High-Accuracy Emissivity Data on the Coatings Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black

  • A. AdibekyanEmail author
  • E. Kononogova
  • C. Monte
  • J. Hollandt
Open Access
TEMPMEKO 2016
Part of the following topical collections:
  1. TEMPMEKO 2016: Selected Papers of the 13th International Symposium on Temperature, Humidity, Moisture and Thermal Measurements in Industry and Science

Abstract

The Physikalisch-Technische Bundesanstalt determined the directional spectral emissivities of several widely used black coatings: Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black. These are and were often applied in different industrial and scientific applications. The measurements are taken angularly resolved over a range from \(10{^{\circ }}\) to \(70{^{\circ }}\). They cover the temperature range typical for the application of the respective coating and a wide wavelength range from \(4~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\). The respective directional total emissivities and hemispherical total emissivities are given as well. The measurements were taken under vacuum at the reduced background calibration facility to achieve low uncertainties and avoid atmospheric interferences. Additionally, some measurements were taken with the emissivity measurement setup in air.

Keywords

Acktar Fractal Black Aeroglaze Z306 Black coating Emissivity Herberts 1534 Nextel 811-21 Uncertainty 

Abbreviations

CCT

Consultative committee for thermometry

DLaTGS

Deuterated l-alanine-doped triglycine sulfate

EMAF

Emissivity measurement in air facility

FDTGS

FIR deuterated triglycine sulfate

FIR

Far-infrared wavelength range

IR-FTS

Infrared Fourier transform spectrometer

ITS-90

International Temperature Scale of 1990

KBr

Potassium bromide

\(\hbox {LN}_{2}\)

Liquid nitrogen

MCT

Liquid nitrogen-cooled mercury cadmium telluride

MIR

Mid-infrared wavelength range

PRT

Platinum resistance thermometer

PTB

Physikalisch-Technische Bundesanstalt

RBCF

Reduced background calibration facility

1 Introduction

Black coatings featuring a high emissivity corresponding to a low reflectivity from the visible to the far-infrared spectral range are widely used in optical instruments to reduce stray light, in control panels or automotive instrumentation to enhance contrast and readability, in thermal applications to control and maximize radiative cooling or heating, in spectroscopy as reference surfaces and radiance standards. For these applications, reliable data of their optical properties are mandatory. However, emissivity data of black coatings published in the literature or in standardization documents often do not provide angularly and/or spectrally resolved information and/or lack uncertainties. The Physikalisch-Technische Bundesanstalt (PTB) determined the directional spectral and directional total emissivities under several angles of observation and also the hemispherical total emissivities of several widely used diffuse black coatings: Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black.

2 Measurements Scheme

The measurements were taken under vacuum (at a pressure of \(10^{-6}\) hPa) at the reduced background calibration facility (RBCF) [1, 2] of PTB to achieve low uncertainties and avoid atmospheric interferences. Furthermore, some of the measurements are compared with measurements obtained with the emissivity measurement in air facility (EMAF) [3], which is routinely operated at PTB for several years. The EMAF successfully took part in an international emissivity comparison organized by the consultative committee for thermometry (CCT) [4].

The measurement scheme for the determination of the emissivity is a comparison of the spectral radiance of the sample which is mounted on a heater and placed inside a temperature-stabilized spherical enclosure against the spectral radiances of two reference blackbodies at different temperatures [1, 2, 3]. The comparison is performed with an infrared Fourier transform spectrometer (IR-FTS). By using two reference blackbodies at different temperatures, the residual background radiation of the instrumentation can be corrected. The “main” reference blackbody is usually operated at a temperature close to the radiance temperature of the sample to achieve a similar signal level. The second reference blackbody is operated at a temperature significantly lower than the temperature of the main blackbody. At the RBCF, this is an liquid nitrogen-cooled (\(\hbox {LN}_{2}\)) blackbody, which is operated at the temperature of boiling nitrogen. At the EMAF, the second reference blackbody is a liquid-cooled blackbody operated at \(18~{^{\circ }}\)C.

3 Measurement Uncertainty

The calculation of the emissivity from the measured detector signals is based on the complete radiation budget, considering every radiation contribution in the optical path. The uncertainty of the directional spectral emissivity is calculated via a Monte–Carlo method [5, 6] based on the final determination equation of the directional spectral emissivity and is spectrally dependent. Furthermore, the uncertainty budget is calculated for each individual measurement, since the relevance of the individual uncertainty contributions depends significantly on the conditions of the measurement and the sample investigated. A detailed description of the data evaluation procedure and the uncertainty calculation is given in [6, 7].

