Investigation of ultra-low insertion speeds in an inelastic artificial cochlear model using custom-made cochlear implant electrodes

  • Silke Hügl
  • Katharina Rülander
  • Thomas Lenarz
  • Omid Majdani
  • Thomas S. Rau
Otology
  • 87 Downloads

Abstract

Purpose

Latest research on cochlear implantations focuses on hearing preservation during insertion of the implant’s electrode array by reducing insertion trauma. One parameter which may influence trauma is insertion speed. The objective of this study was to extend the range of examined insertion speeds to include ultra-low velocities, being lower than manually feasible, and investigate whether these reduce insertion forces.

Methods

24 custom-made cochlear implant test samples were fabricated and inserted into an artificial scala tympani model using 12 different insertion speeds while measuring the resulting insertion forces. Three commercially available slim straight electrode carriers were inserted using the same setup to analyze whether the results are comparable.

Results

Insertions of the test samples using high insertion speeds (2.0/2.8 mm/s) showed significantly higher insertion forces than insertions done with low insertion speeds (0.2 mm/s) or ultra-low insertion speeds (< 0.1 mm/s). The insertions with commercial slim straight electrode arrays showed significantly reduced insertion forces when using a low insertion speed as well.

Conclusions

Slow insertions showed significantly reduced insertion forces. Insertion speeds which are lower than manually feasible showed even lower insertion forces.

Keywords

Electrode carrier Automated insertion Insertion tool Insertion force Cochlea implantation 

Notes

Acknowledgements

The authors would like to thank the German Research Foundation (DFG) for funding this research (support code: MA 4038/9-1). Furthermore, the authors thank Samuel Müller (Institute of Mechatronic Systems, Leibniz University Hannover) for fabrication of the casting mold.

Compliance with ethical standards

Conflict of interest

Thomas Lenarz is a consultant for Cochlear Ltd. Silke Hügl, Katharina Rülander, Omid Majdani and Thomas S. Rau declare that they have no conflict of interest.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

