Springer Nature is making SARS-CoV-2 and COVID-19 research free. View research | View latest news | Sign up for updates

Port site electrosurgical (diathermy) burns during surgical laparoscopy

  • 133 Accesses

  • 28 Citations

Abstract

Background: Direct and capacitive coupling of diathermy current have been reported as causes of occult injury during surgical laparoscopy.

Methods: In order to determine the incidence of electrosurgical injury adjacent to metal and plastic cannulas, skin biopsies at 19 port sites used for monopolar electrosurgery were analyzed for coagulative necrosis. Prior to surgery the cannulas were randomized to either metal or plastic.

Results: Coagulative necrosis was observed at nine electrosurgery port sites compared to only one control (χ2= 4.872; df= 1; 0.05 > p > 0.02). Plastic cannulas afforded no greater protection from skin burns than metal cannulas.

Conclusions: Burns may be the result of direct or capacitive coupling to metal cannulas or capacitive coupling to the skin edge across plastic cannulas. The potential exists for burns to other tissues also in close proximity to a cannula used for electrosurgery.

This is a preview of subscription content, log in to check access.

Author information

Additional information

Received: 12 August 1996/Accepted 26 November 1996

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Willson, P., van der Walt, J., Moxon, D. et al. Port site electrosurgical (diathermy) burns during surgical laparoscopy. Surg Endosc 11, 653–654 (1997). https://doi.org/10.1007/s004649900414

Download citation

  • Key words: Electrosurgery — Surgical diathermy — Capacitive coupling — Direct coupling — Patient burns