Chromatographia

, Volume 82, Issue 1, pp 65–75 | Cite as

Recent Achievements and Future Challenges in Supercritical Fluid Chromatography for the Enantioselective Separation of Chiral Pharmaceuticals

  • Simona Felletti
  • Omar H. Ismail
  • Chiara De Luca
  • Valentina Costa
  • Francesco Gasparrini
  • Luisa Pasti
  • Nicola Marchetti
  • Alberto Cavazzini
  • Martina CataniEmail author
Review
  • 221 Downloads
Part of the following topical collections:
  1. 50th Anniversary Commemorative Issue

Abstract

During the last years, supercritical fluid chromatography (SFC) has attracted a continuously growing number of users. Thanks to the introduction of state-of-the-art equipment, this technique has allowed to run three-to-five times faster separations than in high-performance liquid chromatography (HPLC) on columns packed with particles of comparable dimension, at lower pressure drops and without loss of efficiency. Thanks to its high versatility, its high-throughput screening capability, and “green” character of the mobile phase, SFC has become particularly attractive for the separation of chiral drugs in pharmaceutical industries. In this review, we will consider the latest applications of SFC for the analysis of compounds of pharmaceutical interest and/or with biological activity essentially covering main achievements of the last 3 years. We also focus on some very recent, remarkable applications of SFC in ultrafast enantioseparations on chiral columns of the latest generation. Technical improvements needed on commercial equipment to increase the competitiveness of SFC towards highly efficient enantioseparations are discussed.

Keywords

Supercritical fluid chromatography (SFC) Chiral chromatography Enantioseparations Pharmaceuticals Ultrafast enantioseparations 

Introduction

Most of the molecules that play a key role in living organisms (such as amino acids, nucleic acids, sugars, pharmaceuticals) are chiral [1, 2]. Since biological interactions are strictly stereospecific, the two enantiomers may exhibit a completely different biological activity. It has been demonstrated that, in many cases, while one enantiomer is effectively active as therapeutic, the other one could be totally inactive or even toxic for the human body or the environment. As a consequence, identification of possible impurities and full characterization of chiral active pharmaceutical ingredients (APIs) are crucial steps for the development of drug substances. For this reason, the availability of high-performance analytical methods is fundamental at any stage of the production of pharmaceuticals or biomedical products, whose commercialization is strictly monitored by specific guidelines recommended by regulatory agencies [3, 4, 5].

Chromatography represents the most powerful technique nowadays in use for the separation of chiral compounds for both analytical and preparative purposes. Even if chiral high- (or ultra-high-) performance liquid chromatography (HPLC/UHPLC) still remains the first choice for the separation of enantiomers, during the last years, the attention of separation scientists has moved towards alternative methods, in particular supercritical fluid chromatography (SFC). This technique is based on the same principles as those of LC and, as a matter of fact, it makes use of the same software, hardware, and very similar instrumentation. The main difference is the replacement of common liquids used as mobile phases in LC with mixtures of high-pressurized carbon dioxide (CO\(_2\)) mixed with another solvent (most often methanol or other alcohols). The use of CO\(_2\) above its critical point leads to several advantages from a chromatographic viewpoint. Thanks to a lower viscosity and higher diffusion coefficients, chromatographic separations in SFC can be carried out at high flow rates without remarkable loss of efficiency and with very limited pressure drop along the column.

The development of SFC as a separation method has been somehow slow and discontinuous. The first report on the use of supercritical fluids in chromatography traces back to the 1960s [6], but, at that time, this approach did not attract much attention within the analytical community, having to compete with a well-established and traditional technique as GC. Twenty years later, neat supercritical fluid CO\(_2\) found use, especially with open tubular columns but also with packed ones [7, 8, 9, 10, 11, 12, 13, 14]. Despite promising kinetic performance, the widespread use of SFC has been limited by different factors. First, due to the GC-like conception of using a single fluid as mobile phase, this technique has been restricted mainly to the analysis of nonpolar compounds (the eluting strength of CO\(_2\) is comparable to that of pentane). Other issues come from instrumental limitations such as the lack of UV sensitivity. During the 90s, SFC has started to be considered a competitive chromatographic method, especially for the purification of chiral compounds. Moreover, the introduction of a co-solvent opened the door for the analysis of polar molecules [15].

However, it is after 2010 that SFC has seriously started attracting a growing number of users not only for preparative applications but also for analytical purposes. In that period, latest generation SFC equipments have been commercialized with different technical improvements that enhanced reproducibility, sensitivity, and reliability of the system. As a matter of fact, in state-of-the-art equipments, extra-column variance is still dramatically larger (70–100 \(\upmu\)L\(^2\)) than in modern UHPLC equipment (1–2 \(\upmu\)L\(^2\)). As it will be shown later on, this is a main issue when using very efficient columns of reduced internal diameter packed with fine particles [16, 17, 18].

In general, the use of supercritical fluid CO\(_2\)/co-solvent mixtures has gained popularity over the last years as a “green” alternative to normal-phase or reversed-phase (NP or RP) LC separations. Another advantage is that, differently from LC, polar and hydrophobic stationary phases can be operated in SFC with the same mobile phase, representing a powerful orthogonal method for different analytical applications, especially for chiral separations, as described in past detailed review papers [14, 15, 19].

This review is not intended to be a comprehensive overview of SFC covering all the aspects of this technique from fundamentals of separation to instrumental aspects. Many detailed works [14, 15, 18, 19, 20, 21, 22] have been recently published to which the interested reader is referred to. Scope of this paper is to provide an overview of the most recent advances in chiral SFC, covering literature of the past 3 years. In particular, we will focus on the last applications reported in the literature for rapid high-throughput separation of chiral pharmaceuticals and on the last achievements in ultrafast (sub-minute) chiral SFC separations. A short section on the use of SFC for preparative applications is also included in the last part of the manuscript.

SFC in a Glance

In SFC, the mobile phase is a supercritical fluid. This is a particular state of matter reached when temperature and pressure are near or above the critical point. For pure CO\(_2\), which is the most common supercritical fluid used in chromatography, these values are \(T_\text {c}\) = 31 \(^{\circ }\)C and \(P_\text {c}\) = 74 bar. Fluids exhibit particular properties in supercritical conditions that are intermediate between those of gases and liquids. In particular, density is similar to that of liquids, viscosity is comparable to that of gases, and diffusivity is midway. For these reasons, SFC is considered an intermediate separation technique between gas chromatography (GC) and HPLC. As a marginal remark, the term SFC is also often extended to applications in which temperature is kept below than the critical one, since there is no phase transition when pressure is maintained above 74 bar.

Neat supercritical fluid CO\(_2\) is a nonpolar solvent, comparable to pentane. However, it is rarely used as single eluent. Organic modifiers, such as methanol or other alcohols, are routinely used as co-solvents. Their introduction not only increases the polarity of the mobile phase, hence the solvating power, but it also affects density of the mobile phase. In addition, several different additives (e.g., trifluoroacetic acid, triethylamine, and ammonium acetate) are often added to the mobile phase, enhancing solvating power or favoring ion-pairing with charged analytes. Also water is used sometimes as a ternary mixture, allowing for the elution of the most polar compounds, such as peptides and amphoteric molecules [21].

