Ruthenium dioxide nanoparticles as a high-capacity transducer in solid-contact polymer membrane-based pH-selective electrodes
- 140 Downloads
Abstract
A new approach is presented for the design of ion selective electrodes. Ruthenium dioxide nanoparticles were incorporated into solid-contact electrodes, and their properties were studied for the case of pH-selective electrodes. The use of the RuO2 is shown to significantly improve the potentiometric response, while no redox response is observed. The use of RuO2 results in a Nernstian slope (59 mV/decade) towards hydrogen ions over a wide linear range (pH 2 to 12). The results obtained by chronopotentiometry reveal small resistance, and the capacitance is as high as 1.12 mF. This results in a good stability of the response and in a low potential drift (0.89 μV∙s−1). The electrodes exhibit properties nearly as excellent as those of a glass electrode, but they are much smaller, less fragile, and easy to use.
Schematic representation of the construction of the new kind of electrodes along with calibration and chronopotentiometric plots compared to non-modified GCD/H+-ISM and modified GCD/RuO2/H+-ISM electrodes, respectively. The use of ruthenium dioxide results in a wide analytical pH range (2–12) and in high electrical capacitance (1.12 mF).
Keywords
Ruthenium dioxide Nanoparticles High-capacity transducer Ion-to-electron transducer All-solid-state Solid-contact Ion-selective electrode pH potentiometric sensor Stable potentialAccording to IUPAC and other sources [1, 2, 3] potentiometry is the only accurate method of pH measurement. Measurements of pH are normally performed using the glass electrode due to its high selectivity and wide pH measurement range [4]. However, its construction results in number of limitations and the main one is its fragility what makes the electrode vulnerable to any mechanical damage and large size, which preclude the analysis of small volume samples [5]. Conventional electrodes such as glass electrode are associated with disadvantages caused by the presence of an inner solution. This construction of an electrode requires vertical position, stable temperature and pressure to avoid the phase change of the inner solution [6]. To overcome those shortcomings, the inner solution must have been removed from electrode construction. The first construction of a solid-contact electrode was, so called, coated-wire electrode of a simple construction but poor potential stability and irreproducibility caused by the low double layer capacitance and high resistance of the charge transfer between the ion-selective membrane (ISM) and the electronic conductor [7]. The transduction between the electronic conductor and the ionic conductor (ion – selective membrane) became possible when the internal layer which exhibits both types of transduction (electronic and ionic) was implemented into the electrode construction [8].
Various electroactive materials such as conducting polymers [9] and organic conducting crystals [10] which exhibit high redox capacitance, as well as nanomaterials transducers [11] (such as microfibers, microspheres, nanoparticles and microcapsules), nanoporous gold [12] and carbon black [13, 14] characterized by a high double layer capacitance have been already used as internal layers. Metal oxides also appeared to improve electrodes’ properties. Zeng and Qin [15] used MoO2 as solid-contact layer and observed the impact on electrical parameters, resulting in stable and rapid potential response.
In this paper the new approach was presented by the use of ruthenium oxide nanoparticles as a material for high-capacity transducer to construct a solid contact H+-selective ISE with a PCV membrane. To the best of our knowledge this material has never been used as solid-contact layer in ion selective electrodes’ construction.
Ruthenium dioxide (RuO2) characterized as transition metal dioxide with a rutile structure is of both scientific and technological importance, what results from its unique properties. Ruthenium dioxide exhibits a combination of unique characteristics such as high thermal and chemical stability, low resistivity, and remarkable redox properties [16] what was taken advantage of when designing the ISE.
We are presenting here a universal approach which will be implemented into further studies over ISEs including improving parameters of electrodes and expanding the analytes range. To the best of our knowledge this is the first article concerning this kind of application of ruthenium oxide as solid contact layer in all-solid-state electrodes.
Experimental
Chemicals
Membrane components: hydrogen ionophore V (Calix[4]-aza-crown), sodium tetrakis(4-fluorophenyl)borate dihydrate, 2-Nitrophenyl octyl ether (NPOE) and high molecular weight poly(vinyl chloride) were purchased from Sigma-Aldrich (www.sigmaaldrich.com). Dimethylformamide (DMF) and Tetrahydrofuran (THF) used as solvents were also purchased from Sigma-Aldrich. Ruthenium dioxide (RuO2) used as solid-contact layer was purchased from Acros Organics (www.acros.com). Other chemicals (including TRIS, citric acid, boric acid, sodium hydroxide, potassium hydroxide, lithium hydroxide and sodium chloride) were used for further analysis in form of aqueous solutions. Aqueous solutions were prepared by dissolving salts and acids in distilled and deionized water and then titrated with sodium hydroxide and hydrochloric acid to meet the desired pH values. All chemicals were of analytical grade and were used as received without any further purification.
