Abstract

Transcatheter aortic valve replacement (TAVR) is a minimally-invasive approach for treating severe aortic stenosis. All clinically-used TAVR valves to date utilize chemically-fixed xenograft as the leaflet material. Inherent limitation of the tissue (e.g., calcific degeneration) motivates the search for alternative leaflet material. Here we introduce a novel polymeric TAVR valve that was designed to address the limitations of tissue-valves. In this study, we experimentally evaluated the hemodynamic performance of the valve and compared its performance to clinically-used valves: a gold standard surgical tissue valve, and a TAVR valve. Our comparative testing protocols included: (i) baseline hydrodynamics (ISO:5840-3), (ii) complementary patient-specific hydrodynamics in a dedicated system, and (iii) thrombogenicity. The patient-specific testing system facilitated comparing TAVR valves performance under more realistic conditions. Baseline hydrodynamics results at CO 4–7 L/min showed superior effective orifice area (EOA) for the polymer valve, most-notably as compared to the reference TAVR valve. Regurgitation fraction was higher in the polymeric valve, but within the ISO minimum requirements. Thrombogenicity trends followed the EOA results with the polymeric valve being the least thrombogenic, and clinical TAVR being the most. Hemodynamic-wise, the results strongly indicate that our polymeric TAVR valve can outperform tissue valves.

Keywords

TAVI TAVR Aortic stenosis Heart valve Prosthetic heart valve Valve hydrodynamics Thrombogenicity Medical device 

Abbreviations

AS

Aortic stenosis

CAVD

Calcific aortic valve disease

CO

Cardiac output

DTE

Device thrombogenic emulation

EOA

Effective orifice area

GFP

Gel-filtered platelets

HR

Heart rate

MAP

Mean arterial pressure

PD

Pulse duplicator

PVL

Paravalvular leak

SAVR

Surgical aortic valve replacement

SV

Stroke volume

TAVR

Transcatheter aortic valve replacement

Notes

Acknowledgements

The authors would like to thank Braile Biomédica (Brazil), for providing us with the Inovare valve samples. This project was supported by NIH-NIBIB Quantum Award Phase II-U01EB012487 (DB), NHLBI STTR R41-HL134418 (DB), and the Center for Biotechnology: a New York State Center for Advanced Technology, New York State Department of Economic Development; and corporate support.

Conflict of interest

Author OMR is a consultant for Polynova Cardiovascular Inc. Authors MJS and DB has stock ownership in Polynova Cardiovascular Inc. Authors BK, WCC, MB and GM declare that they have no conflicts of interest.

Supplementary material

10439_2018_2119_MOESM1_ESM.pdf (1.4 mb)
Supplementary material 1 (PDF 1480 kb)
10439_2018_2119_MOESM2_ESM.mp4 (12.6 mb)
Online Video 1 Front (aortic) view of the test valves in the Vivitro PD, at CO of 5 l/min. Supplementary material 2 (MP4 12938 kb)
10439_2018_2119_MOESM3_ESM.mp4 (16.4 mb)
Online Video 2 Endoscopic front (aortic) view of the test valves in the Vivitro PD, at CO of 5 l/min. Supplementary material 3 (MP4 16800 kb)
10439_2018_2119_MOESM4_ESM.mp4 (4.5 mb)
Online Video 3 Angiogram of the 20-mm Polynova polymeric TAVR valve in the patient-specific CAVD model in the Replicator. On the left is the original angiogram. On the right is the subtracted angiogram for better visualization of regurgitation flow. Supplementary material 4 (MP4 4636 kb)
10439_2018_2119_MOESM5_ESM.mp4 (9.1 mb)
Online Video 4 Angiogram of the 19-mm Carpentier-Edwards Perimount Magna Ease SAVR valve in the patient-specific CAVD model in the Replicator. On the left is the original angiogram. On the right is the subtracted angiogram for better visualization of regurgitation flow. Supplementary material 5 (MP4 9299 kb)
10439_2018_2119_MOESM6_ESM.mp4 (5.5 mb)
Online Video 5 Angiogram of the 20-mm Inovare TAVR valve in the patient-specific CAVD model in the Replicator. On the left is the original angiogram. On the right is the subtracted angiogram for better visualization of regurgitation flow. Supplementary material 6 (MP4 5668 kb)

