A robotic flexible endoscope with shared autonomy: a study of mockup cholecystectomy

  • Chengzhi Song
  • Xin Ma
  • Xianfeng Xia
  • Philip Wai Yan Chiu
  • Charing Ching Ning Chong
  • Zheng LiEmail author
2019 EAES Oral
  • 58 Downloads

Abstract

Background

Endoscope is the eye of surgeon in minimally invasive surgery (MIS). Prevailing handheld endoscopes are manually steered, which can cause endoscope-instrument fencing. Robotic endoscopes can reduce the fatigue but could not reduce collisions. Handheld endoscopes with a flexible bending tip can reduce the shaft pivoting and collisions. However, its steering is challenging. In this paper, we present a robotic flexible endoscope with auto-tracking function and compare it with the conventional rigid endoscopes.

Methods

A robotic flexible endoscope (RFE) with shared autonomy is developed. The RFE could either track the instruments automatically or be controlled by a foot pedal. A mockup cholecystectomy was designed to evaluate the performance. Five surgeons were invited to perform the mockup cholecystectomy in an abdominal cavity phantom with a manual rigid endoscope (MRE), a robotic rigid endoscope (RRE), and the RFE. Space occupation, time consumption, and questionnaires based on the NASA task load index were adopted to evaluate the performances and compare the three endoscope systems. An ex vivo experiment was conducted to demonstrate the feasibility of using the RFE in a biological tissue environment.

Results

All surgeons completed the mockup cholecystectomy with the RFE independently. Failure occurred in the cases involving the RRE and the MRE. Inside the body cavity, the space occupied when using the RFE is 17.28% and 23.95% (p < 0.05) of that when using the MRE and the RRE, respectively. Outside the body cavity, the space occupied when using the RFE is 14.60% and 15.53% (p < 0.05) of that by using MRE and RRE. Time consumed in the operations with MRE, RRE, and RFE are 28.3 s, 93.2 s and 34.8 s, respectively. Questionnaires reveal that the performance of the RFE is the best among the three endoscope systems.

Conclusions

The RFE provides a wider field of view (FOV) and occupies less space than rigid endoscopes.

Keywords

Minimally invasive surgery Cholecystectomy Robotic flexible endoscope Shared autonomy 

Notes

Acknowledgements

This work is supported by the Hong Kong Research Grants Council (RGC) with Project Numbers (14212316, 14207017 and 24204818) and CUHK-SJTU Joint Research Collaboration Fund.

Compliance with ethical standards

Disclosures

Philip Waiyan Chiu serves as the scientific advisory board member of Aptorum Co Ltd and scientific advisory board member of EndoMASTER Pte Ltd. Chengzhi SONG, Xin MA, Xianfeng XIA, Charing CN CHONG, and Zheng LI have no conflict of interest or financial ties to disclose.