Density plays a key role on different chromatographic parameters in SFC. First, molecular interactions, and hence retention, but also viscosity, diffusivity, and mobile phase velocity are strongly influenced by changes in density. Density profile along the column is affected by any change in pressure and temperature. A significant variation in density could have a detrimental effect on column efficiency due to the formation of radial temperature gradients when CO\(_2\) is operated under high compressibility conditions [18]. For the reason above, modern SFC instruments are designed to ensure isothermal or adiabatic conditions of the column and a strict control of pressure. Moderate temperature and high pressure ensure a higher density of fluid, further contributing to maintain the conditions away from the critical point (where the situation is the worst). In addition, the use of co-solvents greatly reduces compressibility.

Recent Applications of SFC for the Analysis of Pharmaceuticals Divided by the Type of Chiral Stationary Phase

Polysaccharide-Based CSPs

The most employed class of CSPs used in SFC is that based on immobilized polysaccharide derivatives [14, 15, 23]. The widespread use of cellulose- and amylose-based CSPs can be ascribed especially to their large applicability and their high loadability. This latter characteristic has represented the main reason for the success of these CSPs for preparative purposes in the past [15, 24].

However, in the last few years, 3 or 5 \(\upmu\)m fully porous particles (FPPs) polysaccharide-based CSPs have been mainly used for analytical purposes. Recent works also report the use of 2.5 \(\upmu\)m polysaccharide FPPs [25, 26, 27]. In the following, the main applications of polysaccharide CSPs in SFC will be revised.

High-Throughput Screenings

In a recent paper, different cellulose- and amylose-based 3 \(\upmu\)m FPPs CSPs have been successfully employed for high-throughput screening of 20 pharmaceuticals (including ketoprofen, ibuprofen, and epinephrine) in only 4 min, followed by 2 min of column equilibration [28]. The mobile phases used were CO\(_2\)/methanol or CO\(_2\)/2-propanol mixtures plus a combined additive of trifluoroacetic acid and diethylamine. The simultaneous presence of both the two organic modifiers was found to be beneficial for the enhancement of enantioselectivity.

Retention mechanism of 13 pairs of enantiomers belonging to the same structural family (phenylthiohydantoin-amino acids) has been studied on two different polysaccharide stationary phases (Chiralpak AD-H, amylose-based and Chiralpak OD-H, and cellulose-based) at five different temperatures ranging from 5 up to 40 \(^\circ\)C [29]. The mobile phase used was CO\(_2\)/methanol 90:10 %v/v. Some structural changes seem to affect both CSPs above 20–30 \(^\circ\)C. Below this limit, it was found that retention is substantially unaffected by temperature changes on the cellulose-based column. On the contrary, remarkable differences in retention factors have been observed on the Chiralpak AD-H by changing temperature. According to the authors of this work, this result suggests the presence of more heterogeneous chiral sites on the amylose-based CSP than on the cellulose one. However, another reason that could explain the observed behavior may be the higher rigidity of cellulose-based CSPs with respect to amylose-based ones.

To enhance the possibility to carry out high-throughput analysis, multiple injections in a single experimental run (MISER) chromatographic technique has been applied in SFC with a 10 \(\times\) 4.0 mm (L \(\times\) I.D.) Chiralpak AD-3 column packed with 3 \(\upmu\)m FPPs. This high-throughput method is based on multiple injections within a single chromatographic run to produce a continuous trace of chromatograms. The separation of Tröger’s base enantiomers in an entire 96-well microplate of samples has been performed in 33 min (Fig. 1) [30]. However, this approach is currently limited by the speed of autosamplers. Faster instrument control softwares are required to extensively apply MISER SFC to rapid enantiopurity screenings of a large number of samples.
Fig. 1

a Separation of Tröger’s base enantiomers on a 10 \(\times\) 4.0 mm (L \(\times\) I.D.) Chiralpak AD-3 column packed with 3 \(\upmu\)m FPPs functionalized with an amylose derivative. b Injection of a 96-well plate with MISER approach on the same column. Tröger’s base concentrations between 0.4 and 1 mg/mL. Reproduced with permission from Ref. [30]

Determination of Enantiopurity of APIs and Their Intermediates

One of the most challenging tasks for pharmaceutical industries is the reduction of the number of separation modes required for the analysis of an API and its intermediates. Due to its versatility, SFC appears one of the most appealing chromatographic methods for this purpose.

Barhate et al. have investigated a large number of chromatographic CSPs by both RPLC and SFC for the determination of enantiomeric excess of verubecestat (employed in clinical trials for the treatment of Alzheimer’s disease) and its intermediates [31]. The best results have been obtained in RPLC using a teicoplanin-based CSP made on 2.7 \(\upmu\)m superficially porous particles (SPPs), but cellulose-based chiral columns produced very promising results in SFC for the determination of enantiopurity of the entire verubecestat synthetic route.

Bu et al. have also recently applied SFC for the analysis of poor UV absorbing drugs and synthetic intermediates with charged aerosol detection (CAD) [32]. Enantiomeric excess has been determined under both gradient and isocratic elution conditions, and compared with results obtained with UV detection. A strong correlation between UV and CAD responses under isocratic conditions was observed, while, under gradient conditions, higher absolute errors between the two measures were registered, due to the fact that high amount of organic modifier enhances CAD response of later eluting compounds. In the same work, the practical use of SFC-CAD has been investigated for high-throughput parallel screening of chemo- and bio-catalytic reactions for the quick identification of desired reaction conditions. Twenty-four different hydrolase enzymes were screened in parallel on a well plate and the two isomeric monoacid products were separated on a Chiralpak AD-3 column in less than 2 h.

Comparison Between Chiral SFC and HPLC

Different studies have been recently conducted to compare retention mechanism of enantiomers on polysaccharide-based CSPs in SFC and different LC conditions.

West et al. have compared retention of 24 chiral sulfoxides on seven different polysaccharide CSPs with CO\(_2\)/methanol mixtures as mobile phase, proving that chlorinated cellulose CSPs are better in terms of both retention and enantioselectivity towards molecules containing a chiral sulfur atom [33]. By means of molecular modeling measurements, the authors of this paper demonstrated that molecules that could adopt a folded U-shaped conformation were most efficiently discriminated compared to linear ones. Moreover, SFC was compared to polar organic mode (POM) HPLC. It was found that the chiral selector must adopt a different conformation in the two operating modes. However, SFC outperformed POM HPLC in terms of enantioresolution.

Recently, retention mechanisms of different pairs of enantiomers on polysaccharide CSPs have been explored in both SFC and NPLC conditions. It has been demonstrated that the transposability of methods from NPLC to SFC can be challenging in some cases, mainly due to different interactions (hydrogen bonding and accessibility of chiral cavities) contributing to retention in the two chromatographic modes [34, 35]. Separations in NPLC resulted in shorter retention times and higher enantioresolution for the separation of dihydropyridine derivatives, especially if bearing two chiral centers [36]. Indeed, scope of this work was the investigation of chromatographic conditions transfer from NPLC to SFC (this explains reported larger retention times in SFC). Despite higher retention times, selectivity was not consistently better in SFC, meaning that the additional retention was due to nonspecific interactions of enantiomers with the CSP.

