The role of tactile feedback in grip force during laparoscopic training tasks

Abstract

Background

Laparoscopic minimally invasive surgery has revolutionized surgical care by reducing trauma to the patient, thereby decreasing the need for medication and shortening recovery times. During open procedures, surgeons can directly feel tissue characteristics. However, in laparoscopic surgery, tactile feedback during grip is attenuated and limited to the resistance felt in the tool handle. Excessive grip force during laparoscopic surgery can lead to tissue damage. Providing additional supplementary tactile feedback may allow subjects to have better control of grip force and identification of tissue characteristics, potentially decreasing the learning curve associated with complex minimally invasive techniques.

Methods

A tactile feedback system has been developed and integrated into a modified laparoscopic grasper that allows forces applied at the grasper tips to be felt by the surgeon’s hands. In this study, 15 subjects (11 novices, 4 experts) were asked to perform single-handed peg transfers using these laparoscopic graspers in three trials (feedback OFF, ON, OFF). Peak and average grip forces (newtons) during each grip event were measured and compared using a Wilcoxon ranked test in which each subject served as his or her own control.

Results

After activating the tactile feedback system, the novice subject population showed significant decreases in grip force (p < 0.003). When the system was deactivated for the third trial, there were significant increases in grip force (p < 0.003). Expert subjects showed no significant improvements with the addition of tactile feedback (p > 0.05 in all cases).

Conclusion

Supplementary tactile feedback helped novice subjects reduce grip force during the laparoscopic training task but did not offer improvements for the four expert subjects. This indicates that tactile feedback may be beneficial for laparoscopic training but has limited long-term use in the nonrobotic setting.

This is a preview of subscription content, log in to check access.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10

References

  1. 1.

    Feussner H, Siewert JR (2001) Reduction of surgical access trauma: reliable advantages. Der Chirug Mar 72:236–244

  2. 2.

    Jacobs JK, Goldstein RE (1997) Laparoscopic adrenalectomy: a new standard of care. Ann Surg 225:495–501

  3. 3.

    Lanfranco AR, Castellano AE, Desai JP, Meyers WC (2004) Robotic surgery: a current perspective. Ann Surg 239:14–21

  4. 4.

    Oehler MK (2009) Robot-assisted surgery in gynaecology. Aust N Z J Obstet Gynaecol 49:124–129

  5. 5.

    Kontarinis DA, Son JS, Peine W, Howe RD (1995) A tactile shape sensing and display system for teleoperated manipulation. Proc IEEE Int Conf Robot Autom 1:641–646

  6. 6.

    Franco ML, King CH, Culjat MO, Lewis CE, Bisley JW, Holmes EC, Grundfest WS, Dutson EP (2009) An integrated pneumatic tactile feedback actuator array for robotic surgery. Int J Med Robot 5:13–19

  7. 7.

    King CH, Culjat MO, Franco ML, Lewis CE, Dutson EP, Grundfest WS, Bisley JW (2009) Tactile feedback induces reduced grasping force in robotic surgery. IEEE Trans Haptics 2:103–110

  8. 8.

    Burdea GC (1996) Force and touch feedback for virtual reality. Wiley, New York, pp 3–4

  9. 9.

    Okamura AM (2009) Haptic feedback in robot-assisted minimally invasive surgery. Curr Opin Urol 19:102–107

  10. 10.

    Cavusoglu MC, Williams W, Tendick F, Sastry SS (2003) Robotics for telesurgery: second-generation Berkeley/UCSF laparoscopic telesurgical workstation and looking towards the future applications. Ind Rob 30:22–29

  11. 11.

    Lee S, Lee J, Choi DS, Kim M, Lee CW (1999) The distributed controller architecture for a masterarm and its application to teleoperation with force feedback. In: Proceedings of the IEEE international conference on robotics and automation, vol 1, pp 213–218

  12. 12.

    Brooks FP, Ouh-Young M, Battert JJ, Kilpatrick PJ (1990) Project GROPE—haptic displays for scientific visualization. Comput Graph 24:177–185

  13. 13.

    Cai Y, Ishii M, Sato M. (1996) A human interface device for CAVE size virtual workspace. In: IEEE international conference on systems, man, and cybernetics, Beijing, pp 2084–2089

  14. 14.

    Baumann R, Clavel R. (1998) Haptic interface for virtual reality-based minimally invasive surgery simulation. In: Proceedings of the 1998 IEEE international conference on robotics and automation, Leuven, pp 381–386

  15. 15.

    King CH, Higa AT, Culjat MO, Han SH, Bisley JW, Carman GP, Dutson EP, Grundfest WS (2007) A pneumatic haptic feedback actuator array for robotic surgery or simulation. In: Proceedings of medicine meets virtual reality 15, Long Beach, pp 217–222

  16. 16.

    Bethea BT, Okamura AM, Kitagawa M, Fitton TP, Cattaneo SM, Gott VL, Baumgartner WA, Yuh DD (2004) Application of haptic feedback to robotic surgery. J Laparoendosc Adv Surg Tech 14:191–195

  17. 17.

    Bouzit M, Popescu G, Burdea G, Boian R (2002) The Rutgers Master II-ND force feedback glove. In: 10th symposium on haptic interfaces for virtual environment and teleoperator systems, Orlando, p 145

  18. 18.

    Nakamura M, Jones L (2003) An actuator for the tactile vest: a torso-based haptic device. In: Proceedings of the IEEE symposium on haptic interfaces for virtual environment and teleoperator systems, Los Angeles

  19. 19.