Electrodes preparation
To determine the impact of RuO2 solid contact layer on the all-solid-state pH-selective electrodes’ properties solid-contact electrodes and a coated-disc electrode were developed. The solid contact layers were obtained by dispersion of ruthenium oxide (5 mg) in DMF (1 mL) and oxide’s implementation onto Glassy Carbon Disc (GCD) electrodes using drop casting method. Before casting, the electrodes were cleaned: first by polishing them with alumina slurries (particle size of 0.3 μm and 0.05 μm), then rinsed with water and finally cleaned ultrasonically with water and methanol. Cleaned and dried GCD electrodes were then covered by dropping 20 μl of ruthenium oxide DMF solution in order to prepare RuO2-modified electrodes. Mediation layers were dried in elevated temperature for several minutes. The hydrogen selective membrane cocktail was obtained by dissolution 252 mg of the membrane components in 2 mL of THF. The composition of H+- ISM was as follows: 0.90% (w/w) hydrogen ionophore V, 66% (w/w) o-NPOE, 32.85% (w/w) PVC, and 0.25% (w/w) sodium tetrakis(4-fluorophenyl)borate dihydrate. Then, solid contact layers were covered twice with 30 μL of ion-selective membrane solution in order to obtain solid-contact H+-selective electrodes (GCD/RuO2/H+-ISM). Additionally, the coated disc electrodes (GCD/H+-ISM) were prepared as controls by covering glassy carbon disc electrodes with above mentioned membrane.
Afterwards, all electrodes were left to complete solvent evaporation at room temperature for 24 h. Dried electrodes were conditioned in 0.1 M NaCl solution for 1 day and then the conditioning procedure was repeated prior to every measurement. Three identical electrodes were prepared and all of them were studied.
Electrochemical and electron microscopic measurements
Potentiometric measurements were performed using a 16-channel mV-meter (Lawson Labs, Inc., Malvern, PA) (www.lawsonlabs.com). The reference electrode was an Ag/AgCl electrode with 3 M KCl solution (6.0733.100 Metrohm, Switzerland) (www.metrohm.com) The potentiometric response towards H+ was measured by recording calibration plots for the electrodes in standard solutions buffered with 10 mM citric acid and 10 mM boric acid titrated with sodium hydroxide or hydrochloric acid. The pH range of buffer was 2 to 12. The selectivity measurements were carried out in Tris buffered solutions with the concentration of studied cations of 10−1 M using Fixed Interference Method. Potentiometric measurements were conducted in the presence of glass electrode (6.0150.100 Metrohm, Switzerland).
The chronopotentiometry and electrical impedance spectroscopy measurements were conducted with the use of an Autolab General Purpose Electrochemical System (AUT302N.FRA2-AUTOLAB, Metrohm Autolab, The Netherlands) (www.metrohm-autolab.com). Studied ion-selective electrodes were being connected in sequence as working electrodes into three-electrode cell with an Ag/AgCl/3MKCl electrode as the reference electrode and a glassy carbon rod as the auxiliary one. All data analysis was carried out and interpreted using NOVA 2.1.
Chronopotentiometric studies were carried out in buffer of pH 3. A constant current of +1 nA was applied to the working electrode for 60 s, followed by a − 1 nA current for another 60 s.
The electrochemical impendence spectroscopy (EIS) measurements were performed in buffer of pH 3. Impedance spectra were measured by applying a frequency from 100 kHz to 0.01 Hz using an AC amplitude of 50 mV superimposed on open-circuit potential (OCP).
The morphologies of the ruthenium oxide as well as ion-selective membrane were examined with the use of a scanning electron microscope, SEM model LEO 1530 from LEO Electron Microscopy, Carl Zeiss¸ Germany (www.zeiss.com).
Results and discussion
Choice of material
Ru-based materials are well-known supercapacitors (electric double layer capacitors), which electrical performance is strongly connected to the surface area [17].Because of its unique assets such as high chemical stability, low resistivity, and remarkable redox properties [16], ruthenium dioxide is an excellent material for preparing electrodes [18, 19].In our ion-selective all-solid-state electrodes, ruthenium dioxide nanoparticles were used to create a solid-contact layer placed between ion-selective membrane and electronic conductor.One of the approaches to improve properties of all-solid-state ion-selective electrodes is to implement the material characterized by high electrical capacitance. This can be done in two ways: either by using the material of high surface area or the material which displays high redox capacitance.
