Cardiac implanted electronic devices and MRI safety in 2018—the state of play

  • Ryan Mark Shulman
  • Ben Hunt
Cardiac

Abstract

Traditionally, the presence of cardiac implanted electronic devices (CIEDs) was a contra-indication to magnetic resonance (MR) imaging. Professional groups from around the world are releasing updated guidelines for the imaging of MR-conditional and legacy CIEDs, reflecting increasing evidence that this can be performed safely when strict protocols are followed.

Key Points

The presence of a pacemaker or automatic implanted cardioverter defibrillator is no longer an absolute contraindication to magnetic resonance imaging.

Strict protocols enable diagnostic quality images to be obtained with minimal risk.

Close collaboration among radiologists, cardiologists and device manufacturer representatives is required.

Keywords

Artificial cardiac pacemaker Imaging, magnetic resonance Equipment safety Cardiac imaging techniques 

Abbreviations

AICD

Automatic implanted cardioverter defibrillator

B1+RMS

Derived from the root mean square [RMS] of the flip angle [B1+]

CIED

Cardiac-implanted electronic device

MR

Magnetic resonance

MRI

Magnetic resonance imaging

SAR

Specific absorption rate

1.5T

1.5 Tesla

3T

3 Tesla

Introduction

In 2018, cardiac implanted electronic devices (CIEDs) such as pacemakers and automatic implanted cardioverter defibrillators (AICDs) are no longer an absolute contraindication to magnetic resonance imaging (MRI) in many clinical scenarios. High-quality diagnostic images can now be obtained safely under strict conditions.

The German Roentgen Society [1], British Heart Rhythm Society [2], Heart Rhythm Society [3] and European Society of Cardiology [4], in collaboration with other professional groups, each provide guidelines for magnetic resonance (MR) imaging in the setting of conditional and non-conditional CIEDs. They recommend close cooperation between cardiologists and radiologists.

Safety concerns

Potential safety concerns exist in this setting, which include Reed switch activation, induced currents, heating, torque, antenna effects, device re-programming, ‘power on reset’ and inappropriate therapies [1, 3]. Deaths have occurred in cases of inadvertent MRI of patients with CIEDs, where tailored protocols were not instituted [5]. However, a recent systematic review identified only 19 clinically relevant adverse events from 6237 MRI studies performed under controlled conditions with no fatalities [6].

MR-conditional status

Considerable research and development have resulted in CIEDs that exhibit minimal response to the static magnetic, gradient and RF fields present during MRI. MR-conditional status is granted to leads and generators in specific combinations at specific field strengths. Combinations of components from different vendors (even if individually meeting appropriate criteria) are not granted MR-conditional status.

Conditions for MRI vary between devices and their manufacturers. These should be reviewed on a case-by-case basis and usually include time since device insertion, region being scanned, specific device programming, residual battery power and MRI system parameters and acquired sequences.

MRI in the presence of MR-conditional CIEDs

Institutional protocols should be established that include the CIED managing physician’s determination of MR-conditional status in each case. The manufacturer and model of the generator and leads should be documented as well as the insertion date. A chest radiograph can help to determine the presence of epicardial, abandoned or fractured leads, which can render an otherwise MR-conditional system non-conditional. Prior to scanning, the imaging service should coordinate with device manufacturer representatives to arrange for appropriate CIED reprogramming [7].

If the CIED and MRI system meet MR-conditional requirements, an imaging protocol should be designed that ensures the estimated energy absorbed by the device does not breach mandated thresholds for each sequence. The device should be returned to its standard settings at the completion of the study, clear of the scanner magnetic field.

Advanced Cardiac Life Support-trained staff should be available for the duration of the study from the time the CIED is reprogrammed until reassessed and declared stable to return to unmonitored status. ECG, blood pressure and pulse-oximetry monitoring should occur over this period [1, 3, 8].

MRI can be reasonably performed within the post-insertion exemption period (technically outside the MR-conditional criteria), where the cardiology/radiology consensus concludes that benefit outweighs the risk [9]. In such cases, e.g. with AICD and pacemakers, scanning with the arms by the paient’s side may be useful.

Specific absorption rate (SAR) and B1+RMS

SAR is the traditional metric used for estimating whole-body energy deposition during an MRI study and is still used widely for scanning thresholds. However, it has been found to be variable and potentially unreliable when used to guide imaging protocols in the setting of in-situ CIEDs [10, 11]. B1+RMS (derived from the root mean square [RMS] of the flip angle [B1+]) is reportedly a more accurate estimation of energy deposition. It is increasingly being used to define thresholds for devices at 3T. Techniques used to reduce SAR will also usually reduce B1+RMS (Table 1 [12]).
Table 1.