References

  1. 1.
    Fuchs KH (2002) Minimally invasive surgery. Endoscopy 34(02):154–159CrossRefGoogle Scholar
  2. 2.
    Vitiello V, Lee SL, Cundy TP, Yang GZ (2012) Emerging robotic platforms for minimally invasive surgery. IEEE Rev Biomed Eng 6:111–126CrossRefGoogle Scholar
  3. 3.
    Dakin GF, Gagner M (2003) Comparison of laparoscopic skills performance between standard instruments and two surgical robotic systems. Surg Endosc 17:574–579CrossRefGoogle Scholar
  4. 4.
    Freschi C, Ferrari V, Melfi F, Ferrari M, Mosca F, Cuschieri A (2013) Technical review of the da Vinci surgical telemanipulator. Med Robot Comput Assist Surg 9:396–406CrossRefGoogle Scholar
  5. 5.
    Wang Y (1996) Automated endoscope system for optimal positioning. CIM-INT MANUF 12(4):372CrossRefGoogle Scholar
  6. 6.
    Taylor RH, Funda J, Eldridge B, Gomory S, Gruben K, LaRose D, Talamini M, Kavoussi L, Anderson J (1995) A telerobotic assistant for laparoscopic surgery. IEEE Eng Med Biol Mag 14(3):279–288CrossRefGoogle Scholar
  7. 7.
    Gilbert JM (2009) The EndoAssist™ robotic camera holder as an aid to the introduction of laparoscopic colorectal surgery. Ann R Coll Surg Engl 91(5):389–393CrossRefGoogle Scholar
  8. 8.
    Voros S, Haber GP, Menudet JF, Long JA, Cinquin P (2010) ViKY robotic scope holder: initial clinical experience and preliminary results using instrument tracking. IEEE-ASME T MECH 15(6):879–886Google Scholar
  9. 9.
    Taniguchi K, Nishikawa A, Sekimoto M, Yasui M, Takiguchi S, Seki Y, Monden M, Miyazaki F (2006) Design of a novel wearable laparoscope manipulator: SMART (Synthetic Muscle Actuator based Robotic Technology). Int J Comput Assist Radiol Surg 1:213CrossRefGoogle Scholar
  10. 10.
    Mirbagheri A, Farahmand F, Meghdari A, Karimian F (2011) Design and development of an effective low-cost robotic cameraman for laparoscopic surgery: RoboLens. Sci Iran 18(1):105–114CrossRefGoogle Scholar
  11. 11.
    Kristin J, Geiger R, Kraus P, Klenzner T (2015) Assessment of the endoscopic range of motion for head and neck surgery using the SOLOASSIST endoscope holder. Int J Med Robot 11(4):418–423CrossRefGoogle Scholar
  12. 12.
    Wijsman PJ, Broeders IA, Brenkman HJ, Szold A, Forgione A, Schreuder HW, Consten EC, Draaisma WA, Verheijen PM, Ruurda JP, Kaufman Y (2018) First experience with THE AUTOLAP™ SYSTEM: an image-based robotic camera steering device. Surg Endosc 32(5):2560–2566CrossRefGoogle Scholar
  13. 13.
    Mettler L, Ibrahim M, Jonat W (1998) One year of experience working with the aid of a robotic assistant (the voice-controlled optic holder AESOP) in gynecological endoscopic surgery. Hum Reprod (Oxford, England) 13(10):2748–2750CrossRefGoogle Scholar
  14. 14.
    Arezzo A, Ulmer F, Weiss O, Schurr M, Hamad M, Buess GF (2000) Experimental trial on solo surgery for minimally invasive therapy. Surg Endosc 14(10):955–959CrossRefGoogle Scholar
  15. 15.
    Holländer SW, Klingen HJ, Fritz M, Djalali P, Birk D (2014) Robotic camera assistance and its benefit in 1033 traditional laparoscopic procedures: prospective clinical trial using a joystick-guided camera holder. Surg Technol Int 25:19–23PubMedGoogle Scholar
  16. 16.
    Ozaki R, Kumakiri J, Kikuchi I, Kitade M, Matsuoka S, Jinushi M, Kono A, Takeda S (2011) Beneficial use of ENDOCAMELEON a novel optical devise for single port laparoscopic surgery for large ovarian cyst. J Minim Invasive Gynecol 18(6):S113CrossRefGoogle Scholar
  17. 17.
    Ronald C, Rami L, Marc S, Brett C, Marcos M, Todd W (2014) Improving precision and accuracy in laparoscopy using the ENDOEYE FLEX 3D articulating videoscope. https://www.generalsurgerynews.com/Monographs-Whitepapers/Article/04-14/Improving-Precision-and-Accuracy-in-Laparoscopy-Using-the-ENDOEYE-FLEX-3D-Articulating-Videoscope/27350. Accessed 22 Mar 2014
  18. 18.
    Li Z, Oo MZ, Nalam V, Thang VD, Ren H, Kofidis T, Yu H (2016) Design of a novel flexible endoscope—cardioscope. J Mech Robot 8(5):051014CrossRefGoogle Scholar
  19. 19.
    Ko SY, Kim J, Lee WJ, Kwon DS (2007) Compact laparoscopic assistant robot using a bending mechanism. ADV Robot 21(5–6):689–709CrossRefGoogle Scholar
  20. 20.
    Luo RC, Wang J, Chang CK, Perng YW (2014, August) Surgeon’s third hand: an assistive robot endoscopic system with intuitive maneuverability for laparoscopic surgery. In: 5th Proceedings of the IEEE RAS EMBS International Conference Biomedical Robotics and Biomechatronics, pp 138–143Google Scholar
  21. 21.
    Kazanzides P, Chen Z, Deguet A, Fischer GS, Taylor RH, DiMaio SP (2014, May) An open-source research kit for the da Vinci® Surgical System. In: 2014 IEEE International Conference on Robotics and Automation, pp 6434–6439Google Scholar
  22. 22.
    Li Z, Du R (2013) Design and analysis of a bio-inspired wire-driven multi-section flexible robot. IJARS 10(4):209Google Scholar
  23. 23.
    Li Z, Ren H, Chiu PWY, Du R, Yu H (2016) A novel constrained wire-driven flexible mechanism and its kinematic analysis. MMT 95:59–75Google Scholar
  24. 24.
    Ma X, Song C, Chiu WYP, Li Z (2019) Autonomous flexible endoscope for minimally invasive surgery with enhanced safety. IEEE Robot Autom Lett (RA-L) 4(3):2607–2613CrossRefGoogle Scholar
  25. 25.
    Wang Z, Liu Z, Ma Q, Cheng A, Liu YH, Kim S, Deguet A, Reiter A, Kazanzides P, Taylor RH (2017) Vision-based calibration of dual RCM-based robot arms in human-robot collaborative minimally invasive surgery. IEEE Robot Autom Lett (RA-L) 3(2):672–679CrossRefGoogle Scholar
  26. 26.
    O'Donnell CRD, Thomas Eggemeier F (1986) Workload assessment methodology. In: Boff KR, Kaufman L, Thomas JP (eds) Handbook of perception and human performance. Cognitive processes and performance, vol II. Wiley-interscience, New York, p 42-1–42-43Google Scholar
  27. 27.
    Hart SG, Staveland LE (1988) Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Adv Psychol 52:139–183CrossRefGoogle Scholar
  28. 28.
    Yurko YY, Scerbo MW, Prabhu AS, Acker CE, Stefanidis D (2010) Higher mental workload is associated with poorer laparoscopic performance as measured by the NASA-TLX tool. Simul Healthc 5(5):267–271CrossRefGoogle Scholar