, Volume 24, Issue 5, pp 1112–1117 | Cite as

Ultrahigh field magnetic resonance and colour Doppler real-time fusion imaging of the orbit – a hybrid tool for assessment of choroidal melanoma

  • Uwe WalterEmail author
  • Thoralf Niendorf
  • Andreas Graessl
  • Jan Rieger
  • Paul-Christian Krüger
  • Sönke Langner
  • Rudolf F. Guthoff
  • Oliver Stachs
Head and Neck

Abstract

Objectives

A combination of magnetic resonance images with real-time high-resolution ultrasound known as fusion imaging may improve ophthalmologic examination. This study was undertaken to evaluate the feasibility of orbital high-field magnetic resonance and real-time colour Doppler ultrasound image fusion and navigation.

Methods

This case study, performed between April and June 2013, included one healthy man (age, 47 years) and two patients (one woman, 57 years; one man, 67 years) with choroidal melanomas. All cases underwent 7.0-T magnetic resonance imaging using a custom-made ocular imaging surface coil. The Digital Imaging and Communications in Medicine volume data set was then loaded into the ultrasound system for manual registration of the live ultrasound image and fusion imaging examination.

Results

Data registration, matching and then volume navigation were feasible in all cases. Fusion imaging provided real-time imaging capabilities and high tissue contrast of choroidal tumour and optic nerve. It also allowed adding a real-time colour Doppler signal on magnetic resonance images for assessment of vasculature of tumour and retrobulbar structures.

Conclusions

The combination of orbital high-field magnetic resonance and colour Doppler ultrasound image fusion and navigation is feasible. Multimodal fusion imaging promises to foster assessment and monitoring of choroidal melanoma and optic nerve disorders.

Key points

Orbital magnetic resonance and colour Doppler ultrasound real-time fusion imaging is feasible

Fusion imaging combines the spatial and temporal resolution advantages of each modality

Magnetic resonance and ultrasound fusion imaging improves assessment of choroidal melanoma vascularisation

Keywords

Magnetic resonance imaging Colour Doppler ultrasound Orbital diseases Uveal melanoma Optic nerve 

Notes

Acknowledgement

We thank Esaote S.p.A., Genova, Italy, for providing the ultrasound and fusion imaging system.

The scientific guarantor of this publication is Prof. Oliver Stachs, head of the Ophthalmic Imaging Research Section, Department of Ophthalmology, University of Rostock, Germany. The authors of this manuscript declare relationships with the following companies: Thoralf Niendorf and Jan Rieger are affiliated to MRI.TOOLS GmbH, a spin-off enterprise from the Max-Delbrueck-Center for Molecular Medicine (Berlin, Germany). MRI.TOOLS GmbH had no role in the design and conduct of the study; collection, management, analysis or interpretation of the data; preparation, review or approval of the manuscript; or the decision to submit the manuscript for publication. Esaote S.p.A., Genova, Italy, provided the ultrasound and fusion imaging system. Esaote S.p.A. had no role in the design and conduct of the study; collection, management, analysis or interpretation of the data; preparation, review or approval of the manuscript; or the decision to submit the manuscript for publication. No complex statistical methods were necessary for this paper. Institutional review board approval was obtained. Written informed consent was obtained from all subjects (patients) in this study. Methodology: prospective, observational, performed at one institution.

Supplementary material

Video 1

Real-time 7.0-T MRI and colour Doppler ultrasonography (CDUS) fusion imaging of eye and optic nerve in a healthy volunteer. For denomination of structures please see the figure caption of Fig. 2. Note that the image contrast in the MRI plane shown is lower than in the original MRI data because the MRI plane was reconstructed by the ultrasound system from the imported MRI DICOM 3D data set. The video (file format, .wmv) shows the moving fusion image with gradually changing overlay of the MRI image over the CDUS image in the upper left panel. The blood flow in the central retinal artery (red) and vein (blue), the posterior ciliar arteries and vessels of retina is clearly visible (WMV 14341 kb)