Ruthenium dioxide turned out to exhibit both mentioned properties. Thanks to the nanometric size of its particles (revealed by Scanning Electron Microscopy), oxide is characterized by the high surface area and this was the first reason to consider ruthenium dioxide as promising material to be applied as solid-contact layer in ion-selective electrodes. Besides its favourable morphology, ruthenium dioxide act as an “ion-to-electron” transducer. One of the mechanisms proposed by Fog and Buck in [20] is that in contact with hydrogen ions oxide undergoes redox reaction in which ions and protons are being exchanged. This mechanism decides of oxide’s ability to transduce both ions and electrons through the layer, which simplifies the processes on the interface between ion-selective membrane and electronic conductor.The combination of those two properties: high surface area and high redox capacitance [6] in one material, used as solid-contact layer allowed to obtain electrodes characterized by stable potentiometric response.
One other feature that needs to be taken into consideration when designing the robust ion-selective electrodes is the wettability of chosen material. This is of particular importance in terms of creating the water film between membrane and electronic conductor which may lead to deteriorating membrane adherence [21]. Hydrophilic materials tend to create the water layer, making the electrode prone to damages, whereas hydrophobic materials are said to prevent the occurrence of water layer. In our studies, despite of ruthenium dioxide’s hydrophilic properties (as we presented in [22]) and extensive conditioning (time of conditioning equal to 3 months), during the water layer test we did not observed characteristic features indicating the presence of water film (not shown here). This may be caused by the specific reaction that occurs on the surface of RuO2-layer [20] or/and the high surface area that determines strong adherence between solid-contact layer and membrane, preventing its detaching from electrode’s surface.
The choice of ruthenium dioxide for solid-contact layer in pH-selective electrodes turned out to be appropriate and allowed to obtain pH-sensor which exhibit remarkable analytical and performance parameters described in the paper.
Scanning electron microscopy
SEM images showing the morphology of a ruthenium oxide, b profile of ion-selective membrane (i) and ruthenium oxide (ii) as internal layer casted onto titanium pad (iii) with the estimated thickness of the layers
Potentiometry measurements
Comparison of repeatability of coated wire and RuO2-modified electrodes (based on 6 calibrations conducted over one week n = 6)
Electrode | Slope S ± SD [mV/pH] | Standard potential E0 [mV] | Linear range [M] |
---|---|---|---|
GCD/RuO2/H+-ISM | 59.31 ± 0.15 | 571 ± 2 | 10−2 - 10−12 |
GCD/H+-ISM | 46.86 ± 4.60 | 550 ± 14 | 10−4 - 10−10 |
Mean EMF values with SD recorded over 6 calibrations of (a) GCD/RuO2/H+- ISM electrode and (b) GCD/H+- ISM electrode performed over one week (n = 6 calibrations)
Comparison of reproducibility of coated wire and RuO2-modified electrodes (n = 3, conditioning time equal to 48 h)
Electrode | Slope S ± SD [mV/pH] | Standard potential E0 [mV] | Linear range [M] |
---|---|---|---|
GCD/RuO2/H+-ISM | 59.05 ± 0.18 | 571 ± 1.7 | 10−2 - 10−12 |
GCD/H+-ISM | 48.72 ± 3.3 | 556 ± 12 | 10−4 - 10−10 |
The GCD/RuO2/H+- ISM electrode showed Nernstian response in wide range of pH value (2–12) and the slope value was stable and close to theoretical value with the increasing time of conditioning. The GCD/RuO2/H+-ISM electrode exhibited great repeatability and reproducibility (given by standard deviation values) as it is shown in the Table 1 and Table 2, respectively. For the coated-wire electrode the narrow lineal range of pH was obtained, and the near-Nernstian response was observed from pH = 5 to pH = 10 only after 24 h of conditioning. In further calibrations GCD/H+-ISM electrodes exhibited worse performance.
Electrodes’ response reproducibility tested by measuring EMF in a pH = 4, b pH = 5 and c pH = 6 buffer (i – GCD/RuO2/H+- ISM, ii - GCD/H+- ISM, iii – glass electrode)
Fig. 3 shows that the RuO2-contacted electrode exhibited almost as good stabilization as glass electrode in contrast to coated-wire electrode.