4 Experimental Results on Different Black Coatings

The emissivity measurements of the black coatings were taken at two setups at PTB—under vacuum at the RBCF and in air at the EMAF—by applying the measurement schemes and data, and uncertainty evaluation described shortly above and described in detail in [1, 2, 3, 6, 7]. Two sets of detectors and beamsplitters were used with the IR-FTS according to the wavelength and temperature range of the measurements. In the range from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\), a deuterated l-alanine-doped triglycine sulfate (DLaTGS) or liquid nitrogen-cooled mercury cadmium telluride (MCT) detector in combination with a potassium bromide (KBr) beamsplitter were used. Due to its higher responsivity, the MCT was used for the measurements at lower sample temperatures. In the range from \(14~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\), the emissivity was determined by using a far-infrared deuterated triglycine sulfate (FDTGS) detector in combination with a \(6~\upmu \hbox {m}\) Multilayer Mylar beamsplitter.

Each of the investigated black coatings—Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black—was applied to copper substrates. The disk-shaped substrates have a diameter of 90 mm and a thickness of 5 mm and feature a hole of 2 mm diameter and 50 mm depth from the circumference to accommodate a platinum resistance thermometer (PRT) temperature sensor for monitoring the substrate temperature. The surface of the copper substrates was plane and untreated.

In the following, the individual emissivity results for the four different coatings are presented.

4.1 Nextel 811-21

Nextel 811-21 was in the past provided by 3M and is now available from “Mankiewicz Gebr. & Co” [8] where we obtained it. The substrates were coated according to the manufactures instructions, i.e., spray coating of the primer and spray coating of the Nextel 811-21 coating and subsequent heating at \(120~{^{\circ }}\)C. Some of the emittance properties of Nextel 811-21 are already provided in the literature [9, 10]. Here, the spectral emissivity of the sample Nextel 811-21 was determined under vacuum in the wavelength range from \(5~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at a temperature of \(120~{^{\circ }}\)C under different angles of observation. The resulting spectral curves are plotted in Fig. 1. Some of the measured angles are omitted here for clarity of the figure. The spectrally integrated directional total emissivities are shown in the inset of Fig. 1 with their corresponding standard uncertainties. A theoretical model based on Fresnel equations is fitted to these values and plotted as a solid line. Its integration over the angular range from \(0{^{\circ }}\) to \(90{^{\circ }}\) yields the hemispherical total emissivity. A decrease in the directional spectral emissivities and accordingly of the directional total emissivities toward larger angles is clearly visible, which is typical for high-emitting samples. The numerical values of the directional total and hemispherical total emissivities of Nextel 811-21 in the wavelength range from \(5~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at temperature of \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at temperature of \(120~{^{\circ }}\)C are provided in Table 1. Within the uncertainty of the measurements, no significant spectral change in emissivity can be observed between \(25~{^{\circ }}\)C and \(120~{^{\circ }}\)C.
Fig. 1

Directional spectral emissivity of Nextel 811-21 measured at a temperature of \(120~{^{\circ }}\)C for different angles of observation. The spectrally integrated values of the directional total emissivity with their standard uncertainties as well as the hemispherical total emissivity are shown in the inset

Table 1

Directional total and hemispherical total emissivities of Nextel 811-21 in the wavelength range from \(5~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(120~{^{\circ }}\)C with their corresponding standard uncertainties

Angle

Nextel 811-21

\(u (\varepsilon )\)

Nextel 811-21

\(u (\varepsilon )\)

\(\varepsilon \) (\(25~{^{\circ }}\)C)

(k=1)

\(\varepsilon \) (\(120~{^{\circ }}\)C)

(k=1)

\(10{^{\circ }}\)

0.971

0.012

0.9717

0.0058

\(20{^{\circ }}\)

0.971

0.013

0.9719

0.0057

\(30{^{\circ }}\)

0.971

0.013

0.9685

0.0057

\(40{^{\circ }}\)

0.966

0.012

0.9684

0.0057

\(50{^{\circ }}\)

0.966

0.012

0.9609

0.0058

\(60{^{\circ }}\)