Techniques to reduce B1+RMS [12] (with permission)

Increased RF pulse duration

Use of “low SAR mode”

Increased TR without reducing slice number

Reduced slices for set TR

Reduced ETL

Reduced refocusing angle (FSE sequences)

Reduced flip angle (GRE sequences)

Use of GRE sequences in place of SE or FSE sequences

Radiofrequency (RF); specific absorption rate (SAR); repetition time (TR); echo train length (ETL); fast spin echo (FSE); gradient recalled echo (GRE); spin echo (SE)

MRI in the presence of non-conditional CIEDs

Imaging with legacy CIEDs in situ is inconsistent between institutions, partly because of limited experience as well as uncertainty related to legal ramifications of complications in the setting of ‘off-label’ use. Risks posed by imaging patients with AICD are considered higher than for other CIEDs as most have underlying structural heart disease and more complex interactions between the device and MRI components. Registry data [8] and recent large series [13, 14] have provided evidence that in appropriate circumstances, MRI (including cardiac imaging [15]) can be reasonably performed at 1.5T in patients with non-conditional CIEDs. Strict adherence to an institutional protocol that documents the risk-benefit analysis, pre- and post-imaging CIED settings and imaging sequence design is essential.

Pacemaker dependence should be identified and documented: In those patients identified as dependent, the device should be programmed to asynchronous pacing (VOO/DOO) and output set to maximum to account for possible threshold changes [11]. In those not dependent, devices should be programmed to a non-pacing mode (ODO/OVO) or inhibited mode (VVI/DDI)[1, 2, 3, 13]. For patients with an AICD, all tachycardia detection and therapies should be programmed off [1, 2, 3].

Despite small series with evidence to the contrary, [16, 17] the presence of fractured, abandoned or epicardial leads (not including retained temporary surgical pacing leads) remains a contraindication to MRI scanning. MRI of non-conditional pacemakers implanted before 1998 and AICDs before 2000 is not recommended [18].

Device interrogation clear of the magnetic field should be performed and results documented before returning the patient to their original device settings. Follow-up should occur at a time predetermined by the electrophysiologist or earlier based on their assessment of any adverse incidents or device parameter changes during or after the MRI [1, 2, 3].

Imaging at 3T

Many new MR-conditional devices allow imaging at 3T with high-quality diagnostic imaging achievable even with the device within the imaging field [19]. An MRI system capable of displaying B1+RMS information is a prerequisite for an increasing number of CIEDs at this field strength.

Only small series [20, 21] have demonstrated successful and safe imaging at high field strengths in patients with non-conditional CIEDs. Most were studies imaging the brain and none were performed of the thorax or heart. No clear guidance on the MRI of non-conditional CIEDs at 3T is available at this time.

Summary

Despite opinions to the contrary [22], evidence demonstrates that when strict institutional protocols are adhered to, MRI in the presence of a CIED can be performed with low risk. This requires collaboration among cardiology, radiology and device manufacturers.

Notes

Funding

The authors state that this work has not received any funding.

Compliance with ethical standards

Guarantor

The scientific guarantor of this publication is Ryan Shulman.

Conflict of interest

The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.

Statistics and biometry

No complex statistical methods were necessary for this paper.

Informed consent

Written informed consent was not required.

Ethical approval

Institutional Review Board approval was not required.

Methodology

Not applicable. This manuscript does not detail original research.