References

  1. 1.
    Alavi, S. H., E. M. Groves, and A. Kheradvar. The effects of transcatheter valve crimping on pericardial leaflets. Ann. Thorac. Surg. 97:1260–1266, 2014.CrossRefGoogle Scholar
  2. 2.
    American College of Cardiology/American Heart Association Task Force on Practice Guidelines, and The Society of Cardiovascular, and The Society for Cardiovascular, Interventions, S. Society of Thoracic, R. O. Bonow, B. A. Carabello, C. Kanu, A. C. de Leon, Jr, D. P. Faxon, M. D. Freed, W. H. Gaasch, B. W. Lytle, R. A. Nishimura, P. T. O’Gara, R. A. O’Rourke, C. M. Otto, P. M. Shah, J. S. Shanewise, S. C. Smith, Jr, A. K. Jacobs, C. D. Adams, J. L. Anderson, E. M. Antman, D. P. Faxon, V. Fuster, J. L. Halperin, L. F. Hiratzka, S. A. Hunt, B. W. Lytle, R. Nishimura, R. L. Page, and B. Riegel. ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (writing committee to revise the 1998 Guidelines for the Management of Patients With Valvular Heart Disease): developed in collaboration with the Society of Cardiovascular Anesthesiologists: endorsed by the Society for Cardiovascular Angiography and Interventions and the Society of Thoracic Surgeons. Circulation 114:e84–e231, 2006.Google Scholar
  3. 3.
    Arsalan, M., and T. Walther. Durability of prostheses for transcatheter aortic valve implantation. Nat. Rev. Cardiol. 13:360–367, 2016.CrossRefGoogle Scholar
  4. 4.
    Bezuidenhout, D., D. F. Williams, and P. Zilla. Polymeric heart valves for surgical implantation, catheter-based technologies and heart assist devices. Biomaterials 36:6–25, 2015.CrossRefGoogle Scholar
  5. 5.
    Bianchi, M., G. Marom, R. P. Ghosh, H. A. Fernandez, J. R. Taylor, Jr, M. J. Slepian, and D. Bluestein. Effect of balloon-expandable transcatheter aortic valve replacement positioning: a patient-specific numerical model. Artif. Organs 40(12):E292–E302, 2016.CrossRefGoogle Scholar
  6. 6.
    Claiborne, T. E., G. Girdhar, S. Gallocher-Lowe, J. Sheriff, Y. P. Kato, L. Pinchuk, R. T. Schoephoerster, J. Jesty, and D. Bluestein. Thrombogenic potential of Innovia polymer valves versus Carpentier-Edwards Perimount Magna aortic bioprosthetic valves. ASAIO J. 57:26–31, 2011.CrossRefGoogle Scholar
  7. 7.
    Claiborne, T. E., J. Sheriff, M. Kuetting, U. Steinseifer, M. J. Slepian, and D. Bluestein. In vitro evaluation of a novel hemodynamically optimized trileaflet polymeric prosthetic heart valve. J. Biomech. Eng. 135:021021, 2013.CrossRefGoogle Scholar
  8. 8.
    Claiborne, T. E., M. J. Slepian, S. Hossainy, and D. Bluestein. Polymeric trileaflet prosthetic heart valves: evolution and path to clinical reality. Expert Rev. Med. Devices 9:577–594, 2012.CrossRefGoogle Scholar
  9. 9.
    Claiborne, T. E., M. Xenos, J. Sheriff, W.-C. Chiu, J. S. Soares, Y. Alemu, S. Gupta, S. Judex, M. J. Slepian, and D. Bluestein. Towards optimization of a novel trileaflet polymeric prosthetic heart valve via device thrombogenicity emulation. ASAIO J. 59:275–283, 2013.CrossRefGoogle Scholar
  10. 10.
    Dandeniyage, L. S., R. Adhikari, M. Bown, R. Shanks, B. Adhikari, C. D. Easton, T. R. Gengenbach, D. Cookson, and P. A. Gunatillake. Morphology and surface properties of high strength siloxane poly(urethane-urea)s developed for heart valve application. J. Biomed. Mater. Res. B 2018.  https://doi.org/10.1002/jbm.b.34101.Google Scholar
  11. 11.
    Dandeniyage, L. S., P. A. Gunatillake, R. Adhikari, M. Bown, R. Shanks, and B. Adhikari. Development of high strength siloxane poly(urethane-urea) elastomers based on linked macrodiols for heart valve application. J. Biomed. Mater. Res. B 106(5):1712–1720, 2017.CrossRefGoogle Scholar
  12. 12.
    Dasi, L. P., H. Hatoum, A. Kheradvar, R. Zareian, S. H. Alavi, W. Sun, C. Martin, T. Pham, Q. Wang, P. A. Midha, V. Raghav, and A. P. Yoganathan. On the mechanics of transcatheter aortic valve replacement. Ann. Biomed. Eng. 45:310–331, 2017.CrossRefGoogle Scholar
