Red blood cell tolerance to shear stress above and below the subhemolytic threshold

  • Jarod T. HorobinEmail author
  • Surendran Sabapathy
  • Michael J. Simmonds
Original Paper
  • 36 Downloads

Abstract

Mechanical circulatory support device (MCS) design has improved over the years and yet blood damage (e.g., hemolysis) remains a problem. Accumulating evidence indicates a subhemolytic threshold for red blood cells (RBC)—a threshold at which RBC deformability is impaired prior to hemolysis. The current study aimed to assess the deformability of RBC exposed to supra-physiological shear stresses that are typical of MCS devices and assess whether a method used to estimate an individualized subhemolytic threshold, accurately demarcates whether future application of shear stress was damaging. Suspensions of RBC were “conditioned” with discrete magnitudes of shear stress (5–100 Pa) for specific durations (1–16 s). Cellular deformability was subsequently measured via ektacytometry and a mechanical sensitivity (MS) index was calculated to identify the subhemolytic threshold. Thereafter, fresh RBC suspensions were exposed to a magnitude of shear stress 10 Pa above, 10 Pa below, or matched to a donor’s previously estimated subhemolytic threshold for a given duration (1, 4, 16 s) to ascertain the sensitivity of the subhemolytic threshold. The MS index of RBC was significantly impaired following exposure to 10 Pa above the subhemolytic threshold (p < 0.0001), and significantly enhanced following exposure to 10 Pa below the subhemolytic threshold (p < 0.01). For all shear conditions, there was no significant increase in free hemoglobin. Functional assessments of RBC may be useful when conducting biocompatibility testing of MCS devices, to detect trauma to blood prior to overt cell rupture being induced.

Keywords

Erythrocyte Haemorheology Hemocompatibility Mechanical damage Membrane elasticity 

Notes

Acknowledgements

The authors thank Mr. Kieran Richardson for assistance in data collection.

Compliance with ethical standards

Conflict of interest

The authors declare no conflict of interest.