The measurements confirmed a good reproducibility of GCD/RuO2/H+- ISM electrode, for which registered potential was repetitive when measuring the solution of particular pH value. RuO2 - contacted electrode showed highly reversible potential response towards hydrogen ions.
Redox test
Redox response compared for glass electrode (Open image in new window), RuO2 – modified electrode (Open image in new window
), coated-wire electrode (Open image in new window
) and platinum electrode (Open image in new window
)
There was no redox response detected in RuO2-modified electrodes and a slight change of potential was detected in non-modified ones. The change of potential was caused rather by various pH values in examined solutions than by redox signal. Platinum electrode exhibited near- Nernstian response with the change of log Fe2+/Fe3+ value (61.50 mV/dec) as expected.
Although solid contact layer of RuO2 is an electronic conductor and should exhibit clear redox response, the polymer-based ion-selective membrane is an electronic insulator and eliminates the redox response. Therefore, the examined sensors are not redox sensitive.
Selectivity coefficients
The information about the selectivity of ion-selective electrodes is crucial to determine whether an electrode cannot be used in certain conditions. The value of the selectivity coefficient informs about electrode’s ability to distinguish the primary ion in the presence of interfering ions in the examined solution. Electrode selectivity is determined by the membrane components – mainly ionophore which is responsible for selective recognition and binding of the determined ion. What is important, hydrogen ionophore V used for preparing GCD/RuO2/H+-ISM electrode shows selectivity for divalent ions such as Mg2+ or Ca2+ due to the presence of t-butyl groups in its structure [23]. No influence of these ions on the electrode potential was observed even with their large excess compared to hydrogen ions.
Exemplary potentiometric response of GCD/RuO2/H+-ISM electrode measured in TRIS-HCl buffer of pH range 2–12 containing 0.1 M of NaOH, KOH and LiOH
Interfering cations were examined in form of the hydroxides (NaOH, KOH, LiOH) and their concentration in the analysed solutions was of 10−1 M. Prior to the measurement the ISEs were conditioned in pH = 3 buffered solution for 1 h. To determine the selectivity of studied electrodes, their responses were recorded in buffered TRIS solutions of constant hydroxide concentration, titrated with hydrochloric acid, while the pH was monitored with the glass electrode to achieve solutions of pH range 2 to 12. The selectivity sequence for the studied electrodes was: Li > K > Na. Obtained results were consistent with the results presented in [23] and there was no potentiometric response towards interfering cations observed from pH = 2 to pH = 12.
In summary, the addition of ruthenium oxide as solid contact does not affect the selectivity of investigated H+-selective electrodes, as expected. Similar results were observed when using carbon nanomaterials [13, 25, 26].
Potential stability
EMF stabilization in 10−3 M solutions of KCl and HCl compared for coated-wire (i), solid-contact (ii) and glass (iii) electrodes
Afterwards the electrodes were left for 15 h in 10−3 M HCl solution to determine their potential stability over the time.
Stability of electrodes’ response over: a 15 h, b 1 h measured in 10−3 HCl solution
Studied electrodes have been tested over 3 months and during this period of time all of them displayed stable potentiometric response. After 3 months of intensive tests carried on electrodes, the membrane adherence was not deteriorated, electrodes were not damaged and the potential response was stable, therefore the lifetime of studied GCD/RuO2/H+-ISM electrodes can be estimated at over 3 months.
An overview on recently reported potentiometric methods for determination of hydrogen ions with solid contact electrodes
Electrode/ Solid contact material used for electrode construction | Linear range [M] | Slope [mV/pH] | Standard potential E0 [mV] | Time of response [s] | Ref. |
---|---|---|---|---|---|
Polypyrrole doped with hexacyanoferrate(II) (PPy-Fe(CN)) | 2–12 | 56.9 ± 4.3 | 519 ± 21 | 10 | [27] |
Multi-walled carbon nanotube (MWCNT) | 2.89–9.90 | 58.8 ± 0.4 | – | <10 | [5] |
Polydopamine - carbon nano-onion (CNO – PDA) | 1.50–10.50 | 60.1 ± 0.3 | – | – | [28] |
Polyaniline (PANI) | 2–9 | 52.7 ± 1.1 | – | 60–120 | [29] |
Nano Electrode: Polyaniline deposited onto a carbon fiber (PHNE) | 2.0–12.5 | 60.0 ± 0.5 | – | 6 s (pH 7) | [30] |
Nano Electrode: carbon nanotube deposited onto polyaniline (CFCNEs) | 1–13 | 58 | – | few seconds | [31] |
Polypyrrole (PPy) doped with cobaltabis(dicarbollide) ions ([3,3′-Co(1,2-C2B9H11)2]−) | 3.5–11 | 52 ± 2 | – | <45 | [32] |
Poly(3,4-ethylenedioxythiophene) − poly(styrenesulfonate) (PEDOT(PSS)) | 3–11 | 57.7 ± 0.2 | – | – | [33] |
Ruthenium dioxide (RuO2) | 2–12 | 59.31 ± 0.15 | 571 ± 2 | 5 | this work |
As presented in the table, electrodes with ruthenium dioxide as solid-contact layer are characterized by the linear range complementary to the previous solutions described in literature. Applying ruthenium dioxide allowed however to receive the calibration plot with slope value equal to 59.31 mV/pH which is in very good agreement with the theoretical Nernstian value. Obtained electrode can be also described with the outstanding repeatability represented with the standard deviation values from averaged values of slope and standard potential.