  13. 13.
    Hatoum, H., J. Dollery, S. M. Lilly, J. Crestanello, and L. P. Dasi. Impact of patient-specific morphologies on sinus flow stasis in transcatheter aortic valve replacement: an in vitro study. J. Thorac. Cardiovasc. Surg. 2018.  https://doi.org/10.1016/j.jtcvs.2018.05.086.Google Scholar
  14. 14.
    Hatoum, H., A. Yousefi, S. Lilly, P. Maureira, J. Crestanello, and L. P. Dasi. An in vitro evaluation of turbulence after transcatheter aortic valve implantation. J. Thorac. Cardiovasc. Surg. 2018.  https://doi.org/10.1016/j.jtcvs.2018.05.042.Google Scholar
  15. 15.
    Jesty, J., and D. Bluestein. Acetylated prothrombin as a substrate in the measurement of the procoagulant activity of platelets: elimination of the feedback activation of platelets by thrombin. Anal. Biochem. 272:64–70, 1999.CrossRefGoogle Scholar
  16. 16.
    Kallis, P., J. F. Sneddon, I. A. Simpson, A. Fung, J. R. Pepper, and E. E. Smith. Clinical and hemodynamic evaluation of the 19-mm Carpentier-Edwards supraannular aortic valve. Ann. Thorac. Surg. 54:1182–1185, 1992.CrossRefGoogle Scholar
  17. 17.
    Kamioka, N., J. Wells, P. Keegan, S. Lerakis, J. Binongo, F. Corrigan, J. Condado, A. Patel, J. Forcillo, L. Ogburn, A. Dong, H. Caughron, A. Simone, B. Leshnower, C. Devireddy, K. Mavromatis, R. Guyton, J. Stewart, V. Thourani, P. C. Block, and V. Babaliaros. Predictors and clinical outcomes of next-day discharge after minimalist transfemoral transcatheter aortic valve replacement. JACC Cardiovasc. Interv. 11:107–115, 2018.CrossRefGoogle Scholar
  18. 18.
    Kheradvar, A., E. M. Groves, L. P. Dasi, S. H. Alavi, R. Tranquillo, K. J. Grande-Allen, C. A. Simmons, B. Griffith, A. Falahatpisheh, C. J. Goergen, M. R. Mofrad, F. Baaijens, S. H. Little, and S. Canic. Emerging trends in heart valve engineering: Part I. Solutions for future. Ann. Biomed. Eng. 43:833–843, 2015.CrossRefGoogle Scholar
  19. 19.
    Khoffi, F., and F. Heim. Mechanical degradation of biological heart valve tissue induced by low diameter crimping: an early assessment. J. Mech. Behav. Biomed. Mater. 44:71–75, 2015.CrossRefGoogle Scholar
  20. 20.
    Luscher, T. F. Cutting edge research on transcatheter aortic valve implantation: moving indications, complications, and current outcomes. Eur. Heart J. 39:633–636, 2018.CrossRefGoogle Scholar
  21. 21.
    Martin, C., and W. Sun. Transcatheter valve underexpansion limits leaflet durability: implications for valve-in-valve procedures. Ann. Biomed. Eng. 45:394–404, 2017.CrossRefGoogle Scholar
  22. 22.
    Marwan, M., N. Mekkhala, M. Goller, J. Rother, D. Bittner, A. Schuhbaeck, M. Hell, G. Muschiol, J. Kolwelter, R. Feyrer, C. Schlundt, S. Achenbach, and M. Arnold. Leaflet thrombosis following transcatheter aortic valve implantation. J. Cardiovasc. Comput. Tomogr. 12:8–13, 2018.CrossRefGoogle Scholar
  23. 23.
    Midha, P. A., V. Raghav, R. Sharma, J. F. Condado, I. U. Okafor, T. Rami, G. Kumar, V. H. Thourani, H. Jilaihawi, V. Babaliaros, R. R. Makkar, and A. P. Yoganathan. The fluid mechanics of transcatheter heart valve leaflet thrombosis in the neosinus. Circulation 136:1598–1609, 2017.CrossRefGoogle Scholar
  24. 24.
    Min, J. K., D. S. Berman, and J. Leipsic. Multimodality Imaging for Transcatheter Aortic Valve Replacement. New York: Springer Science & Business Media, 2013.Google Scholar
  25. 25.
    Pinchuk, L., and Y. Zhou. Crosslinked polyolefins for biomedical applicatios and method of making same. In: USPTO, edited by USPTO. Miami: Innovia LLC, 2009.Google Scholar
  26. 26.
    Prawel, D. A., H. Dean, M. Forleo, N. Lewis, J. Gangwish, K. C. Popat, L. P. Dasi, and S. P. James. Hemocompatibility and Hemodynamics of Novel Hyaluronan-Polyethylene Materials for Flexible Heart Valve Leaflets. Cardiovasc. Eng. Technol. 5:70–81, 2014.CrossRefGoogle Scholar
