J. M., Scott, P. A., Turner, N. G., Yue, A. M., & Roberts, P. R. (2009). Targeted left ventricular endocardial pacing using a steerable introducing guide catheter and active fixation pacing lead. Europace, 11, 502–506.

  13. 13.

    Lau, E. W. (2011). Yoked catheter positioning in transseptal endocardial left ventricular lead placement. Pacing and Clinical Electrophysiology, 34(7), 884–93. doi:10.1111/j.1540-8159.2011.03069.x. Epub 2011 Mar 21.

  14. 14.

    Worley, S. J., Leon, A., & Wilkoff, B. L. (2004). Chapter 6-Anatomy and implantation techniques for biventricular devices. In K. A. Ellenbogen, G. N. Kay, & B. L. Wilkoff (Eds.), Device Therapy for Congestive Heart Failure. Philadelphia: Saunders.

  15. 15.

    Worley SJ. Chapter 22-Left ventricular lead implantation. In: EllenbogenKA, KayGN, LauCP, WilkoffBL (eds.): Clinical cardiac pacing, defibrillation and resynchronization therapy, 4th Ed. Saunders, Philadelphia, 2011.

  16. 16.

    Worley, S. J., & Ellenbogen, K. A. (2007). Application of interventional procedures adapted for device implantation: new opportunities for device implanters. Pacing and Clinical Electrophysiology, 30, 938–941.

  17. 17.

    Worley, S. J. (2009). How to use balloons as anchors for CS cannulation, LV lead placement and to regain lost CS or target vein access. Heart Rhythm Journal, 6, 1242–1246.

  18. 18.

    Worley, S. J. (2009). CRT delivery systems based on guide support for LV lead placement. Heart Rhythm Journal, 9, 1383–1387.

  19. 19.

    Koos, R., Sinha, A. M., Markus, K., Breithardt, O. A., Mischke, K., Zarse, M., et al. (2004). Comparison of left ventricular lead placement via the coronary venous approach versus lateral thoracotomy in patients receiving cardiac resynchronization therapy. The American Journal of Cardiology, 94, 59–63.

  20. 20.

    Daoud, E. G., Kalbfleisch, S. J., Hummel, J. D., Weiss, R., Augustini, R. S., Duff, S. B., et al. (2002). Implantation techniques and chronic lead parameters of biventricular pacing dual-chamber defibrillators. Journal of Cardiovascular Electrophysiology, 13, 964–70.

  21. 21.

    Derval, N., Steendijk, P., Gula, L. J., Deplagne, A., Laborderie, J., Sacher, F., et al. (2010). Optimizing hemodynamics in heart failure patients by systematic screening of left ventricular pacing sites. The lateral left ventricular wall and the coronary sinus are rarely the best sites. Journal of the American College of Cardiology, 55, 566–575.

  22. 22.

    Spragg, D. D., Dong, J., Fetics, B. J., Helm, R., et al. (2010). Optimal left ventricular endocardial pacing sites for cardiac resynchronization therapy in patients with ischemic cardiomyopathy. Journal of the American College of Cardiology, 56, 774–781.

  23. 23.

    Elencwajg, B., Cabanillas, N., Cardinali, E., & Trainini, J. (2010). Ventricular resynchronization: a new technique and device for endocardial left ventricular lead placement. Revista Argentina de Cardiología, 78, 143–146.

  24. 24.

    Van Gelder, B. M., Houthuizen, P., & Bracke, F. A. (2011). Transseptal left ventricular endocardial pacing: preliminary experience from a femoral approach with subclavian pull-through. Europace, 13(10), 1454–8. Epub 2011 May 11.

  25. 25.

    Worley SJ, Ellenbogen KA Chapter 23. Interventional techniques for device implantation. In: EllenbogenKA, KayGN, LauCP WilkoffBL (eds.): Clinical Cardiac Pacing, Defibrillation and Resynchronization Therapy, 4th Ed. Saunders, Philadelphia, 2011.

  26. 26.

    Kutyifa, V., Merkely, B., Szilágyi, S., Zima, E., Róka, A., Király, A., Osztheimer, I., Molnár, L., Széplaki, G., & Gellér, L. (2012). Usefulness of electroanatomical mapping during transseptal endocardial left ventricular lead implantation. Europace, 14(4), 599–604. Epub 2011 Dec 22.

Download references

Conflict of interest

S.J.W. receives compensation in various forms from Medtronic, Pressure Products, Guidant, and St. Jude Medical.

Author information

Correspondence to Seth J. Worley.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Video clips demonstrating the steps in this technique are available in the online supporting information of the article (AVI 5671 kb)

ESM 1

Video clips demonstrating the steps in this technique are available in the online supporting information of the article (AVI 5671 kb)

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Patel, M.B., Worley, S.J. Snare coupling of the pre-pectoral pacing lead delivery catheter to the femoral transseptal apparatus for endocardial cardiac resynchronization therapy. J Interv Card Electrophysiol 36, 209–216 (2013). https://doi.org/10.1007/s10840-012-9734-x

Download citation

Keywords

  • Endocardial pacing
  • Transseptal lead placement
  • Snare catheter
  • Deflectable lead delivery catheter
  • SelectSite® catheter
  • Cardiac resynchronization therapy
  • Loop snare