Video 2

Real-time 7.0-T MRI and colour Doppler ultrasonography (CDUS) fusion imaging of orbita in a patient with large choroidal melanoma (CM). For denomination of structures please see the figure caption of Fig. 3. Note that the image contrast in the MRI plane shown is lower than in the original MRI data because the MRI plane was reconstructed by the ultrasound system from the imported MRI DICOM 3D data set. The video (file format, .wmv) shows the CDUS image in the upper left panel, the corresponding MRI plane in the lower right panel and the CDUS-MRI image overlay in the upper right panel. In the moving fusion image the tumour, the retinal detachment and the blood flow in the retinal and tumour vessels and the posterior ciliar arteries are clearly visualized (WMV 25989 kb)

References

  1. 1.
    Coleman DJ, Abramson DH, Jack RL, Franzen LA (1974) Ultrasonic diagnosis of tumors of the choroid. Arch Ophthalmol 91:344–354PubMedCrossRefGoogle Scholar
  2. 2.
    Guthoff RF, Berger RW, Winkler P, Helmke K, Chumbley LC (1991) Doppler ultrasonography of malignant melanomas of the uvea. Arch Ophthalmol 109:537–541PubMedCrossRefGoogle Scholar
  3. 3.
    Wolff-Kormann PG, Kormann BA, Riedel KG et al (1992) Quantitative color Doppler imaging in untreated and irradiated choroidal melanoma. Invest Ophthalmol Vis Sci 33:1928–1933PubMedGoogle Scholar
  4. 4.
    Lieb WE, Cohen SM, Merton DA, Shields JA, Mitchell DG, Goldberg BB (1991) Color Doppler imaging of the eye and orbit. Technique and normal vascular anatomy. Arch Ophthalmol 109:527–531PubMedCrossRefGoogle Scholar
  5. 5.
    Wolff K, Zhivov A, Blanck O, Hildebrandt G, Guthoff R, Walter U (2014) Colour doppler sonography after CyberKnife stereotactic radiosurgery of choroidal melanomas. Klin Monbl Augenheilkd (in press).Google Scholar
  6. 6.
    Mafee MF, Rapoport M, Karimi A, Ansari SA, Shah J (2005) Orbital and ocular imaging using 3- and 1.5-T MR imaging systems. Neuroimaging Clin N Am 15:1–21PubMedCrossRefGoogle Scholar
  7. 7.
    Lemke AJ, Kazi I, Felix R (2006) Magnetic resonance imaging of orbital tumors. Eur Radiol 16:2207–2219PubMedCrossRefGoogle Scholar
  8. 8.
    Buerk BM, Pulido JS, Chiong I et al (2004) Vascular perfusion of choroidal melanoma by 3.0 tesla magnetic resonance imaging. Trans Am Ophthalmol Soc 102:209–215PubMedCentralPubMedGoogle Scholar
  9. 9.
    Zhang Y, San Emeterio Nateras O, Peng Q, Rosende CA, Duong TQ (2012) Blood flow MRI of the human retina/choroid during rest and isometric exercise. Invest Ophthalmol Vis Sci 53:4299–4305PubMedCentralPubMedCrossRefGoogle Scholar
  10. 10.
    Richdale K, Wassenaar P, Teal Bluestein K et al (2009) 7 Tesla MR imaging of the human eye in vivo. J Magn Reson Imaging 30:924–932PubMedCentralPubMedCrossRefGoogle Scholar
  11. 11.
    Christoforidis JB, Wassenaar PA, Christoforidis GA, Ho VY, Knopp MV, Schmalbrock PM (2013) Retrobulbar vasculature using 7-T magnetic resonance imaging with dedicated eye surface coil. Graefes Arch Clin Exp Ophthalmol 251:271–277PubMedCrossRefGoogle Scholar
  12. 12.
    Beenakker JW, van Rijn GA, Luyten GP, Webb AG (2013) High-resolution MRI of uveal melanoma using a microcoil phased array at 7 T. NMR Biomed 26:1864–1869PubMedCrossRefGoogle Scholar
  13. 13.