0.949

0.012

0.9518

0.0058

\(70{^{\circ }}\)

0.912

0.013

0.9168

0.0057

\({\varepsilon }_{hem}\)

0.941

0.012

0.9381

0.0056

The directional spectral emissivity of Nextel 811-21 from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) at a temperature of \(120~{^{\circ }}\)C observed under an angle of \(10{^{\circ }}\) is plotted in Fig. 2. The shaded area around the curve is the range of the standard uncertainty of the measurement. The blue curve in the lower part of the graph and the corresponding right-hand ordinate axis show the spectral distribution of the uncertainty. The numerical values of the directional total and hemispherical total emissivities of Nextel 811-21 in the wavelength range from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) are provided in Table 2.
Fig. 2

Directional spectral emissivity of Nextel 811-21 up to a wavelength of \(100~\upmu \hbox {m}\) measured at a temperature of \(120~{^{\circ }}\)C and under an angle of observation of \(10{^{\circ }}\) with respect to the surface normal. The spectral distribution of the standard uncertainty is shown in the lower part of the plot with the respective scale on the right-hand ordinate axis

Table 2

Directional total and hemispherical total emissivities of Nextel 811-21 in the wavelength range from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) with their corresponding standard uncertainties

Angle

Nextel 811-21

\(u (\varepsilon )\)

\(\varepsilon (120~{^{\circ }}\)C)

(k=1)

\(10{^{\circ }}\)

0.9712

0.0038

\(30{^{\circ }}\)

0.9689

0.0036

\(50{^{\circ }}\)

0.9579

0.0038

\(70{^{\circ }}\)

0.8947

0.0048

\({\varepsilon }_{hem}\)

0.9378

0.0039

Fig. 3

Directional spectral emissivity of Herberts 1534 in the wavelength range from \(5.7~\upmu \hbox {m}\) to \(18.2~\upmu \hbox {m}\) measured at a temperature of \(120~{^{\circ }}\)C in air and at a temperature of \(25~{^{\circ }}\)C under vacuum. Both measurements were taken under an angle of observation of \(10{^{\circ }}\) with respect to the surface normal. The spectral distribution of the standard uncertainty of measurement at \(25~{^{\circ }}\)C under vacuum is shown in the lower half of the plot with the respective scale on the right-hand ordinate axis

4.2 Herberts 1534

Herberts 1534 is no longer in production but was used in the past for the far-infrared wavelength range (FIR) and space applications. There are still a number of instrumentations and blackbodies in use which are coated with this paint. Here, the spectral emissivity of the sample Herberts 1534 was measured under vacuum at a temperature of \(25~{^{\circ }}\)C and in air at a temperature of \(120~{^{\circ }}\)C and is plotted in Fig. 3. The brown curve, with its range of standard uncertainty given as a shaded area, shows the directional spectral emissivity under vacuum. Additionally, the uncertainty is also shown separately with the blue curve in the lower part of Fig. 3 and the corresponding right-hand ordinate axis. The green curve in Fig. 3 shows the directional spectral emissivity measured in air at a temperature of \(120~{^{\circ }}\)C. The directional spectral emissivity of Herberts 1534 measured in the wavelength range from \(14.5~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) and, for comparison, the results obtained in the mid-infrared wavelength (MIR) range are given in Fig. 4. The results of both measurements in the overlapping spectral range from \(14.5~\upmu \hbox {m}\) to \(18~\upmu \hbox {m}\) are consistent within the range of their standard uncertainties. The numerical values of the directional total and hemispherical total emissivities are presented in Tables 3 and 4.
Fig. 4

Directional spectral emissivity of Herberts 1534 measured under vacuum up to a wavelength of \(100~\upmu \hbox {m}\) and compared in the overlapping wavelength range with measurements obtained under vacuum and in air in the MIR range

Table 3

Directional total and hemispherical total emissivities of Herberts 1534 in the wavelength range from \(5~\upmu \hbox {m}\) to \(18.2~\upmu \hbox {m}\) at \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(120~{^{\circ }}\)C with their corresponding standard uncertainties

Angle

Herberts 1534

\(u (\varepsilon )\)

Herberts 1534

\(u (\varepsilon )\)

\(\varepsilon \) (\(25~{^{\circ }}\)C)

(k=1)

\(\varepsilon \) (\(120~{^{\circ }}\)C)