References

  1. Acharya D, Loyaga-Rendon R, Morgan CJ, Sands KA, Pamboukian SV, Rajapreyar I, Tallaj JA (2017) INTERMACS analysis of stroke during support with continuous-flow left ventricular assist devices: risk factors and outcomes. JACC Heart Fail 5(10):703–711.  https://doi.org/10.1016/j.jchf.2017.06.014 CrossRefGoogle Scholar
  2. Baldwin JT, Deutsch S, Geselowitz DB, Tarbell JM (1994) LDA measurements of mean velocity and Reynolds stress fields within an artificial heart ventricle. J Biomech Eng 116(2): 190–200. Retrieved from https://www.ncbi.nlm.nih.gov/pubmed/8078326 CrossRefGoogle Scholar
  3. Baskurt OK, Meiselman HJ (2013) Red blood cell mechanical stability test. Clin Hemorheol Microcirc 55(1):55–62.  https://doi.org/10.3233/CH-131689 CrossRefGoogle Scholar
  4. Baskurt OK, Hardeman MR, Uyuklu M, Ulker P, Cengiz M, Nemeth N, Meiselman HJ (2009) Parameterization of red blood cell elongation index–shear stress curves obtained by ektacytometry. Scand J Clin Lab Investig 69(7):777–788.  https://doi.org/10.3109/00365510903266069 CrossRefGoogle Scholar
  5. Bessis M, Mohandas N, Feo C (1980) Automated ektacytometry: a new method of measuring red cell deformability and red cell indices. Blood Cells 6(3): 315–327. Retrieved from https://www.ncbi.nlm.nih.gov/pubmed/7397390
  6. Chasis JA, Mohandas N (1986) Erythrocyte membrane deformability and stability: two distinct membrane properties that are independently regulated by skeletal protein associations. J Cell Biol 103(2): 343–350. Retrieved from https://www.ncbi.nlm.nih.gov/pubmed/3733870 CrossRefGoogle Scholar
  7. Chen Y, Li D, Li Y, Wan J, Li J, Chen H (2017) Margination of stiffened red blood cells regulated by vessel geometry. Sci Rep 7(1):15253.  https://doi.org/10.1038/s41598-017-15524-0 CrossRefGoogle Scholar
  8. Chien S (1987) Red cell deformability and its relevance to blood flow. Annu Rev Physiol 49:177–192.  https://doi.org/10.1146/annurev.ph.49.030187.001141 CrossRefGoogle Scholar
  9. Deutsch S, Tarbell JM, Manning KB, Rosenberg G, Fontaine AA (2006) Experimental fluid mechanics of pulsatile artificial blood pumps. Annu Rev Fluid Mech 38:65–86.  https://doi.org/10.1146/annurev.fluid.38.050304.092022 CrossRefzbMATHGoogle Scholar
  10. Dodson WR 3rd, Dimitrakopoulos P (2010) Tank-treading of erythrocytes in strong shear flows via a nonstiff cytoskeleton-based continuum computational modeling. Biophys J 99(9):2906–2916.  https://doi.org/10.1016/j.bpj.2010.08.048 CrossRefGoogle Scholar
  11. Filipovic A, Kleinoder H, Pluck D, Hollmann W, Bloch W, Grau M (2015) Influence of whole-body electrostimulation on human red blood cell deformability. J Strength Cond Res 29(9):2570–2578.  https://doi.org/10.1519/JSC.0000000000000916 CrossRefGoogle Scholar
  12. Grau M, Pauly S, Ali J, Walpurgis K, Thevis M, Bloch W, Suhr F (2013) RBC-NOS-dependent S-nitrosylation of cytoskeletal proteins improves RBC deformability. PLoS ONE 8(2):e56759.  https://doi.org/10.1371/journal.pone.0056759 CrossRefGoogle Scholar
  13. Gregory SD, Timms D, Gaddum NR, McDonald C, Pearcy MJ, Fraser JF (2011) In vitro evaluation of a compliant inflow cannula reservoir to reduce suction events with extracorporeal rotary ventricular assist device support. Artif Organs 35(8):765–772.  https://doi.org/10.1111/j.1525-1594.2011.01311.x CrossRefGoogle Scholar
  14. Harboe M (1959) A method for determination of hemoglobin in plasma by near-ultraviolet spectrophotometry. Scand J Clin Lab Investig 11:66–70.  https://doi.org/10.3109/00365515909060410 CrossRefGoogle Scholar
