Advertisement

Surface adhesive and hand-aided needle–assisted biopsy technique (SAHNA)

  • R. Rajakulasingam
  • A. Iqbal
  • S. L. James
  • R. BotchuEmail author
Technical Report
  • 27 Downloads

Abstract

Objective

We describe a novel and efficient CT biopsy technique that we have termed ‘skin adhesive and hand-aided biopsy technique’, shortened to the acronym SAHNA. This technique is ideal for biopsying surface osseous lesions where there is a lack of subcutaneous fat. This avoids the need for the radiologist’s hand to be in close proximity to the CT scanner beam at the time of intervention.

Methods

The SAHNA technique uses a plastic surgical forceps towel clamp to stabilise the biopsy needle just proximal to the lesion after piercing through the skin. The clamp itself is stabilised via a self-adhesive dressing which is stuck down to the skin at its shank.

Results

In our institution, we have already used the SAHNA technique successfully in over 5 cases.

Conclusion

The SAHNA technique is a widely accessible and effective way of stabilising the CT biopsy needle in technically challenging superficial bony lesions.

Keywords

Bone biopsy Superficial CT-guided Clamp Dressing Intervention 

Notes

Compliance with ethical standards

Conflict of interest

The authors declare that there are no conflicts of interest.

References

  1. 1.
    Roberts CC, Morrison WB, Deely DM, et al. Use of a novel percutaneous biopsy localization device: initial musculoskeletal experience. Skelet Radiol. 2007;36(1):53–7.CrossRefGoogle Scholar
  2. 2.
    McLoughlin E, Iqbal A, Patel A, et al. Dual steristrip technique: a novel use of steristrips to reduce operator radiation dose during CT-guided intervention. Skelet Radiol. 2019;48:1617–20.CrossRefGoogle Scholar
  3. 3.
    Nabi M, Teh J. The use of plastic gallipots for balancing needles in CT-guided procedures. Clin Radiol. 2019;74(7):570.CrossRefGoogle Scholar
  4. 4.
    Efstathopoulos EP, Pantos I, Andreou M, et al. Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures. Br J Radiol. 2011;84(997):70–7.CrossRefGoogle Scholar
  5. 5.
    Jungnickel K. Radiation protection during CT-guided interventions. In: Mahnken AH, Ricke J, editors. CT- and MR-guided interventions in radiology. New York: Springer; 2009. p. 35–8.CrossRefGoogle Scholar
  6. 6.
    Body Physics: Motion to metabolism. Davis L. Open Oregon Educational resources. Information taken from following website: https://openoregon.pressbooks.pub/bodyphysics/chapter/body-levers/
  7. 7.
    Kroes MW, Busser WMH, Futterter JJ, et al. Assessment of needle guidance devices for their potential to reduce fluoroscopy time and operator hand dose during C-arm cone-beam computed tomography-guided needle interventions. J Vasc Interv Radiol. 2013;24(6):901–6.CrossRefGoogle Scholar
  8. 8.
    Sarmento S, Pereira JS, Sousa MJ, et al. The use of needle holders in CTF guided biopsies as a dose reduction tool. J Appl Clin Med Phys. 2018;19(1):250–8.CrossRefGoogle Scholar

Copyright information

© ISS 2019

Authors and Affiliations

  • R. Rajakulasingam
    • 1
  • A. Iqbal
    • 1
  • S. L. James
    • 1
  • R. Botchu
    • 1
    Email author
  1. 1.Department of Musculoskeletal RadiologyRoyal Orthopedic Hospital NHS Foundation TrustBirminghamUK

Personalised recommendations