Time of response (t0.95) calculated for RuO2-contacted ISEs following the IUPAC principles equals 5 s. Amongst compared potentiometric methods, the shortest response time can be attributed to nano-electrodes for which the time needed to reach the stable potential equals merely few seconds. In comparison with presented nano-electrodes, studied ISEs displayed comparable time of response.
Despite being characterized by similar analytical parameters, the main difference between nano-sized electrodes and studied ISEs is that the size of nano-sensors can be narrowed to 100 nm, whereas glassy carbon electrodes are at least few centimetres long.
What should be emphasized here is that solid-contact materials presented in the Table 3, which exhibit satisfying analytical parameters such as wide linear range (which allow to measure at least 9 pH values) are the material composites (consist of two varying materials). Ruthenium dioxide applied alone and without any modifications, allows to obtain the ion-selective electrode of parameters that are comparable to the best potentiometric methods described in literature.
Chronopotentiometry measurements
The response of electrodes measured during 1 nA and − 1 nA current flow (orange - GCD/RuO2/H+-ISM, violet - GCD/H+-ISM); inset: the response of GCD/RuO2/H+-ISM electrode recorded during 120 s of forced current flow
The total resistance of the electrode was calculated using the value of potential jump ΔEdc and current I: Rtotal = ΔEdc/2I. The estimated values were as follow: for solid-contact electrode (GCD/RuO2/ H+-ISM) Rtotal = 0.90 ± 0.02 MΩ, for coated-wire electrode (GCD/H+-ISM) Rtotal = 3.20 ± 0.08 MΩ (n = 6).
The potential drift of the electrode was calculated and the capacitance C was obtained with the use of ΔEdc/Δt = I/C equation. The high electrical capacity determines the stability of the potential response (low potential drift ΔEdc/Δt). For RuO2-contact electrodes obtained results for capacitance were approximately 1.12 ± 0.04 mF and potential drift 0.89 ± 0.03 μV∙s−1 (n = 6) while for coated-wire electrodes, the capacitance value was lower (0.89 ± 0.04 μF) and potential drift given by ΔEdc/Δt ratio was much higher (1.13 ± 0.05 mV∙s−1) (n = 6).
Capacitance values compared for GCD/SC/ISM electrodes with various material applied as solid-contact (SC) layers
Electrode symbol | SC material | Capacitance value [μF] | Reference |
---|---|---|---|
GCD/CB/ NO3− -ISM | Carbon Black | 289 | [13] |
GCD/TCNQ/K+-ISM | 7,7,8,8-Tetracyanoquinodimethane | 132 | [34] |
GCD/TCNQ/Na+-ISM | 7,7,8,8-Tetracyanoquinodimethane | 154 | [34] |
GCD/GR/ K+-ISM | Graphene | 91 | [35] |
GCD/PEDOT/Ca2+-ISM | Poly(3,4-ethylenedioxythiophene) | 45 | [9] |
GCD/PEDOT(CNT)/K+-ISM | Poly(3,4-ethylenedioxythiophene) – Carbon Nanotubes | 83 | [36] |
GCD/MoO2/K+-ISM | Molybdenum Dioxide | 86 | [15] |
GCD/GR-TTF/NO3− -ISM | Graphene - Tetrathiafulvalene | 1180 | [37] |
GCD/CIM/K+-ISM | Colloid-Imprinted Mesoporous Carbon | 1000 | [25] |
GCD/CB-GR-FP/K+-ISM | Carbon Black – Graphene - Fluorinated acrylic copolymer | 1471 | [38] |
GCD/PC-SMSs/K+-ISM | Porous Carbon - Sub-Micrometer Spheres | 2803 | [26] |
GCD/CB-TTFCl/Cl−-ISM | Carbon Black – Chloride salt of Tetrathiafulvalene | 2800 | [39] |
GCD/RuO2/ H+-ISM | Ruthenium dioxide | 1120 | this work |
Also, it should be emphasized that electrodes with molybdenum dioxide [15] showed capacitance value that is several time smaller than for RuO2-contacted electrodes described in this paper.
