Advertisement

International Ophthalmology

, Volume 38, Issue 5, pp 1993–2003 | Cite as

Pattern noise (PANO): a new automated functional glaucoma test

  • Sylvain el-Khoury
  • Thomas Hannen
  • Diana Carmen Dragnea
  • Faustin Ngounou
  • Paul-Rolf Preußner
Original Paper
  • 40 Downloads

Abstract

Purpose

To present a newly developed visual field device (pattern noise: PANO) designed to be sensitive to glaucoma defects, cost-effective, material-practical and easy to repair and therefore particularly suited for low-income countries, where glaucoma can be highly prevalent (e.g. sub-Saharan Africa).

Methods

This is primarily a descriptive paper, but it also includes a prospective matched case–control pilot study. Hardware, stimulus, target configuration, testing strategy and result sheet are described. The main outcome measure is the contrast level (range 2–64). Targets are composed of bright/dark pixels flickering with 18 Hz and have a size of 5°. Pixel size is approximated to the hill of vision. Average luminance of targets is constant and equals background luminance.The study was performed in the West Region in Cameroon. Twenty eyes of 20 newly presenting patients with glaucomatous optic disc cupping on funduscopy were compared with 20 eyes of 20 normal patients matched in age and laterality of eye.

Results

Mean age was 32.9 ± 18.8 years for glaucoma patients and 32.2 ± 15.6 years for healthy subjects. Mean contrast threshold was significantly higher in eyes with abnormal disc (16.2 ± 14.3 vs. 4.4 ± 0.8, P = 0.002). Correlation of mean contrast thresholds and cup-to-disc ratio was significant (r = 0.59; P = 0.006). Average examination time was significantly longer for glaucoma eyes compared to healthy eyes (8.2 vs. 6.1 min, P < 0.001), whereas error rate did not differ (4.8 ± 2.5% vs. 4.1 ± 1.8%, P = 0.33).

Conclusion

PANO demonstrated visual field defects in patients with glaucomatous optic disc. Defects correlated significantly with glaucomatous optic nerve head morphological alterations. Healthy eyes obtained normal results. More studies are needed to establish PANO.

Keywords

Glaucoma Visual field Healthcare research 

Notes

Compliance with ethical standards

Conflict of interest

No conflicting relationship exists for any author.