Vera et al. have compared selectivity of a Lux Cellulose-1 towards retention of FMOC-protected amino acids in SFC, NPLC, and RPLC conditions [37]. In terms of retention, SFC lies in the middle between RPLC and NPLC. Although RPLC gave comprehensively the best enantioresolution, the introduction of 2% formic acid as additive in CO\(_2\)/methanol mixture used in SFC provided comparable results in shorter run time, allowing for a better resolution per unit of time.

Chiral SFC–MS Methods

Even if RPLC–MS is still considered the gold standard for the analysis of serum, urine, and plasma samples, in the last years, SFC has been efficiently hyphenated with mass spectrometry giving very promising results.

A reliable SFC–MS/MS method for the separation of amphetamine enantiomers in biological samples has been developed and validated for the first time by Hegstad et al in Ref. [38]. R- and S-amphetamine enantiomers were baseline resolved using a Chiralpak AD-3 column and CO\(_2\)/2-propanol+0.2% cyclohexylamine as mobile phase. This method has been routinely used for the analysis of several human urine samples representing a reliable tool to discriminate between legal use of amphetamine as therapeutic (in most countries, only the S-enantiomer is prescribed) and illegal use (as racemic mixture).

Jenkinson et al. have recently developed a new SFC–MS/MS method for the analysis of metabolites of vitamin D in human serum [39]. The separation has been achieved in 6 min on a Lux Cellulose column using CO\(_2\)/methanol+0.1% formic acid as eluent. Concentrations of metabolites measured on 41 routine human serum samples have been found to be in accordance with those measured by means of UHPLC-MS/MS. In addition, structurally similar metabolites, differing only for the position or direction of an hydroxyl group, have been resolved and quantified by means of the optimized SFC–MS/MS method.

A 3 \(\upmu\)m FPPs Chiralpak IA was efficiently used to separate panthenol enantiomers in cosmetic formulations (such as creams, body lotions, and exfoliants) [40]. Since only the D-enantiomer of panthenol is active as therapeutic, reliable methods to assess enantiopurity of formulated cosmetics are required. The column was employed in SFC conditions (CO\(_2\) with 11% methanol as mobile phase) with both UV and MS detection. The online coupling improved sensitivity (LOQ as low as 0.5 \(\upmu\)g/mL), since underivatized panthenol has a poor signal in UV.

Multidimensional Chromatography

Chiral SFC has also been applied as the second dimension in highly selective multidimensional chromatography approaches to assess purity of chiral APIs. A first (achiral) RPLC dimension is needed to assess the amount of impurities and related substances, while the second (chiral) dimension is used to evaluate the possible presence of undesired enantiomers [41]. Such a system has been applied for the quantitative analysis of an API, its metabolites, and their corresponding enantiomers in a mouse hepatocyte treated sample using Chiralpak IB-3 and AD-3 columns (Fig. 2) [42].
Fig. 2

2D LC-SFC analysis of an API and its metabolite. First dimension: achiral RPLC (top); second dimension: chiral SFC (bottom). Reproduced with permission from Ref. [42]

Other Applications

Immobilized amylose-based CSPs have been proved to be useful for the separation of basic biologically active compounds, whose separation has always been challenging in SFC [43]. The authors of this work investigated the effect of co-solvent, temperature, and backpressure on the enantioseparation of 27 different compounds bearing a basic moiety including amphetamine, cathinone, benzofury, and amino-naphthol derivatives. It was found that adding bases (or the mixture of base and acid) to the mobile phase has a beneficial effect on peak shape and enantioresolution.

Polysaccharide-based CSPs have been also successfully used for the separation of chiral compounds with biological activity such as pesticides containing sulfur or phosphorous atoms [44], fungicides (i.e., fenbuconazole) in foods [45], and herbicides (i.e., napropramide) [46].

Pirkle-type CSPs

Pirkle-type (or brush-type) chiral selectors are among the most versatile, allowing for the separation of a broad range of compounds. They represent the first class of CSPs that has been prepared on sub-2\(\upmu\)m format [47, 48, 49]. One of the main advantages of these CSPs is that they exist in both the enantiomeric versions. Using columns functionalized with the same chiral selector but with opposite configuration, it is possible to reverse the elution order of enantiomers (the so-called “Inverted Chirality Columns Approach”, ICCA). This method has been recently applied by Mazzoccanti et al. to determine the enantiomeric excess of phytocannabinoids in marijuana samples for therapeutic use [50]. Indeed, many problems can be faced when working with Cannabis plant extracts. Not only are they highly complex mixtures but also the minor enantiomer is not always available as reference sample. Moreover, in many cases, it can partially coelute with the main enantiomer. Two complementary—(S,S) and (R,R)—Whelk-O1 CSPs made on sub-2\(\upmu\)m FPPs have been employed in SFC conditions (CO\(_2\)/methanol, 98:2 %v/v) to determine the enantiomeric excess of (-)-\(\Delta ^9\)-THC in medicinal marijuana (Bedrocan®). Thanks to ICCA protocol, besides the major enantiomer ((−)-4 peak in Fig. 3), a not negligible concentration (0.13%) of the (+)-enantiomer ((+)-4 peak in Fig. 3) has been detected. The enantiomeric excess was estimated to be 99.73%.
Fig. 3

Chromatograms of a Bedrocan® ethanol extract analyzed by applying the ICCA protocol. A zoom of the chromatogram between 2 and 6 min is shown in the inset together with the separation of a standard mixture of six-component cannabinoids (dotted chromatogram). The asterisk denotes a chiral unknown impurity. Reproduced from Ref. [50] with permission from The Royal Society of Chemistry

The same sub-2\(\upmu\)m FPP (S,S)-Whelk-O1 CSP has been used for the high-throughput screening of a large library of 129 pharmaceutical compounds with different chemico-physical properties including \(\beta\)-blockers, antidepressants, anticancers, and benzodiazepines, to name but a few [51]. The overall screening was completed in 24 h under fast gradient elution (9 min total analysis time, including column reconditioning) using a mixture of CO\(_2\)/methanol as mobile phase with a success rate of 63%. Even basic racemic mixtures, whose separation is traditionally challenging on Whelk-O1 CSPs prepared on larger particles, have been resolved with the sub-2\(\upmu\)m CSP used in this work.

Macrocyclic Glycopeptide CSPs

Macrocyclic glycopeptide CSPs allow for the separation of underivatized amino acids. This class of CSP has been rarely used in SFC to date.

Recently, however, a 5 \(\upmu\)m teicoplanin CSP (Chirobiotic T2) has been efficiently employed for the separation of D,L-enantiomers of underivatized phenylalanine (Phe), tyrosine (Tyr), and tryptophan (Trp) amino acids. Baseline separations have been obtained in less than 7 min using CO\(_2\) and 40% of organic modifier made of a mixture of methanol/water (90:10 %v/v) [52]. LOD in the range of 0.5–2.0 \(\upmu\)g/mL allowed for the determination of D-enantiomers up to 0.2%. This method has been applied for the determination of enantiopurity of five commercial food supplements confirming the absence of impurities in all of them. The authors have also investigated the possibility to simultaneously determine D,L-Phe and D,L-Tyr by coupling a diol achiral column (first dimension) with the Chirobiotic T2 (second dimension). The separation was obtained in about 15 min.