(k=1)

\(10{^{\circ }}\)

0.917

0.022

0.918

0.024

\(20{^{\circ }}\)

0.915

0.021

0.916

0.027

\(30{^{\circ }}\)

0.918

0.021

0.918

0.025

\(40{^{\circ }}\)

0.916

0.021

0.917

0.024

\(50{^{\circ }}\)

0.887

0.021

0.906

0.024

\(60{^{\circ }}\)

0.883

0.021

0.891

0.024

\(70{^{\circ }}\)

0.843

0.021

0.859

0.025

\({\varepsilon }_{hem}\)

0.880

0.021

0.900

0.024

Table 4

Directional total and hemispherical total emissivities of Herberts 1534 in the wavelength range from \(14.1~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) with their corresponding standard uncertainties

Angle

Herberts 1534

\(u (\varepsilon )\)

\(\varepsilon \) (\(120~{^{\circ }}\)C)

(k=1)

\(10{^{\circ }}\)

0.915

0.013

\(30{^{\circ }}\)

0.928

0.012

\(50{^{\circ }}\)

0.906

0.013

\(70{^{\circ }}\)

0.842

0.014

\({\varepsilon }_{hem}\)

0.888

0.013

Fig. 5

Directional spectral emissivities of three Aeroglaze Z306 samples with thicknesses of \(44~\upmu \hbox {m}\), \(99~\upmu \hbox {m}\) and \(236~\upmu \hbox {m}\) on polished copper substrates under an angle of observation of \(10{^{\circ }}\)

Fig. 6

Directional spectral emissivities of Aeroglaze Z306 sample with thickness of \(99~\upmu \hbox {m}\) under an angle of observation of \(10{^{\circ }}\) at a temperatures of \(25 {^{\circ }}\)C and \(150 {^{\circ }}\)C

Table 5

Directional total and hemispherical total emissivities of Aeroglaze Z306 in the wavelength range from \(5~\upmu \hbox {m}\) to \(16.6~\upmu \hbox {m}\) at \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(150~{^{\circ }}\)C with their corresponding standard uncertainties

Angle

Aeroglaze Z306

\(u (\varepsilon )\)

Aeroglaze Z306

\(u (\varepsilon )\)

\(\varepsilon \) (\(25~{^{\circ }}\)C)

(k=1)

\(\varepsilon \) (\(150~{^{\circ }}\)C)

(k=1)

\(10{^{\circ }}\)

0.924

0.012

0.9638

0.0054

\(20{^{\circ }}\)

0.911

0.013

\(30{^{\circ }}\)

0.919

0.013

0.9615

0.0053

\(40{^{\circ }}\)

0.923

0.012

\(50{^{\circ }}\)

0.922

0.013

0.9472

0.0052

\(60{^{\circ }}\)

0.888

0.015

\(70{^{\circ }}\)

0.818

0.013

0.8579

0.0049

\({\varepsilon }_{hem}\)

0.881

0.013

0.9229

0.0052

Table 6

Directional total and hemispherical total emissivities of Aeroglaze Z306 at three different thicknesses. All measurements were taken at a temperature of \(150~{^{\circ }}\)C and in the wavelength range from \(14.7~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\)

Angle

Aeroglaze Z306, \(44~\upmu \hbox {m}\)

u (\(\varepsilon \))

Aeroglaze Z306, \(99~\upmu \hbox {m}\)

\(u (\varepsilon )\)

Aeroglaze Z306, \(236~\upmu \hbox {m}\)

\(u (\varepsilon )\)

\(\varepsilon \) (\(150~{^{\circ }}\)C)

(k=1)

\(\varepsilon \) (\(150~{^{\circ }}\)C)

(k=1)

\(\varepsilon \) (\(150~{^{\circ }}\)C)

(k=1)

\(10{^{\circ }}\)

0.8793

0.0066

0.9434

0.0102

0.9553

0.0054

\(30{^{\circ }}\)

0.8852

0.0065

0.9422

0.0094

0.9555

0.0054

\(50{^{\circ }}\)

0.8823

0.0066

0.9288

0.0092

0.9402

0.0052

\(70{^{\circ }}\)

0.7966

0.0067

0.8299

0.0082

0.8368

0.0047

\({\varepsilon }_{hem}\)