Obtained capacitance value is comparable to values presented in [37] for GR-TTF -contacted ISE, in [25] for CIM -contacted ISE or in [38] for CB-GR-FP -contacted ISE. However, the values published by [26] for PC-SMSs -contacted ISE and [39] for CB-TTFCl - contacted ISE were significantly higher.
Electrochemical impedance spectroscopy measurements
EIS spectra of GCD/RuO2/H+-ISM (Open image in new window) and GCD/H+-ISM (Open image in new window
) recorded in pH = 3 buffer in the frequency range from 100 kHz to 10 mHz, using an amplitude of 50 mV. Inset: scaled graph of GCD/RuO2/H+-ISM electrode
The characteristic in the low-frequency region is caused by the double-layer capacitance (Cdl) coupled with a charge-transfer resistance (Rct) at the ISM and underlying electronic conductor interface providing information about the easiness for electron/ion transfer at the electrode interface [40]. The Rct can be separated into two components: the electronic and ionic resistances. The electronic resistance includes the electronic conductivity of the RuO2 particles, the electronic contact between particles, and electronic contact between GC material and RuO2. The ionic resistance is the membrane ionic resistance inside the pores of the RuO2 layer. It depends on the membrane conductivity, porosity of the RuO2 layer and its thickness. In the case of very porous layers, as the frequency drops, the AC signal penetrates deeper into the pores and the capacitance increases, which is responsible for the 45° slope seen at high frequencies (seen in the case of GCD/RuO2/H+-ISM electrode).
The low-frequency part of the EIS for GCD with polymeric H + -selective membrane GCD/H+-ISM showed a large semicircle arising from a small capacitance with a large charge-transfer resistance at the blocked GCD│ion-selective PVC membrane interface. The Rct equals 245 MΩ and Cdl expressed as Constant Phase Element CPEdl (Y0) is equal to 0.63 (0,8) μS·s(N). In the case of GCD/RuO2/H+-ISM electrode similar value of CPEdl was obtained 0.84 (0,7) μS·s(N), however as expected the value of Rct has drastically decreased (99 kΩ).
Conclusion
The presence of ruthenium oxide, characterized by high redox capacitance and high surface area, improves tested properties of H+-selective electrode in contrast to coated-wire electrode. Also, designed solid-contact electrodes present nearly as excellent properties as glass electrode and along with their great performance parameters such as small size and durability, (the electrodes) may be considered to be applied instead of mentioned conventional glass electrode with inner solution.The RuO2-based solid-contact pH electrodes show a near Nernstian response to hydrogen ions in the pH range from 2 to 12 with no interferences from monovalent cation (such as Na+, K+ and Li+), which often limit the measuring range of H+- selective electrodes.Despite of the fact that the ruthenium oxide is an electronic conductor, examined sensors do not show the redox response.Modified electrodes exhibit remarkable electrical parameters such as low resistance and potential drift along with high capacitance, which results in great potential stability. Obtained parameters’ values unquestionably confirm the beneficial influence of ruthenium oxide on electrodes’ properties.