Cyclofructan-based CSPs

Recently developed cyclofructan-based 2.7 \(\upmu\)m SPP CSPs have been employed by Armstrong and coworkers to investigate the transposability of chromatographic methods from NPLC to SFC for the enantioseparation of 21 \(\alpha\)-aryl ketones [53]. The mobile phase used in NPLC was a mixture of heptanol/ethanol with percentages ranging from 95:5 to 99:1 %v/v. The same compositions have been transposed to SFC by replacing heptanol with CO\(_2\). 17 of the 21 compounds have been baseline separated in NP conditions, while 10 out to 21 compounds via SFC. Even if the latter allowed for lower analysis time, HPLC provided better resolutions due to greater enantioselectivity values.

Ion-Exchange CSPs

The use of ion-exchange CSPs in SFC is very recent. Lajkó et al. have first used Cinchona alkaloid-based ZWIX(+) and ZWIX(-) CSPs for the enantioseparation of N\(_{\alpha }\)-Fmoc proteinogenic amino acids [54]. The effect of methanol content in the mobile phase and different additives (water, acids, and bases) has been investigated to optimize separation conditions. The addition of water led to the formation of carbonic acid, imparting acidic character to the mobile phase. It was also found that a reduction of temperature has a beneficial effect on enantioresolution, meaning that chiral recognition mechanism is enthalpically controlled.

The chromatographic behavior of the ZWIX(+) column in both HPLC and SFC conditions was compared for the separation of acidic, basic, and zwitterionic species [55]. In general, SFC provided a better enantioresolution than HPLC, but there is evidence that separation mechanism is completely different. By constantly increasing the amount of organic modifier (methanol) in SFC, it was found that the ion-exchange mechanism is strongly influenced by the formation of transient acidic species (carbonic acid mono methyl ester). Finally, it was proved that basic additives are not strictly necessary when using the zwitterionic column, and they could have an effect only on basic analytes.

Towards High-Efficiency Ultrafast SFC Enantioseparations

The recent achievements in particle manufacturing have allowed to prepare very efficient particle formats, such as SPPs or sub-2\(\upmu\)m FPPs, functionalized with chiral selectors. The introduction of these new CSPs packed into columns of very short length has represented a real breakthrough in the field of UHPLC enantioseparations, not only in terms of efficiency (comparable to those of achiral RPLC) but also in terms of speed of separation (analysis time < 1 s) [31, 48, 51, 56, 57, 58, 59, 60, 61, 62, 63, 64, 65, 66, 67].

Since SFC allows to run chromatographic separations at higher flow rates than LC without remarkable loss of efficiency [68], the use of new generation CSPs under these conditions seems to be a promising approach to achieve even faster separations.

Some of the authors of this work have recently obtained very fast enantioseparations using both Teicoplanin and Whelk-O1 CSPs under SFC conditions [51, 59].

In the first case, Teicoplanin was bonded to 1.9 \(\upmu\)m FPPs and packed into a 20 \(\times\) 4.6 mm (L \(\times\) I.D) column operated at 4 mL/min. The enantiomers of Ketorolac have been resolved in less than 70 s on a chiral selector which has been considered a “slow” one [59].

In the second case, using a 50 \(\times\) 4.6 mm (L \(\times\) I.D) column packed with 1.8 \(\upmu\)m FPPs functionalized with Whelk-O1, the separation of abscisic acid enantiomers has been obtained in less than 45 s (flow rate 3.5 mL/min) with a resolution (R\(_s\)) of 2.2 (see Fig. 4a) [51].
Fig. 4

Examples of ultrafast enantioseparations obtained in SFC. a Abscisic acid enantiomers. Column: 50 \(\times\) 4.6 mm (L \(\times\) I.D), Whelk-O1 1.8 \(\upmu\)m FPPs. Flow rate: 3.5 mL/min. Instrument: Waters Acquity UPC\(^2\). Modified with permission from Ref. [51]. b FMOC leucine enantiomers. Column: 30 × 4.6 (L \(\times\) I.D), quinine-based 2.7 \(\upmu\)m SPPs. Flow rate: 20 mL/min. Instrument: Jasco SFC-2000-7. Modified with permission from Ref. [70]. c Warfarin enantiomers. Column: 50 \(\times\) 3 mm (L \(\times\) ID), amylose-based 1.6 \(\upmu\)m FPPs. Flow rate: 3.75 mL/min. Instrument: low-dispersion-modified Agilent 1260 Infinity SFC. Modified with permission from Ref. [72]

Very remarkable results in terms of ultrafast enantioseparations have been reported by Armstrong and coworkers. Sub-minute separations of different pairs of enantiomers of pharmaceutical interest have been obtained on teicoplanin and teicoplanin aglycone CSPs made on 1.9 \(\upmu\)m FPPs packed into 50 \(\times\) 4.6 mm (L \(\times\) ID) columns at a flow rate of 7 mL/min [69]. Moreover, using a 30 \(\times\) 4.6 (L \(\times\) I.D) column packed with 2.7 \(\upmu\)m SPPs functionalized with a quinine derivative, they were able to obtain the separation of different amino acids in 6–8 s with a flow rate of 20 mL/min [70]. An example is shown in Fig. 4b.

However, the use of new generation CSPs in SFC is often partially limited by some instrumental issues. The excessively large extra-column band broadening of current SFC instruments has a detrimental effect on the overall chromatographic performance.

Berger has recently modified a commercial 1260 Infinity SFC from Agilent Technology, by replacing standard tubing (170 \(\upmu\)m ID) and flow cell (13 \(\upmu\)L internal volume) with 120 \(\upmu\)m ID tubing (of shortest possible length) and a 2 \(\upmu\)L internal volume cell [71]. The extra-column dispersion was reduced to about 6–9 \(\upmu\)L\(^2\). With the new configuration, he was able to achieve more than 280,000 plates/m (reduced HETP of 1.93) by employing a prototype 50 \(\times\) 4.6 mm (L \(\times\) ID) column packed with 1.8 \(\upmu\)m Whelk-O1 FPPs. In addition, he reported about the ultrafast separations of 5-methyl 5-phenyl hydantoin enantiomers in roughly 10 s (flow rate 5 mL/min).

Using the same instrumental setup, he has been the first to operate a sub-2\(\upmu\)m-immobilized polysaccharide CSP in SFC conditions [72]. Using a 50 \(\times\) 3 mm (L \(\times\) ID) column packed with an amylose-based CSPs made on 1.6 \(\upmu\)m FPPs, he was able to obtain the ultrafast separation of warfarin enantiomers in less than 10 s (flow rate 3.75 mL/min) with a resolution of 1.5 (see Fig. 4c).