    Graessl A, Schwerter M, Muhle M et al (2013) Design, evaluation and application of a six-channel transceiver array tailored for in vivo human eye imaging at 7.0 T. Proc Intl Soc Mag Reson Med 21:2747Google Scholar
  14. 14.
    Forzoni L, D'Onofrio S, De Beni S et al (2012) Virtual navigator registration procedure for transcranial application. In: Hellmich C, Hamza MH, Simsik D (eds) Proceedings of the IASTED International Conference Biomedical Engineering (BioMed 2012) February 15–17, 2012 Innsbruck. Austria. ACTA, Calgary, pp 496–503Google Scholar
  15. 15.
    Zamboni P, Menegatti E, Viselner G, Morovic S, Bastianello S (2012) Fusion imaging technology of the intracranial veins. Phlebology 27:360–367PubMedCrossRefGoogle Scholar
  16. 16.
    Walter U, Školoudík D (2014) Transcranial sonography (TCS) of brain parenchyma in movement disorders: quality standards, diagnostic applications and novel technologies. Ultraschall Med. doi: 10.1055/s-0033-1356415 Google Scholar
  17. 17.
    Center for Devices and Radiological Health (2008) Information for manufacturers seeking marketing clearance of diagnostic ultrasound systems and transducers. Food and Drug Administration. Silver Spring, MD. http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm070911.pdf. Accessed 29 June 2013
  18. 18.
    Lemke AJ, Alai-Omid M, Hengst SA, Kazi I, Felix R (2006) Eye imaging with a 3.0-T MRI using a surface coil - a study on volunteers and initial patients with uveal melanoma. Eur Radiol 16:1084–1089PubMedCrossRefGoogle Scholar
  19. 19.
    Krueger PC, Stachs O, Hadlich S et al (2012) MR microscopy of the human eye at 7.1 T and correlation with histopathology-proof of principle. Orbit 31:390–393PubMedCrossRefGoogle Scholar
  20. 20.
    Wiener E, Kolk A, Neff A, Settles M, Rummeny E (2005) Evaluation of reconstructed orbital wall fractures: high-resolution MRI using a microscopy surface coil versus 16-slice MSCT. Eur Radiol 15:1250–1255PubMedCrossRefGoogle Scholar
  21. 21.
    Erb-Eigner K, Warmuth C, Taupitz M, Willerding G, Bertelmann E, Asbach P (2013) Impact of magnetic field strength and receiver coil in ocular MRI: a phantom and patient study. Rofo 185:830–837PubMedCrossRefGoogle Scholar
  22. 22.
    Ladd ME (2007) High-field-strength magnetic resonance: potential and limits. Top Magn Reson Imaging 18:139–152PubMedCrossRefGoogle Scholar
  23. 23.
    Erb-Eigner K, Willerding G, Taupitz M, Hamm B, Asbach P (2013) Diffusion-weighted imaging of ocular melanoma. Invest Radiol 48:702–707PubMedCrossRefGoogle Scholar

Copyright information

© European Society of Radiology 2014

Authors and Affiliations

  • Uwe Walter
    • 1
    Email author
  • Thoralf Niendorf
    • 2
    • 3
  • Andreas Graessl
    • 2
  • Jan Rieger
    • 2
    • 3
  • Paul-Christian Krüger
    • 4
  • Sönke Langner
    • 4
  • Rudolf F. Guthoff
    • 5
  • Oliver Stachs
    • 5
  1. 1.Department of NeurologyUniversity of RostockRostockGermany
  2. 2.Berlin Ultrahigh Field Facility, Max-Delbrueck-Center for Molecular MedicineBerlinGermany
  3. 3.MRI.TOOLS GmbHBerlinGermany
  4. 4.Institute for Diagnostic Radiology and NeuroradiologyUniversity of GreifswaldGreifswaldGermany
  5. 5.Department of OphthalmologyUniversity of RostockRostockGermany

Personalised recommendations