References

  1. 1.
    Hoskison E, Mitchell S, Coulson C (2017) Systematic review: radiological and histological evidence of cochlear implant insertion trauma in adult patients. Cochlear Impl Int 18:192–197CrossRefGoogle Scholar
  2. 2.
    Roland PS, Wright CG (2006) Surgical aspects of cochlear implantation: mechanisms of insertional trauma. Cochlear Brainstem Impl 64:11–30Google Scholar
  3. 3.
    Mirsalehi M, Rau TS, Harbach L et al (2017) Insertion forces and intracochlear trauma in temporal bone specimens implanted with a straight atraumatic electrode array. Eur Arch Otorhinolaryngol 5:2131–2140CrossRefGoogle Scholar
  4. 4.
    Avci E, Nauwelaers T, Hamacher V, Kral A (2017) Three-dimensional force profile during cochlear implantation depends on individual geometry and insertion trauma. Ear Hear 38:e168–e179CrossRefGoogle Scholar
  5. 5.
    Zhang J, Bhattacharyya S, Simaan N (2009) Model and parameter identification of friction during robotic insertion of cochlear-implant electrode arrays. IEEE international conference on robotics and automation ICRA, pp 3859–3864Google Scholar
  6. 6.
    Kontorinis G, Lenarz T, Stöver T, Paasche G (2011) Impact of the insertion speed of cochlear implant electrodes on the insertion forces. Otol Neurotol 32:565–570CrossRefGoogle Scholar
  7. 7.
    Rajan GP, Kontorinis G, Kuthubutheen J (2013) The effects of insertion speed on inner ear function during cochlear implantation: a comparison study. Audiol Neurootol 18:17–22CrossRefGoogle Scholar
  8. 8.
    Majdani O, Schurzig D, Hussong A et al (2010) Force measurement of insertion of cochlear implant electrode arrays in vitro: comparison of surgeon to automated insertion tool. Acta Otolaryngol 130:31–36CrossRefGoogle Scholar
  9. 9.
    Rau TS, Hussong A, Leinung M, Lenarz T, Majdani O (2010) Automated insertion of preformed cochlear implant electrodes: evaluation of curling behaviour and insertion forces on an artificial cochlear model. Int J Comput Assist Radiol Surg 5:173–181CrossRefGoogle Scholar
  10. 10.
    Berman AD, Ducker WA, Israelachvili JN (1996) Origin and characterization of different stick-slip friction mechanisms. Langmuir 12(19):4559–4563CrossRefGoogle Scholar
  11. 11.
    Gantz BJ, Turner C, Gfeller KE, Lowder MW (2005) Preservation of hearing in cochlear implant surgery: advantages of combined electrical and acoustical speech processing. Laryngoscope 115:796–802CrossRefGoogle Scholar
  12. 12.
    Büchner A, Schüssler M, Battmer RD, Stöver T, Lesinski-Schiedat A, Lenarz T (2009) Impact of low-frequency hearing. Audiol Neurootol 14(suppl 1):8–13CrossRefGoogle Scholar
  13. 13.
    Von Ilberg CA, Baumann U, Kiefer J, Tillein J, Adunka OF (2011) Electric-acoustic stimulation of the auditory system: a review of the first decade. Audiol Neurootol 16(suppl. 2):1–30CrossRefGoogle Scholar
  14. 14.
    Roland PS, Gstöttner W, Adunka O (2005) Method for hearing preservation in cochlear implant surgery. Oper Tech Otolayngol Head Neck Surg 16(2):93–100CrossRefGoogle Scholar
  15. 15.
    Hüttenbrink KB, Zahnert T, Jolly C, Hofmann G (2002) Movements of cochlear implant electrodes inside the cochlea during insertion: an X-ray microscopy study. Otol Neurotol 23:187–191CrossRefGoogle Scholar
  16. 16.
    Cohen LT, Xu J, Xu SA, Clark GM (1996) Improved and simplified methods for specifying positions of the electrode bands of a cochlear implant array. Am J Otol 17:859–865PubMedGoogle Scholar
  17. 17.
    Todd CA, Naghdy F, Svehla MJ (2007) Force application during cochlear implant insertion: an analysis for improvement of surgeon technique. IEEE Trans Biomed Eng 54:1247–1255CrossRefGoogle Scholar
  18. 18.
    Kontorinis G, Paasche G, Lenarz T, Stöver T (2011) The effect of different lubricants on cochlear implant electrode insertion forces. Otol Neurotol 32:1050–1056CrossRefGoogle Scholar
  19. 19.
    Kobler JP, Dhanasingh A, Kiran R, Jolly C, Ortmaier T (2015) Cochlear dummy electrodes for insertion training and research purposes: fabrication, mechanical characterization, and experimental validation. Bio Med Res Int 2015:574209.  https://doi.org/10.1155/2015/574209 CrossRefGoogle Scholar
  20. 20.
    Roland JT (2005) A model for cochlear implant electrode insertion and force evaluation: results with a new electrode design and insertion technique. Laryngoscope 115:1325–1339CrossRefGoogle Scholar
  21. 21.
    Radeloff A, Unkelbach MH, Mack MG et al (2009) A coated electrode carrier for cochlear implantation reduces insertion forces. Laryngoscope 119:959–963CrossRefGoogle Scholar