Recently, some of the authors of this work have modified a commercial Waters Acquity UPC\(^2\) SFC instrument by a series of technical adjustments including the replacement of (i) standard tubings with shorter and narrower capillaries; (ii) the 8 \(\upmu\)L flow cell with a 3 \(\upmu\)L one; (iii) the injection system with a 200 nL fixed-loop external one; (iv), finally, using an ad hoc designed external column oven [73]. The extra-column variance was reduced from about 85 \(\upmu\)L\(^2\) (original configuration) to slightly more than 2 \(\upmu\)L\(^2\) (optimized configuration) measured at 2.0 mL/min. Kinetic performance of a 50 \(\times\) 4.6 mm (L \(\times\) ID) column packed with 1.8 \(\upmu\)m Whelk-O1 FPPs operated on the optimized SFC instrument has been compared with that obtained on a commercial UHPLC instrument (Waters Acquity I-Class) with extra-column variance of 1 \(\upmu\)L\(^2\). At the minimum of the van Deemter curve, SFC provided a gain of 10% on the efficiency of the second enantiomer (285,000 N/m vs. 260,000 N/m recorded in UHPLC) in roughly 50% shorter analysis time. The expression ultra-high-performance SFC (UHPSFC) can be properly used under these conditions.

In addition, Barhate et al. have demonstrated that, when running ultrafast SFC separations, some unexpected results could be observed [74]. These deviations, not detected in LC, are mostly due to the noise generated by backpressure regulators and the presence of low viscosity eluent inside connection tubings. The latter is responsible for the development of possible turbulent flow inside tubings which could change both retention time and peak shape.

Preparative SFC for the Purification of Chiral Pharmaceuticals

Preparative SFC is routinely used for the purification of chiral drugs in pharmaceutical industries [14]. This technique offers several advantages over preparative LC such as higher productivity, thanks to the possibility of using higher flow rates, lower organic solvent consumption, reduced impact on the environment, and faster solvent removal. In terms of stationary phases, polysaccharide-based ones are the most used for preparative applications owing to their high loadability [75]. However, scale-up from analytical to preparative conditions is more complex in SFC than in LC due to the high compressibility of the mobile phase. Indeed, this may cause possible variations in density, pressure, and temperature that possibly modify the adsorption process on the stationary phase. Besides these variables, also the content of organic modifier is an important parameter that needs to be taken into account. Most of the time, optimization procedures are based on the variation of one of these parameters, while all the others are kept constant. For this reason, ultrafast enantioseparation methods are increasingly required during screening processes. However, different parameters could present interaction effects. To face this problem, chemometric approaches, such as design of experiment (DOE), are increasingly used to find the optimal experimental conditions for purification purposes, taking into account the simultaneous effect of different variables [76, 77, 78].

Perspectives

Due to the unique properties of supercritical fluid CO\(_2\), SFC can be considered not only a “greener” alternative to HPLC but also an orthogonal and, in some cases, more versatile method of separation. This is particularly important for high-throughput screenings at the beginning of the production of new drugs, when the number of unknown impurities could be relevant.

One of the fields in which SFC will be increasingly used is in multidimensional applications, especially RPLC \(\times\) SFC achiral–chiral separations. However, particular attention has to be put on the interface between the first RPLC dimension and the SFC one, especially to avoid the injection of large volumes of water. Different approaches have been already proposed. Particularly interesting is the use of collection loops [79] or active modulators [80] that seem to be able to solve some of the issues encountered in this coupling [81].

Thanks to the introduction of latest generation CSPs made on sub-3\(\upmu\)m SPPs and sub-2\(\upmu\)m FPPs packed into short columns (2–5 cm), the first examples of enantioseparations in less than one minute or even in the order of seconds have been obtained also in SFC. This is a very promising field in which SFC could be expected to emerge as a gold technique. However, as demonstrated in recent works [71, 72, 73], some technical optimizations aimed at the reduction of extra-column band broadening are needed on commercial equipments. This can be obtained not only by replacing standard tubings with small capillaries but also using low-dispersion ovens and flow cells in the order of nanoliters. These improvements will increase competitiveness of UHPSFC towards UHPLC.

From a fundamental point of view, the investigation of mass transfer phenomena in SFC is necessary to understand how diffusion coefficients possibly change with pressure and temperature and their effect on column efficiency. Moreover, due to the lower mobile phase viscosity, turbulent flow effects have been clearly demonstrated through capillaries connecting the injector system to the column and the column to the detector. From an experimental point of view, this is accompanied by a nonlinear dependence of system backpressure on the flow rate (contrary to what happens when the Darcy’s law is applicable). When turbulence is developing, increasingly growing inertial effects become dominant and the relationship between flow and pressure is not linear any longer [82]. The main consequence of turbulence is the improvement in mass transfer [83] even if, on the other hand, experimental findings show that, through a packed bed, turbulence is much more difficult to develop (at least at the flow rates commonly employed in SFC). These findings could be the basis to renew the interest in chiral open tubular columns for SFC applications, since, through them, maintaining of turbulent regime should be possible. These concepts were proposed more than 50 years ago in the fundamental work of J. C. Giddings [83] when, however, technology was not advanced enough to permit their practical realization. The use of open tubular chiral columns on low-dispersion SFC equipments could lead to unmatched kinetic performance in chromatography.