    Fan RE, Culjat MO, King CH, Franco ML, Boryk R, Bisley JW, Dutson EP, Grundfest WS (2008) A haptic feedback system for lower-limb prostheses. IEEE Trans Neural Syst Rehabil Eng 16:270–277

  20. 20.

    Wottawa C, Fan RE, Lewis CE, Jordan B, Culjat MO, Grundfest WS, Dutson EP (2009) Laparoscopic grasper with an integrated tactile feedback system. In: Proceedings of 2009 ICME/IEEE international conference on complex medical engineering, Tempe, pp 1–5

  21. 21.

    Shinohara M, Shimizu Y, Mochizuki A (1998) Three-dimensional tactile display for the blind. IEEE Trans Neural Syst Rehabil Eng 6:249–256

  22. 22.

    Kaczmarek KA, Webster JG, Bach-y-Rita P, Tompkins WJ (1991) Electrotactile and vibrotactile displays for sensory substitution systems. IEEE Trans Biomed Eng 38:1–16

  23. 23.

    Summers IR, Chanter CM (2002) A broadband tactile array on the fingertip. J Acoust Soc Am 112:2118–2126

  24. 24.

    Haga Y, Mizushima M, Matsunaga T, Esashi M (2005) Medical and welfare applications of shape memory alloy microcoil actuators. Smart Mater Struct 14:S266–S272

  25. 25.

    Howe RD, Kontarinis DA, Peine WJ (1995) Shape memory alloy actuator controller design for tactile displays. In: Proceedings of the 34th IEEE conference on decision and control, vol 4, New Orleans, pp 3540–3544

  26. 26.

    Taylor PM, Hosseini-Sianaki A, Varley CJ (1996) An electrorheological fluid-based tactile array for virtual environments. In: IEEE international conference on robotics and automation, vol 1, pp 18–23

  27. 27.

    Bicchi A, Scilingo EP, Sgambelluri N, De Rossi D (2002) Haptic interfaces based on magneto-rheological fluids. In: Proceedings of eurohaptics, Edingburgh, UK, pp 6–11

  28. 28.

    Cohn MB, Lam M, Fearing RS (1992) Tactile feedback for teleoperation. In: Proceedings of the SPIE Telemanipulator Technology 1833, Boston, pp 240–255

  29. 29.

    Asamura N, Yokoyama N, Shinoda H (1998) Selectively stimulating skin receptors for tactile display. IEEE Comput Graph Appl 18:32–37

  30. 30.

    Sato K, Igarashi E, Kimura M (1991) Development of nonconstrained master arm with tactile feedback device. In: ICAR, fifth international conference on advanced robotics: robots in unstructured environments, Pisa, pp 334–338

  31. 31.

    Moy G, Wagner C, Fearing RS (2000) A compliant tactile display for teletaction. In: ICRA 2000 IEEE international conference on robotic automation, Piscataway, pp 3409–3415

  32. 32.

    Rogers CH (1970) Choice of stimulator frequency for tactile arrays. IEEE Trans Man Mach Syst 11:5–11

  33. 33.

    Phillips JR, Johnson KO (1985) Neural mechanisms of scanned and stationary touch. J Acoust Soc Am 77:220–224

  34. 34.

    King CH, Culjat MO, Franco ML, Bisley JW, Dutson E, Grundfest WS (2008) Optimization of pneumatic balloon tactile display for robotic surgery based on human perception. IEEE Trans Biomed Eng 55:2593–2600

  35. 35.

    King CH, Franco ML, Culjat MO, Higa AT, Bisley JW, Dutson E, Grundfest WS (2008) Fabrication and characterization of a balloon actuator array for haptic feedback in robotic surgery. ASME J Med Devices 2:041006-1–041006-7

  36. 36.

    Wottawa C, Fan RE, Bisley JW, Duston EP, Culjat MO, Grundfest WS (2011) Applications of tactile feedback in medicine. Stud Health Technol Inform 163:703–709

  37. 37.

    Culjat MO, King CH, Franco ML, Lewis CE, Bisley JW, Dutson EP, Grundfest WS (2008) A tactile feedback system for robotic surgery. Proc IEEE Eng Med Biol Soc 1:1930–1934

  38. 38.

    Matern U, Waller P (1999) Instruments for minimally invasive surgery: principles of ergonomic handles. Surg Endosc 13:174–182

Download references

Acknowledgments

The authors are grateful for the funding provided by the Telemedicine and Advanced Technology Research Center (TATRC)/Department of Defense under award number W81XWH-07-1-0672, and the funding from the National Science Foundation under award number CBET-0730213. The authors thank Dr. Shane White and Dalene Sederstrom for use of the Instron, as well as Brett Jordan and Will Pannel for assistance with manufacture of the specialized tool handles.

Disclosures

Christopher R. Wottawa, Jeremiah R. Cohen, Richard E. Fan, James W. Bisley, Martin O. Culjat, Warren S. Grundfest, and Erik P. Dutson have no conflicts of interest or financial ties to disclose.

Author information

Correspondence to Christopher R. Wottawa.

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Wottawa, C.R., Cohen, J.R., Fan, R.E. et al. The role of tactile feedback in grip force during laparoscopic training tasks. Surg Endosc 27, 1111–1118 (2013). https://doi.org/10.1007/s00464-012-2612-x

Download citation

Keywords

  • Grip force
  • Laparoscopic surgery training
  • Minimally invasive surgery
  • Tactile feedback
  • Pneumatic actuator