  22. 22.
    Helbig S, Settevendemie C, Mack M, Baumann U, Helbig M, Stöver T (2011) Evaluation of an electrode prototype for atraumatic cochlear implantation in hearing preservation candidates: preliminary results from a temporal bone study. Otol Neurotol 32:419–423CrossRefGoogle Scholar
  23. 23.
    Miroir M, Nguyen Y, Kazmitcheff G, Ferrary E, Sterkers O, Grayeli AB (2012) Friction force measurement during cochlear implant insertion: application to a force-controlled insertion tool design. Otol Neurotol 33:1092–1100PubMedGoogle Scholar
  24. 24.
    Nguyen Y, Miroir M, Kazmitcheff G et al (2012) Cochlear implant insertion forces in microdissected human cochlea to evaluate a prototype array. Audiol Neurootol 17:290–298CrossRefGoogle Scholar
  25. 25.
    Nguyen Y, Kazmitcheff G, De Seta D, Miroir M, Ferrary E, Sterkers O (2014) Definition of metrics to evaluate cochlear array insertion forces performed with forceps, insertion tool, or motorized tool in temporal bone specimens. Bio Med Res Int 2014:532570.  https://doi.org/10.1155/2014/532570 CrossRefGoogle Scholar
  26. 26.
    Todt I, Mittmann P, Ernst A (2014) Intracochlear fluid pressure changes related to the insertional speed of a CI electrode. Bio Med Res Int 2014:507241.  https://doi.org/10.1155/2014/507241 CrossRefGoogle Scholar
  27. 27.
    Todt I, Ernst A, Mittmann P (2016) Effects of different insertion techniques of a cochlear implant electrode on the intracochlear pressure. Audiol Neurootol 21:30–37CrossRefGoogle Scholar
  28. 28.
    De Seta D, Torres R, Russo FY et al (2017) Damage to inner ear structure during cochlear implantation: correlation between insertion force and radio-histological findings in temporal bone specimens. Hear Res 344:90–97CrossRefGoogle Scholar
  29. 29.
    Schurzig D, Webster RJ III, Dietrich MS, Labadie RF (2010) Force of cochlear implant electrode insertion performed by a robotic insertion tool: comparison of traditional versus advance off-stylet techniques. Otol Neurotol 31:1207–1210CrossRefGoogle Scholar
  30. 30.
    Briggs RJ, Tykocinski M, Lazsig R et al (2011) Development and evaluation of the modiolar research array—multi-centre collaborative study in human temporal bones. Cochlear Impl Int 12:129–139CrossRefGoogle Scholar
  31. 31.
    Hussong A, Rau TS, Ortmaier T, Heimann B, Lenarz T, Majdani O (2010) An automated insertion tool for cochlear implants: another step towards atraumatic cochlear implant surgery. Int J Comput Assist Radiol Surg 5:163–171CrossRefGoogle Scholar
  32. 32.
    Balster S, Wenzel GI, Warnecke A, Steffens M, Rettenmaier A, Zhang K, Lenarz T, Reuter G (2013) Optical cochlear implant: evaluation of insertion forces of optical fibres in a cochlear model and of traumata in human temporal bones. Biomed Tech (Berl) 59(1):19–28Google Scholar
  33. 33.
    Nada I, Abdelhamid AN, Negm A (2017) The utilization of round window membrane surface tension in facilitating slim electrodes insertion during cochlear implantation. Eur Arch Otorhinolaryngol 274(9):3327–3334CrossRefGoogle Scholar
  34. 34.
    Ishii T, Takayama M, Takahashi Y (1995) Mechanical properties of human round window, basilar and Reissner’s membranes. Acta Otolaryngol 115(sup519):78–82CrossRefGoogle Scholar
  35. 35.
    Schuster D, Kratchman LB, Labadie RF (2015) Characterization of intracochlear rupture forces in fresh human cadaveric cochleae. Otol Neurotol 36(4):657CrossRefGoogle Scholar
  36. 36.
    Kratchman LB, Schuster D, Dietrich MS, Labadie RF (2016) Force perception thresholds in Cochlear implantation surgery. Audiol Neurootol 21(4):244–249CrossRefGoogle Scholar
  37. 37.
    Kesler K, Dillon NP, Fichera L, Labadie RF (2017) Human kinematics of cochlear implant surgery: an investigation of insertion micro-motions and speed limitations. Otolaryngol Head Neck Surg 157(3):493–498CrossRefGoogle Scholar
  38. 38.
    Rau TS, Hügl S, Salcher R, Lenarz T, Majdani O (2017) Hydraulisch automatisierte Elektrodeninsertion. Presented at the 16th annual meeting of the German Society for Computer and Robot assisted Surgery (CURAC), Hannover, Germany, October 5–7Google Scholar

Copyright information

© Springer-Verlag GmbH Germany, part of Springer Nature 2018

Authors and Affiliations

  1. 1.Department of Otolaryngology and Cluster of Excellence “Hearing4all” EXC 1077/1Hannover Medical School (MHH)HannoverGermany

Personalised recommendations