  15. Helms CC, Marvel M, Zhao W, Stahle M, Vest R, Kato GJ, Kim-Shapiro DB (2013) Mechanisms of hemolysis-associated platelet activation. J Thromb Haemost 11(12):2148–2154.  https://doi.org/10.1111/jth.12422 CrossRefGoogle Scholar
  16. Horobin JT, Sabapathy S, Simmonds MJ (2017) Repetitive supra-physiological shear stress impairs red blood cell deformability and induces hemolysis. Artif Organs 41(11):117–1025.  https://doi.org/10.1111/aor.12890 CrossRefGoogle Scholar
  17. Horobin JT, Watanabe N, Hakozaki M, Sabapathy S, Simmonds MJ (2019) Shear-stress mediated nitric oxide production within red blood cells: a dose-response. Clin Hemorheol Microcirc 71(2):203–214.  https://doi.org/10.3233/CH-189412 CrossRefGoogle Scholar
  18. Koutsiaris AG, Tachmitzi SV, Batis N (2013) Wall shear stress quantification in the human conjunctival pre-capillary arterioles in vivo. Microvasc Res 85:34–39.  https://doi.org/10.1016/j.mvr.2012.11.003 CrossRefGoogle Scholar
  19. Lac E, Barthès-Biesel D (2005) Deformation of a capsule in simple shear flow: effect of membrane prestress. Phys Fluids 17(7):072105.  https://doi.org/10.1063/1.1955127 CrossRefzbMATHGoogle Scholar
  20. Leverett LB, Hellums JD, Alfrey CP, Lynch EC (1972) Red blood cell damage by shear stress. Biophys J 12(3):257–273.  https://doi.org/10.1016/S0006-3495(72)86085-5 CrossRefGoogle Scholar
  21. Malinauskas RA (1997) Plasma hemoglobin measurement techniques for the in vitro evaluation of blood damage caused by medical devices. Artif Organs 21(12): 1255–1267. Retrieved from https://www.ncbi.nlm.nih.gov/pubmed/9423977 CrossRefGoogle Scholar
  22. McNamee AP, Sabapathy S, Singh I, Horobin J, Guerrero J, Simmonds MJ (2016) Acute free-iron exposure does not explain the impaired haemorheology associated with haemochromatosis. PLoS ONE 11(1):e0146448.  https://doi.org/10.1371/journal.pone.0146448 CrossRefGoogle Scholar
  23. McNamee AP, Horobin JT, Tansley GD, Simmonds MJ (2017) Oxidative stress increases erythrocyte sensitivity to shear-mediated damage. Artif Organs.  https://doi.org/10.1111/aor.12997 CrossRefGoogle Scholar
  24. McNamee AP, Richardson K, Horobin J, Kuck L, Simmonds MJ (2019) Susceptibility of density-fractionated erythrocytes to subhaemolytic mechanical shear stress. Int J Artif Organs 42(3):151–157.  https://doi.org/10.1177/0391398818790334 CrossRefGoogle Scholar
  25. Meram E, Yilmaz BD, Bas C, Atac N, Yalcin O, Meiselman HJ, Baskurt OK (2013) Shear stress-induced improvement of red blood cell deformability. Biorheology 50(3–4):165–176.  https://doi.org/10.3233/BIR-130637 CrossRefGoogle Scholar
  26. Papaioannou TG, Stefanadis C (2005) Vascular wall shear stress: basic principles and methods. Hellenic J Cardiol 46(1): 9–15. Retrieved from http://www.ncbi.nlm.nih.gov/pubmed/15807389
  27. Paul R, Apel J, Klaus S, Schugner F, Schwindke P, Reul H (2003) Shear stress related blood damage in laminar couette flow. Artif Organs 27(6): 517–529. Retrieved from http://www.ncbi.nlm.nih.gov/pubmed/12780506 CrossRefGoogle Scholar
  28. Piper R, Carr PJ, Kelsey LJ, Bulmer AC, Keogh S, Doyle BJ (2018) The mechanistic causes of peripheral intravenous catheter failure based on a parametric computational study. Sci Rep 8(1):3441.  https://doi.org/10.1038/s41598-018-21617-1 CrossRefGoogle Scholar
  29. Rey J, Buffet PA, Ciceron L, Milon G, Mercereau-Puijalon O, Safeukui I (2014) Reduced erythrocyte deformability associated with hypoargininemia during Plasmodium falciparum malaria. Sci Rep 4:3767.  https://doi.org/10.1038/srep03767 CrossRefGoogle Scholar
  30. Secomb TW, Pries AR (2007) Basic principles of hemodynamics. In: Baskurt OK, Hardeman MR, Rampling MW, Meiselman HJ (eds) Handbook of hemorheology and hemodynamics. IOS Press, AmsterdamGoogle Scholar