Notes

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

References

  1. 1.
    Nikolai LN, McClure EL, Macleod SL, Wong CS (2006) Stereoisomer quantification of the beta-blocker drugs atenolol, metoprolol, and propranolol in wastewaters by chiral high-performance liquid chromatography-tandem mass spectrometry. J Chromatogr A 1131:103–109CrossRefGoogle Scholar
  2. 2.
    Sanganyado E, Lu Z, Fu Q, Schlenk D, Gan J (2017) Chiral pharmaceuticals: a review on their environmental occurrence and fate processes. Water Res 124:527–542CrossRefGoogle Scholar
  3. 3.
    Liu JT, Liu RH (2002) Enantiomeric composition of abused amine drugs: chromatographic methods of analysis and data interpretation. J Biochem Biophys Methods 54:115–146CrossRefGoogle Scholar
  4. 4.
    Nguyen LA, He H, Pham-Huy C (2006) Chiral drugs: an overview. J Biochem Biophys Methods 2:85–100Google Scholar
  5. 5.
    International conference on harmonization, IQ32A (R2): Impurities in new drug substances (2006)Google Scholar
  6. 6.
    Klesper E, Corwin AH, Turner DA (1962) High pressure gas chromatography above critical temperatures. J Org Chem 27:700–701CrossRefGoogle Scholar
  7. 7.
    Novotny M, Springston SR, Paeden PA, Fjeldsted JC, Lee ML (1981) Capillary supercritical fluid chromatography. Anal Chem 53:407A–411ACrossRefGoogle Scholar
  8. 8.
    Paeden PA, Fjeldsted JC, Springston SR, Lee ML, Novotny M (1982) Instrumental aspects of capillary supercritical fluid chromatography. Anal Chem 54:1090–1093CrossRefGoogle Scholar
  9. 9.
    Springston SR, Novotny M (1981) Kinetic optimization of capillary supercritical fluid chromatography using carbon dioxide as the mobile phase. Chromatographia 14:679–684CrossRefGoogle Scholar
  10. 10.
    Paeden PA, Lee ML (1983) Theoretical treatment of resolving power in open tubular column supercritical fluid chromatography. J Chromatogr 259:1–16CrossRefGoogle Scholar
  11. 11.
    Gere DR, Board R, McManigill D (1982) Supercritical fluid chromatography with small particle diameter packed columns. Anal Chem 54:736–740CrossRefGoogle Scholar
  12. 12.
    Crowther JB, Henion JD (1985) Supercritical fluid chromatography of polar drugs using small-particle packed columns with mass spectrometric detection. Anal Chem 57:2711–2716CrossRefGoogle Scholar
  13. 13.
    Schwarz HE (1987) LC-GC 5:14–22 Kindly provide article title for reference [13].Google Scholar
  14. 14.
    Desfontaine V, Guillarme D, Francotte E, Nováková L (2015) Supercritical fluid chromatography in pharmaceutical analysis. J Pharm Biomed Anal 113:56–71CrossRefGoogle Scholar
  15. 15.
    Lemasson E, Bertin S, West C (2016) Use and practice of achiral and chiral supercritical fluid chromatography in pharmaceutical analysis and purification. J Sep Sci 39:212–233CrossRefGoogle Scholar
  16. 16.
    Perrenoud AGG, Veuthey JL, Guillarme D (2012) Comparison of ultra-high performance supercritical fluid chromatography and ultra-high performance liquid chromatography for the analysis of pharmaceutical compounds. J Chromatogr A 1266:158–167CrossRefGoogle Scholar
  17. 17.
    Berger TA, Berger BK (2013) Separation of natural food pigments in saponified and un-saponified paprika oleoresin by ultra high performance supercritical fluid chromatography (UHPSFC). Chromatographia 76:591–601CrossRefGoogle Scholar
  18. 18.
    Poole, C.F. (ed.) (2017) Supercritical fluid chromatography, 1st Edition. Elsevier, New YorkGoogle Scholar
  19. 19.
    Kalíková K, Šlechtová T, Vozka J, Tesařová E (2014) Supercritical fluid chromatography as a tool for enantioselective separation; a review. Anal Chim Acta 821:1–33CrossRefGoogle Scholar
  20. 20.
    Lesellier E, West C (2015) The many faces of packed column supercritical fluid chromatography—a critical review. J Chromatogr A 1382:2–46CrossRefGoogle Scholar
  21. 21.
    Tarafder A (2016) Metamorphosis of supercritical fluid chromatography to sfc: an overview. TrAC 81:3–10Google Scholar
  22. 22.
    Vozka J, Kalíková K, Roussel C, Armstrong DW, Tesařová E (2013) An insight into the use of dimethylphenyl carbamate cyclofructan 7 chiral stationary phase in supercritical fluid chromatography: the basic comparison with HPLC. J Sep Sci. 81:1711–1719CrossRefGoogle Scholar
  23. 23.
    Mangelings D, Heyden YV (2008) Chiral separations in sub- and supercritical fluid chromatography. J Sep Sci 31:1252–1273CrossRefGoogle Scholar
  24. 24.
    Cavazzini A, Kaczmarski K, Szabelski P, Zhou D, Liu X, Guiochon G (2002) Modeling of the separation of the enantiomers of 1-phenyl-1-propanol on cellulose tribenzoate. J Chromatogr A 73:5705–5715Google Scholar
  25. 25.
    Li L (2016) Direct enantiomer determination of methorphan by HPLC-MS and SFC-MS. Forensic Chem 2:82–85CrossRefGoogle Scholar
  26. 26.
    Carnes S, O’Brien S, Szewczak A, Tremeau-Cayel L, Rowe WF, McCord B, Lurie IS (2017) Comparison of ultra high performance supercritical fluid chromatography, ultra high performance liquid chromatography, and gas chromatography for the separation of synthetic cathinones. J Sep Sci 40:3545–3556CrossRefGoogle Scholar
  27. 27.
    Hegstad S, Havnen H, Helland A, Falch BMH, Spigset O (2017) Enantiomeric separation and quantification of citalopram in serum by ultra-high performance supercritical fluid chromatography-tandem mass spectrometry. J Chromatogr B 1061–1062:103–109CrossRefGoogle Scholar
  28. 28.
    Nováková L, Douša M (2017) General screening and optimization strategy for fast chiral separations in modern supercritical fluid chromatography. Anal Chim Acta 950:199–210CrossRefGoogle Scholar
  29. 29.
    Khater S, Canault B, Azzimani T, Bonnet P, West C (2018) Thermodynamic enantioseparation behavior of phenylthiohydantoin-amino acid derivatives in supercritical fluid chromatography on polysaccharide chiral stationary phases. J Sep Sci 41:1450–1459CrossRefGoogle Scholar
  30. 30.
    Kawatzki K, Biba M, Regalado EL, Welch CJ (2016) Miser chiral supercritical fluid chromatography for high throughput analysis of enantiopurity. J. Chromatogr. A 1429:374–379CrossRefGoogle Scholar
  31. 31.
    Barhate CL, Lopez DA, Makarov AA, Bu X, Morris WJ, Lekhal A, Hartman R, Armstrong DW, Regalado EL (2018) Macrocyclic glycopeptide chiral selectors bonded to core-shell particles enables enantiopurity analysis of the entire verubecestat synthetic route. J Chromatogr A 1539:87–92CrossRefGoogle Scholar
  32. 32.
    Bu X, Regalado EL, Cuff J, Schafer W, Gong X (2016) Chiral analysis of poor uv absorbing pharmaceuticals by supercritical fluid chromatography-charged aerosol detection. J Supercrit Fluids 116:20–25CrossRefGoogle Scholar
  33. 33.
    West C, Konjaria ML, Shashviashvili N, Lemasson E, Bonnet P, Kakava R, Volonterio A, Chankvetadze B (2017) Enantioseparation of novel chiral sulfoxides on chlorinated polysaccharide stationary phases in supercritical fluid chromatography. J Chromatogr A 1499:174–182CrossRefGoogle Scholar
  34. 34.