0.8512

0.0064

0.8971

0.0101

0.9129

0.0052

Fig. 7

Directional spectral emissivity of Acktar Fractal Black from \(5~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) measured at a temperature of \(120~{^{\circ }}\)C and under an angle of observation of \(10{^{\circ }}\) with respect to the surface normal. The spectral distribution of the standard uncertainty is shown in the lower half of the plot with the respective scale on the right-hand ordinate axis

Fig. 8

Directional spectral emissivity of Acktar Fractal Black under different angles of observation with respect to the surface normal

4.3 Aeroglaze Z306

The Aeroglaze Z306 is an absorptive polyurethane coating which is often used in aerospace applications and which is well suited for vacuum conditions. Some of its emittance properties can be found in [11, 12]. In preparation of the measurements of Aeroglaze Z306, three copper substrates were spray coated with Aeroglaze according to the instructions given in the European Cooperation for Space Standardization document ECSS-Q-70-25A [13] with thicknesses of \(44~\upmu \hbox {m}\), \(99~\upmu \hbox {m}\) and \(236~\upmu \hbox {m}\). The measurements were taken at a temperature of \(150~{^{\circ }}\)C and are shown in Fig. 5 for an angle of observation of \(10{^{\circ }}\). A significant decrease in emissivity above \(22~\upmu \hbox {m}\) can be seen for the sample with a coating thickness of \(44~\upmu \hbox {m}\). The sample with the coating thickness of \(236~\upmu \hbox {m}\) shows a nearly constant (spectrally flat) emissivity. The clearly visible modulation in the emissivity of all three samples for longer wavelengths is explained by an onset of transparency of the coating toward longer wavelengths and multiple beam interference within the coating due to the highly reflective copper substrate. The observed modulation period is inversely proportional to the optical thickness of the coating.

The Aeroglaze Z306 coating of all three thicknesses is fully opaque in the MIR range, but the directional spectral emissivity shows a slight temperature dependence as shown in Fig. 6. The directional spectral emissivity at \(25~{^{\circ }}\)C is shown as the brown curve with the corresponding uncertainty range plotted as the shaded area around the curve. The green curve in Fig. 6 illustrates the directional spectral emissivity at \(150~{^{\circ }}\)C.

Tables 5 and 6 provide the numerical values of the directional total and hemispherical total emissivities of Aeroglaze Z306.
Table 7

Directional total and hemispherical total emissivities of Acktar Fractal Black in the wavelength range from \(5.7~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(25~{^{\circ }}\)C and from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(120~{^{\circ }}\)C with their corresponding standard uncertainties

Angle

Acktar Fractal Black

\(u (\varepsilon )\)

Acktar Fractal Black

\(u (\varepsilon )\)

\(\varepsilon \) (\(25~{^{\circ }}\)C)

(k=1)

\(\varepsilon \) (\(120~{^{\circ }}\)C)

(k=1)

\(10{^{\circ }}\)

0.963

0.017

0.9603

0.0036

\(20{^{\circ }}\)

0.962

0.018

0.9604

0.0037

\(30{^{\circ }}\)

0.962

0.018

0.9565

0.0043

\(40{^{\circ }}\)

0.958

0.018

0.9513

0.0037

\(50{^{\circ }}\)

0.948

0.017

0.9432

0.0038

\(60{^{\circ }}\)

0.919

0.017

0.9134

0.0038

\(70{^{\circ }}\)

0.847

0.017

0.8409

0.0039

\({\varepsilon }_{hem}\)

0.924

0.017

0.9104

0.0038

Table 8

Directional total and hemispherical total emissivities of Acktar Fractal Black with their corresponding standard uncertainties. The measurements were taken at a temperature of \(120~{^{\circ }}\)C and in the wavelength range from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\)

Angle

Acktar Fractal Black

\(u (\varepsilon )\)

\(\varepsilon \) (\(120~{^{\circ }}\)C)

(k=1)

\(10{^{\circ }}\)

0.928

0.018

\(30{^{\circ }}\)

0.930

0.018

\(50{^{\circ }}\)

0.911

0.018

\(70{^{\circ }}\)

0.780

0.018

\({\varepsilon }_{hem}\)