Notes
References
- 1.Gomes HI, Mayes WM, Rogerson M, Stewart DI, Burke IT (2016) Alkaline residues and the environment: a review of impacts, management practices and opportunities. J Clean Prod 112(4):3571–3582CrossRefGoogle Scholar
- 2.Alam AU, Qin Y, Nambiar S, Yeow JTW, Howlader MMR, Hu NH, Deen MJ (2018) Polymers and organic materials-based pH sensors for healthcare applications. Prog Mater Sci 96:174–216CrossRefGoogle Scholar
- 3.Buck RP, Rondinini S, Covington AK, Baucke FGK, Brett CMA, Camões MF, Milton MJT, Mussini T, Naumann R, Pratt KW, Spitzer P, Wilson GP (2002) Measurement of pH: definition, standards, and procedures. Pure Appl Chem 74(11):2169–2200CrossRefGoogle Scholar
- 4.Haber F, Klemensiewicz Z (1909) Über elektrische Phasengrenzekräfte. Z Phys Chem 67:385–431Google Scholar
- 5.Crespo GA, Desta DG, Macho S, Rius FX (2009) Solid-contact pH-selective electrode using multi-walled carbon nanotubes. Anal Bioanal Chem 395(7):2371–2376CrossRefGoogle Scholar
- 6.Zdrachek E, Bakker E (2019) Potentiometric sensing. Anal Chem 91:2–26CrossRefGoogle Scholar
- 7.Cattrall RW, Freiser H (1971) Coated wire ion-selective electrodes. Anal Chem 43:1905–1906CrossRefGoogle Scholar
- 8.Kisiel A, Michalska A, Maksymiuk K (2018) Rectifying effect for ion-selective electrodes with conducting polymer solid contact. Synth Met 246:246–253CrossRefGoogle Scholar
- 9.Ocana C, Abramova N, Bratov A, Lindfors T, Bobacka J (2018) Calcium-selective electrodes based on photo-cured polyurethane-acrylate membranes covalently attached to methacrylate functionalized poly(3,4-ethylenedioxythiophene) as solid-contact. Talanta 186:279–285CrossRefGoogle Scholar
- 10.Pięk M, Paczosa-Bator B, Piech R (2016) The complex crystal of NaTCNQ–TCNQ supported on different carbon materials as ion-to-electron transducer in all-solid-state sodium-selective electrode. J Electrochem Soc 163:573–579CrossRefGoogle Scholar
- 11.Yin T, Qin W (2013) Applications of nanomaterials in potentiometric sensors. Trends Anal Chem 51:79–86CrossRefGoogle Scholar
- 12.Yin T, Pan D, Qin W (2014) Ion-selective electrodes with colloid-imprinted mesoporous carbon as solid contact. Anal Chem 86(22):11038–11044CrossRefGoogle Scholar
- 13.Paczosa-Bator B (2014) Effects of type of nanosized carbon black on the performance of an all-solid-state potentiometric electrode for nitrate. Microchim Acta 181(9–10):1093–1099CrossRefGoogle Scholar
- 14.Ivanova NM, Podeshvo IV, Goikhman MY, Yakimanskii AV, Mikhelson KN (2013) Potassium-selective solid contact electrodes with poly(amidoacid) Cu(I) complex, electron-ion exchanging resin and different sorts of carbon black in the transducer layer. Sensors Actuators B Chem 186:589–596CrossRefGoogle Scholar
- 15.Zeng XZ, Qin W (2017) A solid-contact potassium-selective electrode with MoO2microspheres as ion-to-electron transducer. Anal Chim Acta 982:72–77CrossRefGoogle Scholar
- 16.Vishnu Prataap RK, Arunachalam R, Pavul Raj R, Mohan S, Péter L (2018) Effect of electrodeposition modes on ruthenium oxide electrodes for supercapacitors. Curr Appl Phys 18:1143–1148CrossRefGoogle Scholar
- 17.Li Q, Zheng S, Xu Y, Xue H, Pang H (2017) Ruthenium based materials as electrode materials for supercapacitors. Chem Eng J 333:505–518CrossRefGoogle Scholar
- 18.Mihell JA, Atkinson JK (1998) Planar thick-film pH electrodes based on ruthenium dioxide hydrate. Sensors Actuators B Chem 48:505–511CrossRefGoogle Scholar
- 19.Maurya DK, Sardarinejad A, Alameh K (2013) High-sensitivity pH sensor employing a sub-micron ruthenium oxide thin-film in conjunction with a thick reference electrode. Sensors Actuators A: Physical 203:300–303CrossRefGoogle Scholar
- 20.Fog A, Buck RP (1984) Electronic semiconducting oxides as pH sensors. Sensors Actuators 5:137–146CrossRefGoogle Scholar
- 21.Fibbioli M, Morf WE, Badertscher M, De Rooij NF, Pretsch E (2000) Potential drifts of solid-contacted ion-selective electrodes due to zero-current ion fluxes through the sensor membrane. Electroanalysis 12:1286–1292CrossRefGoogle Scholar