    Khater S, Lozac’h MA, Adam I, Francotte E, West C (2016) Comparison of liquid and supercritical fluid chromatography mobile phases for enantioselective separations on polysaccharide stationary phases. J Chromatogr A 1467:463–472CrossRefGoogle Scholar
  35. 35.
    Kalíková K, Martínková M, Schmidt MG, Tesařová E (2018) Cellulose tris-(3,5-dimethylphenylcarbamate)-based chiral stationary phase for the enantioseparation of drugs in supercritical fluid chromatography: comparison with HPLC. J Sep Sci 41:1471–1478CrossRefGoogle Scholar
  36. 36.
    Hoguet V, Charton J, Hecquet PE, Lakhmi C, Lipka E (2018) Supercritical fluid chromatography versus high performance liquid chromatography for enantiomeric and diastereoisomeric separations on coated polysaccharides-based stationary phases: Application to dihydropyridone derivatives. J Chromatogr A 1549:39–50CrossRefGoogle Scholar
  37. 37.
    Vera CM, Shock D, Dennis GR, Farrell W, Shalliker RA (2017) Comparing the selectivity and chiral separation of D- and L-fluorenylmethyloxycarbonyl chloride protected amino acids in analytical high performance liquid chromatography and supercritical fluid chromatography; evaluating throughput, economic and environmental impact. J Chromatogr A 1493:10–18CrossRefGoogle Scholar
  38. 38.
    Hegstad S, Havnen H, Helland A, Spigset O, Frost J (2018) Enantiomeric separation and quantification of R/S-amphetamine in urine by ultra-high performance supercritical fluid chromatography tandem mass spectrometry. J Chromatogr B 1077–1078:7–12CrossRefGoogle Scholar
  39. 39.
    Jenkinson C, Taylor A, Storbeck KH, Hewison M (2018) Analysis of multiple vitamin d metabolites by ultra-performance supercritical fluid chromatography-tandem mass spectrometry (UPSFC-MS/MS). J Chromatogr B 1087–1088:43–48CrossRefGoogle Scholar
  40. 40.
    Khater S, West C (2015) Development and validation of a supercritical fluid chromatography method for the direct determination of enantiomeric purity of provitamin B5 in cosmetic formulations with mass spectrometric detection. J Pharm Biomed Anal 102:321–325CrossRefGoogle Scholar
  41. 41.
    Venkatramani CJ, Al-Sayah M, Li G, Goel M, Girotti J, Zang L, Wigman L, Yehl P, Chetwyn N (2016) Simultaneous achiral-chiral analysis of pharmaceutical compounds using two-dimensional reversed phase liquid chromatography-supercritical fluid chromatography. Talanta 148:548–555CrossRefGoogle Scholar
  42. 42.
    Goel M, Larson E, Venkatramani CJ, Al-Sayah M (2018) Optimization of a two-dimensional liquid chromatography-supercritical fluid chromatography-mass spectrometry (2D-LC-SFS-MS) system to assess in-vivo inter-conversion of chiral drug molecules. J Chromatogr B 1084:89–95CrossRefGoogle Scholar
  43. 43.
    Geryk R, Kalíková K, Schmid MG, Tesařova E (2016) Enantioselective separation of biologically active basic compounds in ultra-performance supercritical fluid chromatography. Anal Chim Acta 932:98–105CrossRefGoogle Scholar
  44. 44.
    Zhang L, Miao Y, Lin C (2018) Enantiomeric separation of six chiral pesticides that contain chiral sulfur/phosphorus atoms by supercritical fluid chromatography. J Sci Sep 41:1460–1470CrossRefGoogle Scholar
  45. 45.
    Tao Y, Zheng Z, Yu Y, Xu J, Liu X, Wu X, Dong F, Zheng Y (2018) Supercritical fluid chromatographytandem mass spectrometry-assisted methodology for rapid enantiomeric analysis of fenbuconazole and its chiral metabolites in fruits, vegetables, cereals, and soil. Food Chem 241:32–39CrossRefGoogle Scholar
  46. 46.
    Zhao L, Xie J, Guo F, Liu K (2018) Enantioseparation of napropamide by supercritical fluid chromatography: effects of the chromatographic conditions and separation mechanism. Chirality 30:661–669CrossRefGoogle Scholar
  47. 47.
    Cavazzini A, Marchetti N, Guzzinati R, Pierini M, Ciogli A, Kotoni D, D’Acquarica I, Villani C, Gasparrini F (2014) Enantioseparation by ultra-high-performance liquid chromatography. Trends Anal Chem 63:95–103CrossRefGoogle Scholar
  48. 48.
    Cancelliere G, Ciogli A, D’Acquarica I, Gasparrini F, Kocergin J, Misiti D, Pierini M, Ritchie H, Simone P, Villani C (2010) Transition from enantioselective high performance to ultra-high performance liquid chromatography: a case study of a brush-type chiral stationary phase based on sub-5-micron to sub-2-micron silica particles. J Chromatogr A 1217:990–999CrossRefGoogle Scholar
  49. 49.
    Catani M, Ismail OH, Felletti S, Gasparrini F, Pasti L, Costa V, Cavazzini A (2017) Pirkle-type chiral stationary phases for ultra-high performance ultra-fast enantioseparations. Am Pharmaceut Rev 20Google Scholar
  50. 50.
    Mazzoccanti G, Ismail OH, D’Acquarica I, Villani CM, Wilcox CM, Cavazzini A, Gasparrini F (2017) Cannabis through the looking glass: chemo- and enantio-selective separation of phytocannabinoids by enantioselective ultra high performance supercritical fluid chromatography. Chem Commun 53:12262–12265CrossRefGoogle Scholar
  51. 51.
    Sciascera L, Ismail O, Ciogli A, Kotoni D, Cavazzini A, Botta L, Szczerba T, Kocergin J, Villani C, Gasparrini F (2015) Expanding the potential of chiral chromatography for high-throughput screening of large compound libraries by means of sub-2 \(\upmu\)m Whelk-O1 stationary phase in supercritical fluid conditions. J Chromatogr A 1383:160–168CrossRefGoogle Scholar
  52. 52.
    Sánchez-Hernández L, Bernal JL, del Nozal MJ, Toribio L (2016) Chiral analysis of aromatic amino acids in food supplements usingsubcritical fluid chromatography and chirobiotic T2 column. J Supercrit Fluids 107:519–525CrossRefGoogle Scholar
  53. 53.
    Breitbach AS, Lim Y, Xu QL, Kürti YL, Armstrong DW, Breitbach ZS (2016) Enantiomeric separations of \(\alpha\)-aryl ketones with cyclofructan chiral stationary phases via high performance liquid chromatography and supercritical fluid chromatography. J Chromatogr A 1427:45–54CrossRefGoogle Scholar
  54. 54.
    Lajkó G, Ilisz I, Tóth G, Fülöp F, Lindner W, Péter A (2015) Application of Cinchona alkaloid-based zwitterionic chiral stationaryphases in supercritical fluid chromatography for theenantioseparation of n\(_{\alpha }\)-protected proteinogenic amino acids. J Chromatogr A 1415:134–145CrossRefGoogle Scholar
  55. 55.
    Wolrab D, Frühauf P, Gerner C, Kohout M, Lindner W (2017) Consequences of transition from liquid chromatography to supercritical fluid chromatography on the overall performance of a chiral zwitterionic ion-exchanger. J Chromatogr A 1517:165–175CrossRefGoogle Scholar
  56. 56.
    Rocchi S, Fanali C, Fanali S (2015) Use of a novel sub-2 \(\upmu\)m silica hydride vancomycin stationary phase in nano-liquid chromatography. ii. Separation of derivatized amino acid enantiomers. Chirality 27:767–772CrossRefGoogle Scholar
  57. 57.
    Barhate CL, Breitbach ZS, Costa Pinto E, Regalado EL, Welch CJ, Armstrong DW (2015) Ultrafast separation of fluorinated and desfluorinated pharmaceuticals using highly efficient and selective chiral selectors bonded to superficially porous particles. J Chromatogr A 1426:241–247CrossRefGoogle Scholar