0.895

0.018

4.4 Acktar Fractal Black

Acktar Fractal Black is a highly emitting coating of the company “Acktar” [14]. Measurements of its reflectivity and other characteristics of this coating can be found in [14, 15], but direct emissivity measurements are missing. Here, the directional spectral emissivity of Acktar Fractal Black was measured under vacuum from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at \(120~{^{\circ }}\)C. The measurement under an angle of observation of \(10{^{\circ }}\) is plotted in Fig. 7. The spectral distribution of the uncertainty is also shown in the lower half of Fig. 7 with the corresponding ordinate axis on the right-hand side. The directional spectral emissivity was also measured under different angles of observation from \(10{^{\circ }}\) to \(70{^{\circ }}\) and is plotted in Fig. 8. Tables 7 and 8 provide the numerical values of the directional total and hemispherical total emissivities of Acktar Fractal Black.

4.5 Total and Hemispherical Emissivities

The integrated quantities, total and hemispherical emissivity, of the four black coatings are presented here for an overview, depending on temperature and the integrated wavelength range (MIR: \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\), FIR: \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\)). The directional total emissivities with their standard uncertainties are plotted in Figs. 9 and 10 and the hemispherical total emissivities in Fig. 11.
Fig. 9

Directional total emissivities of the investigated coatings in the wavelength range from \(4~\upmu \hbox {m}\) to \(20~\upmu \hbox {m}\) at the temperatures of \(120~{^{\circ }}\)C, respectively, \(150~{^{\circ }}\)C with their corresponding standard uncertainties

Fig. 10

Directional total emissivities of the investigated coatings in the wavelength range from \(20~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) at the temperatures of \(120~{^{\circ }}\)C, respectively, \(150~{^{\circ }}\)C with their corresponding standard uncertainties

Fig. 11

Hemispherical total emissivities of the investigated coatings with their corresponding standard uncertainties. The results are grouped according to the integrated spectral ranges and measurement temperatures

5 Conclusion

Highly accurate measurements of directional spectral, directional total and hemispherical total emissivities for four widely used black coatings—Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black—were taken by PTB in the wavelengths range from \(4~\upmu \hbox {m}\) to \(100~\upmu \hbox {m}\) at different temperatures in the range from \(25~{^{\circ }}\)C to \(150~{^{\circ }}\)C.

The achieved absolute standard uncertainty (\(k=1\)) is calculated according to [5, 6, 7]. It is 0.005 or better for measurements at high temperatures (\(T=120~{^{\circ }}\)C and \(T=150~{^{\circ }}\)C) and it is better than 0.022 (for some measurements better than 0.011) for measurements at \(25~{^{\circ }}\)C. It should be noted that the achieved uncertainty significantly depends on the experimental conditions and the investigated coating.

The results achieved in this work are traceable to the International Temperature Scale of 1990 (ITS-90) due to the use of two radiance standards in terms of two reference blackbodies which are directly linked to the ITS-90. Furthermore, some of the measurements at the RBCF were validated by comparing them with measurements performed at EMAF. EMAF has successfully participated in CCT-S1, the supplementary comparison of CCT for spectral emissivity [4]. All emissivity results obtained at the RBCF and at the EMAF are consistent within the uncertainties of the measurements.

6 Note

The spectral emissivity data of the investigated coatings is made available as an electronic supplement to this article on the journals website.

Notes

Acknowledgements

The authors would like to thank the two anonymous reviewers for their valuable suggestions to the paper.