- 22.Lenar N, Paczosa-Bator B, Piech R, Królicka A (2019) Poly(3-octylthiophene-2,5-diyl) - Nanosized ruthenium dioxide composite material as solid-contact layer in polymer membrane-based K+-selective electrodes. Electrochim Acta. https://doi.org/10.1016/j.electacta.2019.134718 CrossRefGoogle Scholar
- 23.Liu XJ, Peng B, Liu F, Qin Y (2007) Potentiometric liquid membrane pH sensors based on calix[4]-aza-crowns. Sensors Actuators B Chem 125:656–663CrossRefGoogle Scholar
- 24.Kuruoglu D, Canel E, Memon S, Yilmaz M, Kilic E (2003) Hydrogen ion-selective poly(vinyl chloride) membrane electrode based on a calix[4]arene. Anal Sci 19:217–221CrossRefGoogle Scholar
- 25.Hu J, Zou XU, Stein A, Bühlmann P (2014) Ion-selective electrodes with colloid-imprinted mesoporous carbon as solid contact. Anal Chem 86:7111–7118CrossRefGoogle Scholar
- 26.Ye J, Li F, Gan S, Jiang Y, An Q, Zhang Q, Niu L (2015) Using sp 2-c dominant porous carbon sub-micrometer spheres as solid transducers in ion-selective electrodes. Electrochem Commun 50:60–63CrossRefGoogle Scholar
- 27.Michalska A, Hulanicki A, Lewenstam A (1994) All solid-state hydrogen ion-selective electrode based on a conducting poly(pyrro1e) solid contact. Analyst 119:2417–2420CrossRefGoogle Scholar
- 28.Zuaznabar-Gardona JC, Fragoso A (2018) A wide-range solid state potentiometric pH sensor based on poly-dopamine T coated carbon nano-onion electrodes. Sensors Actuators B Chem 273:664–671CrossRefGoogle Scholar
- 29.Lindfors T, Ervelä S, Ivaska A (2003) Polyaniline as pH-sensitive component in plasticized PVC membranes. J Electroanal Chem 560:69–78CrossRefGoogle Scholar
- 30.Zhang X, Ogorevc B, Wang J (2002) Solid-state pH nanoelectrode based on polyaniline thin film electrodeposited onto ion-beam etched carbon fiber. Anal Chim Acta 452:1–10CrossRefGoogle Scholar
- 31.Kaempgen M, Roth S (2006) Transparent and flexible carbon nanotube/polyaniline pH sensors. J Electroanal Chem 586:72–76CrossRefGoogle Scholar
- 32.Zine N, Bausells J, Teixidor F, Viñas C, Masalles C, Samitier J, Errachid A (2006) All-solid-state hydrogen sensing microelectrodes based on novel PPy[3,3′-Co(1,2-C2B9H11)2] as a solid internal contact. Mater Sci Eng C 26:399–404CrossRefGoogle Scholar
- 33.Guzinski M, Jarvis JM, D’Orazio P, Izadyar A, Pendley BD, Lindner E (2017) Solid-contact pH sensor without CO2 interference with a Superhydrophobic PEDOT-C14 as solid contact: the ultimate “water layer” test. Anal Chem 89:8468–8475CrossRefGoogle Scholar
- 34.Paczosa-Bator B, Pięk M, Piech R (2015) Application of nanostructured TCNQ to potentiometric ion-selective K+ and Na+ electrodes. Anal Chem 87(3):1718–1725CrossRefGoogle Scholar
- 35.Li F, Ye J, Zhou M, Gan S, Zhang Q, Han D, Niu L (2012) All-solid-state potassium-selective electrode using graphene as the solid contact. Analyst 137(3):618–623CrossRefGoogle Scholar
- 36.Mousavi Z, Bobacka J, Lewenstam A, Ivaska A (2009) Poly(3,4-ethylenedioxythiophene) (PEDOT) doped with carbon nanotubes as ion-to-electron transducer in polymer membrane-based potassium ion-selective electrodes. J Electroanal Chem 633:246–252CrossRefGoogle Scholar
- 37.Pięk M, Piech R, Paczosa-Bator B (2016) All-solid-state nitrate selective electrode with graphene/tetrathiafulvalene nanocomposite as high redox and double layer capacitance solid contact. Electrochim Acta 210:407–414CrossRefGoogle Scholar
- 38.Paczosa-Bator B (2015) Ion-selective electrodes with superhydrophobic polymer/carbon nanocomposites as solid contact. Carbon 95:879–887CrossRefGoogle Scholar
- 39.Pięk M, Paczosa-Bator B, Smajdor J, Piech R (2018) Molecular organic materials intermediate layers modified with carbon black in potentiometric sensors for chloride determination. Electrochim Acta 283:1753–1762CrossRefGoogle Scholar
- 40.Bobacka J (1999) Potential stability of all-solid-state ion-selective electrodes using conducting polymers as ion-to-electron transducers. Anal Chem 71(21):4932–4937CrossRefGoogle Scholar
Copyright information
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.