The value of virtual touch tissue imaging quantification in the differential diagnosis between benign and malignant breast lesions

  • Wen-tao Kong
  • Wei-jun Zhou
  • Yin Wang
  • Xiao-min Zhuang
  • Min WuEmail author
Original Article–Breast & Thyroid
  • 7 Downloads

Abstract

Purpose

To evaluate the value and diagnostic performance of virtual touch tissue imaging quantification (VTIQ) and to determine the optimum cut-off value for differential diagnosis between benign and malignant breast lesions.

Methods

Conventional ultrasonography (US) and VTIQ were performed in 454 patients with 466 breast lesions with a Siemens Acuson S3000 ultrasound machine. All lesions were assessed by an ultrasound Breast Imaging Reporting and Data System (BI-RADS) and confirmed by histopathology. The maximum, mean, and minimum shear wave velocity (SWV) values were quantitatively measured in m/s within the regions of interest (ROIs) and ranged from 0.5 to 10 m/s. The sensitivity, specificity, accuracy, and area under the receiver operating curve (AUC) of the VTIQ, BI-RADS, and combined data were compared.

Results

Among the 466 breast lesions, 266 were benign and 200 were malignant. All of the SWV values of the malignant lesions were significantly greater than those of the benign ones (P < 0.05). The optimal cut-off values for SWVmax, SWVmin, SWVmean, and SWVmax/SWVmin obtained from ROC analysis were 5.37 m/s, 3.08 m/s, 4.04 m/s, and 1.83, respectively. Logistic regression analysis revealed that BI-RADS was an independent risk factor for the differential diagnosis of breast lesions, whereas SWV values were not independent risk factors.

Conclusions

VTIQ is useful in the differential diagnosis between benign and malignant breast lesions. The combination of VTIQ and ultrasonic BI-RADS can improve the diagnostic performance.

Keywords

Virtual touch tissue imaging quantification Breast imaging reporting and data system Shear wave velocity Diagnostic performance 

Notes

Funding

This study was funded by the National Natural Science Foundation of China (81671701) and the Nanjing Medical Science and Technique Development Foundation.

Compliance with ethical standards

Conflict of interest

There are no conflicts of interest.

Ethical statement

All procedures followed were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national) and with the Helsinki Declaration of 1964 and later versions. Informed consent was obtained from all patients for being included in the study.

References

  1. 1.
    Ferlay J, Soerjomataram I, Dikshit R, et al. Cancer incidence and mortality worldwide: sources, methods and major patterns in GLOBOCAN 2012. Int J Cancer. 2015;136:E359–86.CrossRefGoogle Scholar
  2. 2.
    Hooley RJ, Scoutt LM, Philpotts LE. Breast ultrasonography: state of the art. Radiology. 2013;268:642–59.CrossRefGoogle Scholar
  3. 3.
    Ricci P, Maggini E, Mancuso E, et al. Clinical application of breast elastography: state of the art. Eur J Radiol. 2014;83:429–37.CrossRefGoogle Scholar
  4. 4.
    Li DD, Guo LH, Xu HX, et al. Acoustic radiation force impulse elastography for differentiation of malignant and benign breast lesions: a meta-analysis. Int J Clin Exp Med. 2015;8:4753–61.Google Scholar
  5. 5.
    Bai M, Du L, Gu J, et al. Virtual touch tissue quantification using acoustic radiation force impulse technology: initial clinical experience with solid breast masses. J Ultrasound Med. 2012;31:289–94.CrossRefGoogle Scholar
  6. 6.
    Xie J, Wu R, Xu HX, et al. Relationship between parameters from virtual touch tissue quantification (VTQ) imaging with clinicopathologic prognostic factors in women with invasive ductal breast cancer. Int J Clin Exp Pathol. 2014;7:6644–62.Google Scholar
  7. 7.
    Kim YS, Park JG, Kim BS, et al. Diagnostic value of elastography using acoustic radiation force impulse imaging and strain ratio for breast tumors. J Breast Cancer. 2014;17:76–82.CrossRefGoogle Scholar
  8. 8.
    Ianculescu V, Ciolovan LM, Dunant A, et al. Added value of Virtual Touch IQ shear wave elastography in the ultrasound assessment of breast lesions. Eur J Radiol. 2014;83:773–7.CrossRefGoogle Scholar
  9. 9.
    Liu H, Zhao LX, Xu G, et al. Diagnostic value of virtual touch tissue imaging quantification for benign and malignant breast lesions with different sizes. Int J Clin Exp Med. 2015;8:13118–26.Google Scholar
  10. 10.
    Golatta M, Schweitzer-Martin M, Harcos A, et al. Evaluation of virtual touch tissue imaging quantification, a new shear wave velocity imaging method, for breast lesion assessment by ultrasound. Biomed Res Int. 2014;2014:960262.CrossRefGoogle Scholar
  11. 11.
    Barr RG, Zhang Z. Effects of precompression on elasticity imaging of the breast: development of a clinically useful semiquantitative method of precompression assessment. J Ultrasound Med. 2012;31:895–902.CrossRefGoogle Scholar
  12. 12.
    Li XL, Xu HX, Bo XW, et al. Value of virtual touch tissue imaging quantification for evaluation of ultrasound breast imaging-reporting and data system category 4 lesions. Ultrasound Med Biol. 2016;42:2050–7.CrossRefGoogle Scholar
  13. 13.
    Tozaki M, Saito M, Benson J, et al. Shear wave velocity measurements for differential diagnosis of solid breast masses: a comparison between virtual touch quantification and virtual touch IQ. Ultrasound Med Biol. 2013;39:2233–45.CrossRefGoogle Scholar
  14. 14.
    Li DD, Xu HX, Guo LH, et al. Combination of two-dimensional shear wave elastography with ultrasound breast imaging reporting and data system in the diagnosis of breast lesions: a new method to increase the diagnostic performance. Eur Radiol. 2016;26:3290–300.CrossRefGoogle Scholar
  15. 15.
    Yao MH, Wu R, Xu G, et al. A novel two-dimensional quantitative shear wave elastography to make differential diagnosis of breast lesions: comprehensive evaluation and influencing factors. Clin Hemorheol Microcirc. 2016;64:223–33.CrossRefGoogle Scholar
  16. 16.
    Giannotti E, Vinnicombe S, Thomson K, et al. Shear-wave elastography and greyscale assessment of palpable probably benign masses: is biopsy always required? Br J Radiol. 2016;89:20150865.CrossRefGoogle Scholar
  17. 17.
    Kapetas P, Pinker-Domenig K, Woitek R, et al. Clinical application of acoustic radiation force impulse imaging with virtual touch IQ in breast ultrasound: diagnostic performance and reproducibility of a new technique. Acta Radiol. 2017;58:140–7.CrossRefGoogle Scholar

Copyright information

© The Japan Society of Ultrasonics in Medicine 2019

Authors and Affiliations

  1. 1.Department of Ultrasound, Drumtower HospitalMedical College of Nanjing UniversityNanjingChina
  2. 2.Department of Thyroid and Breast Surgery, Drumtower HospitalMedical College of Nanjing UniversityNanjingChina

Personalised recommendations