Supplementary material

10765_2017_2212_MOESM1_ESM.pdf (302 kb)
Nextel 811-21 at 25 °C from 5 to 20 μm (pdf 302 KB)
10765_2017_2212_MOESM2_ESM.pdf (369 kb)
Nextel 811-21 at 120 °C from 4 to 20 μm (pdf 368 KB)
10765_2017_2212_MOESM3_ESM.pdf (153 kb)
Nextel 811-21 at 120 °C from 20 to 100 μm (pdf 153 KB)
10765_2017_2212_MOESM4_ESM.pdf (297 kb)
Herberts 1534 at 25 °C from 5 to 18.2 μm (pdf 297 KB)
10765_2017_2212_MOESM5_ESM.pdf (438 kb)
Herberts 1534 at 120 °C from 4 to 20 μm (pdf 437 KB)
10765_2017_2212_MOESM6_ESM.pdf (154 kb)
Herberts 1534 at 120 °C from 20 to 100 μm (pdf 153 KB)
10765_2017_2212_MOESM7_ESM.pdf (290 kb)
Aeroglaze Z306 at 25 °C from 5 to 16.7 μm (pdf 289 KB)
10765_2017_2212_MOESM8_ESM.pdf (268 kb)
Aeroglaze Z306 at 150 °C from 4 to 20 μm (pdf 268 KB)
10765_2017_2212_MOESM9_ESM.pdf (133 kb)
Aeroglaze Z306 with thickness of 44 μm at 150 °C from 20 to 100 μm (pdf 132 KB)
10765_2017_2212_MOESM10_ESM.pdf (133 kb)
Aeroglaze Z306 with thickness of 99 μm at 150 °C from 20 to 100 μm (pdf 132 KB)
10765_2017_2212_MOESM11_ESM.pdf (133 kb)
Aeroglaze Z306 with thickness of 236 μm at 150 °C from 20 to 100 μm (pdf 132 KB)
10765_2017_2212_MOESM12_ESM.pdf (267 kb)
Acktar Fractal Black at 25 °C from 5.7 to 20 μm (pdf 266 KB)
10765_2017_2212_MOESM13_ESM.pdf (368 kb)
Acktar Fractal Black at 120 °C from 4 to 20 μm (pdf 367 KB)
10765_2017_2212_MOESM14_ESM.pdf (153 kb)
Acktar Fractal Black at 120 °C from 20 to 100 μm (pdf 153 KB)

References

  1. 1.
    C. Monte, B. Gutschwager, S.P. Morozova, J. Hollandt, Int. J. Thermophys. (2009). doi: 10.1007/s10765-008-0442-9 CrossRefGoogle Scholar
  2. 2.
    A. Adibekyan, C. Monte, M. Kehrt, B. Gutschwager, J. Hollandt, Int. J. Thermophys. (2015). doi: 10.1007/s10765-014-1745-7 CrossRefGoogle Scholar
  3. 3.
    C. Monte, J. Hollandt, High Temp. High Press. 39, 151–164 (2010)Google Scholar
  4. 4.
    Key and supplementary comparisons, CCT-S1, http://kcdb.bipm.org
  5. 5.
    Joint Committee for Guides in Metrology (JCGM), Evaluation of measurement data-guide to the expression of uncertainty in measurement, http://www.bipm.org/en/committees/jc/jcgm/. Accessed 2008
  6. 6.
    C. Monte, J. Hollandt, Metrologia (2010). doi: 10.1088/0026-1394/47/2/s14 CrossRefGoogle Scholar
  7. 7.
    A. Adibekyan, High-accuracy spectral emissivity measurement for industrial and remote sensing applications. (2016), http://elpub.bib.uni-wuppertal.de/edocs/dokumente/fbc/physik/diss2016/adibekyan
  8. 8.
    Companies website: http://www.nextel-coating.com/
  9. 9.
    J. Lohrengel, R. Todtenhaupt, PTB-Mitteilungen 106, 259–265 (1996)Google Scholar
  10. 10.
    M. Battuello, S. Clausen, J. Hameury, P. Bloembergen, Conference proceedings. TEMPMEKO, 601–606 (1999)Google Scholar
  11. 11.
    M. Persky, Rev. Sci. Instrum. (1999). doi: 10.1063/1.1149739 CrossRefGoogle Scholar
  12. 12.
    P.G. Maloney, P. Smith, V. King, C. Billman, M. Winkler, E. Mazur, Appl. Opt. (2010). doi: 10.1364/AO.49.001065 CrossRefGoogle Scholar
  13. 13.
    ECSS, European Space Agency, The application of the black coating Aeroglaze Z306, ECSS-Q-70-25A (1999)Google Scholar
  14. 14.
    Technical information on companies website: http://www.acktar.com/category/BlackOpticalCoating
  15. 15.
    Y. Salomon, N. Sternberg, I. Gouzman, G. Lempert, E. Grossman, D. Katsir, R. Cotostiano, T. Minton, in Proceedings of the International Symposium on Materials in a Space Environment, 2009Google Scholar

Copyright information

© The Author(s) 2017

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Authors and Affiliations

  • A. Adibekyan
    • 1
    Email author
  • E. Kononogova
    • 1
  • C. Monte
    • 1
  • J. Hollandt
    • 1
  1. 1.Physikalisch-Technische Bundesanstalt (PTB)BerlinGermany

Personalised recommendations