  58. 58.
    Kotoni D, Ciogli A, D’Acquarica I, Kocergin J, Szczerba T, Ritchie H, Villani C, Gasparrini F (2012) Enantioselective ultra-high and high performance liquid chromatography: a comparative study of columns based on the Whelk-O1 selector. J Chromatogr A 1269:226–241CrossRefGoogle Scholar
  59. 59.
    Ismail OH, Ciogli A, Villani C, Martino MD, Pierini M, Cavazzini A, Bell DS, Gasparrini F (2016) Ultra-fast high-efficiency enantioseparations by means of a teicoplanin-based chiral stationary phase made on sub-2 \(\upmu\)m totally porous silica particles of narrow size distribution. J Chromatogr A 1427:55–68CrossRefGoogle Scholar
  60. 60.
    Ismail OH, Catani M, Pasti L, Cavazzini A, Ciogli A, Villani C, Kotoni D, Gasparrini F, Bell DS (2016) Experimental evidence of the kinetic performance achievable with columns packed with the new 1.9 \(\upmu\)m fully porous particles Titan C\(_{18}\). J Chromatogr A 1454:86–92CrossRefGoogle Scholar
  61. 61.
    Ismail OH, Pasti L, Ciogli A, Villani C, Kocergin J, Anderson S, Gasparrini F, Cavazzini A, Catani M (2016) Pirkle-type chiral stationary phase on coreshell and fully porousparticles: Are superficially porous particles always the better choice toward ultrafast high-performance enantioseparations? J Chromatogr A 1466:96–104CrossRefGoogle Scholar
  62. 62.
    Kharaishvili Q, Jibuti G, Farkas T, Chankvetadze B (2016) Further proof to the utility of polysaccharide-based chiral selectors in combination with superficially porous silica particles as effective chiral stationary phases for separation of enantiomers in high-performance liquid chromatography. J Chromatogr A 1467:163–168CrossRefGoogle Scholar
  63. 63.
    Catani M, Ismail OH, Gasparrini F, Antonelli M, Pasti L, Marchetti N, Felletti S, Cavazzini A (2017) Recent advancements and future directions of superficially porous chiral stationary phases for ultrafast high-performance enantioseparations. Analyst 142:555–566CrossRefGoogle Scholar
  64. 64.
    Ismail OH, Antonelli M, Ciogli A, Villani C, Cavazzini A, Catani M, Felletti S, Bell DS, Gasparrini F (2017) Future perspectives in high efficient and ultrafast chiral liquidchromatography through zwitterionic teicoplanin-based 2\(\upmu\)m superficially porous particles. J Chromatogr A 1520:91–102CrossRefGoogle Scholar
  65. 65.
    Catani M, Felletti S, Ismail OH, Gasparrini F, Pasti L, Marchetti N, Luca CD, Costa V, Cavazzini A (2018) New frontiers and cutting edge applications in ultra high performance liquid chromatography through latest generation superficially porous particles with particular emphasis to the field of chiral separations. Anal Bioanal Chem 410:2457–2465CrossRefGoogle Scholar
  66. 66.
    Patel DC, Breitbach ZS, Wahab MF, Barhate CL, Armstrong DW (2015) Gone in seconds: praxis, performance and peculiarities of ultrafast chiral liquid chromatography with superficially porous particles. Anal Chem 87:9137–9148CrossRefGoogle Scholar
  67. 67.
    Spudeit DA, Dolzan MD, Breitbach ZS, Barber WE, Micke GA, Armstrong DW (2014) Superficially porous particles vs. fully porous particles for bonded high performance liquid chromatographic chiral stationary phases: Isopropyl cyclofructan 6. J Chromatogr A 1363:89–95CrossRefGoogle Scholar
  68. 68.
    Regalado EL, Welch CJ (2015) Pushing the speed limit in enantioselective supercritical fluid chromatography. J Sep Sci 38:2826–2832CrossRefGoogle Scholar
  69. 69.
    Barhate CL, Wahab MF, Breitbach ZS, Bell DS, Armstrong DW (2015) High efficiency, narrow particle size distribution, sub-2 \(\upmu\)m based macrocyclic glycopeptide chiral stationary phases in HPLC and SFC. Anal Chim Acta 898:128–137CrossRefGoogle Scholar
  70. 70.
    Patel DC, Breitbach ZS, Yu J, Nguyen KA, Armstrong DW (2017) Quinine bonded to superficially porous particles for high-efficiency and ultrafast liquid and supercritical fluid chromatography. Anal Chim Acta 963:164–174CrossRefGoogle Scholar
  71. 71.
    Berger TA (2016) Kinetic performance of a 50 mm long 1.8 \(\upmu\)m chiral column in supercritical fluid chromatography. J Chromatogr A 1459:136–144CrossRefGoogle Scholar
  72. 72.
    Berger TA (2017) Preliminary kinetic evaluation of an immobilized polysaccharide sub-2 \(\upmu\)m column using a low dispersion supercritical fluid chromatograph. J Chromatogr A 1510:82–88CrossRefGoogle Scholar
  73. 73.
    Ismail OH, Losacco GL, Mazzoccanrti G, Ciogli A, Villani C, Catani M, Pasti L, Anderson S, Cavazzini A, Gasparrini F (2018) Unmatched kinetic performance in enantioselective supercritical fluid chromatography by combining latest generation Whelk-O1 chiral stationary phases with a low-dispersion in-house modified equipment. Anal Chem.  https://doi.org/10.1021/acs.analchem.8b01907
  74. 74.
    Barhate CL, Wahab MF, Tognarelli DJ, Berger TA, Armstrong DW (2016) Instrumental idiosyncrasies affecting the performance of ultrafast chiral and achiral sub/supercritical fluid chromatography. Anal Chem 88:8664–8672CrossRefGoogle Scholar
  75. 75.
    Speybrouck D, Lipka E (2016) Preparative supercritical fluid chromatography: a powerful tool for chiral separations. J Chromatogr A 1467:33–55CrossRefGoogle Scholar
  76. 76.
    Galea C, Mangelings D, Vander Heyden Y (2015) Method development for impurity profiling in SFC: the selection of a dissimilar set of stationary phases. J Pharm Biomed Anal 111:333–343CrossRefGoogle Scholar
  77. 77.
    Enmark M, Ȧsberg D, Leek H, Öhlén K, Klarqvist M, Samuelsson J, Fornstedt T (2015) Evaluation of scale-up from analytical to preparative supercritical fluid chromatography. J Chromatogr A 1425:280–286CrossRefGoogle Scholar
  78. 78.
    Landagaray E, Vaccher C, Yous S, Lipka E (2016) Design of experiments for enantiomeric separation in supercritical fluid chromatography. J Pharm Biomed Anal 120:297–305CrossRefGoogle Scholar
  79. 79.
    Sun M, Sandahl M, Turner C (2018) Comprehensive on-line two-dimensional liquid chromatography \(\times\) supercritical fluid chromatography with trapping column-assisted modulation for depolymerised lignin analysis. J Chromatogr A 1541:21–30CrossRefGoogle Scholar
  80. 80.
    Gargano AFG, Duffin M, Navarro P, Schoenmakers PJ (2016) Reducing dilution and analysis time in online comprehensive two-dimensional liquid chromatography by active modulation. Anal Chem 88:1785–1793CrossRefGoogle Scholar
  81. 81.
    Francois I, Pereira AD, Lynen F, Sandra P (2008) Construction of a new interface for comprehensive supercritical fluid chromatography x reversed phase liquid chromatography (SFC \(\times\) RPLC). J Sep Sci 1541:21–30Google Scholar
  82. 82.
    Pauw RD, Choikhet K, Desmet G, Broeckhoven K (2014) Occurrence of turbulent flow conditions in supercritical fluid chromatography. J Chromatogr A 1361:277–285CrossRefGoogle Scholar
  83. 83.
    Giddings JC (1965) Dynamics of Chromatography. Marcel Dekker, New YorkGoogle Scholar