References

  1. 1.
    Sommer T, Bauer W, Fischbach K et al (2017) MR imaging in patients with cardiac pacemakers and implantable cardioverter defibrillators. Rofo 189:204–217CrossRefGoogle Scholar
  2. 2.
    Lowe MD, Plummer CJ, Manisty CH, Linker NJ, British Heart Rhythm Society (2015) Safe use of MRI in people with cardiac implantable electronic devices. Heart 101:1950–1953CrossRefGoogle Scholar
  3. 3.
    Indik JH, Gimbel JR, Abe H et al (2017) 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm 14:e97–e153CrossRefGoogle Scholar
  4. 4.
    Brignole M, Auricchio A, Baron-Esquivias G et al (2013) 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Eur Heart J 34:2281–2329CrossRefGoogle Scholar
  5. 5.
    Irnich W, Irnich B, Bartsch C, Stertmann WA, Gufler H, Weiler G (2005) Do we need pacemakers resistant to magnetic resonance imaging? Europace 7:353–365CrossRefGoogle Scholar
  6. 6.
    Di Leo G, Zanardo M, Schiaffino S, Secchi F, Sardanelli F (2017) Safety of MRI in patients with conditional or non-conditional pacemaker or other implantable cardiac electronic devices (ICED): a systematic review. Paper presented at: 103rd Scientific Assembly and Annual Meeting RSNA; 2017 Nov 26 - Dec 1; Chicago, IL, USAGoogle Scholar
  7. 7.
    Expert Panel on MR Safety, Kanal E, Barkovich AJ et al (2013) ACR guidance document on MR safe practices: 2013. J Magn Reson Imaging 37:501–530CrossRefGoogle Scholar
  8. 8.
    Russo RJ, Costa HS, Silva PD et al (2017) Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med 376:755–764CrossRefGoogle Scholar
  9. 9.
    Friedman HL, Acker N, Dalzell C et al (2013) Magnetic resonance imaging in patients with recently implanted pacemakers. Pacing Clin Electrophysiol 36:1090–1095CrossRefGoogle Scholar
  10. 10.
    Baker KB, Tkach JA, Nyenhuis JA et al (2004) Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating. J Magn Reson Imaging 20:315–320CrossRefGoogle Scholar
  11. 11.
    Gimbel JR, Bailey SM, Tchou PJ, Ruggieri PM, Wilkoff BL (2005) Strategies for the safe magnetic resonance imaging of pacemaker-dependent patients. Pacing Clin Electrophysiol 28:1041–1046CrossRefGoogle Scholar
  12. 12.
    Faulkner W (2016) New MRI safety labels & devices. Signals 5:1Google Scholar
  13. 13.
    Nazarian S, Hansford R, Rahsepar AA et al (2017) Safety of magnetic resonance imaging in patients with cardiac devices. N Engl J Med 377:2555–2564CrossRefGoogle Scholar
  14. 14.
    Mason S, Osborn JS, Dhar R et al (2017) Real world MRI experience with nonconditional and conditional cardiac rhythm devices after MagnaSafe. J Cardiovasc Electrophysiol 28:1468–1474CrossRefGoogle Scholar
  15. 15.
    Do DH, Eyvazian V, Bayoneta AJ et al (2018) Cardiac magnetic resonance imaging using wideband sequences in patients with nonconditional cardiac implanted electronic devices. Heart Rhythm 15:218–225CrossRefGoogle Scholar
  16. 16.
    Hwang YM, Kim J, Lee JH et al (2016) Cardiac implantable electronic device safety during magnetic resonance imaging. Korean Circ J 46:804–810CrossRefGoogle Scholar
  17. 17.
    Higgins JV, Gard JJ, Sheldon SH et al (2014) Safety and outcomes of magnetic resonance imaging in patients with abandoned pacemaker and defibrillator leads. Pacing Clin Electrophysiol 37:1284–1290CrossRefGoogle Scholar
  18. 18.
    Miller JD, Nazarian S, Halperin HR (2016) Implantable electronic cardiac devices and compatibility with magnetic resonance imaging. J Am Coll Cardiol 68:1590–1598CrossRefGoogle Scholar
  19. 19.
    Shulman RM, Moodley M, Vaishnav M, Wright J (2018) Cardiac magnetic resonance with conditional pacemaker at three Tesla (3T) field strength. Cardiovascular Imaging Asia 2:28–30CrossRefGoogle Scholar
  20. 20.
    Naehle CP, Meyer C, Thomas D et al (2008) Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology 249:991–1001CrossRefGoogle Scholar
  21. 21.
    Gimbel JR (2008) Magnetic resonance imaging of implantable cardiac rhythm devices at 3.0 tesla. Pacing Clin Electrophysiol 31:795–801CrossRefGoogle Scholar
  22. 22.
    Delfino JG, Viohl I, Woods TO (2017) Risks of MRI in patients with a pacemaker or defibrillator. N Engl J Med 376:2495CrossRefGoogle Scholar

Copyright information

© European Society of Radiology 2018

Authors and Affiliations

  1. 1.Queensland X-RayGold Coast Private HospitalSouthportAustralia
  2. 2.Department of Medical ImagingGold Coast University HospitalSouthportAustralia
  3. 3.Griffith UniversitySouthportAustralia
  4. 4.Department of Cardiology and ElectrophysiologyPindara Private HospitalBenowaAustralia

Personalised recommendations