  27. 27.
    Rahmani, B., S. Tzamtzis, R. Sheridan, M. J. Mullen, J. Yap, A. M. Seifalian, and G. Burriesci. In vitro hydrodynamic assessment of a new transcatheter heart valve concept (the TRISKELE). J. Cardiovasc. Transl. Res. 10:104–115, 2017.CrossRefGoogle Scholar
  28. 28.
    Rodriguez-Gabella, T., P. Voisine, R. Puri, P. Pibarot, and J. Rodes-Cabau. Aortic bioprosthetic valve durability: incidence, mechanisms, predictors, and management of surgical and transcatheter valve degeneration. J. Am. Coll. Cardiol. 70:1013–1028, 2017.CrossRefGoogle Scholar
  29. 29.
    Rosenhek, R., T. Binder, G. Maurer, and H. Baumgartner. Normal values for Doppler echocardiographic assessment of heart valve prostheses. J. Am. Soc. Echocardiogr. 16:1116–1127, 2003.CrossRefGoogle Scholar
  30. 30.
    Rotman, O. M., B. Kovarovic, C. Sadasivan, L. Gruberg, B. B. Lieber, and D. Bluestein. Realistic vascular replicator for TAVR procedures. Cardiovasc. Eng. Technol. 2018.  https://doi.org/10.1007/s13239-018-0356-z.Google Scholar
  31. 31.
    Scherman, J., D. Bezuidenhout, C. Ofoegbu, D. F. Williams, and P. Zilla. Tavi for low to middle income countries. Eur. Heart J. 38:1182–1184, 2017.CrossRefGoogle Scholar
  32. 32.
    Sheriff, J., D. Bluestein, G. Girdhar, and J. Jesty. High-shear stress sensitizes platelets to subsequent low-shear conditions. Ann. Biomed. Eng. 38:1442–1450, 2010.CrossRefGoogle Scholar
  33. 33.
    Sheriff, J., T. E. Claiborne, P. L. Tran, R. Kothadia, S. George, Y. P. Kato, L. Pinchuk, M. J. Slepian, and D. Bluestein. Physical characterization and platelet interactions under shear flows of a novel thermoset polyisobutylene-based co-polymer. ACS Appl. Mater. Interfaces 7:22058–22066, 2015.CrossRefGoogle Scholar
  34. 34.
    Thourani, V. H., S. Kodali, R. R. Makkar, H. C. Herrmann, M. Williams, V. Babaliaros, R. Smalling, S. Lim, S. C. Malaisrie, and S. Kapadia. Transcatheter aortic valve replacement versus surgical valve replacement in intermediate-risk patients: a propensity score analysis. Lancet 387:2218–2225, 2016.CrossRefGoogle Scholar
  35. 35.
    Vahanian, A., H. Baumgartner, J. Bax, E. Butchart, R. Dion, G. Filippatos, F. Flachskampf, R. Hall, B. Iung, J. Kasprzak, P. Nataf, P. Tornos, L. Torracca, A. Wenink, and Task Force on the Management of Valvular Hearth Disease of the European Society of Cardiology and E. S. C. C. F. P. Guidelines. Guidelines on the management of valvular heart disease: the Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology. Eur. Heart J. 28:230–268, 2007.Google Scholar
  36. 36.
    Wang, M., A. P. Furnary, H. F. Li, and G. L. Grunkemeier. Bioprosthetic aortic valve durability: a meta-regression of published studies. Ann. Thorac. Surg. 104:1080–1087, 2017.CrossRefGoogle Scholar
  37. 37.
    Yin, W., Y. Alemu, K. Affeld, J. Jesty, and D. Bluestein. Flow-induced platelet activation in bileaflet and monoleaflet mechanical heart valves. Ann. Biomed. Eng. 32:1058–1066, 2004.CrossRefGoogle Scholar
  38. 38.
    Yousefi, A., D. L. Bark, and L. P. Dasi. Effect of arched leaflets and stent profile on the hemodynamics of tri-leaflet flexible polymeric heart valves. Ann. Biomed. Eng. 45(2):464–475, 2016.CrossRefGoogle Scholar

Copyright information

© Biomedical Engineering Society 2018

Authors and Affiliations

  • Oren M. Rotman
    • 1
  • Brandon Kovarovic
    • 1
  • Wei-Che Chiu
    • 1
  • Matteo Bianchi
    • 1
  • Gil Marom
    • 1
    • 2
  • Marvin J. Slepian
    • 3
  • Danny Bluestein
    • 1
    Email author
  1. 1.Department of Biomedical EngineeringStony Brook UniversityStony BrookUSA
  2. 2.School of Mechanical EngineeringTel Aviv UniversityTel AvivIsrael
  3. 3.Department of Biomedical EngineeringUniversity of ArizonaTucsonUSA

Personalised recommendations