  31. Selgrade BP, Truskey GA (2012) Computational fluid dynamics analysis to determine shear stresses and rates in a centrifugal left ventricular assist device. Artif Organs 36(4):E89–E96.  https://doi.org/10.1111/j.1525-1594.2011.01416.x CrossRefGoogle Scholar
  32. Simmonds MJ, Meiselman HJ (2016) Prediction of the level and duration of shear stress exposure that induces subhemolytic damage to erythrocytes. Biorheology 53(5–6):237–249.  https://doi.org/10.3233/BIR-16120 CrossRefGoogle Scholar
  33. Simmonds MJ, Connes P, Sabapathy S (2013) Exercise-induced blood lactate increase does not change red blood cell deformability in cyclists. PLoS ONE 8(8):e71219.  https://doi.org/10.1371/journal.pone.0071219 CrossRefGoogle Scholar
  34. Simmonds MJ, Atac N, Baskurt OK, Meiselman HJ, Yalcin O (2014) Erythrocyte deformability responses to intermittent and continuous subhemolytic shear stress. Biorheology 51(2–3):171–185.  https://doi.org/10.3233/BIR-140665 CrossRefGoogle Scholar
  35. Simmonds MJ, Suriany S, Ponce D, Detterich JA (2018) Red blood cell mechanical sensitivity improves in patients with sickle cell disease undergoing chronic transfusion after prolonged, subhemolytic shear exposure. Transfusion 58(12):2788–2796.  https://doi.org/10.1111/trf.14901 CrossRefGoogle Scholar
  36. Simon HA, Ge L, Sotiropoulos F, Yoganathan AP (2010) Numerical investigation of the performance of three hinge designs of bileaflet mechanical heart valves. Ann Biomed Eng 38(11):3295–3310.  https://doi.org/10.1007/s10439-010-0086-3 CrossRefGoogle Scholar
  37. Thamsen B, Blumel B, Schaller J, Paschereit CO, Affeld K, Goubergrits L, Kertzscher U (2015) Numerical analysis of blood damage potential of the HeartMate II and HeartWare HVAD rotary blood pumps. Artif Organs 39(8):651–659.  https://doi.org/10.1111/aor.12542 CrossRefGoogle Scholar
  38. Ulker P, Sati L, Celik-Ozenci C, Meiselman HJ, Baskurt OK (2009) Mechanical stimulation of nitric oxide synthesizing mechanisms in erythrocytes. Biorheology 46(2):121–132.  https://doi.org/10.3233/BIR-2009-0532 CrossRefGoogle Scholar
  39. Ulker P, Meiselman HJ, Baskurt OK (2010) Nitric oxide generation in red blood cells induced by mechanical stress. Clin Hemorheol Microcirc 45(2–4):169–175.  https://doi.org/10.3233/CH-2010-1293 CrossRefGoogle Scholar
  40. Ulker P, Yaras N, Yalcin O, Celik-Ozenci C, Johnson PC, Meiselman HJ, Baskurt OK (2011) Shear stress activation of nitric oxide synthase and increased nitric oxide levels in human red blood cells. Nitric Oxide 24(4):184–191.  https://doi.org/10.1016/j.niox.2011.03.003 CrossRefGoogle Scholar
  41. Ulker P, Gunduz F, Meiselman HJ, Baskurt OK (2013) Nitric oxide generated by red blood cells following exposure to shear stress dilates isolated small mesenteric arteries under hypoxic conditions. Clin Hemorheol Microcirc 54(4):357–369.  https://doi.org/10.3233/CH-2012-1618 CrossRefGoogle Scholar
  42. Watanabe N, Shimada T, Hakozaki M, Hara R (2018) Visualization of erythrocyte deformation induced by supraphysiological shear stress. Int J Artif Organs 41(12):838–844.  https://doi.org/10.1177/0391398818793387 CrossRefGoogle Scholar
  43. Wiedeman MP (1963) Dimensions of blood vessels from distributing artery to collecting vein. Circ Res 12: 375–378. Retrieved from http://www.ncbi.nlm.nih.gov/pubmed/14000509 CrossRefGoogle Scholar

Copyright information

© Springer-Verlag GmbH Germany, part of Springer Nature 2019

Authors and Affiliations

  1. 1.Menzies Health Institute QueenslandGriffith UniversityGold CoastAustralia
  2. 2.Biorheology Research LaboratoryGriffith UniversityGold CoastAustralia

Personalised recommendations