References

  1. 1.
    World Health Organization (2007) Global initiative for the elimination of avoidable blindness: action plan 2006–2011. http://www.who.int/iris/handle/10665/43754
  2. 2.
    Quigley HA, Broman AT (2006) The number of people with glaucoma worldwide in 2010 and 2020. Br J Ophthalmol 90(3):262–267CrossRefPubMedCentralGoogle Scholar
  3. 3.
    Öhnell H, Heijl A, Brenner L, Anderson H, Bengtsson B (2016) Structural and Functional Progression in the Early Manifest Glaucoma Trial. Ophthalmology. doi: 10.1016/j.ophtha.2016.01.039 CrossRefPubMedPubMedCentralGoogle Scholar
  4. 4.
    Quigley HA, Dunkelberger GR, Green WR (1989) Retinal ganglion cell atrophy correlated with automated perimetry in human eyes with glaucoma. Am J Ophthalmol 107(5):453–464CrossRefPubMedCentralGoogle Scholar
  5. 5.
    Quigley HA, Addicks EM, Green WR (1982) Optic nerve damage in human glaucoma: III. Quantitative correlation of nerve fiber loss and visual field defect in glaucoma, ischemic neuropathy, papilledema, and toxic neuropathy. Arch Ophthalmol 100(1):135–146CrossRefPubMedCentralGoogle Scholar
  6. 6.
    Harwerth RS, Carter-Dawson L, Shen F, Smith EL, Crawford MLJ (1999) Ganglion cell losses underlying visual field defects from experimental glaucoma. Invest Ophthalmol Vis Sci 40(10):2242–2250PubMedPubMedCentralGoogle Scholar
  7. 7.
    Olsen AS, Alberti M, Serup L, la Cour M, Damato B, Kolko M (2016) Glaucoma detection with damato multifixation campimetry online. Eye 30:731CrossRefPubMedCentralGoogle Scholar
  8. 8.
    Ianchulev T, Pham P, Makarov V, Francis B, Minckler D (2005) Peristat: a computer-based perimetry self-test for cost-effective population screening of glaucoma. Curr Eye Res 30(1):1–6CrossRefPubMedCentralGoogle Scholar
  9. 9.
    Johnson CA, Thapa SS, Robin AL (2014) PO412. Use of an iPad tablet to perform visual field screening in Nepal. Paper presented at the American Academy of Ophthalmology (AAO) annual meeting, McCormick Place, Chicago, USAGoogle Scholar
  10. 10.
    Quigley HA (1996) Number of people with glaucoma worldwide. Br J Ophthalmol 80(5):389–393CrossRefPubMedCentralGoogle Scholar
  11. 11.
    Resnikoff S, Pascolini D, Etya’ale D, Kocur I, Pararajasegaram R, Pokharel GP, Mariotti SP (2004) Global data on visual impairment in the year 2002. Bull World Health Organ 82(11):844–851PubMedPubMedCentralGoogle Scholar
  12. 12.
    Committee for Development Policy (2015) The least developed country category. 2015 country snapshot. https://www.un.org/development/desa/dpad/wp-content/uploads/sites/45/LDC_Fact_Sheet_booklet.pdf
  13. 13.
    Preußner P-R, Großmann A, Ngounou F, Kouogan G, Tamon J (2009) Glaucoma screening in western Cameroon. Graefe’s Arch Clin Exp Ophthalmol 247(12):1671–1675CrossRefGoogle Scholar
  14. 14.
    Buhrmann RR, Quigley HA, Barron Y, West SK, Oliva MS, Mmbaga BBO (2000) Prevalence of glaucoma in a rural east African population. Invest Ophthalmol Vis Sci 41(1):40–48PubMedPubMedCentralGoogle Scholar
  15. 15.
    Ntim-Amponsah CT, Amoaku WMK, Ofosu-Amaah S, Ewusi RK, Idirisuriya-Khair R, Nyatepe-Coo E, Adu-Darko M (2004) Prevalence of glaucoma in an African population. Eye 18(5):491–497CrossRefPubMedCentralGoogle Scholar
  16. 16.
    Kloevekorn S (2014) Test acer aspire E1-572G-54204G50Mnkk notebook. http://www.notebookcheck.net/Review-Acer-Aspire-E1-572G-54204G75Mnkk-Notebook.99958.0.html
  17. 17.
    Johnson CA, Cioffi GA, Van Buskirk EM (1999) Frequency doubling technology perimetry using a 24-2 stimulus presentation pattern. Optom Vis Sci 76(8):571–581CrossRefPubMedCentralGoogle Scholar
  18. 18.
    Armaly MF (1971) Visual field defects in early open angle glaucoma. Trans Am Ophthalmol Soc 69:147PubMedPubMedCentralGoogle Scholar
  19. 19.
    Drance SM (1972) The glaucomatous visual field. Br J Ophthalmol 56(3):186CrossRefPubMedCentralGoogle Scholar
  20. 20.
    Harrington DO (1964) The Bjerrum scotoma. Trans Am Ophthalmol Soc 62:324PubMedPubMedCentralGoogle Scholar
  21. 21.
    Garway-Heath DF, Ruben ST, Viswanathan A, Hitchings RA (1998) Vertical cup/disc ratio in relation to optic disc size: its value in the assessment of the glaucoma suspect. Br J Ophthalmol 82(10):1118–1124. doi: 10.1136/bjo.82.10.1118 CrossRefPubMedPubMedCentralGoogle Scholar
  22. 22.
    Harizman N, Oliveira C, Chiang A et al (2006) The ISNT rule and differentiation of normal from glaucomatous eyes. Arch Ophthalmol 124(11):1579–1583. doi: 10.1001/archopht.124.11.1579 CrossRefPubMedPubMedCentralGoogle Scholar
  23. 23.
    Bach M, Hawlina M, Holder GE, Marmor MF, Meigen T, Miyake Y (2000) Standard for pattern electroretinography. Doc Ophthalmol 101(1):11–18CrossRefPubMedCentralGoogle Scholar
  24. 24.
    Aulhorn E, Koest G (1988) Rauschfeldkampimetrie. Eine neuartige perimetrische Untersuchungsmethode. Klin Monatsblaetter fuer Augenheilkd 192:284–288CrossRefGoogle Scholar
  25. 25.
    Drum B, Massof R, O’Leary D, Quigley H, Breton M, Krupin T, Leight J, Mangat-Rai J (1987) Pattern discrimination perimetry: a new concept in visual field testing. In: Greve EL, Heijl A (eds) Seventh international visual field symposium, Amsterdam, September 1986. Springer, Dordrecht, pp 433–440. doi: 10.1007/978-94-009-3325-5_55 CrossRefGoogle Scholar
  26. 26.
    Green DG (1970) Regional variations in the visual acuity for interference fringes on the retina. J Physiol 207(2):351–356CrossRefPubMedCentralGoogle Scholar
  27. 27.
    Curcio CA, Sloan KR, Kalina RE, Hendrickson AE (1990) Human photoreceptor topography. J Comp Neurol 292(4):497–523CrossRefPubMedCentralGoogle Scholar
  28. 28.
    Johnson CA, Adams CW, Lewis RA (1988) Fatigue effects in automated perimetry. Appl Opt 27(6):1030–1037CrossRefPubMedCentralGoogle Scholar
  29. 29.
    Searle AE, Wild JM, Shaw DE, O’Neill EC (1991) Time-related variation in normal automated static perimetry. Ophthalmology 98(5):701–707CrossRefPubMedCentralGoogle Scholar
  30. 30.
    De Tarso Ponte Pierre-Filho P, Schimiti RB, De Vasconcellos JPC, Costa VP (2006) Sensitivity and specificity of frequency-doubling technology, tendency-oriented perimetry, SITA Standard and SITA fast perimetry in perimetrically inexperienced individuals. Acta Ophthalmol Scand 84(3):345–350CrossRefGoogle Scholar
  31. 31.
    Teresa CM, Andrea P, Stefano A, Vincenzo R, Maria RS (2007) The influence of learning effect on frequency doubling technology perimetry (matrix). J Glaucoma 16(3):297–301CrossRefGoogle Scholar
  32. 32.
    Tahir HJ, Murray IJ, Parry NR, Aslam TM (2014) Optimisation and assessment of three modern touch screen tablet computers for clinical vision testing. PLoS ONE 9(4):e95074CrossRefPubMedCentralGoogle Scholar
  33. 33.
    Black J, Jacobs R, Phillips G, Chen L, Tan E, Tran A, Thompson B (2013) An assessment of the iPad as a testing platform for distance visual acuity in adults. BMJ Open 3(6):e002730CrossRefPubMedCentralGoogle Scholar

Copyright information

© Springer Science+Business Media B.V. 2017

Authors and Affiliations

  • Sylvain el-Khoury
    • 1
  • Thomas Hannen
    • 1
  • Diana Carmen Dragnea
    • 1
  • Faustin Ngounou
    • 2
  • Paul-Rolf Preußner
    • 1
  1. 1.Department of OphthalmologyUniversity Medical Centre MainzMainzGermany
  2. 2.Presbyterian Eye HospitalBafoussamCameroon

Personalised recommendations